Qing Zhu1,2, Ping Ye1, Jinyao Wang2, Zhaohua Chang1,2. 1. Shanghai Institute for Minimally Invasive Therapy, School of Medical Instrument and Food Engineering, University of Shanghai for Science and Technology, Shanghai, PR China. 2. Shanghai MicroPort Endovascular MedTech (Group) Co., Ltd, Shanghai, PR China.
Abstract
OBJECTIVE: This study evaluated the effectiveness and safety of a novel endograft with a biodegradable coating at both ends in preventing edge stenosis in swine iliac arteries. The biodegradable coating was composed of polylactide and paclitaxel. METHODS: Four types of endograft were implanted in the iliac arteries of healthy swine: an endograft without coating (control group) and endografts with polylactide and paclitaxel coating containing 0.1, 0.3, or 3.6 μg/mm2 of paclitaxel. The edge stenosis of these endografts in swine iliac arteries was assessed using angiographic image data at 30, 90, and 180 days after the operation. After terminal angiography, histologic evaluation of the treated arteries was performed. The treated sections of iliac arteries and blood samples were obtained at 1, 7, 30, 90, and 180 days for pharmacokinetic analysis. RESULTS: The results of angiographic and histologic evaluation demonstrated that intimal hyperplasia contributed to edge stenosis and polylactide-paclitaxel coating effectively inhibited edge stenosis. At 30 days, edge stenosis was observed at both the proximal and distal edges of the endograft without coating. At 90 days, edge stenosis was detected for the endograft coated with 0.1 μg/mm2 paclitaxel, and ectasia dilation occurred at the proximal and distal edges of the endograft coated with 3.6 μg/mm2 paclitaxel. No edge stenosis or other adverse effects were observed at 90 and 180 days for the endograft coated with 0.3 μg/mm2 paclitaxel. In addition, for the endograft coated with 0.3 μg/mm2 paclitaxel, a pharmacokinetic analysis showed that the paclitaxel concentration of treated segments decreased from 14 264 ± 1020 ng/g at day 1 to 80 ± 70 ng/g at day 90, and 20 ± 40 ng/g at day 180. The plasma paclitaxel concentration was low at day 1 and no longer detected after 7 days. CONCLUSIONS: Polylactide and paclitaxel coating containing 0.3 μg/mm2 paclitaxel at both ends of endografts effectively and safely inhibits edge stenosis in swine iliac arteries.
OBJECTIVE: This study evaluated the effectiveness and safety of a novel endograft with a biodegradable coating at both ends in preventing edge stenosis in swine iliac arteries. The biodegradable coating was composed of polylactide and paclitaxel. METHODS: Four types of endograft were implanted in the iliac arteries of healthy swine: an endograft without coating (control group) and endografts with polylactide and paclitaxel coating containing 0.1, 0.3, or 3.6 μg/mm2 of paclitaxel. The edge stenosis of these endografts in swine iliac arteries was assessed using angiographic image data at 30, 90, and 180 days after the operation. After terminal angiography, histologic evaluation of the treated arteries was performed. The treated sections of iliac arteries and blood samples were obtained at 1, 7, 30, 90, and 180 days for pharmacokinetic analysis. RESULTS: The results of angiographic and histologic evaluation demonstrated that intimal hyperplasia contributed to edge stenosis and polylactide-paclitaxel coating effectively inhibited edge stenosis. At 30 days, edge stenosis was observed at both the proximal and distal edges of the endograft without coating. At 90 days, edge stenosis was detected for the endograft coated with 0.1 μg/mm2 paclitaxel, and ectasia dilation occurred at the proximal and distal edges of the endograft coated with 3.6 μg/mm2 paclitaxel. No edge stenosis or other adverse effects were observed at 90 and 180 days for the endograft coated with 0.3 μg/mm2 paclitaxel. In addition, for the endograft coated with 0.3 μg/mm2 paclitaxel, a pharmacokinetic analysis showed that the paclitaxel concentration of treated segments decreased from 14 264 ± 1020 ng/g at day 1 to 80 ± 70 ng/g at day 90, and 20 ± 40 ng/g at day 180. The plasma paclitaxel concentration was low at day 1 and no longer detected after 7 days. CONCLUSIONS: Polylactide and paclitaxel coating containing 0.3 μg/mm2 paclitaxel at both ends of endografts effectively and safely inhibits edge stenosis in swine iliac arteries.
Type of Research: Swine modelKey Findings: Edge stenosis of endografts at 6 months could be decreased by using endografts with polylactide and paclitaxel coating containing 0.3 μg/mm2 paclitaxel at both ends in swine iliac arteries.Take Home Message: Polylactide and paclitaxel coating at both ends of endografts effectively and safely inhibits edge stenosis.Peripheral arterial disease (PAD) is commonly caused by occlusive lesions in lower extremity vessels. Endovascular treatment is often the first choice for patients with PAD. The bare nitinol stent has been used for the treatment of superficial femoral artery (SFA) lesions, and the 1-year primary patency rate has been reported to be between 63% and 80% in lesions with a length of 5 to 11 cm.1, 2, 3 The major limitation of the nitinol stent is the ingrowth of intimal hyperplasia through stent interstices, which is associated with stent fractures in mobile SFA lesions. An endograft is used to inhibit the development of in-stent restenosis, and early clinical data suggest that the use of an endograft can result in a higher short-term patency rate than angioplasty in lesions in this challenging territory.5, 6, 7 The primary patency rate at 12 months was 78.4% for a heparin-bonded endograft in the treatment of long-segment femoropopliteal arterial lesions, whereas the 3-year primary patency rate did not significantly differ between patients treated with an endograft and those who received the bare nitinol stent.Edge stenosis can cause failure of the primary patency of an endograft. A clinical study reported that three of four patients presenting with acute endograft thrombosis had underlying edge stenosis. A retrospective study of patients who received self-expanding polytetrafluoroethylene (ePTFE) stent grafts between November 2001 and December 2011 demonstrated that 88 patients had a total of 115 edge stenoses, most of which occurred within the first year after implantation; the average time to stenosis discovery was 10.7 months. A 78-year-old patient who was implanted with a 6 × 250-mm Viabahn endograft in the SFA suddenly died owing to acute myocardial infarction 23 months later. Histologic analysis revealed neointimal proliferation not in the middle but at the edges of the stent graft and polymorphic smooth muscle cell (SMC) proliferation in the abundant extracellular matrix.Paclitaxel has a potent inhibitory effect on the proliferation and migration of SMCs., Lin et al used a paclitaxel drug balloon to expand the edge immediately after implantation of the ePTFE stent graft. The results showed that, compared with no treatment of the stent graft, drug balloon dilatation decreased the incidence of edge stenosis (P = .021) and target lesion revascularization (P = .010), and the 1-year patency rate was significantly higher (92.1% vs 76.4%; P =. 042). Baek et al demonstrated that coating paclitaxel onto the terminal parts of hemodialysis grafts markedly decreased neointimal hyperplasia at the sites of graft-vessel anastomoses. Although various attempts have been made to prevent edge stenosis in endografts, the use of biodegradable polymers and a paclitaxel coating on the endograft to prevent edge stenosis has not yet been investigated. Because of their biodegradability, polymers provide a solution for controlled drug release, which is critical for drug delivery at the right time for novel endografts with a coating.,In this study, we coated both ends of endografts with polylactide and different amounts of paclitaxel to evaluate the coating's effectiveness and safety in preventing edge stenosis in swine iliac arteries. Because an endograft was designed with a coating on both ends to decrease the paclitaxel dose, the effect of paclitaxel doses on edge stenosis was investigated.In accordance with several commercially available drug eluting stents, the following drug doses were used in this study:Low dose (0.1 μg/mm2), below the dose of the Eluvia stent (0.167 μg/mm2; Boston Scientific, Marlborough, Mass);High dose (3.6 μg/mm2), above the dose of the Zilver PTX stent (3 μg/mm2; Cook Medical, Bloomington, Ind); andIntermediate dose (0.3 μg/mm2), three times the low drug dose.
Methods
Materials and preparation of polylactide and paclitaxel-coated endografts
Paclitaxel was purchased from Sigma-Aldrich (St Louis, Mo), and polylactide was obtained from Durect (Cupertino, Calif). Endografts were supplied by Shanghai MicroPort Endovascular MedTech, (Shanghai, China). The endografts were composed of a self-expanding nitinol stent and ePTFE fabric. The stent was bonded to ePTFE fabric. All other chemicals and reagents used were of analytical grade.Polylactide and paclitaxel in a fixed ratio were dissolved in n-propyl acetate to prepare the coating solution. Before coating, each endograft was inspected for contaminants and particles. The coating solution was also examined for particles. The endograft was mounted at the end of a rod-shaped metal holder. Flow of the coating solution was achieved using a 10-mL syringe and syringe pump. Solution flow, endograft rotation, and translation were controlled using computer software. The coating solution was sprayed on abluminal surfaces on each end of the endograft (5 mm for each end; Fig 1) by using a drug spraying machine. The paclitaxel drug density was 0.1, 0.3, and 3.6 μg/mm2 in terms of the endograft surface area. All grafts were sterilized in ethylene oxide before implantation. The qualitative and quantitative assessment of paclitaxel after sterilization was performed through high-performance liquid chromatography (HPLC) and the Chinese Pharmacopeia method.
Fig 1
A, Schematic of the endograft and treated arteries. Both ends of the endograft were coated (5 mm each). The proximal and distal ends of treated arteries were sampled, and the length of each samples was 15 mm. B, An actual photo of the endograft.
A, Schematic of the endograft and treated arteries. Both ends of the endograft were coated (5 mm each). The proximal and distal ends of treated arteries were sampled, and the length of each samples was 15 mm. B, An actual photo of the endograft.
Morphologic characterization
Surface morphologic changes of the polylactide and paclitaxel coated endografts were analyzed using a scanning electron microscope (FEI, Hillsboro, Ore). Four samples obtained from the control group (endografts without coating) and the experimental group (coated endografts) containing 0.1, 0.3, and 3.6 μg/mm2 paclitaxel were evaluated by randomly examining the surface morphology.
In vitro release test
For the in vitro release test, we used 45% ethanol/water as the release medium. Samples of polylactide and paclitaxel coated endografts were soaked in brown vials with 10 mL of the release medium, and the vials were placed in a shaker at 37 °C. At the designated time, the release medium was completely removed from the vials and stored for analysis. The medium was replaced with 10 mL of fresh release medium. The amount of paclitaxel released into the medium was measured through HPLC (Agilent 1200 Series; Agilent Technologies, Santa Clara, Calif) with a 4.6 × 250.0-mm C18 reversed-phase column. The mobile phase used was methanol:acetonitrile:water (23:36:41) under isocratic conditions at flow rate of 1 mL/min, and the ultraviolet detector was set at 227 nm. The retention time of paclitaxel was 22 minutes.
Experimental animals and operation technique
To select an appropriate dose of paclitaxel for preventing edge stenosis, an in vivo study was performed using different types of endografts (Fig 2, A). A total of 20 healthy swine (n = 20) received one of the four types of graft: an endograft without coating (n = 3), an endograft with 0.1 μg/mm2 paclitaxel (n = 7), an endograft with 0.3 μg/mm2 paclitaxel (n = 7), or an endograft with 3.6 μg/mm2 paclitaxel (n = 3). The endografts had a length of 40 mm and a diameter of 8 to 10 mm and were implanted under the guidance of digital subtraction angiography to obtain a graft to artery ratio of 1.1 to 1.2. The animals were maintained in standard animal care facilities at Gateway Medical Innovation Center. This study was reviewed and approved by the Institutional Animal Care and Use Committee of Gateway Medical Innovation Center. The operation procedures conformed to the Guidelines for the Care and Use of Laboratory Animals (National Institutes of Health publication No. 85-23, revised 1996).
Fig 2
In vivo study design. A, Preventing edge stenosis in the in vivo study for the endograft in a swine iliac artery. ∗If any endograft group exhibited edge stenosis on angiogram during follow-up, the group was terminated, and all animals in the group were sacrificed for histologic evaluation. •Angiogram and histology. ○Angiogram. B, In vivo pharmacokinetic study of the endograft in the swine iliac artery.
In vivo study design. A, Preventing edge stenosis in the in vivo study for the endograft in a swine iliac artery. ∗If any endograft group exhibited edge stenosis on angiogram during follow-up, the group was terminated, and all animals in the group were sacrificed for histologic evaluation. •Angiogram and histology. ○Angiogram. B, In vivo pharmacokinetic study of the endograft in the swine iliac artery.
Angiographic evaluation
To assess the edge stenosis of treated segments, digital subtraction angiography was performed via common carotid cutdown, an 8F sheath inserted, as a survival procedure at 30, 90, and 180 days after the operation. A quantitative vascular analysis was performed at different time points by using QAngio-XA software, version 7.3 (Medis, Medical Imaging Systems, Leiden, the Netherlands). The minimal lumen diameter of the treated segments and reference vessel diameter (RVD) at the proximal and distal ends were measured postoperatively (30, 90, and 180 days). The RVD was taken from the proximal and distal portions of the treated site, which was an unaffected segment of the endograft. The percentage of diameter stenosis (DS%) was calculated as follows: (1 – [minimal lumen diameter/RVD] × 100%). Edge stenosis was defined as a DS% of more than 50%.
Tissue sampling and histologic evaluation
If any endograft group exhibited edge stenosis on angiography evaluation during follow-up, the follow-up of this group was terminated and all animals in the group were sacrificed for histologic evaluation. After the terminal angiogram, the implanted endografts were excised along with adjacent vessels. The sample processing and histologic analysis were performed using PharmaLegacy Biomedical Technology (Shanghai, China). The endografts were immersion fixed in 10% neutral buffered formalin for at least 24 hours, dehydrated with gradient alcohol, infiltrated with xylene, embedded in methyl methacrylate, and then placed in an oven at 33 °C to 35 °C until they had completely hardened. Sections (50 μm) were cut perpendicular to the direction of blood flow at the proximal, middle, and distal portions of the endografts. One section was obtained at the center of the endograft, and the others were obtained at the proximal and distal sides of the endograft. All sections were stained with hematoxylin and eosin, and a histologic evaluation was performed. The scores for neointimal proliferation and inflammation were obtained as reported elsewhere., The inflammation score was as follows: 0 (no inflammatory cells surrounding the strut), 1 (light, noncircumferential lymphohistiocytic infiltrate surrounding strut), 2 (localized, moderate cellular aggregate surrounding the strut noncircumferentially), 3 (localized, moderate to dense cellular aggregate surrounding the strut noncircumferentially), and 4 (circumferential dense lymphohistiocytic cell infiltration of the strut). The neointimal proliferation scoring ranged from 0 to 5 in 0.5-unit increments. Any tissue ingrowth overlying the endograft lumen was considered to be neointimal hyperplasia, with minimal lining of the endograft being 0.5 and total occlusion of the lumen by hyperplasia arbitrarily scored as 5. An exemplary slide was used to define the score of 0.5 before scoring started. The degree of hyperplasia in the hematoxylin and eosin sections was scored by three independent observers who were blinded to the experimental group.
In vivo pharmacokinetic evaluation
In vivo pharmacokinetic evaluation was performed for endografts with appropriate doses to prevent the edge stenosis (Fig 2, B). The pharmacokinetic evaluation was performed for 10 pigs. Treated arteries and blood samples were retrieved at 1, 7, 30, 90, and 180 days. Each time point had two pigs, and each pig had two treated arteries. Two samples were from the proximal and distal portions of each artery (Fig 1, A); thus, eight samples were obtained for each time point. The organ samples (spleen and kidneys) were examined for drug distribution. The artery and organ samples were tested by Shanghai ChemPartner. An HPLC-tandem mass spectrometric method was developed and validated for quantifying the paclitaxel in the pig artery and organ homogenate over the concentration ranges of 1.00 to 1000.00 ng/mL (artery) and 0.10 to 100.00 ng/mL (organ). Sample preparation involved the extraction of the analyte and internal standard from the pig artery and organ homogenate through liquid-liquid extraction. The analytes were chromatographed on an reversed-phase HPLC column (Waters Xbridge-C18 [Milford, Mass], 3.5 μm, 2.1 × 50.0 mm) and detected through tandem mass spectrometry using a turbo ion spray interface in the positive ionization mode. Samples were freshly prepared or kept frozen at −70 °C before analysis, and 100 μL of an artery or organ homogenate sample was used for analysis.The cumulative percentage of paclitaxel released in vivo was calculated as follows: (1 – [the amount of residual drug/the amount of initial drug loaded] × 100%). The amount of residual drug and initial drug loading were quantitated through HPLC.
Statistical analysis
One-way analysis of variance was used to determine statistically significant differences at a P value of less than .05. Experimental data collected in this study are presented as the mean ± standard deviation.
Results
Characterization of polylactide and paclitaxel coated endografts
The time of chromatography peaks for endografts after sterilization was consistent with that for standard substance (Supplementary Fig 1), indicating that they both contained the same compound. The total amount of paclitaxel loaded at the ends of the endograft (5 mm each) for 0.1, 0.3, and 3.6 μg/mm2 paclitaxel was 38.5 ± 5.8, 90.1 ± 12.7, and 1129.0 ± 89.7 μg, respectively. The density of paclitaxel was 0.13 ± 0.02, 0.30 ± 0.04, and 3.70 ± 0.24 μg/mm2, respectively.
Supplementary Fig 1
The high-performance liquid chromatography (HPLC) chromatogram of paclitaxel after sterilization for standard substance (A) and the endograft with 0.1 μg/mm2(B), 0.3 μg/mm2(C), and 3.6 μg/mm2(D) paclitaxel. ∗The test solution from the sample of endograft with 3.6 μg/mm2 paclitaxel was diluted five times to get better HPLC chromatogram spectra.
The surface morphology of the endografts before and after coating was compared to determine whether the surface features were different (Fig 3). We observed that the coating was thin and porous when the endograft was coated with 0.1 μg/mm2 paclitaxel but thick and dense when endograft was coated with 0.3 or 3.6 μg/mm2 paclitaxel. This finding suggested that the coating on the endograft surface was denser when a higher dose of paclitaxel was used.
Fig 3
Scanning electron microscope images of the surface of the endograft without coating (A) and with the coating containing 0.1 μg/mm2(B), 0.3 μg/mm2(C), and 3.6 μg/mm2(D) paclitaxel (original magnification ×2000).
Scanning electron microscope images of the surface of the endograft without coating (A) and with the coating containing 0.1 μg/mm2(B), 0.3 μg/mm2(C), and 3.6 μg/mm2(D) paclitaxel (original magnification ×2000).
In vitro release profile of paclitaxel
The in vitro release profiles of paclitaxel within 48 hours for the endografts coated with 0.1, 0.3, and 3.6 μg/mm2 paclitaxel are displayed in Fig 4. For the endograft coated with 0.1 μg/mm2 paclitaxel, the initial amount of paclitaxel released was more than 40% of the amount loaded, and the drug was released gradually until the amount was undetectable at 24 hours. By contrast, for the endograft coated with 0.3 and 3.6 μg/mm2 paclitaxel, the initial amount of paclitaxel released was less than 40%, and the cumulative amount of drug released was more than 60% at 48 hours.
Fig 4
In vitro release curves (presented as the mean ± standard deviation) for the endograft with polylactide and paclitaxel coating containing 0.1 μg/mm2 (squares), 0.3 μg/mm2 (circles), and 3.6 μg/mm2 (triangles) paclitaxel.
In vitro release curves (presented as the mean ± standard deviation) for the endograft with polylactide and paclitaxel coating containing 0.1 μg/mm2 (squares), 0.3 μg/mm2 (circles), and 3.6 μg/mm2 (triangles) paclitaxel.
Endograft deployment and angiographic evaluation
Adult domestic pigs weighing between 75 and 85 kg were pretreated with aspirin (325 mg) and clopidogrel (75 mg) by mouth for 3 days before and on the day of implantation. Aspirin (100 mg) was then administered daily until the day of euthanasia. Heparin was used for intraoperative anticoagulation at a dose of 100 IU/kg. All grafts were deployed successfully without complication and all animals remained alive for the duration of the study. The weight of the animals at end of study was 92.0 ± 4.5 kg.At 30 days, the follow-up was terminated for the control group, and 3 pigs were humanely killed. At 90 days, the follow-up was terminated for the 0.1 μg/mm2 and 3.6 μg/mm2 paclitaxel groups. The number of humanely killed pigs was 7 for the 0.1 μg/mm2 paclitaxel group and 3 for the 3.6 μg/mm2 paclitaxel group. At 180 days, the follow-up was terminated for the 0.3 μg/mm2 paclitaxel group and 7 pigs were sacrificed.The angiographic data for all endografts obtained at 30, 90, and 180 days are shown in Table I and Fig 5. At 30 days, edge stenosis was observed at the proximal and distal edges of the endograft without coating. At 90 days, edge stenosis was discovered at the proximal and distal edges of the endograft coated with 0.1 μg/mm2 paclitaxel, although the DS% was 45 ± 7%. Artery ectasia occurred at the proximal and distal edges of the endograft coated with 3.6 μg/mm2 paclitaxel. At 180 days, no edge stenosis was yet noted for the endograft coated with 0.3 μg/mm2 paclitaxel.
At 30 days, the follow-up was terminated for control group. At 90 days, the follow-up was terminated for the 0.1 μg/mm2 group and the 3.6 μg/mm2 group. At 180 days, the follow-up was terminated for the 0.3 μg/mm2 group.
P < .05 vs the control group.
P < .05 vs the 0.1μg/mm2 group.
Fig 5
Representative angiography images of the endograft without coating (control group) and with the coating containing 0.1, μg/mm2 0.3 μg/mm2, and 3.6 μg/mm2 paclitaxel at 30, 90, and 180 days.
Angiographic data at 30, 90 and 180 daysa%DS, Percentage of diameter stenosis; MLD, minimal lumen diameter; RVD, reference vessel diameter.At 30 days, the follow-up was terminated for control group. At 90 days, the follow-up was terminated for the 0.1 μg/mm2 group and the 3.6 μg/mm2 group. At 180 days, the follow-up was terminated for the 0.3 μg/mm2 group.P < .05 vs the control group.P < .05 vs the 0.1μg/mm2 group.Representative angiography images of the endograft without coating (control group) and with the coating containing 0.1, μg/mm2 0.3 μg/mm2, and 3.6 μg/mm2 paclitaxel at 30, 90, and 180 days.
Histologic evaluation
Treated arterial cross-sections of the control group (30 days) and different drug dose groups (90 and 180 days) are shown in Fig 6. Intimal hyperplasia at the proximal and distal portions of the endograft was evident in the control group and was also found but less severe in the 0.1 μg/mm2 paclitaxel group. The 0.3 μg/mm2 and 3.6 μg/mm2 paclitaxel groups showed significantly less neointimal formation than the control group. To better observe edge stenosis, magnified histologic images were obtained and are shown in Supplementary Fig 2; scores for inflammation and intimal hyperplasia are listed in Table II. It was found that the scores for inflammation and intimal hyperplasia were higher in the control group than in the paclitaxel coating groups.
Fig 6
Cross-sections of the middle, and distal portions of the endograft without coating (control group) and with the coating containing 0.1 μg/mm2, 0.3 μg/mm2, and 3.6 μg/mm2 paclitaxel at termination. (Stain: hematoxylin and eosin.)
Supplementary Fig 2
The magnification histologic images (original magnification ×40) for the endograft without coating (A) and with the coating containing 0.1 μg/mm2(B), 0.3 μg/mm2(C), and 3.6 μg/mm2(D) of paclitaxel. (Stain: hematoxylin and eosin.)
Table II
Paclitaxel concentration of the endograft coated with 0.3 μg/mm2 paclitaxel in treated arteries and plasma
Days
Treated artery, ng/g
Plasma, ng/mL
1
14 264 ± 1020
0.2
7
4160 ± 700
BQL
30
588 ± 320
BQL
90
80 ± 70
BQL
180
20 ± 40
BQL
BQL, Below the quantification limit. The detection range of paclitaxel plasma is 0.10-100.00 ng/mL.
Cross-sections of the middle, and distal portions of the endograft without coating (control group) and with the coating containing 0.1 μg/mm2, 0.3 μg/mm2, and 3.6 μg/mm2 paclitaxel at termination. (Stain: hematoxylin and eosin.)Paclitaxel concentration of the endograft coated with 0.3 μg/mm2 paclitaxel in treated arteries and plasmaBQL, Below the quantification limit. The detection range of paclitaxel plasma is 0.10-100.00 ng/mL.In vivo pharmacokinetic evaluation was performed for 10 pigs treated with the endograft coated with 0.3 μg/mm2 paclitaxel. The paclitaxel concentrations in the artery and plasma are shown in Table III. The treated arterial paclitaxel concentration decreased from 14 264 ± 1020 ng/g at day 1 to 80 ± 70 ng/g at day 90, and 20 ± 40 ng/g at day 180. The plasma paclitaxel concentration was 0.2 ng/mL at day 1, and was not detectable after 7 days. The paclitaxel concentration in the spleen and kidneys was 0.6 and 1.1 ng/g at day 1, respectively, and was not detectable after 7 days.
Table III
Scores for inflammation and neointimal proliferation at termination
Control
0.1 μg/mm2
0.3 μg/mm2
3.6 μg/mm2
(n = 3)
(n = 7)
(n = 7)
(n = 3)
Inflammation
Proximal
2.0 ± 0.6
3.0 ± 1.2
1.2 ± 1.2
1.1 ± 1.0
Middle
2.0 ± 1.0
2.2 ± 1.1
0.5 ± 0.5a
1.0 ± 0.8
Distal
2.2 ± 1.0
2.3 ± 0.8
0.6 ± 0.9
1.5 ± 0.8
Neointimal proliferation
Proximal
2.2 ± 1.0
1.8 ± 1.3
0.9 ± 1.0
0.7 ± 1.7
Middle
2.0 ± 0.9
1.2 ± 1.2
0.2 ± 0.5a
0.3 ± 1.1
Distal
2.0 ± 1.2
2.0 ± 1.3
0.5 ± 0.8a
0.5 ± 0.5a
P < .05 vs control group.
Scores for inflammation and neointimal proliferation at terminationP < .05 vs control group.The cumulative percentage of paclitaxel released in vivo is shown in Supplementary Fig 3. At day 1, the amount of paclitaxel released in vivo was 50 ± 8%, and at 180 days, all the paclitaxel had been released.
Supplementary Fig 3
The cumulative percentage paclitaxel released in vivo for the endograft with polylactide acid and paclitaxel coating containing 0.3 μg/mm2 paclitaxel at 1, 7, 30, 90, and 180 days.
Discussion
In this study, our results indicated that the polylactide and paclitaxel coating containing 0.3 μg/mm2 paclitaxel at both ends of the endograft (5 mm each) was effective in decreasing edge stenosis of the endograft at 180 days in swine iliac arteries. For a low paclitaxel dose of 0.1 μg/mm2, edge stenosis was observed, whereas for a high dose of 3.6 μg/mm2 artery ectasia was found in the treated segments at 90 days. These results suggest that paclitaxel and polylactide coating have effects on suppressing edge stenosis of endograft.In line with the clinical observations of previous studies,, our angiographic and histologic evaluations demonstrated that neointimal proliferation mainly occurred at the edges of the endograft. Intimal hyperplasia contributed to edge stenosis and a progressive decrease in the extent of neointimal formation at both ends of the endograft with an increasing paclitaxel dose.Paclitaxel-coated devices improve the patency of lower extremity revascularization in patients with PAD. However, a meta-analysis by Katsanos et al suggested an increase in late mortality for patients treated with paclitaxel-coated devices. This finding spurred considerable debate in the interventional community in terms of the safety of paclitaxel. Although the US Food and Drug Administration has responded to recent, favorable, long-term safety data on paclitaxel-based devices used in PAD, it continues to maintain that additional follow-up is needed.In the present study, artery ectasia occurred for the endograft with a high drug dose and was observed for a paclitaxel-eluting stent approved by the US Food and Drug Administration; however, the paclitaxel dose of the approved stent was considerably low (0.167 μg/mm2), which can be attributed to more than 12 months of sustained drug release. Therefore, it is important to study the amount and release time of coating drug.Paclitaxel is a highly lipophilic drug that is rapidly taken up by cells and exerts an inhibitory effect on PDGF-induced human arterial SMC proliferation.,25, 26, 27 The mode of action is the stabilization of polymerized microtubules and enhancement of the microtubular assembly. Ishihara et al found that the mechanism of restenosis after the endograft implantation was neointimal hyperplasia at the edge of the endograft. Neointimal hyperplasia is largely a result of proliferation and migration of SMCs and extracellular matrix formation within 1 to 6 months; thus, the drug release must have a long-lasting effect after endograft implantation.As illustrated in Fig 4, an analysis of the in vitro release profiles of the endografts demonstrated that for a low paclitaxel dose (0.1 μg/mm2), the initial release was significantly faster, and the residual amount was undetectable after 24 hours. The drug release time was possibly too short to inhibit neointimal hyperplasia, resulting in edge stenosis at 90 days. The endografts coated with 0.3 and 3.6 μg/mm2 paclitaxel had a lower rate of drug release, the initial amount of paclitaxel released was less than 40%, and the cumulative amount of drug released was more than 60% at 48 hours. Both coatings should have a long drug release period. The high paclitaxel dose (3.6 μg/mm2) was more than 10 times higher than that for the endograft coated with 0.3 μg/mm2 paclitaxel. A high dose can elicit a profound toxic response from SMCs in media and trigger the loss of vascular tone with consequent distention, which may have been responsible for the artery ectasia at 90 days after the operation. By comparing the in vitro release profiles of different types of endografts, it was revealed that the release kinetics of paclitaxel from the polylactide coating depended on the drug loadings.Drug release from biodegradable polymer systems can be controlled by (1) diffusion of drug from the polymeric matrix, (2) dissolution of drug into the release medium, and (3) biodegradation of polymeric chains. For the endograft coated with 0.1 μg/mm2 paclitaxel, the coating was thin and porous; thus, the release medium could easily penetrate the coating, resulting in rapid drug release. Drug release was primarily controlled by drug dissolution; thus, a rapid initial release was observed. For the endograft with 0.3 and 3.6 μg/mm2 paclitaxel, the coating was thick and dense; thus, the release medium could not easily penetrate the coating, resulting in slow release of the drug. Drug release was primarily controlled by slow diffusion of the drug through the coating; thus, the drug release rate was lower compared with that for the endografts coated with 0.1 μg/mm2 of paclitaxel.In our study, the pharmacokinetic analysis of arteries treated with an endograft coated with 0.3 μg/mm2 of paclitaxel revealed that the drug level in the treated vessel decreased over time, and at 180 days, the drug level was 20 ng/g, which was still within the therapeutic range reported from cell culture experiments (effective range, 10 nmol/L [7 ng/g] to 50 μmol/L [42 700 ng/g]), and no adverse effect or systemic toxicity of paclitaxel was observed at 180 days.The residual amount of paclitaxel in the endografts was quantified through HPLC at each time point. The in vitro/in vivo correlation in this study is illustrated in Fig 7. The in vivo release of paclitaxel from the endografts retrieved from swine iliac arteries was closely predicted by the in vitro release model in 45% ethanol/water medium.
Fig 7
In vitro/in vivo correlation for the endograft with 0.3 μg/mm2 paclitaxel and polylactide coating in swine iliac artery models.
In vitro/in vivo correlation for the endograft with 0.3 μg/mm2 paclitaxel and polylactide coating in swine iliac artery models.For the endograft without coating, histologic analysis (Fig 6) showed severe intimal hyperplasia contributing to edge stenosis, which is consistent with reported clinical observations made by Ishihara et al. When the endograft is used in clinical practice, some measures should be taken to prevent edge stenosis, such as using paclitaxel at both ends of the endograft.This study has several limitations. First, this study was performed in a healthy swine model. Because the atherosclerotic plaque component can affect drug uptake and retention, a healthy swine model does not precisely simulate responses to the coated endograft in humans. Second, although the endograft with polylactide and paclitaxel coating containing 0.3 μg/mm2 of paclitaxel was safe and effective in this study, the sample size was still small; therefore, more experimental animals should be considered.
Conclusions
This study demonstrates that endografts with polylactide and paclitaxel coating at both ends are effective and safe in inhibiting edge stenosis in swine iliac arteries. The amount of paclitaxel in the coating and its release profile are important variables in suppressing neointimal hyperplasia and limiting of drug toxicity. The optimum drug dose of paclitaxel at both ends of the endograft and without apparent toxicity was 0.3 μg/mm2.
Author contributions
Conception and design: QZ, PY, ZCAnalysis and interpretation: QZ, PY, JW, ZCData collection: QZ, JWWriting the article: QZ, JWCritical revision of the article: QZ, PY, ZCFinal approval of the article: QZ, PY, JW, ZCStatistical analysis: QZ, ZCObtained funding: Not applicableOverall responsibility: ZC
Authors: John F LaDisa; Lars E Olson; Ismail Guler; Douglas A Hettrick; Said H Audi; Judy R Kersten; David C Warltier; Paul S Pagel Journal: J Appl Physiol (1985) Date: 2004-02-06
Authors: John R Laird; Barry T Katzen; Dierk Scheinert; Johannes Lammer; Jeffrey Carpenter; Maurice Buchbinder; Rajesh Dave; Gary Ansel; Alexandra Lansky; Ecaterina Cristea; Tyrone J Collins; Jeffrey Goldstein; Michael R Jaff Journal: Circ Cardiovasc Interv Date: 2010-05-18 Impact factor: 6.546
Authors: Bahar Golchehr; Suzanne Holewijn; Rombout R Kruse; Laurens A van Walraven; Clark J Zeebregts; Michel M P J Reijnen Journal: Catheter Cardiovasc Interv Date: 2015-06-23 Impact factor: 2.692
Authors: T Kuji; T Masaki; K Goteti; L Li; S Zhuplatov; C M Terry; W Zhu; J K Leypoldt; R Rathi; D K Blumenthal; S E Kern; A K Cheung Journal: Kidney Int Date: 2006-05-03 Impact factor: 10.612
Authors: A W Heldman; L Cheng; G M Jenkins; P F Heller; D W Kim; M Ware; C Nater; R H Hruban; B Rezai; B S Abella; K E Bunge; J L Kinsella; S J Sollott; E G Lakatta; J A Brinker; W L Hunter; J P Froehlich Journal: Circulation Date: 2001-05-08 Impact factor: 29.690
Authors: Hans Krankenberg; Michael Schlüter; Hermann J Steinkamp; Karlheinz Bürgelin; Dierk Scheinert; Karl-Ludwig Schulte; Erich Minar; Patrick Peeters; Marc Bosiers; Gunnar Tepe; Bernhard Reimers; Felix Mahler; Thilo Tübler; Thomas Zeller Journal: Circulation Date: 2007-06-25 Impact factor: 29.690
Authors: Richard R Saxon; Michael D Dake; Robert L Volgelzang; Barry T Katzen; Gary J Becker Journal: J Vasc Interv Radiol Date: 2008-04-10 Impact factor: 3.464