| Literature DB >> 34659825 |
Iveta Nemcova1,2, Jiri Pasta1,2, Katerina Hladikova1,2, Martin Komarc2, Darina Pospisilova1, Pavel Nemec1,2, Jan Tesar1,2, Vladimir Kratky2,3, Martin Sin1,2.
Abstract
PURPOSE: To evaluate a 12-year follow-up of myopic patients after iris-fixated phakic intraocular lenses (IF pIOLs) implantation. Setting. Ophthalmology Department, Military University Hospital in Prague (Czech Republic).Entities:
Year: 2021 PMID: 34659825 PMCID: PMC8516571 DOI: 10.1155/2021/7027793
Source DB: PubMed Journal: J Ophthalmol ISSN: 2090-004X Impact factor: 1.909
Baseline characteristics of patients.
| Parameters at baseline | Mean ± SD [range] |
|---|---|
| No. of eyes | 85 |
| No. of patients | 46 |
| Age | 29.3 ± 5.5 [18; 43] |
| Gender male/female (%) | 28/72 |
| Sphere (D) | 9.37 ± 2.87 [−16.75; −2] |
| Cylinder (D) | 1.37 ± 1.44 [0; 6,75] |
| CDVA, Snellen decimal scale | 0.91 ± 0.17 [0.4; 1.13] |
| IOP (mmHg) | 14.8 ± 3.0 [8.5; 22.5] |
| ACD from endothelium (mm) | 3.30 ± 0.23 [2.75; 3.77] |
| Mean keratometry (D) | 43.8 ± 1.6 [39.4; 47.1] |
| Mean axial length (mm) | 26.99 ± 1.27 [24.99; 31.39] |
| Mean corneal pachymetry ( | 521 ± 31,95 [453; 585] |
| ECD | 2588 ± 285 [1481; 3215] |
|
|
|
| Verisyse (Abbott) | 40 |
| Verisyse Toric (Abbott) | 17 |
| Veriflex (Abbott) | 28 |
SD: standard deviation; No.: number; D: dioptres; CDVA: corrected distance visual acuity; IOP: intraocular pressure; ACD: anterior chamber depth; ECD: endothelial cells density.
Refractive results.
| Time after the pIOL implantation | No. of eyes | Sphere (D) |
| Cylinder (D) |
|
|---|---|---|---|---|---|
| 6 months | 80 | 0.14 ± 0.61 [−1.5; 1] | <0.001 | 0.70 ± 0.48 [0; 2] | 0.001 |
| 1 year | 75 | 0.04 ± 0.67 [−2.0; 1.5] | 0.139 | 0.64 ± 0.48 [0; 2] | 0.623 |
| 2 years | 61 | −0.06 ± 0.67 [−2.25; 1.25] | 0.019 | 0.72 ± 0.67 [0; 2] | 0.304 |
| 5 years | 60 | −0.34 ± 0.84 [−3.5; 1.25] | 0.001 | 0.86 ± 0.51 [0; 2] | 0.019 |
| 12 years | 42 | −0.42 ± 1.08 [−5.5; 1.50] | 0.024 | 0.87 ± 0.55 [0; 2.5] | 0.425 |
No.: number; D: dioptres; p value: probability value.
Postoperative visual acuity results (Snellen decimal scale).
| Time after the pIOL implantation | No. of eyes | UDVA |
| CDVA |
|
|---|---|---|---|---|---|
| 6 months | 80 | 0.94 ± 0.15 [0.55; 1.25] | 0.040 | 0.98 ± 0.12 [0.65; 1.25] | 0.001 |
| 1 year | 75 | 0.95 ± 0.15 [0.50; 1.25] | 0.666 | 0.99 ± 0.10 [0.70; 1.25] | 0.281 |
| 2 years | 61 | 0.93 ± 0.13 [0.40; 1.00] | 0.972 | 0.99 ± 0.07 [0.75; 1.25] | 0.729 |
| 5 years | 60 | 0.93 ± 0.17 [0.30; 1.10] | 0.972 | 1.00 ± 0.07 [0.8; 1.25] | 0.020 |
| 12 years | 42 | 0.86 ± 0.21 [0.30; 1.05] | 0.005 | 1.00 ± 0.06 [0.9; 1.25] | 0.063 |
IF pIOL: iris-fixated phakic intraocular lens; UDVA: uncorrected distance visual acuity; CDVA: corrected distance visual acuity.
Figure 1Cumulative distance visual acuity.
Efficacy and safety indices.
| Time (years after the implantation) | 1 | 2 | 5 | 12 |
|---|---|---|---|---|
| Efficacy indices (EI) | 1.06 | 1.03 | 1.03 | 0.96 |
| Mean postoperative UDVA/mean preoperative CDVA | — | — | — | — |
| Safety indices (SI) | 1.10 | 1.10 | 1.10 | 1.10 |
| Mean postoperative CDVA/mean preoperative CDVA | — | — | — | — |
UCVA: uncorrected distance visual acuity; CDVA: corrected distance visual acuity.
Figure 2Change in corrected distance visual acuity.
Intraocular pressure.
| Time after the IF pIOL implantation | No. of eyes | IOP (mmHg) |
|---|---|---|
| 0 months | 85 | 14.8 ± 3.0 [8.5; 22.5] |
| 1 months | 82 | 14.5 ± 2.4 [11.5; 22.0] |
| 2 years | 61 | 14.8 ± 2.8 [10.0; 22.5] |
| 5 years | 60 | 15.2 ± 3.0 [9.0; 21.5] |
IF pIOL: iris-fixated phakic intraocular lens; No.: number; IOP: intraocular pressure.
Figure 3Mean endothelial cell density (ECD, cells/mm2) from the preoperative status to 12-year postoperative status in eyes implanted with iris-fixated phakic intraocular lenses.
Endothelial cell density after the IF pIOL Implantation.
| Time after the IF pIOL implantation | Number of eyes | ECD (cells/mm2) | EC loss (%) | Corrected EC loss (%) |
|---|---|---|---|---|
| 1 year | 75 | 2430 ± 312 [1421; 3221] | 6.6 | 6.0 |
| 2 years | 61 | 2369 ± 262 [1715; 3009] | 9.3 | 8.1 |
| 5 years | 60 | 2175 ± 298 [1286; 2937] | 15.8 | 12.8 |
| 12 years | 42 | 2091 ± 312 [1196; 2674] | 19.1 | 11.9 |
IF pIOL: iris-fixated phakic intraocular lens; No. number; ECD: endothelial cells density; EC loss: endothelial cells loss; endothelial cells loss corrected to physiological endothelial cells loss.