A Novel Endoscopic Mouthpiece for COVID-19 Prevention.

Takuto Hikichi1, Nakamura Jun1, Koichi Hamada2, Daiki Nemoto3.   

Abstract

Entities:  

Year:  2021        PMID: 34649419      PMCID: PMC8831421          DOI: 10.5946/ce.2021.172

Source DB:  PubMed          Journal:  Clin Endosc        ISSN: 2234-2400


× No keyword cloud information.
During the present coronavirus disease 2019 (COVID-19) pandemic, gastrointestinal endoscopy is regarded as a risky procedure because virus transmission can occur via an airborne route. To reduce the spread of aerosolized droplets, several devices have been developed [1-7]. This novel mouthpiece consists of a conventional mouthpiece made of polypropylene and a hat-shaped aerosol droplet guard, which are connected together. (Fig. 1A and B). The aerosol droplet guard, which is made of a thermoplastic elastomer, has a width of 90 mm and a height of 50 mm, and can prevent aerosolized droplets from being ejected from the mouth (Fig. 1C). The center of the aerosol droplet guard has four slits that are 6 mm long and are oriented to allow the scope to pass through them. (Fig. 1D), while also acting as valves to prevent aerosolized droplet diffusion.
Fig. 1.

Appearance of the new mouthpiece. (A, B) Side view. The mouthpiece is connected to a hat-shaped aerosol droplet guard. (C, D) Front view. The guard prevents aerosolized droplets from exiting the mouth when the patient coughs. The red arrow shows the slit that the scope is passed through.

The effectiveness of the new mouthpiece in preventing aerosolized droplet diffusion was compared to that of a conventional mouthpiece. As an experimental model, simulated saliva was sprayed from the mouth of a mannequin using a cough machine. The simulated saliva was prepared by adding 76 g of glycerin to 1 L of water [8]. With the endoscope inserted into the mouthpiece, saliva was sprayed in the following conditions: duration, 0.2 sec; velocity, 10.7 m/sec; volume, 75.4 mL; and pressure, 0.08 MPa [8]. Three technical repeats were completed, and the diffusion of the aerosolized droplets was photographed by a camera in a darkened room with a Polarion light. Although there was significant aerosolized droplet diffusion when using the conventional mouthpiece (Fig. 2A), it was only minimally detectable with the new mouthpiece (Fig. 2B).
Fig. 2.

A comparison of the spread of aerosolized droplets during coughing. (A) Significant droplet diffusion was observed when using a conventional mouthpiece. (B) Significant reduction in the amount of diffused droplets when using the new mouthpiece.

The aerosol droplet guard not only prevents diffusion in the area from where the scope is inserted, but also blocks diffusion from the gap between the mouthpiece and the mouth, which is a distinct feature not found in conventional mouthpieces. As the new mouthpiece has not yet been applied clinically, its effect on respiratory and circulatory dynamics during endoscopy is still unknown. However, when used in a virtual clinical setting, there was no apparent breathing discomfort (Fig. 3A and B).
Fig. 3.

The virtual clinical application of the mouthpiece. (A) Front view. The hat-shaped aerosol droplet guard is larger than a person’s mouth. (B) The aerosol-droplet guard is located 16.5 mm from the mouth and therefore does not affect respiration.

In conclusion, this new mouthpiece can protect clinicians from infection when carrying out endoscopic procedures, especially in the wake of the COVID-19 pandemic.
  7 in total

1.  Commercially available endoscopy facemasks to prevent aerosolizing spread of droplets during COVID-19 outbreak.

Authors:  Alejandro Bojórquez; Francisco Javier Zozaya Larequi; Maria Teresa Betés; José Carlos Súbtil; Miguel Muñoz-Navas
Journal:  Endosc Int Open       Date:  2020-06-02

2.  Shielding for patients using a single-use vinyl-box under continuous aerosol suction to minimize SARS-CoV-2 transmission during emergency endoscopy.

Authors:  Hideki Kobara; Noriko Nishiyama; Tsutomu Masaki
Journal:  Dig Endosc       Date:  2020-06-03       Impact factor: 7.559

3.  Endoscopic shield: barrier enclosure during the endoscopy to prevent aerosol droplets during the COVID-19 pandemic.

Authors:  Ryota Sagami; Hidefumi Nishikiori; Takao Sato; Kazunari Murakami
Journal:  VideoGIE       Date:  2020-05-11

4.  "Double-surgical-mask-with-slit" method: reducing exposure to aerosol generation at upper gastrointestinal endoscopy during the COVID-19 pandemic.

Authors:  Nikolaos Lazaridis; Alexandros Skamnelos; Alberto Murino; Rocio Chacchi Cahuin; Nikolaos Koukias; Edward J Despott
Journal:  Endoscopy       Date:  2020-09-23       Impact factor: 10.093

5.  Novel device for preventing diffusion of aerosol droplets from subjects undergoing esophagogastroduodenoscopy during COVID-19 pandemic.

Authors:  Hiroyuki Endo; Tomoyuki Koike; Atsushi Masamune
Journal:  Dig Endosc       Date:  2020-07-22       Impact factor: 7.559

6.  Use of a Glove-Covered Mouthpiece during Upper Endoscopy to Prevent COVID-19 Transmission.

Authors:  Sho Sasaki; Jun Nishikawa; Isao Sakaida
Journal:  Clin Endosc       Date:  2020-08-21

7.  Coronavirus disease outbreak: a simple infection prevention measure using a surgical mask during endoscopy.

Authors:  Hirotsugu Maruyama; Akira Higashimori; Kei Yamamoto; Akinobu Nakata; Yuki Ishikawa-Kakiya; Masafumi Yamamura; Yasuhiro Fujiwara
Journal:  Endoscopy       Date:  2020-08-20       Impact factor: 10.093

  7 in total
  2 in total

1.  Simple and innovative methods of minimizing risk of aerosol generation during endoscopy.

Authors:  Sumitro Kosasih; Adli Metussin; Norwani Basir; Vui Heng Chong
Journal:  Clin Endosc       Date:  2022-05-11

2.  A Novel Aerosol-Exposure Protection Mask for Patients During Upper Endoscopy.

Authors:  Soo-Jeong Cho
Journal:  Clin Endosc       Date:  2022-03-29
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.