Abstract
BACKGROUND: This study aims to report the results of a pioneering clinical study using the single-port transaxillary robotic thyroidectomy (START) for 200 patients with thyroid tumor and to introduce our novel two-step retraction method.Entities:
Keywords: Robotic single-port thyroidectomy; Robotic single-port transaxillary thyroidectomy; Robotic thyroidectomy; Robotic transaxillary thyroidectomy; START; da vinci SP
Mesh:
Year: 2021 PMID: 34741206 PMCID: PMC8921151 DOI: 10.1007/s00464-021-08837-9
Source DB: PubMed Journal: Surg Endosc ISSN: 0930-2794 Impact factor: 4.584
Fig. 1Two-step retraction operative procedure. A Robotic console view after the first docking. B Widening of the skin flap with the da Vinci SP. C Working space creation between the two heads of the SCM. D Finding thyroid beneath the strap muscles. E The second docking creating sufficient working space. F: Completion of the second docking
Clinical characteristics of patients (n = 200) and pathologic result of patients with thyroid cancer (n = 190)
| Variables | Mean ± SD | Range |
|---|---|---|
| Age (years) | 34.7 ± 7.9 | 13–58 |
| Sex ratio (male:female) | 2:198 | |
| Extent of thyroidectomy (lobectomy: BTT) | 177:23 | |
| Pathologic type (malignant:benign) | 190:10 | |
| BMI | 22.2 ± 3.7 | 15.9–37.0 |
| Postoperative hospital stay (days) | 3.0 ± 0.1 | 2–4 |
| Tumor size (cm) | 0.9 ± 0.8 | 0.2–5 |
| Pathology, | ||
| Papillary carcinoma | 187 (98.4) | |
| Follicular carcinoma | 3 (1.6) | |
| Multifocality, n (%) | 32 (16.6) | |
| Bilaterality, n (%) | 20 (10.4) | |
| Central lymph node, n | ||
| Retrieved | 3.6 ± 3.3 | 0–18 |
| Metastatic | 0.8 ± 1.3 | 0–6 |
T stage, n (%) T1/T2/T3/T4 | 171/13/5/1 | |
N stage N0/N1a | 117/73 | |
TNM stage, n (%) I/II/III/IV | 190/0/0/0 |
Differences in thyroid lobectomy approaches
| Lobectomy with conventional skin flap ( | Lobectomy with double-docking method ( | ||
|---|---|---|---|
| Total OP time (min) | 115.3 ± 17.3 | 116.7 ± 23.2 | 0.700 |
| Working space (min) | 34.1 ± 7.5 | 28.9 ± 10.9 | 0.007 |
| Docking (min) | 5.2 ± 1.7 | 3.8 ± 2.0 | 0.000 |
| Console (min) | 48.0 ± 14.2 | 60.0 ± 18.5 | 0.000 |
N number, OP operation
Fig. 2Cosmetic outcome 6 months after surgery
Fig. 3Various version of Chung’s retractors. Currently, the minimal sized Chung’s retractor (right) is used for two-step method
Fig. 4Comparison of the extent of skin flap dissection and the cosmetic outcome. A Comparison between conventional skin flap and SP skin flap before docking. B Widened SP skin flap after the 1st docking in the two-step method
Fig. 5Spherical working space required for the SP system