| Literature DB >> 34719801 |
Tomohisa Okada1, Shinya Handa2, Bill Ding2, Shin-Ichi Urayama1, Koji Fujimoto1, Atsushi Shima1, Daisuke Yoshii3, Takashi Ayaki3, Nobukatsu Sawamoto4, Ryosuke Takahashi3, Hirotaka Onoe1, Tadashi Isa1, Labros Petropoulos2.
Abstract
PURPOSE: To demonstrate the capability of insertable inductively coupled volumetric coils for MR microscopy in a human 7T MR system.Entities:
Keywords: connection-free; human 7 Tesla; inductive coupling; knee coil; microimaging; unplugged insertable coil
Mesh:
Year: 2021 PMID: 34719801 PMCID: PMC9297907 DOI: 10.1002/mrm.29062
Source DB: PubMed Journal: Magn Reson Med ISSN: 0740-3194 Impact factor: 3.737
FIGURE 1Views of insertable coil arrangements. (A) A D26 coil view with a trimmer capacitor and the other from the opposite side. (B) Two orthogonal views of the D64 coil. These limited views are provided, because only cupper elements exist in the views orthogonal to the presented ones of the D26 coil, and the opposite views of the D64 coil are the same
FIGURE 2Equivalent circuit diagrams of D26 (A) and D64 (B) coils. The D26 coil is a four‐turn solenoid coil (parts: C1, C2 = 9 pF, fixed capacitors [R15S series, Johanson Technology Inc., Camarillo, CA, US]; C3, C4 = 8.2 pF, C5–C9 = 4.3 pF, fixed capacitors [ATC100B series, American Technical Ceramics, Huntington Station, NY, US]; L1, L2 = 22 nH [Midi Spring Air Core Inductors, Coilcraft Inc., Cary, IL, US]; VC1: 2–6pF Trimmer Capacitor [JZ060HV, Knowles Voltronics, Cazenovia, NY, US]) and the D64 coil is a quadrature saddle coil (parts: C1–C4 = 14.2pF, C5–C8 = 11.1 pF, C9–C12 = 19 pF, C13–C20 = 14 pF, fixed capacitors [ATC100B series, American Technical Ceramics, Huntington Station, NY, US]; L1–L4 = 9.85 nH [Micro Spring Air Core Inductors, Coilcraft Inc., Cary, IL, US]). All diodes used in the D26 and D64 coils are fast MRI protection diode (UMX9989SM, Microsemi, Irvine, CA, US)
FIGURE 3(A) Relative SNR maps. SNR gains versus knee coil alone were around 12‐fold and 2.5‐fold at the center for D26 and D64 coils, respectively. (B) Parallel imaging factors and g‐factor maps. The parallel imaging factors are presented above the g‐factor maps. The directions of acceleration are also presented in parenthesis (LR: left‐right; AP: anterior‐posterior; and SI: superior‐inferior directions). Relatively small g‐factors were observed with the acceleration factor of 2, especially in LR and AP directions compared with SI direction. The parallel factor of 3 showed maximum g‐factor of around 3 or higher. The acceleration factor of 2 in two directions (2 × 2) had maximum g‐factors of less than 2
FIGURE 4Brain specimen of a neuronal intranuclear inclusion disease patient with many amyloid plaques. On the left are ‐weighted images of a brain specimen acquired using the D26 coil with isotropic 50 μm resolution in TR/TE/FA (ms/ms/degrees) of 200/24/24 (A) and 100/24/18 (B). Numerous tiny low‐signal dots are observed. Similar contrast was attained when TR was cut in half. (C) Amyloid‐β immunohistochemistry shows numerous amyloid plaques. (D) Iron staining shows iron deposition at the amyloid plaques. This iron deposition was considered to be the cause of the low‐signal dots. (A‐D) Scale bars: 1 mm
FIGURE 5A whole brain imaging of a macaque specimen acquired using the D64 solenoid coil in isotropic 160 μm resolution with the acceleration factors of 1, 2, 3, and 2 × 2 (from left to right). No foldover artifact was observed. With the acceleration factor of 2, the image is comparable to that without acceleration, but noise was slightly higher in 3 and 2 × 2 conditions, especially at the image center