Nick Stanojcic1,2, Christopher Hull3, David P S O'Brart1,2. 1. Department of Ophthalmology, 8945Guy's and St Thomas' NHS Foundation Trust, London, UK. 2. King's College London, London, UK. 3. School of Health Science, Division of Optometry and Visual Sciences, City University, London, UK.
We thank Drs. Kanclerz and Grzybowski for their interest in our work.In our review, we extensively discussed the published literature pertaining to the effects
of lens glistenings on visual function. We agree that forward light scatter from glistenings
will cause reduced image contrast However, the key question is whether the forward light
scatter is sufficient to cause disability glare. As noted by Drs Kanclerz and Grzybowski,
laboratory models have demonstrated correlations between glistenings density and size and
light scatter, including our own publication,
but clinical studies reporting visual impairment with glistenings occurrence are in
the minority and have limitations. Indeed, the retrospective cohort study of Henriksen et al.,
referenced in the letter, was a small pilot study, lacking methodology relating to
image acquisition and analysis, such as ambient illumination level and digital/analogue
camera settings, and with glistenings quantification and analysis undertaken using
operator-dependent, general purpose, image processing software (‘ImageJ’). In addition, the
authors stated themselves, that the study was underpowered to detect the true effect.We would like to direct Drs. Kanclerz and Grzybowski to our recent paper, undertaken to
further address such issues and which is referenced and discussed in the review.
In a prospective methodology we evaluated forward light scatter in a cohort of
patients with glistenings. Using a newly developed, defined, reproducible, standardized
8-point ordinal scale of glistenings density and an array of computerized visual function
tests (Advanced Vision and Optometric Tests, City Occupational, London, UK), performed under
strictly controlled ambient conditions and based on the same scientific principle as the
C-Quant test (Oculus, Optigeräte, Germany), we found no association between glistening
grades and visual function including straylight parameter or integrated straylight parameter.
This study we feel further adds to the evidence that glistenings, unless present in
extremely large amounts, have a minimal effect on visual performance in vivo.We do, however, recognise the limitations of light scatter testing that is currently
available to undertake such in vivo investigations. In their study Colin and Orignac could
only obtain valid results in about 50% of cases with the C-quant test
This same test was used by Henriksen et al. although no mention was made in their
paper of the percentage of successfully completed tests.
It is of note that in our own study.
approximately 20% of participants were not able to complete our light scatter test It
must also be recognized that evaluation of glistenings themselves and their visual
perturbations in vivo is much more challenging than in vitro. In our experience, the precise
quantification of glistenings in vivo has proven difficult and dependent on multiple factors
often not discussed in studies, such as ambient illumination, slit lamp parameters and image
acquisition parameters.In our review, we have not asserted that lens glistenings do not cause intraocular light
scatter, but that convincing in-vivo evidence does not yet exist The paper by Labuz et al.,
cited by Drs Kanclerz and Grzybowski, demonstrated that to produce a straylight value
equivalent to the 70-year-old crystalline lens would require 400 glistenings/mm2
of 15 micron diameter, a figure rising to 3000 glistenings/mm2 if they are of 5
micron diameter.
These levels are rarely seen in modern clinical practice and support the balance of
evidence that glistenings have a negligible effect on visual function at levels found in
modern IOLs.Regarding the comments concerning IOL calcification, Neuhann et al.
in their excellent paper, which is cited and discussed in our review, proposed three
possible routes for IOL calcification: primary calcification which is related to the IOL
itself (e.g. the polymer, manufacturing or packaging process); secondary, that is not only
dependent on the IOL but also associated with pre-existing disease, which may involve
breakdown of blood aqueous barrier; and false positive calcification or pseudo-calcification
that occurs due to misdiagnosis of tissue artefacts or incorrect use of special stains. We
disagree that we only reviewed the risk factors for secondary calcification; we discussed
and high-lighted in detail and cited what has been published in multiple publications and
hypothesised regarding IOL calcification relating to the IOL itself including issues with
polymers, manufacturing, and packaging.