Taeho Kim1, Zhen Ji1, Benjamin Lewis1, Eric Laugeman1, Alex Price1, Yao Hao1, Geoffrey Hugo1, Nels Knutson1, Bin Cai1,2, Hyun Kim1, Lauren Henke1.
Abstract
PURPOSE: Ethos adaptive radiotherapy (ART) is emerging with AI-enhanced adaptive planning and high-quality cone-beam computed tomography (CBCT). Although a respiratory motion management solution is critical for reducing motion artifacts on abdominothoracic CBCT and improving tumor motion control during beam delivery, our institutional Ethos system has not incorporated a commercial solution. Here we developed an institutional visually guided respiratory motion management system to coach patients in regular breathing or breath hold during intrafractional CBCT scans and beam delivery with Ethos ART.Entities:
Keywords: Ethos; adaptive radiotherapy; respiratory motion management; visual guidance
Mesh:
Year: 2021 PMID: 34697865 PMCID: PMC8803298 DOI: 10.1002/acm2.13441
Source DB: PubMed Journal: J Appl Clin Med Phys ISSN: 1526-9914 Impact factor: 2.102
FIGURE 1Components of the visually guided respiratory motion management system: (a) respiratory motion sensor and interface, (b) in‐room visual display system, and (c) visual guidance displayed on the Ethos bore; the components are: (1) air bag, (2) air tube, (3) pressure sensor, (4) projector, (5) adjustable stand, (6) extension screen, and (7) visual guidance display
FIGURE 2Components of respiratory motion trace management systems: (a) Logger Pro and (b & c) in‐house visual guidance software with (b) wave mode and (c) bar mode; the components are: (1) respiratory motion trace of the wave mode, (2) guiding window (2a: upper boundary and 2b: lower boundary), (3) patient information, (4) display setup menu, (5) display mode selection, (6) respiratory motion trace of the bar mode, and (7) breath‐hold timer
Summary of respiratory motion management techniques for three patients undergoing Ethos adaptive radiotherapy; P1 includes the number of EBHs
| Fx1 | Fx2 | Fx3 | Fx4 | Fx5 | |
|---|---|---|---|---|---|
| P1 | VG‐EBH (21) | VG‐EBH (18) | VG‐EBH (16) | VG‐EBH (15) | VG‐EBH (13) |
| P2 | VG‐SB | VG‐SB | VG‐EBH/VG‐SB | VG‐EBH/VG‐SB | VG‐EBH/VG‐SB |
| P3 | VG‐EBH/VG‐FB | VG‐EBH/VG‐FB | VG‐EBH/VG‐FB | VG‐EBH/VG‐FB | VG‐EBH/VG‐FB |
Abbreviations: P, patient; VG‐EBH, visual guidance with exhalation breath hold; VG‐FB, visual guidance with free breathing; VG‐SB, visual guidance with shallow breathing.
FIGURE 3Example of respiratory trace analyses: (a) breath holds of P1 and (b) aligned (red curves) and average (yellow curve) free‐breathing traces of P3; (c) dose map on cone‐beam computed tomography (CBCT) and (d) dose volume histogram of an adapted plan for P1 are presented (dotted lines from the reference plan and solid lines from the adapted plan)
Summary of respiratory motion traces for three patients treated with Ethos adaptive radiotherapy; P1 includes the respiratory amplitude of the average respiratory cycle in kPa, breath‐hold amplitude variation normalized by the respiratory amplitude of the average respiratory cycle (normalized breath‐hold amplitude variation), and minimum breath‐hold time in seconds. P2 and P3 include the respiratory amplitude of the average respiratory cycle in kPa, respiratory amplitude variation normalized by the respiratory amplitude of the average respiratory cycle (normalized respiratory amplitude variation), and respiratory period in seconds
| Fx1 | Fx2 | Fx3 | Fx4 | Fx5 | Mean ± Std | |
|---|---|---|---|---|---|---|
| P1 | 0.44 kPa | 0.22 kPa | 0.22 kPa | 0.16 kPa | 0.11 kPa | 0.23 ± 0.13 kPa |
| 1.9 ± 1.6 (%) | 6.4 ± 5.5 (%) | 5.4 ± 4.2 (%) | 7.8 ± 6.1 (%) | 9.9 ± 7.5 (%) | 6.3 ± 3.0 (%) | |
| 17.7 s | 23.7 s | 20.7 s | 21.8 s | 23.5 s | 21.5 ± 2.4 s | |
| P2 | 0.17 kPa | 0.16 kPa | 0.069 kPa | 0.15 kPa | 0.16 kPa | 0.14 ± 0.04 kPa |
| 31 ± 28 (%) | 22 ± 25 (%) | 28 ± 25 (%) | 18 ± 13 (%) | 26 ± 22 (%) | 25 ± 5 (%) | |
| 3.9 ± 0.94 s | 3.1 ± 0.65 s | 3.5 ± 0.78 s | 3.3 ± 0.34 s | 3.6 ± 0.72 s | 3.5 ± 0.3 s | |
| P3 | 0.1 kPa | 0.037 kPa | 0.044 kPa | 0.14 kPa | 0.14 kPa | 0.09 ± 0.05 kPa |
| 15 ± 12 (%) | 24 ± 18 (%) | 21 ± 15 (%) | 20 ± 13 (%) | 16 ± 13 (%) | 19 ± 4 (%) | |
| 4.3 ± 0.47 s | 3.6 ± 0.50 s | 4.1 ± 0.75 s | 3.7 ± 0.46 s | 3.9 ± 0.55 s | 3.9 ± 0.3 s |
Summary of intrafractional cone‐beam computed tomography (CBCT) shifts (lateral, vertical, and longitudinal in centimeters) from three patients treated with Ethos adaptive radiotherapy; the first verification CBCT shifts (shaded) were applied after adapted plan generation on the initial CBCT. – Indicates no CBCT
| Shift# | Fx1 | Fx2 | Fx3 | Fx4 | Fx5 | |
|---|---|---|---|---|---|---|
| P1 | 1a | 0.33, −0.28, 0.07 | 1.27, −0.45, −0.21 | 0.22, 0, 0 | 0.32, 0, 0.08 | 0.61, 0, 0 |
| 1b | – | 0, 0, 0 | – | 0.2, 0.36, −1.03 | – | |
| 2 | 0.56, −0.43, 1.4 | −0.52, 0.33, 0.15 | 0, 0, 0 | −0.15, −0.05, 0.5 | −0.26, 0, 0 | |
| 3 | 0, 0, 0 | 0.3, −0.15, −0.27 | 0, 0, 0 | 0, 0, 0 | 0, 0, 0 | |
| P2 | 1 | 0, 0, 0 | 0, 0, 0 | 0, 0, 0 | 0, 0, 0 | 0, 0, 0 |
| 2 | 0, 0, 0 | 0, 0, 0 | 0, 0, 0 | 0, 0, 0 | 0, 0, 0 | |
| P3 | 1a | 0, 0, 0 | 0, 0, 0 | 0, 0, 0 | 0, 0, 0 | −0.06, 0.06, 0 |
| 1b | 0, 0, 0 | – | – | – | – | |
| 2 | 0, 0, 0 | 0, 0, 0 | 0, 0, 0 | 0, 0, 0 | 0, 0, 0 |
FIGURE 4Intrafractional shifts of Patient 1 during Ethos ART: (a) 3D shift magnitude and (b) longitudinal shifts for Ethos ART with breath holds; 3D shift magnitude = (x = lateral, y = vertical, and z = longitudinal shift in cm)