Andrea Padoan1, Chiara Cosma2, Francesco Bonfante3, Foscarina Della Rocca4, Francesco Barbaro5, Claudia Santarossa4, Luigi Dall'Olmo6, Matteo Pagliari3, Alessio Bortolami3, Annamaria Cattelan5, Vito Cianci4, Daniela Basso1, Mario Plebani7.
Abstract
BACKGROUND: Studies evaluating neutralizing antibody (NAb) after BNT162b2 vaccine are scarce. We therefore compared NAb using the plaque reduction neutralization test (PRNT) in vaccinated subjects, with those from five chemiluminescent (CLIA) assays, two targeting ACE and S-RBD interaction.Entities:
Keywords: Antibody; BNT162b2; COVID-19; Comirnaty; Immunoassays; Immunological response; SARS-CoV-2 vaccine; Serology
Mesh:
Substances:
Year: 2021 PMID: 34755646 PMCID: PMC8553360 DOI: 10.1016/j.cca.2021.10.028
Source DB: PubMed Journal: Clin Chim Acta ISSN: 0009-8981 Impact factor: 3.786
Characteristics of the chemiluminescent SARS-CoV-2 Antibodies assays investigated in this study, given by the manufacturers.
| Manufacturer | Roche Diagnostics | Diasorin Inc | Snibe diagnostics | Yhlo | |
|---|---|---|---|---|---|
| Anti-SARS-CoV-2 S | Liaison SARS-CoV-2 Trimeric-S IgG | SARS-CoV-2 S-RBD IgG | SARS-CoV-2 Neutralizing Antibody | iFlash-2019-nCov Neutralizing Antibody | |
| Cobas e analysers | LIAISON XL Analyzer | Maglumi series | Maglumi series | iFlash analysers | |
| Elettro-ChemiLuminescent (ECLIA) | Chemiluminescent immunoassay (CLIA) | Chemiluminescent immunoassay (CLIA) | Chemiluminescent immunoassay (CLIA) | Chemiluminescent immunoassay (CLIA) | |
| Total Antibodies (included IgG) | IgG antibodies | IgG antibodies | Neutralizing antibodies | Neutralizing antibodies | |
| Spike RBD portion | Trimeric Spike protein | Spike RBD portion | Spike RBD-ACE2 protein interaction | Spike RBD-ACE2 protein interaction | |
| kBAU/L | kBAU/L | kBAU/L | mg/L | kAU/L | |
| < 0.8 Negative | < 33.8 Negative | < 4.33 Negative | < 0.3 Negative | < 10 Negative | |
| 169 | 148 | 171 | 150 | 35 | |
| 376 | 331 | 386 | 343 | 104 | |
Calculated including samples at t0, t12 and t28.
Fig. 1Reciprocal of plaque reduction neutralization test results at low (PRNT50) or high (PRNT90) stringency thresholds and age (panels A and B) or subdivided by gender (C and D).
Fig. 2Dot-plots of PRNT50 and PRNT90 results (panels A and B) and Liaison SARS-CoV Trimeric-S IgG (panel C), Elecsys SARS-CoV-2 S-RBD Ab (panel D), Maglumi SARS-CoV-2 S-RBD IgG (E), Maglumi SARS-CoV-2 Neutralizing Ab (NAb) (F) and iFlash nCoV-2019 NAb (G), at different times from the first Comirnaty/BNT162b2 inoculum.
Fig. 3Comparison of PRNT50 (panels A, B, C, D, E) and PRNT90 (panels F, G, H, I, L) results and immunoassay results. A) and F), Liaison SARS-CoV Trimeric-S IgG; B) and G) Elecsys SARS-CoV-2 S-RBD Ab; C) and H) Maglumi SARS-CoV-2 S-RBD IgG; D) and I) Maglumi SARS-CoV-2 Neutralizing Ab (NAb); E) and L) iFlash nCoV-2019 NAb.
Fig. 4Dot-plots and pairwise comparison of PRNT50 and PRNT90 results and immunoassay results.
Correlation coefficients from multivariate linear regression analyses, obtained using CLIA assays, PRNT50 or PRNT90 results, age and gender.
| CLIA assays | PRNT50 correlation coefficient from multivariate analysis ( | PRNT90 correlation coefficient from multivariate analysis ( |
|---|---|---|
| Trimeric-S IgG, Liaison | 0.836 | 0.799 |
| S-RBD IgG, Maglumi | 0.820 | 0.828 |
| S-RBD Ab, Elecsys | 0.833 | 0.850 |
| hACE-S-RBD NAb, iFlash | 0.818 | 0.851 |
| ACE-S-RBD NAb, Maglumi | 0.839 | 0.872 |
Optimal thresholds identified by Youden index for each CLIA assays with respect to different PRNT50 levels by non-parametric ROC analyses. The best combination of Sensitivity and Specificity found by Youden index is also reported for any PRNT50 levels and CLIA assay.
| PRNT50 levels | CLIA assays | Optimal Threshold | AUC | Sensitivity/Specificity (%) |
|---|---|---|---|---|
| 20 | Trimeric-S IgG, Liaison | 535 kBAU/L | 0.962 | 87.0/93.0 |
| 40 | 606 kBAU/L | 0.980 | 90.0/97.0 | |
| 160 | 860 kBAU/L | 0.987 | 95.2/97.6 | |
| 320 | 1335 kBAU/L | 0.984 | 92.0/96.0 | |
| 20 | S-RBD IgG, Maglumi | 90 kBAU/L | 0.984 | 92.0/95.0 |
| 40 | 94 kBAU/L | 0.989 | 97.0/95.0 | |
| 160 | 360 kBAU/L | 0.990 | 96.0/96.0 | |
| 320 | 585 kBAU/L | 0.986 | 97.0/94.0 | |
| 20 | S-RBD Ab, Elecsys | 23 kBAU/L | 0.973 | 0.91/0.94 |
| 40 | 33 kBAU/L | 0.984 | 0.94/0.94 | |
| 160 | 205 kBAU/L | 0.990 | 0.98/0.95 | |
| 320 | 435 kBAU/L | 0.984 | 0.93/0.93 | |
| 20 | ACE-S-RBD NAb, iFlash | 18 kAU/L | 0.958 | 0.98/0.96 |
| 40 | 32 kAU/L | 0.981 | 0.91/0.95 | |
| 160 | 96 kAU/L | 1.000 | 1.00/1.00 | |
| 320 | 138 kAU/L | 1.000 | 1.00/1.00 | |
| 20 | ACE-S-RBD NAb, Maglumi | 0.27 mg/L | 0.981 | 0.93/0.95 |
| 40 | 0.38 mg/L | 0.989 | 0.95/0.96 | |
| 160 | 0.96 mg/L | 0.995 | 0.99/0.95 | |
| 320 | 1.90 mg/L | 0.995 | 0.96/0.98 |