Aleš Fidler1,2, Andrej Steyer3, Damjan Manevski4, Rok Gašperšič5,6.
Abstract
BACKGROUND: It is well recognized that dental procedures represent a potential way of infection transmission. With the COVID-19 pandemic, the focus of dental procedure associated transmission has rapidly changed from bacteria to viruses. The aim was to develop an experimental setup for testing the spread of viruses by ultrasonic scaler (USS) generated dental spray and evaluate its mitigation by antiviral coolants.Entities:
Keywords: antimicrobial(s); infection control; microbiology; non-surgical periodontal therapy; virology
Mesh:
Substances:
Year: 2021 PMID: 34730843 PMCID: PMC8661989 DOI: 10.1002/JPER.21-0335
Source DB: PubMed Journal: J Periodontol ISSN: 0022-3492 Impact factor: 4.494
FIGURE 1Experimental set‐up for virus transmission evaluation. (A) The schematic diagram of the virus transmission tunnel (VTT). (B) Platform with ultrasonic scaler (USS) handpiece holder and groove, (C) semi‐cylindrical groove with USS tip, and blunt needle for virus suspension delivery, (D) Top view on the virus transmission tunnel, showing (1) USS device, (2) platform with USS handpiece and groove, (3) three consecutively positioned 48‐well cell culture plates, and (4) air sampler. (E) USS tip, positioned in the groove, during the experiment
Chemical characteristics of antiviral coolants, used in suspension test and virus transmission prevention test
| 0.25% NaOCl | EOW | |
|---|---|---|
| pH | 11.2 | 7.34 |
| Eh (mV) | 507 | 765 |
| Free chlorine (mg/L) | 220 | 22 |
Suspension test for the saline, 0.25% NaOCl and EOW coolants. The table presents the estimated mean log_10 TCID50/mL viral concentration at different time points (0, 5, 30, 60, 300, 600 seconds after the start of the experiment). Saline was evaluated only at 0 and 600 seconds
| 0 s | 5 s | 30 s | 60 s | 300 s | 600 s | |
|---|---|---|---|---|---|---|
| Saline | 6.45 | 6.56 | ||||
|
| 0.76 | |||||
| 0.25% NaOCl | 6.45 | <2.8 | <2.8 | <2.8 | <2.8 | <2.8 |
|
| 0.003 | 0.003 | 0.003 | 0.003 | 0.003 | |
| EOW | 6.45 | <2.8 | <2.8 | <2.8 | <2.8 | <2.8 |
|
| 0.003 | 0.003 | 0.003 | 0.003 | 0.003 |
T‐tests comparing the viral concentration at any time point to the one at time 0s have also been conducted. The corresponding P‐values are given for every coolant.
FIGURE 2The cytopathogenic and cytotoxic ef in settled droplet sample, obtained from three different coolants: saline, 0.25% NaOCl, and EOW. When a cytotoxic effect was observed, the cytopathogenic could not be evaluated