Developing an ultra-performance liquid chromatography-tandem mass spectrometry for detecting aldosterone in human plasma.

Wenbin Lin1, Zhenrong Yao1, Yuzhe Li2, Zhihao Liao2, Jiahao Xiao3, Yonghong Chen4, Pinning Feng1, Wenjia Gan1.   

Abstract

BACKGROUND: Accurately measuring plasma aldosterone concentration is difficult but meaningful for primary aldosteronism (PA) diagnosis.
METHODS: In this study, we developed an ultra-performance liquid chromatography-tandem mass spectrometry (UPLC-MS/MS) method for plasma aldosterone detection, evaluated its performance according to guidelines issued by CLSI, including detection limit, linearity, precision, and compared it with chemiluminescence immunoassay. Then, a reference range of plasma aldosterone in young people was established by using this method.
RESULTS: The lower limit of quantitation (LOQ) was 10 pg/ml. The mean recovery rates of analyte added to serum were 100.07-102.05% in different concentrations. The linearity range was 20-2000 pg/ml. Inter-assay CVs were 2.20-3.97% at aldosterone concentrations of 65.66-854.75 pg/ml. The regression equation of UPLC-MS/MS (x) and chemiluminescence immunoassay (y) was y = 1.002x + 65.854 (r = 0.9456, n = 237). The reference range of plasma aldosterone detected by UPLC-MS/MS was 11.30-363.82 pg/ml in young people in South China, and there was no statistically significant difference in plasma aldosterone concentration between two genders.
CONCLUSION: In conclusion, UPLC-MS/MS can rapidly and accurately detect plasma aldosterone and is appropriate for clinical application.
© 2021 The Authors. Journal of Clinical Laboratory Analysis published by Wiley Periodicals LLC.

Entities:  

Keywords:  UPLC-MS/MS; aldosterone; performance; plasma; primary aldosteronism

Mesh:

Substances:

Year:  2021        PMID: 34590736      PMCID: PMC8605128          DOI: 10.1002/jcla.24029

Source DB:  PubMed          Journal:  J Clin Lab Anal        ISSN: 0887-8013            Impact factor:   2.352


INTRODUCTION

Autonomously excessive aldosterone secretion which is caused by aldosterone‐producing adenoma and adrenal cortical hyperplasia is the etiology of primary aldosteronism (PA). PA is mainly prevalent in secondary hypertension, which accounts for about 10–30% of refractory hypertension in different studies. , , , Too much aldosterone causes not only high blood pressure through sodium and water retention but also organ damage through inflammation and fibrosis. So, PA patients are more likely to suffer from cardiovascular complications, such as ventricular hypertrophy, cerebral infarction, arrhythmia, etc. Fortunately, surgical resection and aldosterone antagonists can effectively treat PA when the etiology is assured. Hence, plasma aldosterone measurement is significant for PA, even hypertension, diagnosis. Aldosterone is extremely difficult to be accurately measured because of its trace and numerous analogs in plasma. At present, conventional tests based on radioimmunoassay and chemiluminescence immunoassay (CLIA) are unsatisfactory due to their defects in safety and accuracy, because radioimmunoassay uses radioisotope‐labeled antibody to detect aldosterone, and CLIA does not distinguish structural analogs of aldosterone in plasma in some cases. Mass spectrum has great advantages in trace detection and anti‐interference, so gas or liquid chromatography‐tandem mass spectrometry (GC/LC‐MS) has been applied in aldosterone detection. , Sample needs complex derivatization before GC‐MS detection, which makes it difficult for clinical application. Compared with GC‐MS, LC‐MS requires only simple sample pretreatment, therefore, aldosterone measurements based on LC‐MS will be popular in clinical laboratories. In this study, we developed an aldosterone detection method based on ultra‐performance liquid chromatography‐tandem mass spectrometry (UPLC‐MS/MS). We determined its methodological characteristics and a reference interval in young people by using this method. Besides, correlation and bias between this method and our daily‐used CLIA method were also analyzed. Our results indicate UPLC‐MS/MS has extremely high accuracy in aldosterone detection and requires simple sample pretreatment, which is suitable for clinical laboratory. Our study is one of the few studies on the detection of aldosterone by using UPLC‐MS/MS. , It lays foundation for clinical application of UPLC‐MS/MS in aldosterone detection.

MATERIALS AND METHODS

Recruitment of healthy volunteers

This study was approved by the medical ethics committee of the hospital (batch number: [2018]054), and the agreement from every participant was obtained. A total of 237 healthy volunteers were recruited in the First Affiliated Hospital of Sun Yat‐sen University. Inclusion criteria were (1) normal hypertension (systolic pressure: 90–140 mmHg; diastolic pressure: 60–90 mmHg, 1 mmHg = 0.133 kPa); (2) no obviously cardiac, liver, and renal dysfunction; (3) no drug intake in recent three months, including (a) diuretics, such as spironolactone, amiloride, aminophenylpyridine, etc.; (b) central nervous system drugs, such as clonidine, α‐methyldopa, etc.; (c) non‐steroidal anti‐inflammatory drugs; (d) anti‐hypertensive drugs, such as angiotensin‐converting enzyme inhibitors, angiotensin receptor antagonists, renin inhibitors, dihydropyridine calcium channel antagonists; (4) no smoking; (5) no pregnancy; (6) no recent surgery; (7) no hormone replacement therapy; (8) no family history of hypertension and diabetes.

Sample collection

Before sample collection, participants were asked to have a full‐night sleep (>8 h). After that, participants were asked to stand or walk for 2 h and seat for 15 min. Then, blood sample in upright position was collected between 9:00 and 10:00 AM. Specimen was collected in EDTA‐K2 anticoagulant tube. Plasma was separated by centrifugation at room temperature, 3000 g for 5 min. Then, plasma was split into Eppendorf tubes and store at −20°C before use. Another blood sample from the same volunteer was collected in a procoagulant tube at the same time to check the baseline characteristics, such as blood sodium and potassium concentrations, liver and kidney function, etc.

Sample detection

Pretreatment was needed before UPLC‐MS/MS detection. 15 μl internal standard (10 ng/ml d4‐aldosterone, Sigma‐Aldrich) was added into 200 μl plasma for each sample. Then, 200 μl of 0.1 M zinc sulfate solution‐methanol (v/v, 5:5) and 450 μl of 0.05% phosphoric acid (v/v) were added. Sample was vortexed for 90 sec and centrifuged at 18,000 g for 10 min. 200 μl methanol and pure water were respectively used to active SPE column (Oasis MAX, 30 μm). Then, 625 μl supernatant was added into the SPE column. The column was washed with 200 μl of 0.05% phosphoric acid (v/v), 0.1% ammonia methanol‐water (v/v, 1:9), and pure water, respectively. Aldosterone was washed by 50 μl methanol‐water solution (v/v, 7:3) before UPLC‐MS/MS detection. The UPLC‐MS/MS system consisted of a Waters ACQUITYTM UPLC system and a Waters XevoTM TQ‐S tandem triple quadrupole mass spectrometer. 20 μl eluent was injected into UPLC system. C18 column (CORTECS, 2.1*100 mm, 1.6 μm particles) was used to separate. The mobile phase consisted of methanol and pure water. The flow rate was 0.3 ml/min. Elution gradients are shown in Table 1. Parameters for mass spectrometer setup and detected ion pairs are shown in Tables 2 and 3. The ion source of the mass spectrometer worked in the negative ion model.
TABLE 1

Elution gradients

GradientTime (min)Methanol (%)Water (%)
103070
20.53070
33.57030
44.5955
55955
65.53070
783070
TABLE 2

Mass spectrometer setup parameters

ParametersValue
Capillary voltage2.5kV
Source temperature150°C
Desolvation temperature600°C
TABLE 3

Detected ion‐pairs

Compound nameParent (m/z)Daughter (m/z)
ALD359189
ALD359297
d4‐ALD363190

Abbreviation: ALD, aldosterone.

Elution gradients Mass spectrometer setup parameters Detected ion‐pairs Abbreviation: ALD, aldosterone.

Methodological evaluation

Limit of detection (LOD) and limit of quantitation (LOQ)

Methodological evaluation and comparison were performed according to guidelines published by the Clinical and Laboratory Standards Institute (CLSI), including C50, C62‐A, C28‐A2, and CLIA'88. For the sensitivity evaluation, a series of low aldosterone concentration samples were repeatedly tested. LOD was defined as signal to noise ratio (S/N) >3 and coefficient of variation (CV) <20%. LOQ was defined as S/N >10 and CV <20%. Each sample was tested six times.

Linearity

A series of aldosterone standards with different concentrations (20–2000 pg/ml) which contained internal standards were prepared. Each sample was tested twice, and average peak area ratio of aldosterone to internal standard was calculated. Linear regression analysis was done by using aldosterone standard concentrations (x‐axis) and peak area ratios of aldosterone to internal standard (y‐axis), and R 2 should be over 0.99.

Recovery

50, 200, and 1000 pg/ml aldosterone were added into blood sample with known aldosterone concentration. Each sample was tested three times. Recovery concentration was calculated by the detected concentration subtracting the known concentration. Recovery rate was defined as the percentage of the recovery concentration to the added concentration.

Precision

A total of samples with different aldosterone concentrations (65.66, 234.14, and 854.75 pg/ml) were used. Each sample was measured twice at a time, 6 times a day for three consecutive days. CVs were calculated and compared with TEa (14.7%) given by CLIA’88. The CV of intra‐assay precision should be no more than a quarter of TEa, and the CV of inter‐assay precision should be no more than a third of TEa.

Methodological comparison

A total of 237 samples were detected by UPLC‐MS/MS and CLIA, respectively. Bias and regression coefficient were analyzed by Bland‐Altman analysis.

Reference range establishment

Aldosterone in plasma collected from 193 healthy volunteers (21–30 years) was measured by UPLC‐MS/MS as previously described. Statistical difference was tested in different genders. Reference range was determined by percentile method or mean ± 2 times standard deviation (SD).

Statistical analysis

SPSS v23.0 (IBM) was used in the whole statistical analysis. Aldosterone concentrations were shown in mean ± SD or indicated. CV was calculated by dividing SD by mean. Normal distribution was checked by Kolmogorov‐Smirnov test. Outliers were excluded by Mahalanobis distance method. Data with normal distribution were analyzed by the Student t test and the Pearson correlation analysis to detect statistical difference and correlation between methods and genders. If the data were non‐normal distribution, the Mann–Whitney U test and the Spearmen correlation analysis were used. p < 0.05 was defined as statistically significant difference.

RESULTS AND DISCUSSION

Method validation

Selectivity

First, we determined the selectivity of our method. A chromatogram graphical of aldosterone is shown in Figure 1. The chromatographic retention times of aldosterone and internal standard were about 2.5 min both, and no obvious abnormal peak was observed next to aldosterone and internal standard, which showed high selectivity of our method.
FIGURE 1

A chromatogram graphical of aldosterone

A chromatogram graphical of aldosterone

LOD, LOQ, and linearity

After detecting a series of low aldosterone concentration samples, we found that the LOD for aldosterone detected by the developed UPLC‐MS/MS method was 5 pg/ml, and the LOQ was 10 pg/ml. Then, we determined the linearity by detecting a serious of aldosterone standards with different concentrations. We found the linearity range was 20–2000 pg/ml. The linear regression equation between the peak area ratio (y) and the aldosterone concentration (x) was y = 0.0104861x + 0.00268166 with an R 2 = 0.9993, which met the methodological requirement. Compared with previously reported MS methods for aldosterone detection, we found the LOQ of our method was comparable with previously published UPLC‐MS/MS methods (10 and 10.83 pg/ml), , while the linearity range was smaller than Tan et al. reported method (10–5000 pg/ml) but larger than Hinchliffe et al. reported method (0–1153 pg/ml). Then, we compared our developed method with other reported LC‐MS methods. Our method was comparable with them in LOD/LOQ, but had a larger linearity range and used less sample volume, which may be more suitable for clinical use. ,

Recovery and precision tests

Recovery rates of the developed method were determined as 100.07%, 102.05%, and 101.95% after repeatedly measuring samples added with three different concentrations of aldosterone. For precision evaluation, results were shown in Table 4. The maximum intra‐assay and inter‐assay CVs were 2.75% and 3.97%, respectively, which appeared at the low aldosterone concentration. For medium and high aldosterone concentrations, intra‐assay and inter‐assay CVs were much less than those for the low concentration. Results of the recovery and precision tests all meet the requirements of the C62‐A and CLIA’88 and are better than other reported LC‐MS/MS methods, , , which means the developed method has a satisfactory repeatability and reproducibility.
TABLE 4

Results of precision evaluation

SampleTimesMean (pg/ml)Intra‐run CV (%)Inter‐run CV (%)
A3665.662.753.97
B36234.142.703.39
C36854.751.702.20
Results of precision evaluation For methodological comparison, the same batch of plasma samples were detected by the UPLC‐MS/MS and CLIA methods respectively, and Mann‐Whitney U test was used to compare the results, because both aldosterone concentrations detected by UPLC‐MS/MS and CLIA were non‐normal distribution. There was a statistically significant difference between the results of two methods (p < 0.01). The linear regression equation of UPLC‐MS/MS (x) and CLIA (y) detected aldosterone concentrations was y = 1.002x + 65.854 with an R 2 = 0.8941. The average bias between aldosterone concentrations detected by these two methods was −66.07 pg/ml, which is consistent with previous findings that there is an obvious negative bias between LC‐MS/MS methods and immunoassays. The SD of the bias was 31.27. There were 10 samples that exceeded the 95% consistency limit (Figure 2). These results indicate that the correlation between the two methods is good, although there is a fixed bias between them. These results lead us to believe that the developed UPLC‐MS/MS method is sensitive and accurate, which might be suitable for clinical application.
FIGURE 2

Methodological comparison between UPLC‐MS/MS and CLIA. (A) Scatter plot of aldosterone concentrations was detected by two specified methods. Dotted line: the line of identity. (B) Bland‐Altman plot with method difference expressed as absolute concentration. Solid line: −4.78, −66.07, and −127.36 pg/ml (from top to bottom)

Methodological comparison between UPLC‐MS/MS and CLIA. (A) Scatter plot of aldosterone concentrations was detected by two specified methods. Dotted line: the line of identity. (B) Bland‐Altman plot with method difference expressed as absolute concentration. Solid line: −4.78, −66.07, and −127.36 pg/ml (from top to bottom)

Reference range of UPLC‐MS/MS detected aldosterone concentration

A total of 193 young volunteers (21–30 years old) met the inclusion criteria. Plasma aldosterone concentrations of the volunteers were detected by the UPLC‐MS/MS method. The aldosterone concentrations were also non‐normal distribution in this section. Three outliers were eliminated from the results based on the Mahalanobis distance method. 72 cases were male in the rest of cases. Before establishing the reference range, we examined the correlation between plasma aldosterone concentration and gender by using Mann–Whitney U test. We found that there was no correlation between them, which is in accordance with our previous study. So, we established a uniform reference range for both male and female. The reference range of UPLC‐MS/MS‐detected plasma aldosterone concentration was determined by 2.5% and 97.5% percentile values, which was 11.30–363.82 pg/ml. Although the lower limit of the reference range is similar to that reported in other studies, the upper limit of the reference range in our study is larger than that reported in other studies. , , We think it may be due to different sample sources and collecting method. Next, we will collect samples from other age groups, identify the correlation between aldosterone concentration and age, and establish a reference range in a wider age range.

CONCLUSION

We developed an aldosterone detection method based on UPLC‐MS/MS and evaluated its methodological characteristics. We think the aldosterone detection method based on UPLC‐MS/MS is suitable for clinical application. We also used this method to establish a reference range of plasma aldosterone in young people.

CONFLICT OF INTERESTS

The authors declare that they have no competing interests.
  22 in total

1.  Prevalence of primary hyperaldosteronism in resistant hypertension: a retrospective observational study.

Authors:  Stella Douma; Konstantinos Petidis; Michael Doumas; Panagiota Papaefthimiou; Areti Triantafyllou; Niki Kartali; Nikolaos Papadopoulos; Konstantinos Vogiatzis; Chrysanthos Zamboulis
Journal:  Lancet       Date:  2008-06-07       Impact factor: 79.321

2.  Development and validation of a method using supported liquid extraction for aldosterone determination in human plasma by LC-MS/MS.

Authors:  Cécile Meunier; Dominique Blondelle; Patrice Faure; Jean-Philippe Baguet; Caroline Le Goff; Olivier Chabre; Véronique Ducros
Journal:  Clin Chim Acta       Date:  2015-05-13       Impact factor: 3.786

3.  High efficiency of the aldosterone-to-renin ratio in precisely detecting primary aldosteronism.

Authors:  Wenjia Gan; Wenbin Lin; Jianjun Ouyang; Yuzhe Li; Dubo Chen; Zhenrong Yao; Pinning Feng
Journal:  J Hum Hypertens       Date:  2018-09-21       Impact factor: 3.012

4.  Quantitation and clinical evaluation of plasma aldosterone by ultra-performance liquid chromatography-mass spectrometry.

Authors:  Xin Tan; Fengjuan Li; Xue Wang; Yuan Wang
Journal:  J Chromatogr A       Date:  2019-08-13       Impact factor: 4.759

5.  Prevalence of primary aldosteronism and association with cardiovascular complications in patients with resistant and refractory hypertension.

Authors:  Mirko Parasiliti-Caprino; Chiara Lopez; Nunzia Prencipe; Barbara Lucatello; Fabio Settanni; Giuseppe Giraudo; Denis Rossato; Giulio Mengozzi; Ezio Ghigo; Andrea Benso; Mauro Maccario
Journal:  J Hypertens       Date:  2020-09       Impact factor: 4.844

6.  Prevalence of and risk factors for primary aldosteronism among patients with resistant hypertension in China.

Authors:  Xiaojing Sang; Yiran Jiang; Weiqing Wang; Li Yan; Jiasheng Zhao; Yongde Peng; Wei Gu; Gang Chen; Wei Liu; Guang Ning
Journal:  J Hypertens       Date:  2013-07       Impact factor: 4.844

7.  Measurement of serum aldosterone in picomolar level by LC-MS/MS using charge-tagged technique.

Authors:  Bonnie Mei-Wah Fong; Tak-Shing Siu; Sidney Tam
Journal:  Anal Bioanal Chem       Date:  2015-09-15       Impact factor: 4.142

8.  The Unrecognized Prevalence of Primary Aldosteronism: A Cross-sectional Study.

Authors:  Jenifer M Brown; Mohammed Siddiqui; David A Calhoun; Robert M Carey; Paul N Hopkins; Gordon H Williams; Anand Vaidya
Journal:  Ann Intern Med       Date:  2020-05-26       Impact factor: 25.391

9.  Reference intervals for plasma concentrations of adrenal steroids measured by LC-MS/MS: Impact of gender, age, oral contraceptives, body mass index and blood pressure status.

Authors:  Graeme Eisenhofer; Mirko Peitzsch; Denise Kaden; Katharina Langton; Christina Pamporaki; Jimmy Masjkur; George Tsatsaronis; Anastasios Mangelis; Tracy A Williams; Martin Reincke; Jacques W M Lenders; Stefan R Bornstein
Journal:  Clin Chim Acta       Date:  2017-05-04       Impact factor: 3.786

10.  Developing an ultra-performance liquid chromatography-tandem mass spectrometry for detecting aldosterone in human plasma.

Authors:  Wenbin Lin; Zhenrong Yao; Yuzhe Li; Zhihao Liao; Jiahao Xiao; Yonghong Chen; Pinning Feng; Wenjia Gan
Journal:  J Clin Lab Anal       Date:  2021-09-30       Impact factor: 2.352