Hemodynamic alternations following stent deployment and post-dilation in a heavily calcified coronary artery: In silico and ex-vivo approaches.

Peshala T Gamage1, Pengfei Dong2, Juhwan Lee3, Yazan Gharaibeh3, Vladislav N Zimin4, Luis A P Dallan4, Hiram G Bezerra5, David L Wilson3, Linxia Gu6.   

Abstract

In this work, hemodynamic alterations in a patient-specific, heavily calcified coronary artery following stent deployment and post-dilations are quantified using in silico and ex-vivo approaches. Three-dimensional artery models were reconstructed from OCT images. Stent deployment and post-dilation with various inflation pressures were performed through both the finite element method (FEM) and ex vivo experiments. Results from FEM agreed very well with the ex-vivo measurements, interms of lumen areas, stent underexpansion, and strut malapposition. In addition, computational fluid dynamics (CFD) simulations were performed to delineate the hemodynamic alterations after stent deployment and post-dilations. A pressure time history at the inlet and a lumped parameter model (LPM) at the outlet were adopted to mimic the aortic pressure and the distal arterial tree, respectively. The pressure drop across the lesion, pertaining to the clinical measure of instantaneous wave-free flow ratio (iFR), was investigated. Results have shown that post-dilations are necessary for the lumen gain as well as the hemodynamic restoration towards hemostasis. Malapposed struts induced much higher shear rate, flow disturbances and lower time-averaged wall shear stress (TAWSS) around struts. Post-dilations mitigated the strut malapposition, and thus the shear rate. Moreover, stenting induced larger area of low TAWSS (<0.4 Pa) and lager volume of high shear rate (>2000 s-1), indicating higher risks of in-stent restenosis (ISR) and stent thrombosis (ST), respectively. Oscillatory shear index (OSI) and relative residence time (RRT) indicated the wall regions more prone to ISR are located near the malapposed stent struts.
Copyright © 2021 Elsevier Ltd. All rights reserved.

Entities:  

Keywords:  Calcified artery; Computational fluid dynamics (CFD); Finite element method (FEM); Hemodynamics; Optical coherence tomography (OCT); Oscillatory shear index; Percutaneous coronary intervention (PCI); Post-dilation; Relative residence time; Restenosis; Shear rate; Stent expansion; Stent thrombosis; Wall shear stress

Mesh:

Year:  2021        PMID: 34715552      PMCID: PMC8642291          DOI: 10.1016/j.compbiomed.2021.104962

Source DB:  PubMed          Journal:  Comput Biol Med        ISSN: 0010-4825            Impact factor:   4.589


  68 in total

1.  Alterations in wall shear stress predict sites of neointimal hyperplasia after stent implantation in rabbit iliac arteries.

Authors:  John F LaDisa; Lars E Olson; Robert C Molthen; Douglas A Hettrick; Phillip F Pratt; Michael D Hardel; Judy R Kersten; David C Warltier; Paul S Pagel
Journal:  Am J Physiol Heart Circ Physiol       Date:  2005-01-14       Impact factor: 4.733

2.  Computational study of fluid mechanical disturbance induced by endovascular stents.

Authors:  Taewon Seo; Levanto G Schachter; Abdul I Barakat
Journal:  Ann Biomed Eng       Date:  2005-04       Impact factor: 3.934

3.  Computational analysis of the radial mechanical performance of PLLA coronary artery stents.

Authors:  R G Pauck; B D Reddy
Journal:  Med Eng Phys       Date:  2014-10-19       Impact factor: 2.242

4.  Computed high concentrations of low-density lipoprotein correlate with plaque locations in human coronary arteries.

Authors:  Ufuk Olgac; Joseph Knight; Dimos Poulikakos; Stefan C Saur; Hatem Alkadhi; Lotus M Desbiolles; Philippe C Cattin; Vartan Kurtcuoglu
Journal:  J Biomech       Date:  2011-07-01       Impact factor: 2.712

Review 5.  How Do We Treat Complex Calcified Coronary Artery Disease?

Authors:  Paul N Fiorilli; Saif Anwaruddin
Journal:  Curr Treat Options Cardiovasc Med       Date:  2016-12

6.  Coronary revascularization trends in the United States, 2001-2008.

Authors:  Andrew J Epstein; Daniel Polsky; Feifei Yang; Lin Yang; Peter W Groeneveld
Journal:  JAMA       Date:  2011-05-04       Impact factor: 56.272

7.  Three-dimensional computational fluid dynamics modeling of alterations in coronary wall shear stress produced by stent implantation.

Authors:  John F LaDisa; Ismail Guler; Lars E Olson; Douglas A Hettrick; Judy R Kersten; David C Warltier; Paul S Pagel
Journal:  Ann Biomed Eng       Date:  2003-09       Impact factor: 3.934

8.  Prevalence of uncontrolled risk factors for cardiovascular disease: United States, 1999-2010.

Authors:  Cheryl D Fryar; Te-Ching Chen; Xianfen Li
Journal:  NCHS Data Brief       Date:  2012-08

9.  Hemodynamics in Idealized Stented Coronary Arteries: Important Stent Design Considerations.

Authors:  Susann Beier; John Ormiston; Mark Webster; John Cater; Stuart Norris; Pau Medrano-Gracia; Alistair Young; Brett Cowan
Journal:  Ann Biomed Eng       Date:  2015-07-16       Impact factor: 3.934

10.  Optical Coherence Tomography-Based Modeling of Stent Deployment in Heavily Calcified Coronary Lesion.

Authors:  Pengfei Dong; Hozhabr Mozafari; David Prabhu; Hiram G Bezerra; David L Wilson; Linxia Gu
Journal:  J Biomech Eng       Date:  2020-05-01       Impact factor: 2.097

View more
  1 in total

1.  Upper Airway Flow Dynamics in Obstructive Sleep Apnea Patients with Various Apnea-Hypopnea Index.

Authors:  Shengmao Lin; Thyagaseely Sheela Premaraj; Peshala T Gamage; Pengfei Dong; Sundaralingam Premaraj; Linxia Gu
Journal:  Life (Basel)       Date:  2022-07-19
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.