Plasma membrane topography has been shown to strongly influence the behavior of many cellular processes such as clathrin-mediated endocytosis, actin rearrangements, and others. Recent studies have used three-dimensional (3D) nanostructures such as nanopillars to imprint well-defined membrane curvatures (the "nano-bio interface"). In these studies, proteins and their interactions were probed by two-dimensional fluorescence microscopy. However, the low resolution and limited axial detail of such methods are not optimal to determine the relative spatial position and distribution of proteins along a 100 nm-diameter object, which is below the optical diffraction limit. Here, we introduce a general method to explore the nanoscale distribution of proteins at the nano-bio interface with 10-20 nm precision using 3D single-molecule super-resolution (SR) localization microscopy. This is achieved by combining a silicone-oil immersion objective and 3D double-helix point spread function microscopy. We carefully adjust the objective to minimize spherical aberrations between quartz nanopillars and the cell. To validate the 3D SR method, we imaged the 3D shape of surface-labeled nanopillars and compared the results with electron microscopy measurements. Turning to transmembrane-anchored labels in cells, the high quality 3D SR reconstructions reveal the membrane tightly wrapping around the nanopillars. Interestingly, the cytoplasmic protein AP-2 involved in clathrin-mediated endocytosis accumulates along the nanopillar above a specific threshold of 1/R (the reciprocal of the radius) membrane curvature. Finally, we observe that AP-2 and actin preferentially accumulate at positive Gaussian curvature near the pillar caps. Our results establish a general method to investigate the nanoscale distribution of proteins at the nano-bio interface using 3D SR microscopy.
Plasma membrane topography has been shown to strongly influence the behavior of many cellular processes such as clathrin-mediated endocytosis, actin rearrangements, and others. Recent studies have used three-dimensional (3D) nanostructures such as nanopillars to imprint well-defined membrane curvatures (the "nano-bio interface"). In these studies, proteins and their interactions were probed by two-dimensional fluorescence microscopy. However, the low resolution and limited axial detail of such methods are not optimal to determine the relative spatial position and distribution of proteins along a 100 nm-diameter object, which is below the optical diffraction limit. Here, we introduce a general method to explore the nanoscale distribution of proteins at the nano-bio interface with 10-20 nm precision using 3D single-molecule super-resolution (SR) localization microscopy. This is achieved by combining a silicone-oil immersion objective and 3D double-helix point spread function microscopy. We carefully adjust the objective to minimize spherical aberrations between quartz nanopillars and the cell. To validate the 3D SR method, we imaged the 3D shape of surface-labeled nanopillars and compared the results with electron microscopy measurements. Turning to transmembrane-anchored labels in cells, the high quality 3D SR reconstructions reveal the membrane tightly wrapping around the nanopillars. Interestingly, the cytoplasmic protein AP-2 involved in clathrin-mediated endocytosis accumulates along the nanopillar above a specific threshold of 1/R (the reciprocal of the radius) membrane curvature. Finally, we observe that AP-2 and actin preferentially accumulate at positive Gaussian curvature near the pillar caps. Our results establish a general method to investigate the nanoscale distribution of proteins at the nano-bio interface using 3D SR microscopy.
The cell-to-material
interface
is often a key determinant of successful applications for tissue engineering
and biomedical implants. Material properties, including chemical functionalization,
surface topography, and bulk stiffness, collectively set instructive
signals for cell behavior.[1,2] Nanoscale surface topography
is particularly interesting as it is widely tunable and has been shown
to significantly affect cellular responses. In the past decades, nanofabrication
emerged as a powerful tool to precisely engineer nanostructures, i.e., the nano–bio interface, to control cell behavior.
For example, nanopillars have been shown to reduce focal adhesions
and membrane tension;[3,4] nanogratings and nanofibers induce
cell alignment and neural development,[5,6] and nanopores
accelerate stem cell differentiation.[7,8] Nanopillars
made of different materials were also developed into electrical and
optical sensors for measuring live cell activities.[9,10] Due
to its biological significance, there is great interest in visualizing
the nano–bio interface, especially with respect to the membrane
shape around nanotopography and intracellular proteins at the interface.
By varying the diameter of the nanopillars, which changes the shape
of the membrane at the interface, recent studies show that nanopillars
locally activate clathrin-mediated endocytosis and the polymerization
of actin fibers in a curvature-dependent manner.[11,12] However, the visualization of the membrane shape at the interface
and the quantification of the curvature values are technically challenging,
especially in three dimensions (3D).Ultrastructural analysis
by electron microscopy remains a powerful
approach for visualizing the nano–bio interface. Transmission
electron microscopy (TEM) has been used to visualize the membrane
shape around nanopillars and to measure the gap distance at the interface.
Focused-ion-beam milling and scanning electron microscopy (FIB-SEM)
imaging can be more advantageous by allowing selective opening and
imaging of the interface at desired locations.[13]However, as is well-known, electron microscopy gives
a snapshot
of the physical effects with very high spatial resolution and general
cellular context but without specific identification of nondiscernable
proteins, while fluorescence microscopy gives molecular specificity
without going beyond the optical diffraction limit (DL) of 250–300
nm. Recent work in correlated single-molecule localization and cryogenic
electron tomography has shown important progress in combining these
two modalities.[14,15] Previous studies have used fluorescence
2D DL experiments to study the behavior of specific proteins at the
nano–bio interface. However, the analysis of the nanoscale
distribution of the proteins around and along the entire nanopillar
in 3D and at 10–20 nm resolution is not possible from the low
resolution 2D DL images. While some experiments have probed the nano–bio
interface at higher resolutions than the DL, more information would
be obtained about the nanoscale distribution along the pillars with
10–20 nm resolution in 3D.[16,17] To address
this optically, we describe a method to use super-resolution (SR)
fluorescence microscopy in three dimensions (3D) to precisely explore
the positions of probe molecules down to ∼10 nm in cells interacting
with fabricated quartz nanopillars. Our approach required optical
optimization and the use of control imaging tests to validate the
procedures. We show that membrane-anchored labels as well as proteins
that preferentially accumulate on curved membranes such as AP-2 and
actin interact with the curvature constraints of the nanopillar substrate
in different ways. This work shows the utility of the 3D SR approach
and should stimulate further use of the method to quantitatively characterize
the nano–bio interface.
Results and Discussion
Nanopillar Imaging Strategy
to Mitigate Spherical Aberrations
Recent studies used a conventional-oil
immersion objective (CIO)
in 2D DL fluorescence imaging experiments to image the behavior of
cytoplasmic proteins such as AP-2 and actin on quartz nanopillars.[11,12] Moving from 2D DL to 3D SR imaging on the quartz nanopillar substrate
is not straightforward. The CIO is optimized for imaging through glass
rather than quartz substrates because the refractive index of the
immersion oil matches the glass refractive index (n = 1.52 for CIO and glass). As a result, minimal refraction is experienced
by light at the oil–glass interface (see Figure S1A for a sketch of the usual configuration). In general,
any refraction from index mismatch between the coverslip and the sample
will lead to spherical aberrations that deteriorate image quality.
Specifically, the problem at hand can be affected by the refractive
index mismatches between the cell (n ∼ 1.40),
the quartz substrate and nanopillars (n = 1.45),
and the immersion oil (n = 1.52). Due to the limited
spatial resolution of 2D DL images, the effect on image quality or
the resolution is not very significant or observable in most images
of the interface. However, the 3D localization precision and localization
accuracy will degrade rapidly in the presence of spherical aberrations
and will deteriorate the final SR reconstruction image quality.[18−24]To mitigate spherical aberrations arising from imaging cells
on quartz nanopillar substrates with 3D SR,
we repurposed a silicone-oil immersion objective (SIO) with a correction
collar in order to more carefully address the refractive index mismatches.
Similar to water immersion (WI) objectives that index match to aqueous
media with a carefully chosen glass substrate thickness, the SIOs
strive to use the approximate match between the silicone immersion
oil (n = 1.40) and the index in the cell cytoplasm.
As a result of the objective design, the optical path mismatch induced
by a glass coverslip can be reduced by carefully adjusting the correction
collar (see Figure S1B for a schematic
of a cell imaged with the usual configuration for a SIO). The SIO
has been employed successfully in various microscopy methods.[25,26] Here, we show that the adjustable SIO can also work well with the
cells-on-quartz nanopillar 3D SR imaging problem.Figure A depicts
the situation when using a SIO compared to the CIO for a simple case
where a cell is imaged on a flat quartz coverslip instead of glass.
With the CIO, light refracts between the cell cytoplasm and the coverslip
and between the coverslip and oil (Figure A left). Ultimately, the wavefront distortion
from the refraction results in the spherical aberrations[27] that degrade localization precision and accuracy.
However, with the calibrated SIO, the light experiences minimal refraction
at the interfaces (Figure A right), and there is less wavefront distortion, less spherical
aberration, and higher image quality. The only issue here is whether
the SIO can be corrected sufficiently for our quartz substrate, given
the fact that the objective design assumes glass coverslips.
Figure 1
Comparison
of SIO and CIO. (A) Cartoon schematic depicting how
light rays refract using the CIO (left) and the SIO on quartz (Qtz)
substrates. (B). Images of 200 nm beads immobilized in 5% agarose
on quartz coverslips imaged with both objectives (top). Calibration
bar depicts ADC counts on the EMCCD camera. Cross sections (orange
and blue lines) are fit to a Gaussian (bottom). The full width at
half maximum (purple double arrow) is clearly smaller for the case
with the SIO (C) DL (left) and 2D SR reconstructions (middle and right)
of FBP-17 labeled cells imaged with the CIO and SIO grown on glass
and quartz substrates, respectively. Calibration bar depicts the number
of localizations in each bin of the histogram reconstruction (bin
size 32 nm). Orange and blue boxes are magnified images of the 2D
SR reconstructions. White arrows show examples of tubule invaginations.
Comparison
of SIO and CIO. (A) Cartoon schematic depicting how
light rays refract using the CIO (left) and the SIO on quartz (Qtz)
substrates. (B). Images of 200 nm beads immobilized in 5% agarose
on quartz coverslips imaged with both objectives (top). Calibration
bar depicts ADC counts on the EMCCD camera. Cross sections (orange
and blue lines) are fit to a Gaussian (bottom). The full width at
half maximum (purple double arrow) is clearly smaller for the case
with the SIO (C) DL (left) and 2D SR reconstructions (middle and right)
of FBP-17 labeled cells imaged with the CIO and SIO grown on glass
and quartz substrates, respectively. Calibration bar depicts the number
of localizations in each bin of the histogram reconstruction (bin
size 32 nm). Orange and blue boxes are magnified images of the 2D
SR reconstructions. White arrows show examples of tubule invaginations.The correction collar adjustment of the SIO is
first carefully
tested on a 200 μm thick quartz coverslip, a thickness similar
to that used to fabricate nanopillars in later experiments. To do
this, we imaged 200 nm poly(styrene) fluorescent beads immobilized
in 5% agarose on 200 μm thick quartz coverslips at various collar
adjustments, and the extent of spherical aberrations at each adjustment
was assessed by quantifying the peak intensity of the bead image spots.
Spherical aberrations will reduce peak intensity because the photon
distribution is spread over a larger spatial scale.[28,29] The peak intensity of the observed spot is a common metric used
to assess the extent of spherical aberrations present in the image
and has been used in approaches such as adaptive optics.[30,31] We set the correction collar to the adjustment with the maximum
peak intensity (Methods and Figure S1C). We confirmed this by calibrating the SIO on several
coverslips and observed that the adjustment was always the same. This
adjustment was used for all imaging experiments involving the 200
μm thick quartz substrates.Figure B depicts
representative beads on 200 μm thick quartz substrates imaged
with both the CIO and SIO. The bead imaged with the SIO has a more
tightly focused photon distribution and thus higher peak intensity
compared to the bead imaged with the CIO. Horizontal cross sections
fit to a Gaussian (plots below bead images) also show that the full
width at half maximum is smaller for the bead imaged with the SIO.We conclude that fewer spherical aberrations are present, leading
to an expected superior performance with the SIO compared to the CIO.
The quantification of the spherical aberrations over many beads imaged
with both objectives further confirms the approach (Figure S1D).Imaging on quartz substrates with a SIO
can then be extended to
cellular SR imaging. We first chose to verify that the SR image quality
from cells adhering to quartz substrates using the SIO are comparable
to the quality from standard imaging approaches: imaging cells on
glass substrates with the CIO (see Figure A right and S1A for both configurations). We overexpressed the protein Formin-binding
protein 17 (FBP17) in U2OS cells and grew cells on glass and quartz
substrates. FBP17 is a Bin/amphiphysin/Rvs (BAR) domain protein, a
class of proteins that are banana-shaped, bind preferentially to regions
of membrane curvature, and can also induce membrane curvature.[32−34] When overexpressed in cells, FBP17 forms tubule invaginations that
are not resolvable using DL imaging.[35] The
ability to resolve the tubule invaginations will be used to compare
the image quality of both imaging modalities.FBP17 was expressed
with a green fluorescent protein (GFP) domain
fusion and was labeled with GFP nanobodies covalently labeled with
Alexa Fluor 647 (AF647, Figure S1E for
controls). 2D Stochastic Optical Resolution Microscopy (STORM,[36] dSTORM[37]) SR data
was acquired by imaging fixed cells close to the coverslip with high
laser intensity in a blinking buffer. We also call this type of SR
microscopy SMACM for single-molecule active control microscopy[38] as a more general term for any mechanism that
forces the concentration of emitting molecules in single frames down
to a sparse level allowing for single-molecule localizations and subsequent
SR reconstruction. The data was then processed to yield 2D SR reconstructions. Figure C (first column)
depicts the DL images of labeled FBP17 cells in both objective/substrate
combinations. Due to the diffraction limit, the features of the invaginations
are not easily apparent. The second and third columns show the 2D
SR reconstructions of the cells first on a large scale and then as
magnified images, respectively. The features of the tubule invaginations
in the SR reconstructions are much more clearly observed and, qualitatively,
the invaginations appear similar in both situations. In addition,
the median localization precision is 10 nm (Figure S1F for distributions) for both cases, and quantification of
the diameters of the invaginations (Figure S1G) is similar for both configurations and is in close agreement to
the literature.[39,40] Thus, as the image quality is
both qualitatively and quantitatively similar for both imaging configurations,
2D SR imaging of cells on quartz substrates with the silicone immersion
objectives retains the image quality found using standard SR imaging
approaches.
3D SR Microscopy of Surface-Labeled Nanopillars
Using a Double-Helix
PSF Microscope
Next, the imaging configuration combining
a quartz substrate and SIO must be adapted for 3D SR imaging of the
nano–bio interface. Even with standard open-aperture widefield
microscopy, the shape of the detected single-molecule spots, regarded
here as the point spread function (PSF), changes as a function of
defocus. However, the extraction of the 3D position from the shape
changes is challenging. The shape of the standard PSF is symmetric
above and below the focus, resulting in the potential redundancy of
the Z-position. In addition, the shape quickly blurs
400 nm away from the focus in both directions, and the determination
of Z requires high signal-to-noise.[41] As cells and the nanopillars may extend in the axial direction
several μm, the relatively short imaging range of ∼800
nm is not desirable.To circumvent these issues, we used PSF
engineering approaches to more optimally extract the Z-position, as described in many previous studies.[42−44] In PSF engineering,
we insert a simple transmissive phase mask in the Fourier plane (conjugate
to the back focal plane) of the microscope. As the Fourier plane is
usually found close to the back of the objective in many microscopes,
it can be challenging to access. Thus, we used simple 4f emission
processing optics outside the microscope to relay the collected fluorescence
light from the usual image plane to a new image plane four focal lengths
away. Now, there is easy access to the Fourier plane as illustrated
in Figure A. The phase
mask imparts a phase delay in the collected fluorescence light that
modulates the shape of the PSF after the light is focused on the camera
detector. Here, we chose to insert a double-helix phase mask[42] in the Fourier plane of our microscope (phase
pattern Figure A inset),
which has been used in previous 3D SR imaging experiments with cells.[22,45,46] The modified PSF now has two
lobes and is termed the double-helix PSF (DHPSF).[42] The axial range of this DHPSF design is 2 μm; the
shape is asymmetric above and below the focus, and it changes rapidly
with Z, circumventing the issues of the standard
PSF for 3D imaging and facilitating precise Z-position
estimation. The XY position is extracted by fitting
the DHPSF spot on the camera (see Figure D for examples) to a double-Gaussian function
and finding the midpoint of the fit. As the angle between the lobes
rotates as a function of the Z-position of the emitter,
a carefully calibrated curve measured prior to data acquisition connects
the lobe angle to the Z-position. We note that no
scanning of the objective or the stage occurs during data acquisition.
The focus is simply set to one position, and we acquire DHPSF images
of all emitters over the entire axial range of the DHPSF and extract
the Z-position in post-processing. This localization
estimation procedure has been shown to provide localization errors
independent of Z, as opposed to other approaches
to 3D such as astigmatism.[44]
Figure 2
Schematic of
microscope and representative images of single molecules
imaged with the DHPSF. (A) Cartoon depiction of our DHPSF microscope.
The excitation light entering the objective in the epifluorescence
configuration is used to image surface-labeled nanopillars. The emission
is relayed through 4f emission optics onto the camera detector. The
DHPSF mask (mask shown by arrow) is inserted in the Fourier plane
for 3D imaging. Calibration bar in units of radians. (B) Representative
field of view of experimental data showing three emitters. As the
lobe angle for all three emitters is varied, the Z-positions are different. (C) This plot depicts the calibration curve
that correlates the lobe angle to the Z-position.
(D) Selected images reveal experimental DHPSFs extracted from a SMACM
data set of the surface-labeled nanopillars. The emitters were fit
to extract the Z-position (Z-position
shown in top right of image).
Schematic of
microscope and representative images of single molecules
imaged with the DHPSF. (A) Cartoon depiction of our DHPSF microscope.
The excitation light entering the objective in the epifluorescence
configuration is used to image surface-labeled nanopillars. The emission
is relayed through 4f emission optics onto the camera detector. The
DHPSF mask (mask shown by arrow) is inserted in the Fourier plane
for 3D imaging. Calibration bar in units of radians. (B) Representative
field of view of experimental data showing three emitters. As the
lobe angle for all three emitters is varied, the Z-positions are different. (C) This plot depicts the calibration curve
that correlates the lobe angle to the Z-position.
(D) Selected images reveal experimental DHPSFs extracted from a SMACM
data set of the surface-labeled nanopillars. The emitters were fit
to extract the Z-position (Z-position
shown in top right of image).The shape of the DHPSF is highly sensitive to aberrations in the
microscope. With the extreme index mismatch of the quartz–water
interface, imaging with the CIO produces aberrations so severe that
the shape of the DHPSF degrades substantially (Figure S2A left). The two lobes are no longer visible and
cannot be localized for precise 3D position estimation. In contrast,
the SIO mitigates these aberrations and maintains the expected two
lobe shape (Figure S2A right). Thus, using
the SIO and a DHPSF microscope, in principle, we can now image proteins
in 3D near the quartz nanopillars. However, a certain degree of spherical
aberration may still be present and may potentially be Z-dependent. For instance, the light closer to the top of the nanopillars
experiences more refraction than the light near the bottom of the
pillars. Therefore, we chose a model system to first benchmark the
image quality of our 3D SR reconstructions. We will compare the dimensions
and shapes extracted from 3D SR reconstructions of covalently surface-labeled
nanopillars and from scanning electron microscopy (SEM) images of
the nanopillars. If the image quality is high, the shapes and dimensions
of the 3D SR reconstructions should be similar to the shapes and dimensions
from the SEM images.Fabrication of the nanopillars is described
in the next paragraph,
so we continue with 3D SR DHPSF imaging of covalently labeled quartz
nanopillars here. To optically image these pillars, we first surface-labeled
them by modifying the surface with free amine groups and then covalently
attaching AF647 fluorophores using NHS chemistry with low nonspecific
binding (Methods and Figure S2D for controls). We then collected many frames containing
emitters bound to many pillars and within the axial range of the DHPSF. Figure B is a representative
small field of view (FOV) in a frame of our acquired data where we
see three emitters with various lobe angles and, thus, different Z-positions. Figure C depicts the calibration curve that connects the lobe angle
to the Z-position of a poly(styrene) fluorescent
bead on the surface; this was acquired by moving to known Z-positions using a precise motorized piezo stage prior
to SMACM data acquisition. Using this curve and double-Gaussian fits,
we extract the 3D position of the emitter along the nanopillars for
all the blinking single molecules from the camera image stack. Figure D showcases selected
emitters at different Z-positions. As the heights
of the pillars are 884 ± 72 nm (mean ± standard deviation),
the entire axial range of the DHPSF is not fully utilized here. The
total axial range of the emitters in Figure D covers the entire height of the pillar
selected.The other crucial aspect of our experiments is the
fabrication
of the quartz nanopillars, which follows previous work with some modifications.
One modification utilized nanopillars fabricated with photolithography
and chemical etching methods rather than electron beam lithography,
in order to easily add a design containing a reference marker in an
array pattern (Figure A, described in Methods and Figure S2B). The reference marker allowed a specific correlation
between a pillar in the 3D SR reconstructions and the same pillar
in the SEM image. To reduce the diameter of the nanopillars to below
the optical diffraction limit, we employed wet etching after the dry
etching step. We used buffer oxide etchant for the quartz substrate,
which provides a generally isotropic process to remove material from
the substrate and shrink material to a given dimension. This fabrication
process resulted in tapering elliptical pillars (see Figure S2C for details on the height and diameter measurements)
that have an indentation or dimple close to the coverslip. The tapering
stems from differences in the rate of chemical etching along the pillar,
while the dimple is a residual from the photolithography step of the
fabrication process. All nanopillar substrates used in the experiments
described later are fabricated with this same protocol and have similar
dimensions (see Figure S2C for details
on the height and diameter measurements).
Figure 3
Comparison between 3D
SR reconstructions of surface-labeled nanopillars
and SEM images. (A) SEM images of a patterned array of nanopillars.
The reference marker (E9) is clearly visible in the top left image
of the nanopillars with a 30° tilt. The magnified images to the
right of the image show an individual nanopillar; top with the tilt
and bottom a top-down view. The dimple at the coverslip and elliptical
shape of the pillars are clearly visible. (B) Top left is a cartoon
depiction demonstrating imaging surface-labeled nanopillars. Below,
the bright puncta in an array of surface-labeled nanopillars is visible
in the DL image. Right: Two orientations of the 3D SR reconstructions
show an array of nanopillars. Color encodes Z-position.
CS in calibration bar refers to the position of the coverslip. (C)
Magnified images of an individual pillar (∗ in B) at various
orientations. (D) 100 nm Z-slices of the pillar (C)
from the bottom of the coverslip to the top of the nanopillar. Clear
elliptical-shaped rings are visible. (E) 250 nm Z-slices of the 3D data shown as an XY projection
at the center of an individual nanopillar fit to an ellipse. The orange
dashed line is one axis extracted from the fit and may be compared
to diameters extracted from SEM images. (F) Histogram depicts the
difference in diameter between the 3D SR reconstructions and the SEM
images of the nanopillars.
Comparison between 3D
SR reconstructions of surface-labeled nanopillars
and SEM images. (A) SEM images of a patterned array of nanopillars.
The reference marker (E9) is clearly visible in the top left image
of the nanopillars with a 30° tilt. The magnified images to the
right of the image show an individual nanopillar; top with the tilt
and bottom a top-down view. The dimple at the coverslip and elliptical
shape of the pillars are clearly visible. (B) Top left is a cartoon
depiction demonstrating imaging surface-labeled nanopillars. Below,
the bright puncta in an array of surface-labeled nanopillars is visible
in the DL image. Right: Two orientations of the 3D SR reconstructions
show an array of nanopillars. Color encodes Z-position.
CS in calibration bar refers to the position of the coverslip. (C)
Magnified images of an individual pillar (∗ in B) at various
orientations. (D) 100 nm Z-slices of the pillar (C)
from the bottom of the coverslip to the top of the nanopillar. Clear
elliptical-shaped rings are visible. (E) 250 nm Z-slices of the 3D data shown as an XY projection
at the center of an individual nanopillar fit to an ellipse. The orange
dashed line is one axis extracted from the fit and may be compared
to diameters extracted from SEM images. (F) Histogram depicts the
difference in diameter between the 3D SR reconstructions and the SEM
images of the nanopillars.
Comparing SEM Images and 3D SR Reconstructions
With
the fabricated and labeled nanopillars and the 3D DHPSF SR imaging
in hand, the two imaging modalities may be directly compared as shown
in Figure . Figure B depicts the DL
and 3D SR reconstructions of the same nanopillar array in Figure A. The DL image shows
bright spots in the same array pattern as the SEM image. As the molecules
along the shaft of the fluorescently labeled 3D nanopillar will all
project onto a 2D image corresponding to the focal depth of ∼700–800
nm, the apparent density of the fluorophore at the nanopillars will
be higher compared to the coverslip. Thus, the bright spots in the
DL image correspond to the positions of labeled nanopillars.At the right of Figure B, the surface-labeled nanopillars were then imaged with the DHPSF
3D microscope. Any emitters that were poorly localized (see Methods) were filtered and removed. Further, we
merged molecules to correct for overcounting (Methods and Figure S3A). After filtering, the
median XY localization precision is 12 nm, while
the median Z precision is 27 nm; both precisions
are an order of magnitude better than that possible in DL imaging
(see Figure S3B for histograms of localization
precisions). The XY projection of the 3D SR reconstructions
(Figure B first of
two right images) show an array pattern of regions containing a high
density of localizations showing the location of the nanopillars.
The color scale encodes the Z-position in the reconstruction.
As the coverslip is also labeled, the teal color in the reconstruction
indicates the location of the coverslip (see Figure S3C for an additional 3D SR reconstruction). We have also shifted
the Z-positions such that Z = 0
at the coverslip to provide an experimentally relevant reference plane.
All 3D SR reconstructions described later include this shift.Critically, the array patterns of the nanopillars from the DL image
and the 3D SR reconstructions are in good agreement with the same
pattern from the SEM images. In addition, rotation of the 3D SR reconstruction
to show a 3D perspective (far right Figure B) shows cylindrical, pillar-like structures
similar to that in the SEM images. Figure C shows a magnification of one specific pillar
at four different orientations (pillar zoomed in marked with ∗
in Figure B). Qualitatively,
the 3D SR results are quite similar to the SEM image. The pillar is
straight and does not have any distorting shapes. As only the outer
surface of the nanopillars are labeled with fluorophores, XY projections at various Z-positions should
appear as hollow elliptical rings given sufficiently high 3D precision. Figure D depicts 100 nm Z-slices from the bottom to the top of the nanopillar illustrated
in Figure C. For all
these slices except at the top of the nanopillar, elliptical hollow
rings are evident. As the top cap of the pillar is also labeled, the
top slice, as expected, is not a hollow ring but an ellipse filled
with localizations. These Z-slices further underscore
the high image quality in our 3D SR reconstructions. Furthermore,
the Z-slices at the bottom (Z =
0 nm) show a dense number of localizations that surround the pillars.
At Z = 100 nm, we observe a dark void roughly ∼1
μm in diameter surrounding the labeled nanopillar. Interestingly,
this dark void is surrounded by localizations that do not stem from
the nanopillar. These localizations are from the coverslip, and the
appearance of the dark void 100 nm away from the bottom arises due
to the dimple found at the bottom of the nanopillar. The many qualitative
observations described here establish that the nanopillars from the
3D SR reconstructions have the same features and shapes as the pillars
in the SEM images.To quantitatively verify the high image quality,
we compared the
measured diameters of the pillars in the 3D SR reconstructions and
the SEM images. First, we extracted the diameter at the axial halfway
point in the SEM images by inspection of the SEM micrograph. To extract
the diameters in the 3D SR reconstructions, we found the center of
the pillars along Z and then took a 250 nm thick Z-slice at the center (see Methods and Figure S4A for details). We then
fit the XY projection of the localizations in this
slice to an ellipse (Figure E) and used the major and minor axes of the ellipse fit as
estimations for the diameter. Only the diameter from the fit that
was at the same orientation of the SEM image in Figure A was used. Using the SEM reference marker
as a guide, we then took the difference between the diameters from
the 3D SR reconstruction and the SEM image of the same pillar for
all the pillars analyzed (Figure F, n = 31 pillars analyzed). The mean
difference was 8.0 ± 4.7 nm (mean ± standard error of the
mean), indicating the diameters of our 3D SR reconstruction do not
differ strongly from the SEM measurement, now benchmarked by correlative
imaging.As the pillars taper, we hypothesized the number of
3D single-molecule
localizations would decrease as a function of distance away from the
coverslip. To test this, we extracted 50 nm thick Z-slices from the bottom of the coverslip to the top of the pillar
(Figure A and Methods for more details). The 50 nm Z-slices over many pillars (n = 16) were binned into
a histogram (Figure B), which shows that, as the distance away from the coverslip increases,
the number of localizations decreases, as we expect. To verify the
behavior of this distribution, we simulated nanopillars with single-molecule
localizations randomly decorating the surface of the pillars (Figure C). The extent of
tapering, the diameters, and the heights of the nanopillars were similar
to the experimentally measured dimensions (see Methods for details). Critically, the probability that a localization was
found at a specific region of the pillar was determined by the surface
area at that section of the pillar to mimic the expected local behavior
of the surface attachment. The positions of the simulated localizations
were distorted in the XY and Z direction
by Gaussian kicks with σ values equivalent to the median XY and Z localization precisions to simulate
realistic experimental conditions.
Figure 4
Quantification of the number of molecules
along the nanopillars.
(A) Cartoon depiction demonstrating the protocol to extract the number
of molecules along the nanopillars from the 3D SR reconstructions
of surface-labeled nanopillars. Localizations are projected onto the Z axis in 50 nm Z-slices from the bottom
to the top of the nanopillar. (B) Compiled histogram of n = 16 nanopillars demonstrating how the number of molecules varies
along the nanopillar. As the pillars taper, we see the count decrease
closer to the top. (C) 3D density plot of a simulated tapering nanopillar.
Each localization is colored on the basis of the local density. (D)
Compiled histogram of n = 16 simulated nanopillars.
Distribution very similar to the experimental distribution in (B).
Values in bins of the histogram are mean ± standard error of
the mean (SEM). To ensure similar Y axis scaling
for comparison, histograms have been normalized such that each bin
represents the probability of finding a molecule at that location.
Quantification of the number of molecules
along the nanopillars.
(A) Cartoon depiction demonstrating the protocol to extract the number
of molecules along the nanopillars from the 3D SR reconstructions
of surface-labeled nanopillars. Localizations are projected onto the Z axis in 50 nm Z-slices from the bottom
to the top of the nanopillar. (B) Compiled histogram of n = 16 nanopillars demonstrating how the number of molecules varies
along the nanopillar. As the pillars taper, we see the count decrease
closer to the top. (C) 3D density plot of a simulated tapering nanopillar.
Each localization is colored on the basis of the local density. (D)
Compiled histogram of n = 16 simulated nanopillars.
Distribution very similar to the experimental distribution in (B).
Values in bins of the histogram are mean ± standard error of
the mean (SEM). To ensure similar Y axis scaling
for comparison, histograms have been normalized such that each bin
represents the probability of finding a molecule at that location.Figure C depicts
a 3D scatter plot of simulated nanopillar molecule localizations (see Figure S4B for an additional orientation) color
encoded by their local density calculated from a kernel density estimation
algorithm. Qualitatively, the 3D scatter plot reveals that the number
of localizations decreases closer to the top for the simulated nanopillar,
similar to the experimental nanopillars. In addition, as expected,
the local density is nearly uniform everywhere because it is not biased
by the surface area. At the boundaries, the local density decreases
because there are 0 localizations above and below the cap and very
bottom of the pillar, respectively. We then extracted 50 nm Z-slices along the pillar using the same methodology as
the experimental measurement. Compiling the Z-slices
over 16 simulated pillars (same number of experimental pillars used
in this analysis) and binning the data into a histogram leads to the
distribution in Figure D (see Figure S4C for the distribution
of the simulated pillars without Gaussian kicks added). The shape
of the distribution of the simulated pillars very closely resembles
the shape of the distribution from the experimental data. This analysis
verifies that the pillars taper with the measured number of molecules
decreasing in an expected manner.
Cell Membrane–Nanopillar
Interactions Revealed from 3D
SR Reconstructions
Turning to cellular imaging, we first
applied our method on a simple cellular system: the plasma membrane
itself. Previous FIB-SEM and TEM approaches have imaged the membrane–nanopillar
interface to reveal that the membrane can wrap around the pillars
tightly.[12,13] Thus, unlike proteins, which may exhibit
a biologically more complex behavior near the nanopillars, the membrane
is a relatively simple cellular system to first demonstrate that our
3D SR method is applicable to imaging the nano–bio interface
in cells.To label the membrane, we overexpressed a transmembrane
domain of platelet-derived growth factor receptor (PDGFR) linked to
a SNAP tag (Methods) in U-2 OS cells that
were seeded onto the nanopillar substrate. The SNAP tag reaction was
then used to attach AF647 with low nonspecific binding (see Figure S5A for controls). With this approach,
the transmembrane domain is a single α helix that simply serves
to anchor the fluorophore to the membrane. DL images (Figure A) show a labeled membrane
with bright spots that are the nanopillars in the expected array pattern.
After 3D DHPSF imaging, postprocessing, filtering, and merging the
data, the XY projection of the 3D SR reconstruction
(Figure A, middle
and right columns) also shows highly dense regions of localizations
in an array corresponding to the locations of the nanopillars (median XY precision: 10 nm; median Z precision:
20 nm; see Figure S5B for the distributions).
This allows us to conclude that the area surrounding the nanopillar
is the cellular membrane. As we image near the growing edge of the
adhering cell whose membrane spreads on the substrate, we can observe
many membrane protrusions. When the reconstruction is projected at
a different orientation, the nanopillars appear as cylindrical pillar
like structures, as expected (Figure A; see Figure S5C for an
additional 3D SR reconstruction).
Figure 5
3D SR reconstructions of transmembrane-labeled
cells. (A) Top left
depicts a cartoon where we image the membrane at the nanopillars.
Below the cartoon, the DL image shows bright puncta corresponding
to the positions of the nanopillars in the DL image. The two 3D SR
reconstructions at the right at different orientations reveal the
membrane wrapping around the nanopillar array. Color encodes Z-position, and CS refers to the coverslip axial position.
(B) Magnified images of a single pillar (∗ in A) at various
orientations. The membrane is visibly wrapping around this pillar.
(C) 100 nm Z-slices from the bottom to the top of
the nanopillar in (C). Bottom right values indicate Z-position. Clear hollow elliptical rings are visible in the slices.
(D) Two orientations of an individual pillar with a possible endocytosis
event at the pillar. The white arrow points to the possible vesicle
in both orientations. (E) Diameters at the center of the nanopillars
extracted from surface-labeled and transmembrane-labeled 3D SR reconstructions.
The distribution from the transmembrane-labeled cells is shifted toward
higher values.
3D SR reconstructions of transmembrane-labeled
cells. (A) Top left
depicts a cartoon where we image the membrane at the nanopillars.
Below the cartoon, the DL image shows bright puncta corresponding
to the positions of the nanopillars in the DL image. The two 3D SR
reconstructions at the right at different orientations reveal the
membrane wrapping around the nanopillar array. Color encodes Z-position, and CS refers to the coverslip axial position.
(B) Magnified images of a single pillar (∗ in A) at various
orientations. The membrane is visibly wrapping around this pillar.
(C) 100 nm Z-slices from the bottom to the top of
the nanopillar in (C). Bottom right values indicate Z-position. Clear hollow elliptical rings are visible in the slices.
(D) Two orientations of an individual pillar with a possible endocytosis
event at the pillar. The white arrow points to the possible vesicle
in both orientations. (E) Diameters at the center of the nanopillars
extracted from surface-labeled and transmembrane-labeled 3D SR reconstructions.
The distribution from the transmembrane-labeled cells is shifted toward
higher values.Figure B shows
various perspective orientations of an individual pillar in the 3D
reconstruction from Figure A (denoted by the ∗ in Figure A). We clearly see that the membrane wraps
around the pillar tightly as it appears contiguous with the nanopillar.
The membrane imaging also shows that the nanopillar is elliptical
in the cross section and is perpendicular to the surface. To benchmark
the image quality of the reconstructions, we extracted 100 nm Z-slices from the bottom to the top of the membrane-wrapped
nanopillar (Figure C). Similar to the surface-labeled nanopillar reconstructions with
sufficiently high 3D precision, the Z-slices appear
as hollow elliptical rings as we labeled the membrane surrounding
the outer surface of the pillar. Figure C also shows that the hollow elliptical rings
change at the top cap region where we observe an ellipse filled with
localizations. These Z-slices confirm the capability
to observe membrane-labeled cells hugging the nanopillars.Interestingly,
for several nanopillars, we observed a bulge-like
feature at certain Z-positions along the pillar. Figure D (white arrow on
left panel) shows one such example where a bulge is prominently protruding
out from the membrane wrapped nanopillar. A 100 nm Z-slice of the image on the right shows a hollow elliptical ring,
the membrane wrapped nanopillar, attached to another hollow ring.
We hypothesize this attached hollow ring may be a vesicle forming
and budding off from the nanopillar. Vesicles have previously been
observed to bud off nanopillars from TEM images,[12] but the observation of such an effect is not easily possible
with 2D DL imaging approaches.Next, we probed how tightly the
membrane wraps around the nanopillars.
In previous work, some investigators have assumed that the membrane
wraps tightly around the nanopillar, so that the measurement of the
membrane diameter is equivalent to the diameter of the nanopillar
itself.[11,12] Those studies have probed the influence
of membrane curvature on the cell and curvature-sensing proteins indirectly
by using nanopillar diameter as a proxy. In fact, a study of how the
membrane wraps around the nanopillars with EM imaging can be extremely
challenging. For instance, while FIB-SEM methods have investigated
the interaction between the membrane and nanopillars,[13] successful FIB-SEM imaging involves cross sectioning the
interface between hard (pillars) and soft (cell) materials. Cross
sectioning between materials of different physical properties may
induce artifacts. With the 3D SR method presented here, a fluorescence
imaging technique can now be applied to directly probe the difference
in the measured diameters between the nanopillar and the membrane.To compare and extract the diameters by the two approaches, similar
to Figure E,F, we
found the center axial positions of the nanopillar regions in our
surface-labeled and membrane-labeled 3D SR reconstructions. From this
location, a 250 nm Z-slice was projected onto the XY plane and fit to an ellipse. In our previous analysis,
we extracted a diameter derived from a single axis of the fit to match
the SEM orientation (see Methods for more
details). Here, we now simply average the diameters extracted from
the two axes of the fit. The averaged diameters over many nanopillars
(n = 31 for surface-labeled nanopillars and n = 27 for membrane-labeled nanopillars) are shown in the
histogram of Figure E. The distributions show that the membrane-labeled distribution
is shifted toward larger diameters. The membrane-labeled reconstructions
have diameters of 302 ± 4.2 nm (mean ± SEM), while the surface-labeled
nanopillars have diameters of 250.7 ± 3.9 nm. The membrane and
cytoplasmic proteins bound to the membrane will therefore appear to
have larger diameters than the nanopillar itself. This roughly 50
nm difference suggests ∼25 nm gap distance between the cell
membrane and the nanopillar surface, which agrees very well with previous
measurements from FIB-SEM images.[13] Thus,
our 3D SR method allows for relatively fast and simple imaging of
the membrane at the nano–bio interface and illustrates that
the difference between the diameters of the membrane and that of the
underlying nanopillar can be observed.
3D SR Reconstructions Find
That AP-2 Distributes Away from the
Nanopillar Base
With successful application of our 3D SR
method to the labeled plasma membrane, we next imaged a more complex
biological system at the nano–bio interface: the spatial distribution
of an intracellular protein AP-2. AP-2 is a multi-subunit protein
complex that is a critical adaptor protein for clathrin-mediated endocytosis.[47−49] AP-2 has previously been shown to accumulate near nanopillars fabricated
with high degrees of curvature.[12] To image
AP-2, we used primary and secondary antibodies (see Methods and Figure S6A for controls)
where the secondary has a blinking AF647 dye attached.Figure A shows the DL image
of AP-2 labeled cells grown on the quartz nanopillars. Similar to
our previous results, the bright puncta in a specific patterned array
show the location of the nanopillar and reveal that AP-2 accumulates
at the nanopillars. The bright puncta are resolved as elliptical rings
as shown in the XY projection of the 3D SR reconstruction
in Figure A, middle
image, and in the magnified images (see Figure S6B for additional reconstructions). In addition, we observe
unstructured features colored with teal in the reconstruction. The
magnified image (Figure A) shows an example of one unstructured feature more clearly. These
structures, also termed plaques, form on the cellular surface next
to the coverslip and have been observed in previous SR experiments
imaging AP-2.[50] Thus, these plaques give
us a reference for the coverslip position.
Figure 6
3D SR reconstructions
of cells grown on nanopillars containing
labeled AP-2 proteins. (A) Cartoon in the top left depicts imaging
the cytoplasmic curvature-sensing protein AP-2 at the nano–bio
interface. Below the cartoon, we see bright puncta corresponding to
the positions of the nanopillars in the DL image. The 3D SR reconstruction
to the right shows a large field and AP-2 distributing around the
pillars (orange inset) and also on the coverslip (magenta inset). Z-position encoded by color and CS refers to the position
of coverslip in the calibration bar. (B) XZ projection
of an individual pillar. White arrows point to AP-2 distributing on
the coverslip (right arrow) and near the nanopillar (left arrow).
AP-2 on the pillar appears to distribute away from the pillar bottom.
(C) 100 nm Z-slices from the bottom to the top of
the nanopillars. Red dots indicate positions of the nanopillar. AP-2
protein appears to form sectors of rings near the pillars and with
few localizations at the pillar bottom (coverslip).
3D SR reconstructions
of cells grown on nanopillars containing
labeled AP-2 proteins. (A) Cartoon in the top left depicts imaging
the cytoplasmic curvature-sensing protein AP-2 at the nano–bio
interface. Below the cartoon, we see bright puncta corresponding to
the positions of the nanopillars in the DL image. The 3D SR reconstruction
to the right shows a large field and AP-2 distributing around the
pillars (orange inset) and also on the coverslip (magenta inset). Z-position encoded by color and CS refers to the position
of coverslip in the calibration bar. (B) XZ projection
of an individual pillar. White arrows point to AP-2 distributing on
the coverslip (right arrow) and near the nanopillar (left arrow).
AP-2 on the pillar appears to distribute away from the pillar bottom.
(C) 100 nm Z-slices from the bottom to the top of
the nanopillars. Red dots indicate positions of the nanopillar. AP-2
protein appears to form sectors of rings near the pillars and with
few localizations at the pillar bottom (coverslip).Strikingly, AP2 does not distribute uniformly on the surface
of
nanopillars like the membrane marker. The rings at the location of
the nanopillars do not have AP-2 localizations encoded in the color
teal, indicating that AP-2 does not localize around or near the bottom
of the pillar. Figure B is an XZ projection of a region close to one nanopillar,
demonstrating the behavior of AP-2 along the pillar. Adjacent to the
pillar, there is a plaque located on the coverslip. However, on the
pillar
(located near the left white arrow), we observe that the majority
of AP-2 is distributed at a higher axial position on the nanopillar
compared to the position of the plaque.Figure C further
confirms this result by showing 100 nm Z-slices from
the bottom to the top of various nanopillars, where the red dots indicate
the positions of the nanopillars. Close to the bottom (Z = 0 to 200 nm), we observe very few AP-2 molecules distributed at
the pillars. Instead, we observe the plaques that adhere to the coverslip.
As the distance away from the coverslip increases, AP-2 molecules
start to localize on the nanopillars forming rings and sectors of
rings. Again, 3D SR imaging at the nano–bio interface shows
features not observable with conventional 2D DL imaging. Only with
sufficient resolution and 3D information can we observe that AP-2
prefers to distribute away from the coverslip at higher Z-positions along the nanopillar shaft. This result will be explored
in more quantitative detail below.
3D SR Reconstructions of
Actin Molecules, Fibers, and Bundles
Distributing at the Nanopillars
Next, we investigated the
nanoscale distribution of actin molecules at the nano–bio interface.
Actin is a well-known cytoskeletal protein with many functions, for
example, polymerization to form fibers that are critical for processes
such as cell motility,[51] cell division,[52] and clathrin-mediated endocytosis.[53] In addition, the polymerization of actin fiber
is curvature sensitive and has been shown to reorganize upon changes
in membrane curvature.[11] To label actin,
we used phalloidin, a small molecule that binds specifically to actin
fibers, linked with AF647 (see Methods and Figure S7A for controls).Figure A, left, shows the DL image
of an actin-labeled cell. We observe bright puncta organized in an
array corresponding to actin fibers distributing at the nanopillars
as previously reported. We also observe actin fibers throughout the
cell. Figure A (middle
and right) shows an XY projection of the 3D SR reconstruction
(see Figure S7B for an additional reconstruction).
We clearly observe actin fibers that are located at various Z-positions in the reconstruction. In addition, we see regions
where actin is accumulating around the nanopillars. To observe these
features more clearly, Figure A, right, also shows a 300 nm thick Z-slice
at the coverslip. The white arrows depict the actin fibers and the
regions where actin accumulates around the nanopillars.
Figure 7
3D SR reconstructions
of labeled actin in cells incubated on nanopillars.
(A) DL image (top left) shows a cell where the bright puncta correspond
to regions of the nanopillars. The XY projection
of the 3D SR reconstruction (middle) shows nanopillar regions and
actin fibers. The 300 nm Z-slice close to the coverslip
(right) more clearly shows the nanopillar regions and actin fibers
(white arrows). (B) 100 nm Z-slices from the bottom
to the top of the nanopillars. We observe the actin fibers and elliptical
hollow rings at the nanopillars. (C) Magnified images of actin at
specific pillars and Z-positions shown in (B) (magenta,
teal, and orange boxes). The actin at these locations appears to form
hair-like structures that are difficult to observe with DL imaging.
3D SR reconstructions
of labeled actin in cells incubated on nanopillars.
(A) DL image (top left) shows a cell where the bright puncta correspond
to regions of the nanopillars. The XY projection
of the 3D SR reconstruction (middle) shows nanopillar regions and
actin fibers. The 300 nm Z-slice close to the coverslip
(right) more clearly shows the nanopillar regions and actin fibers
(white arrows). (B) 100 nm Z-slices from the bottom
to the top of the nanopillars. We observe the actin fibers and elliptical
hollow rings at the nanopillars. (C) Magnified images of actin at
specific pillars and Z-positions shown in (B) (magenta,
teal, and orange boxes). The actin at these locations appears to form
hair-like structures that are difficult to observe with DL imaging.We extracted 100 nm Z-slices from
the 3D SR data
from the bottom to the top of various nanopillars to more clearly
visualize the behavior of actin around the pillars. As expected, in
these Z-slices, we see hollow rings that are generally
surrounded by actin molecules. Nearby, actin fibers are prominently
seen located by the pillars at Z = 500 nm. Strikingly,
a few of the nanopillars appear to be surrounded by a hair-like structure
as well. Potentially, this hair-like structure arises from actin bundles
accumulating at the nanopillars or from fibers that wrap around the
pillars (Figure C).
These hair-like features on the pillars are large (1–2 μm),
but the fine features are much more challenging to clearly observe
in the DL image because they seem to occur only at certain Z-positions.
AP-2 Molecules Distribute along Nanopillars
at Increased 1/R Membrane Curvature Regions
The 3D SR reconstructions
of the membrane, AP-2, and actin reveal features that now may be quantified.
We consider the number of molecules along the pillars and compare
the behavior of the various labels and biomolecules. For instance,
AP-2 appears to distribute closer to the top of the pillars, in contrast
to the membrane. To quantify the number of localizations for the membrane,
AP-2, and actin reconstructions, we projected 3D localizations of
the molecules surrounding the nanopillars into 50 nm thick Z-slices, similar to our previous analysis with the surface-labeled
nanopillars (Figure A depicts a cartoon describing the analysis). For each of the labeled
targets, the 50 nm Z-slices over many pillars of
different heights and diameters were all binned and compiled into
histograms.Figure A shows the distribution of molecules along the nanopillars
for the membrane (n = 37 pillars analyzed). Given
that the pillars taper, the surface area available to the membrane
will thereby decrease, and thus, we would expect that the number of
molecules would gradually decrease toward the top of the pillar, maximizing
near the bottom of the pillar. While the number of molecules does
gradually decrease closer to the top of the nanopillar, we observe
that the count near the very bottom (Z = 0–100
nm) is low and progressively increases to a maximum of around Z = 200 nm. This result is consistent with previous findings[54] where the membrane did not fully wrap near the
bottom of the nanopillar. Instead, the membrane rises from flat regions
near the coverslip until the membrane encounters a nanopillar where
it then wraps around the pillar.
Figure 8
Quantifying molecular positions and the
variation of 1/R curvature along the nanopillars.
(A) Distribution of molecules
along the nanopillars for transmembrane-labeled cells (n = 37 pillars) quantified by projecting localizations in 50 nm Z-slices along the pillar. We observe very few molecules
near the bottom of the pillar, and the distribution peaks around Z = 200 nm. As the pillars taper, the count decreases closer
to the top. (B) Distribution of the molecules along the nanopillars
for AP-2 (n = 49 pillars). The histogram peaks around
500 nm and begins to decrease. Very few molecules are observed from Z = 0 to Z = 200 nm. (C) Distribution of
the molecules along the nanopillars for actin (n =
27 pillars). Similar to the membrane distribution, very few molecules
are found near the coverslip. The distributions seem to peak near
the middle and decrease closer to the top. To ensure similar Y axis scaling for comparison, histograms in (A), (B), and
(C) have been normalized such that each bin represents the probability
of finding a molecule at that location. (D) The 1/R curvature along the nanopillars is calculated using the SEM images
(left), the surface-labeled 3D SR reconstructions (middle), and the
transmembrane-labeled 3D SR reconstructions (right). The curves for
the SEM and surface-labeled nanopillar panels are similar. The double
headed red arrow on the membrane curvature panel shows the regions
where AP-2 preferentially distributes at the nanopillar. The dashed
black line shows where the nanopillar cap is located, and the 1/R curvature was not analyzed in this region. Values in curves
are mean ± SEM.
Quantifying molecular positions and the
variation of 1/R curvature along the nanopillars.
(A) Distribution of molecules
along the nanopillars for transmembrane-labeled cells (n = 37 pillars) quantified by projecting localizations in 50 nm Z-slices along the pillar. We observe very few molecules
near the bottom of the pillar, and the distribution peaks around Z = 200 nm. As the pillars taper, the count decreases closer
to the top. (B) Distribution of the molecules along the nanopillars
for AP-2 (n = 49 pillars). The histogram peaks around
500 nm and begins to decrease. Very few molecules are observed from Z = 0 to Z = 200 nm. (C) Distribution of
the molecules along the nanopillars for actin (n =
27 pillars). Similar to the membrane distribution, very few molecules
are found near the coverslip. The distributions seem to peak near
the middle and decrease closer to the top. To ensure similar Y axis scaling for comparison, histograms in (A), (B), and
(C) have been normalized such that each bin represents the probability
of finding a molecule at that location. (D) The 1/R curvature along the nanopillars is calculated using the SEM images
(left), the surface-labeled 3D SR reconstructions (middle), and the
transmembrane-labeled 3D SR reconstructions (right). The curves for
the SEM and surface-labeled nanopillar panels are similar. The double
headed red arrow on the membrane curvature panel shows the regions
where AP-2 preferentially distributes at the nanopillar. The dashed
black line shows where the nanopillar cap is located, and the 1/R curvature was not analyzed in this region. Values in curves
are mean ± SEM.Figure B is the
histogram for the AP-2 labeled cells (n = 49 pillars
analyzed). The distribution significantly differs from the membrane-labeled
distribution and reflects the behavior we observed in the 3D SR reconstructions
of AP-2 labeled cells (Figure ). We observe that the number of molecules is very low from Z = 0 to Z = 200 nm. From Z = 200 nm, the distribution gradually rises until it reaches a maximum
at Z = 500 nm. Further up the nanopillar, we would
expect that the number of molecules might decrease above Z = 700 nm as the cap is approached. While the number does indeed
decrease, strikingly, it decreases at a rate much smaller than the
rate found in the distribution derived from membrane-labeled cells.
This result reveals that AP-2 does not homogeneously distribute on
the membrane along the nanopillars. Instead, AP-2 appears to prefer
the middle and top rather than the bottom of the nanopillars.Figure C depicts
the histogram for actin-labeled cells (n = 27 pillars
analyzed). Similar to the membrane, the number of molecules from Z = 0 to Z = 100 nm is low. The distribution
gradually increases until a maximum is reached around Z = 450 nm. The number of molecules decreases from the maximum, although
similar to the AP-2 labeled cells, the rate of decrease is much slower
compared to the rate for membrane-labeled cells in Figure A. In general, the actin and
membrane distributions are similar except the rate decreases closer
to the top of the nanopillars.As the nanopillars taper, their
2-dimensional curvature, the reciprocal
of the radius of a cross section at fixed Z (which
we term 1/R curvature), increases from the bottom
to the top. (For the nearly cylindrical structure of the pillar shaft,
the axial curvature is approximately zero.) Thus, we hypothesized
that it is this strong variation in 1/R curvature
along the pillars that drives AP-2 to preferentially accumulate at
regions away from the bottom of the nanopillar. To first assess the
extent of tapering, we extracted the diameter along the pillars from
the SEM images to calculate the 1/R curvature (see Methods and Figure S8A for additional details). For simplicity, we only extracted the diameters
at the orientation shown in the side view magnification in Figure A. As the very top
or the cap of the nanopillar is approximately shaped as a hemisphere,
the 2D 1/R curvature metric is not well-suited for
the 3D surface at the top. Thus, we have excluded calculations of
1/R curvature near the top of the nanopillar for
all our results.Figure D shows
how the 1/R curvature changes along the pillar. We
clearly observe that, as the Z-position increases,
the curvature increases as well. However, as we described earlier,
the pillar curvature is not exactly equivalent to the membrane curvature.
Before extracting the membrane curvature, as a positive control, we
first extracted the 1/R curvature along the pillar
for our surface-labeled 3D SR reconstructions to compare to the 1/R curvature measurements from the SEM images (left and middle
panels). To extract the curvature, we first measured the diameters
of the pillars by fitting ellipses to projected 250 nm Z-slices that were centered at various axial points along the pillar
(see Methods and Figure S8B), similar to our analysis above. From the ellipse fit,
we extracted the diameter that was the same diameter measured in the
side view SEM images in Figure A. Using the diameters and repeating this analysis procedure
over many pillars, the 1/R curvature was calculated
and plotted as a function of position along the pillar in Figure D (middle). We can
clearly see that the plots and absolute values of 1/R curvature from the SEM images and SR reconstructions are reasonably
similar, underscoring the good correspondence between the two methods.Next, we calculated the 1/R plasma membrane curvature
at the nano–bio interface using the same analysis protocol
used to calculate the plot in Figure D with a slight modification. Instead of selecting
for a diameter from either the major or minor axes of the ellipse
fit, the diameters were simply averaged to calculate the 1/R curvature for the membrane. Figure D (right) reveals how the membrane curvature
varies along the pillar, increasing near the top due to the tapering.
In addition, the 1/R membrane curvature is also larger
than the 1/R curvature of the nanopillars themselves,
as expected from the separation between the nanopillar and the membrane
in Figure E. Critically,
the membrane curvature plot reveals that AP-2, in particular, does
not appear to preferentially accumulate at low degrees of membrane
curvature near the bottom of the pillar (Figure B). Only after the membrane curvature increases
its value above a specific threshold (shown by the red line in the
right panel) do we observe AP-2 beginning to preferentially accumulate
at the nanopillar.
Positive Gaussian Curvature at the Pillar
Caps Increases the
Relative Number of AP-2 and Actin Molecules
The analysis
presented above clearly reveals that increased 1/R curvature leads to a higher number of AP-2 molecules at higher axial
positions along the pillar. While the curvature analysis at the cap
of the nanopillar was excluded, the distributions in Figure B,C near the cap differ from
the membrane distribution. Thus, we investigated how AP-2 and actin
behave at the cap more closely. The behavior near the cap, however,
is obfuscated by variations in the membrane surface area, which may
be regarded as the property that is locally sensed by these proteins.
For instance, closer to the top of the pillar, the membrane surface
area is smaller, and thus, the number of molecules will decrease.
To account for these effects, we normalized the distributions in Figure B,C by the membrane
surface area found at each Z-position of the histograms.
To do this, we first assumed that the number of molecules in each Z-position in the membrane distribution in Figure A is proportional to the membrane
surface, which can also be calculated at the cap region. Then, each Z-position along the pillar in the AP-2 and actin distributions
was divided by the molecular count found at each Z-position in the membrane distribution.The distribution in Figure A depicts the normalized
distribution for AP-2 along the entire pillar. Similar to the unnormalized
distribution, the relative number of AP-2 molecules is close to zero
from Z = 0 to Z = 200 nm, rises
to a local maximum at Z = 500 nm, and then plateaus
until Z = 750 nm with the relative number being nearly
double. Strikingly, after Z = 750 nm, the distribution
rapidly increases and maximizes at the cap of the nanopillar where
the number of AP-2 molecules is now nearly 8 times greater than the
number of membrane molecules. In addition, Figure B depicts the normalized histogram for actin,
which is similar but has variations. First, from Z = 0 nm to Z = 750 nm, the number of actin molecules
is roughly equivalent to the number of membrane molecules at these
positions. However, past 750 nm, the relative number increases sharply
until it maximizes at the top of the pillar. At the cap of the pillar,
the number of actin molecules is five times greater than the number
of membrane molecules.
Figure 9
Quantifying number of molecules normalized to the membrane
surface
area and Gaussian curvature analysis along the nanopillars. (A) Distribution
of molecules of AP-2 in cells along the nanopillars normalized to
the membrane surface area (n = 49). We observe very
few relative molecules near the bottom of the coverslip. The relative
number is around two from Z = 400 to Z = 800 nm. Near the top of the pillars, the relative number increases
substantially. (B) Distribution of actin molecules in cells near the
nanopillars normalized to the membrane surface area (n = 27). The number of molecules relative to the membrane is nearly
constant until around Z = 800 nm. The relative number
increases greatly after this point. (C) Representative mesh surface
derived from 3D localization data from the transmembrane-labeled 3D
SR reconstruction. (D) Gaussian curvature along the nanopillar. Curvature
is nearly 0 but eventually increases to relatively large positive
values closer to the cap. Values are mean ± SEM.
Quantifying number of molecules normalized to the membrane
surface
area and Gaussian curvature analysis along the nanopillars. (A) Distribution
of molecules of AP-2 in cells along the nanopillars normalized to
the membrane surface area (n = 49). We observe very
few relative molecules near the bottom of the coverslip. The relative
number is around two from Z = 400 to Z = 800 nm. Near the top of the pillars, the relative number increases
substantially. (B) Distribution of actin molecules in cells near the
nanopillars normalized to the membrane surface area (n = 27). The number of molecules relative to the membrane is nearly
constant until around Z = 800 nm. The relative number
increases greatly after this point. (C) Representative mesh surface
derived from 3D localization data from the transmembrane-labeled 3D
SR reconstruction. (D) Gaussian curvature along the nanopillar. Curvature
is nearly 0 but eventually increases to relatively large positive
values closer to the cap. Values are mean ± SEM.The rapid increase in the number of AP-2 and actin molecules
at
the cap of the nanopillar indicates that the 3D shape of the pillar
near the top influences the behavior of these proteins. We hypothesize
that, since the hemielliptical shape of the cap induces a positive
Gaussian curvature, it is this property that influences the behavior
of AP-2 and actin near the cap region. Gaussian curvature is a mathematical
metric that is used to describe the curvature of 3D surfaces (see Methods). Spherical objects such as spheres and
ellipsoids have positive Gaussian curvature, while saddle surfaces
feature negative Gaussian curvature. The surfaces of cylinders have
zero Gaussian curvature. For our nanopillars, we would expect that,
as the body of the pillar is a truncated cone, the Gaussian curvature
is 0 along the shaft of the nanopillar while the hemielliptical cap
has a positive Gaussian curvature.We used approaches previously
described to extract the Gaussian
curvature along the nanopillars.[46] Briefly,
the screened Poisson surface reconstruction[55] algorithm was applied to the 3D localizations of the nanopillars
to create a 3D triangulated surface mesh using the software MeshLab[56] (Methods and Figure S9A). We created surface meshes using
both the surface-labeled (n = 15) and membrane-labeled
(n = 14) 3D SR reconstructions. Figure C shows one example of a surface
mesh derived from a membrane-labeled reconstruction (see Figure S9B for surface meshes from surface-labeled
reconstructions). From the surface, we can clearly see a pillar structure
that has a hemielliptical cap, as expected. The surface mesh additionally
reflects the positions of the 3D localizations and is hollow (Figure S9C).Gaussian curvature was then
calculated along the nanopillar as
described previously.[57] The Gaussian curvature
values along the mesh were projected into Z-slices
spaced equidistant along the pillar. The Gaussian curvature at the
various Z-slices was then calculated over many different
pillars and averaged. As expected, the Gaussian curvature from the
meshes of the surface-labeled pillars (n = 15, Figure S9D) is zero along the body and increases
to positive values at the cap. In addition, we analyzed the Gaussian
curvature from our simulated nanopillars and found (n = 15, Figure S9E) a similar behavior
to the results from the surface-labeled nanopillars, further confirming
our hypothesis that the Gaussian curvature would be highly positive
near the cap. Finally, Figure D reveals the variation in Gaussian curvature from surfaces
of the membrane-labeled reconstructions. We clearly see close to zero
Gaussian curvature along the body and highly positive Gaussian curvature
near the cap. The highly positive values at the cap correlate well
with the increased relative number of AP-2 and actin molecules near
the cap in Figure A,B. This result indicates that the positive Gaussian curvature may
be the driver, which increases the number of curvature-sensing proteins.Our results show that 2D and 3D variants of membrane curvature
metrics, 1/R curvature and Gaussian curvature, may
influence the behavior of specific curvature-sensing proteins. Previous
studies have assumed that the 1/R curvature is critical
for the behavior of curvature-sensing proteins at the nano–bio
interface. Here, we show that the shape of the pillar is also critical
for the behavior of these proteins. The positive Gaussian curvature
at the top induces changes that may increase the number and behavior
of the curvature-sensing proteins at the cap. Thus, future studies
should systematically study the effects of 1/R and
Gaussian curvature separately for various curvature-sensing proteins
to understand the contribution of each variant of curvature. This
conclusion is enabled by using our described 3D SR method that allows
one to probe 3D localizations of specific proteins at the nanoscale
on nanofabricated quartz surfaces.
Conclusion
We
have shown that having 3D SR information about the positions
of the plasma membrane and various curvature-sensing proteins can
lead to more information about the cellular behaviors near the nanopillars
seeking to define a particular nano–bio interface. This arises
from several factors: first, the higher spatial resolution; second,
the 3D character of the information; finally, since the acquisition
method involves measuring the localization of many single molecules
sampling the object of interest, it is then possible to apply powerful
statistical methods from point-based image analysis to extract further
insight. While previous studies were mostly focused on the importance
of 1/R curvature effects, here, we show distinct
effects of 1/R curvature vs Gaussian curvature in
affecting protein localizations. This ability can then be applied
in future studies in combination with the assessment of cellular signaling
changes or with two-color 3D to compare positions of different biomolecules.
By directly demonstrating the power of the technique on realistic
cellular imaging problems, we expect this approach to be widely applicable
to other cellular imaging problems where nanoscale objects nearby
drive the cellular response and behavior.
Methods
Nanopillar
Fabrication
Quartz substrates were ordered
from Technical Glass Products with dimensions of 1 in. × 1 in.
and 200 μm thickness (Technical Glass Product 1 × 1 ×
0.2). These substrates are provided with a thickness that can vary
by as much as ±50 μm, but within a batch, the thickness
variation
is much smaller. The substrates were cleaned by isopropyl alcohol
and sonication to remove any surface particles. The substrates were
then dehydrated at 180 °C on a hot plate for 2 min and then coated
with hexamethyldisilazane (HMDS) to promote resist adhesion. The arrays
of nanopillars
were fabricated on these prepared substrates using a photolithography
and wet etching technique. In the photolithography step, a positive
photoresist (Shipley 3612) was spin coated on the substrate and exposed
by a mask-less aligner (Heidelberg MLA-150) to form arrays of circular
holes. A chromium mask was later deposited onto the patterned substrates
using electron-beam evaporation, and excess resist was removed by
acetone, resulting in arrays of circular chromium disks for pillar
fabrication. Vertical pillars were subsequently fabricated by anisotropic
reactive ion etching with a mixture of CHF3, C4F8, and Ar (Versaline
LL-ICP Oxford etcher) to produce a tapering vertical profile. The
substrates were then immersed in Chromium Etchant 1020 (Transene)
to remove the chrome mask. Nanopillars were subsequently modified
through a wet etching technique to shrink the pillar diameters. The
final nanopillar dimensions were precisely controlled by submerging
the substrates in a Buffered Oxide Etchant 20:1 (Transene). The final
nanopillars were characterized with scanning electron microscopy.
SEM Imaging of Nanopillars
The SEM images were taken
using a FEI Nova (NanoSEM 450). Since quartz is a nonconductive substrate,
the imaging was operated with very low voltage (2 kV). The images
were taken with an Everhart–Thornley detector in the field-free
mode at lower magnification and in the immersion mode with the through-lens
detector for high resolution imaging at high magnification.
DNA Vectors
The vector for the eGFP-FBP17 protein fusion,
pEGFP-C1-FBP17, was a gift from Pietro De Camilli (Addgene plasmid
# 22229). To construct the vector for cell surface tethered SnapTag
expression, the DNA fragment encoding SnapTag is cloned into pDisplay
(Invitrogen, V66020) between restriction sites BglII and SalI.
Cell Culture and Transfection
U-2 OS cells (ATCC HTB-96)
were maintained in a 37 °C, 5% CO2 atmosphere in complete
cell culture medium (Dulbecco’s Modified Eagle’s Medium,
DMEM) (Sigma-Aldrich, D6429) supplied with 10% (v/v) fetal bovine
serum (FBS) (Sigma-Aldrich, F4135), 100 units/mL penicillin, and 100
μg/mL streptomycin (Gibco, 15140122). For microscopic imaging,
U-2 OS cells were detached using TrypLE Express enzyme (Gibco, 12604013)
and plated on gelatin (Sigma-Aldrich, G9391)-coated nanopillar substrates
in complete cell culture medium. Before cell plating, quartz nanopillar
substrates were first treated by air plasma (Harrick Plasma) for 15
min and then incubated in phosphate-buffered saline (PBS) with 0.1
mg/mL poly-l-lysine (Sigma-Aldrich, P5899) for 1 h. Afterward,
the nanopillar substrates were washed with PBS three times and incubated
in PBS with 0.5% (v/v) glutaraldehyde (Sigma-Aldrich, G6257). After
further washing with PBS (3×), the substrates were then incubated
in PBS with 0.5% gelatin (Sigma-Aldrich, G9391) for 1 h at 37 °C.
The coated substrates were finally washed with PBS (3×) and treated
with 1 mg/mL sodium borohydride (Sigma-Aldrich, 452882) in PBS for
5 min to eliminate autofluorescence.For the expression of GFP-FBP17
and cell surface-tethered SnapTag, cells were transfected with DNA
vectors by electroporation. U-2 OS cells were grown in 6-well plates
(Corning, 353046). For transfection, one well of the cells was detached
from the culture plate using TrypLE Express enzyme and spun down at
300 relative centrifugal force (RCF) for 3 min. The supernatants were
removed as completely as possible, leaving cell pellets that were
then resuspended in an electroporation mix containing 100 μL
of Electroporation buffer II (88 mM KH2PO4 and
14 mM NaHCO3, pH 7.4), 2 μL of Electroporation buffer
I (360 mM ATP + 600 mM MgCl2), and 1 μg of DNA vector.
The electroporation was executed in a 2 mm-gap electroporation cuvette
(Invitrogen, P45050) by Amaxa Nucleofector II following the manufacturer’s
protocol. The cells were recovered from electroporation in 650 μL
of complete cell culture medium for 5 min at room temperature (RT)
and were plated on nanopillar substrates. The cells were grown for
24 h before the next treatment.
Labeling
Surface Labeling
of Nanopillars
The nanopillar substrates
were cleaned with 1 M potassium hydroxide (Sigma-Aldrich, 221473)
solution for 15 min at RT. The substrates were washed with nanopure
water five times and air-dried. The substrates were then treated with
air plasma for 15 min and attached to plastic dishes, which have a
hole punched in the bottom, with silicone sealant. The substrates
were rinsed with anhydrous methanol (Sigma-Aldrich, 322415) and incubated
in 2 mL of a mixture containing anhydrous methanol, glacial acetic
acid (Sigma-Aldrich, 695092), and (3-aminopropyl) triethoxysilane
(Sigma-Aldrich, A3648) in a v/v/v ratio of 100:5:3 for 30 min. The
substrates were washed with anhydrous methanol five times and then
with nanopure water three times. Afterward, the substrates were rinsed
with 0.1 M sodium bicarbonate (Sigma-Aldrich, S5761) solution (pH
8) and incubated in 0.1 M sodium bicarbonate solution (pH 8) with
2 μM Alexa Fluor 647 NHS Ester (Invitrogen, A37573) for 30 min.
The substrates were washed with PBS five times before imaging.
Membrane
Labeling
The pDisplay-SnapTag transfected
cells were cultured in complete cell culture medium for 24 h before
labeling. The cells were incubated in a labeling solution, 5% CO2-balanced complete cell culture medium with 5 μM SNAP-Surface
Alexa Fluor 647 (NEB, S9136S), for 15 min at 37 °C. Before being
added to the cells, the dye and medium were mixed thoroughly by pipetting
up and down ten times. The cells were then quickly washed with CO2-balanced complete cell culture medium five times and fixed
with 4% paraformaldehyde (PFA) (Sigma-Aldrich, 158127) in PBS for
15 min at RT. The samples were washed with PBS three times before
imaging.
GFP Nanobody Labeling
Anti-GFP nanobody
(Chromotek,
gt-250) was diluted in 0.2 M sodium bicarbonate solution at pH 8.2
to a final concentration of ∼60 μM. Alexa Fluor 647 NHS
Ester stock solution (1 mg/mL in DMSO) was added into diluted anti-GFP
nanobody to a final dye concentration of ∼120 μM. The
mixture of nanobody and dye was incubated for 1 h at 25 °C. Then,
free dyes were removed from the solution using a Zeba spin desalting
column (Thermo Scientific, 89882). As the extinction coefficients
are known, the concentrations of purified nanobody and conjugated
AF647 dye were determined using light absorption at 280 and 647 nm
wavelengths, respectively. The degree of labeling (the average number
of dye molecules per protein) was calculated to be 1.08.
Immunofluorescence
Labeling
The cells were fixed with
4% PFA in PBS for 15 min at RT. The fixed cells were washed with PBS
three times and permeabilized with 0.1% triton-X (Sigma-Aldrich, T9284)
in PBS for 15 min at RT. The samples were washed with PBS three times
and then blocked in PBS with 5% (w/v) bovine serum albumin (BSA, Sigma-Aldrich
A3059) overnight at 4 °C.To label GFP-FBP17, the samples
were incubated in PBS with 5% BSA and 1 nM AF647-conjugated GFP nanobody
for 2 h at RT. The samples were then washed with PBS containing 0.1%
triton-X and 5% BSA three times for 15 min each and PBS five times
for 2 min each before imaging.To label endogenous AP2 complexes,
the samples were incubated in
PBS with 5% BSA and Anti-alpha Adaptin primary antibody (Abcam, ab2730,
1:250) for 2 h at RT and washed with PBS containing 0.1% triton-X
and 5% BSA three times for 15 min each. The samples were then incubated
with goat antimouse secondary antibody, Alexa Fluor 647 (Invitrogen
A32728, 1:500) in PBS with 5% BSA, for 2 h at RT. The samples were
finally washed with 0.1% triton-X and 5% BSA in PBS five times for
5 min each and with PBS three times before imaging.
Actin Labeling
with Phalloidin
Cells seeded on chips
were fixed with 4% PFA for 15 min and subsequently washed three times
with PBS. Cells were incubated with 5% BSA in PBS for 1 h. Following
incubation, 330 nM of phalloidin conjugated AF647 (Cell Signaling
Technology, 8940) was added to the solution for 15 min. Then, samples
were washed once with PBS before imaging.
Super-Resolution
Microscopy
Optical Setup
All data and images were acquired using
our custom built widefield double-helix PSF inverted microscope. We
imaged our samples using a 1 W 647 nm continuous wave (CW) laser (MPB
Communications) for 3D SR imaging and a 100 mW 641 nm CW (Coherent
Cube) for 2D SR imaging and calibration of the silicone oil objective.
In our setup, the 647 nm excitation laser was first passed through
a cleanup excitation bandpass (631/36) filter (Semrock, FF01-631/36-25)
and then a quarter wave plate (Thorlabs, WPQSM05-633) for circular
polarization. The size of the laser beam was then magnified twice
using two pairs of lenses before entering the backport of the microscope
(Olympus IX71). A Köhler lens placed before the backport was
used to focus the light at the back focal plane of the objective for
widefield imaging. Inside the microscope, a dichroic mirror (Semrock,
Di01-R405/488/561/635-25x36) was used to relay the light through the
objective. The standard oil immersion objective (UPlanSAPo 100x/1.4 oil, Olympus) with immersion oil (Zeiss, 444960-0000-000)
was used for imaging samples on the glass substrates. The silicone
oil immersion objective (UPlanSAPo 100x/1.35 silicone
oil, Olympus) with silicone immersion oil (Olympus, Z-81114) was used
for imaging samples on quartz substrates and for imaging nanopillar
samples. The samples were mounted on a motorized XY stage (Physik Instrumente, U-780.DOS) and a precision XYZ piezo stage (Physik Instrumente, P-545.3C8). The emitted fluorescence
light was collected using the objective, transmitted through the dichroic,
and then focused using the tube lens (f = 180 mm)
inside the microscope to the standard intermediate image plane position.
The emitted fluorescence was relayed using two lenses (f = 90 mm for both lenses) in a 4f optical configuration
to access the Fourier plane of the microscope, enabling the insertion
of the double-helix phase mask for 3D SR imaging. The double-helix
phase mask (emission wavelength of 665 nm, diameter of 2.8 mm, fabricated
with fused silica, Double-Helix Optics LLC) was removed for DL imaging
or 2D SR experiments. Emission filters (ET700/75 bandpass filter,
Chroma, ZET647 notch filter, Chroma, 680/60 bandpass filter, Omega)
for imaging with the 647 nm laser were placed in the 4f emission pathway to remove the reflected laser light and Raman scattering.
The emission filters for imaging with the 641 nm laser were changed
slightly (680/60 bandpass filter, Omega, 655 long-pass, Chroma). The
light was eventually focused by the second lens in the 4f optical pathway onto a Si electron multiplying charged-coupled device
camera for data and image acquisition (iXon897, Andor).
Calibration
of Silicone Oil Objective and Comparison with Standard
Oil Objective
To calibrate the correction collar of the SIO,
a dilute concentration of 200 nm poly(styrene) fluorescent beads (Thermo
Scientific, T7280) was immobilized in 5% (weight/volume) agarose (Invitrogen,
16520050) on a flat 200 μm thick quartz coverslip (the same
substrate used to fabricate the nanopillars). The beads were imaged
in our microscope without the DHPSF mask inserted. The bead images
were collected at an exposure of 50 ms and an EM gain of 200 on our
microscope at ∼1W/cm2. We fit each bead image to
a 2D Gaussian with least-squares regression using Matlab and extracted
the peak intensity. The correction collar was set to the adjustment
yielding the highest peak intensity. Although this was at the end
of the adjustment range, the results were good. For comparison between
objective performances, beads were immobilized in 5% agarose on the
200 μm thick quartz coverslip and a 160 μm thick glass
substrate.
2D SR Data Acquisition and Image Reconstruction
FBP17-labeled
U-2 OS cells were imaged on flat glass and quartz coverslips using
the CIO and SIO, respectively. For both imaging configurations, the
exposure time was 50 ms and the EM gain was 200. Both samples were
imaged in an oxygen scavenger reductant blinking buffer to allow emitters
to be confined to a long-lived dark state in order to ensure sparsity.
The buffer consists of 100 mM tri(hydroxymethyl)aminomethane-HCl (Thermo
Fisher), 10% (weight/volume) glucose (BD Difco), 2 μL/mL catalase,
560 μg/mL glucose oxidase, and 10 mM of cysteamine (all Sigma-Aldrich).
The focus was set close to the coverslip near the thin edge of a cell
whose membrane was spreading on the substrate in order to reduce background
from out of the focus emitters. A DL image at lower intensity (∼1W/cm2) was acquired before data acquisition. After increasing the
laser intensity to ∼1.8 kW/cm2, emitters were shelved
to the dark state for roughly 30 s. After this time period, blinking
single molecules that were not overlapping were observed and data
acquisition began. Roughly 40 000 frames were acquired for
one cellular sample.The data was processed in ThunderSTORM,
a free ImageJ plugin. The emitters were coarsely detected with a standard
maximum intensity approach. Each emitter was fit to a 2D Gaussian.
The precision was calculated using Mortensen’s equation.[41] Any emitters with poor precision (>20 nm)
and
whose σ value from the Gaussian fit was poor (>200 nm) were
removed. The localizations were binned in 2D histograms with a bin
width of 32 nm for visualization. The widths of the tubules in the
reconstructions were calculated using approaches previously described.[45]
3D SR Data Acquisition and Image Reconstruction
Prior
to 3D SR imaging of the samples, 200 nm poly(styrene) fluorescent
beads immobilized on the surface of a flat 200 μm thick quartz
coverslip with 1% (weight/volume) poly(vinyl alcohol) (PVA) were imaged
with our DHPSF microscope with the phase mask installed in the Fourier
plane. The beads were imaged over an axial range of 2 μm in
step sizes of 50 nm at 50 ms exposure time and an EM gain of 200.
We used our fine piezo XYZ stage and custom Matlab
code to move the stage. This Z-scan calibration yields
a curve relating lobe angle to Z-position that is
later used to extract the Z-position of the emitters.
Samples were incubated with dilute 200 nm poly(styrene) fluorescent
beads for 8 min prior to imaging. The bead solution was removed, and
the sample was washed three times. These beads typically stick to
the coverslip and provide fiducials for drift correction in post-processing.
The samples were then incubated in a modified blinking buffer solution
consisting of 100 mM Tris-HCl, 10% glucose, 2 μL/mL catalase,
560 μg/mL glucose oxidase, and 40 mM of cysteamine and imaged.
Nanopillar regions were first found by illuminating the sample with
white light without the phase mask installed. In this configuration,
the reference markers and arrays of the pillars were clearly visible.
Then, we fluorescently imaged the sample at low intensity until well-labeled
samples (either nanopillars or cells on the nanopillars) were visible
with a fiducial in the field of view. A DL image was first taken at
∼1W/cm2. Then, the focus was set at the coverslip
using a fiducial on the coverslip as a reference. From that point,
we moved the focus 500 nm upward using the XYZ piezo
stage. We inserted the phase mask and set the laser intensity between
2.86 and 15.9 kW/cm2. After 30 s of shelving, we acquired
the data of blinking DHPSFs. The exposure time was 35 ms, and the
EM gain was 200. We acquired approximately 70 000–100 000
frames of data.The data was processed by fitting the emitters
to a double-Gaussian function using Easy-DHPSF,[58] a freely available software in Matlab designed for localizing
DHPSF emitters. The calibration curve described above was used to
extract the Z-positions. The precision was calculated
from the detected photons using a formula calibrated for our microscope
using previously described approaches.[59] After processing, poorly localized emitters (XY precision > 30 nm or Z precision > 40 nm
or lobe
distance > 8 pixels) were removed. Axial positions of the emitters
were shifted such that Z = 0 at the coverslip. The
localized single-molecule positions were rendered with the Vutara
SRX program (Bruker), a software package designed for 3D SR visualization.
Localizations were merged to correct for overcounting (see description
below). In the reconstruction, each localization was blurred by a
Gaussian with a σ of 50 nm in X, Y, and Z to reflect the localization precision in
the measurements. When displaying our images in a series of Z-slices, the slice thickness was chosen on the basis of
the emitter density. A 100 nm thick Z-slice yielded
a sufficient number of localizations to clearly observe the hollow
rings. The Z-position was encoded by the color. For
visualization of the 3D density plots, each localization was assigned
a color on the basis of its local density, which was calculated from
a kernel density estimation algorithm.
Correction of Overcounting
Overcounting was corrected
by first isolating clusters where the localizations in the clusters
were temporally adjacent to one another. These clusters were identified
by encoding the temporal information in each localization by color.
Then, clusters were extracted if all the localizations in the clusters
were encoded in a similar color. As the localizations in these clusters
are neighboring each other both temporally and spatially, these clusters
(termed pseudoclusters) are indicative of emitters being in consecutive
frames or blinking on and off over the data acquisition. The mean
off frames (number of frames between two localizations adjacent temporally)
for all the clusters was 8.9. To correct for overcounting as exhibited
by these pseudoclusters, we applied a spatial and temporal threshold.
This threshold ensured that any molecules within a certain spatial
radius and temporal distance are merged into one molecule. We varied
both the temporal and spatial threshold and observed the effect on
the percentage of molecules that were merged for the surface-labeled
nanopillar reconstructions. After 20 off frames, the percentage of
merged molecules did not change significantly, so we set the temporal
threshold for all our reconstructions at 20 frames. Setting the XY distance at 50 nm and Z distance at
100 nm, we observed that the majority of the pseudoclusters disappeared
and were merged. Setting the spatial thresholds to larger values degraded
the image quality as localization density throughout the reconstruction
decreased. Spatial thresholds that were set lower resulted in the
appearance of the pseudoclusters. Therefore, we set the spatial thresholds
to a XY distance of 50 nm and a Z distance of 100 nm for all our 3D SR reconstructions.
Quantification
and Analysis
Diameter and Curvature Measurements
Using ImageJ, diameters
at the center of the nanopillars were extracted from SEM images by
measuring the distance of a cross section through the pillar. Only
one side of the elliptical pillars was imaged for these measurements.
For the surface-labeled nanopillars, individual pillars were isolated
in Vutara SRX. The center of the pillar was found, and a 250 nm projection
of the localizations onto the XY plane was computed.
These projections were fit to an ellipse using least-squares regression.
The axes of the ellipse fit were used as estimations for pillar diameter.
The axis of the fit that was on the same side as that for the SEM
image was extracted to compare to the SEM measurements. Note, not
all surface-labeled nanopillar reconstructions were analyzed. Occasionally,
the fitting of the ellipse failed as there were not a sufficient number
of localizations in the projection to obtain a good fit. These pillars
from the surface-labeled reconstructions and SEM images were excluded
in the analysis. The membrane diameter from the 3D SR reconstructions
was measured with the same protocol as described above. The diameters
from axes of the fit of the ellipse for the membrane and surface-labeled
reconstructions were averaged. Then, these averaged diameters were
compared with each other yielding the results in Figure F.To extract curvature
from the SEM images, cross sections were measured from the bottom
to the top of the nanopillars. The centers of these projections were
located at 20 equidistant points along the pillar to the top. These
cross sections provided the diameters of the pillars, and the reciprocal
of the radius was calculated to extract (1/R) the
curvature along the pillars. To calculate the curvature from the 3D
SR reconstructions, points along individual nanopillars spaced evenly
apart were found. These points served as the center for a 250 nm projection
of the localization onto the XY plane. The projection
was fit to an ellipse to extract the diameters and hence the curvature.
For the surface-labeled nanopillar reconstructions, only the diameter
found at the same side as the SEM image was used to calculate the
curvature. For the membrane reconstructions, the diameters of the
fit from both axes were averaged before the curvature was calculated.
As pillars with many different heights were analyzed, the curvature
at each equidistant point was averaged. The horizontal axis in Figure D is normalized to
the height from 0 to 1 to reflect the variation in heights.
Quantifying
the Number of Molecules along the Nanopillars
To calculate
the number of molecules along the nanopillars in our
3D SR reconstructions, we first determined the axial position of the
coverslip. The coverslip was clearly apparent in the surface-labeled
reconstructions. The bottom section of the membrane, the plaques,
and the actin fibers were used as the reference axial position of
the coverslip for the cellular reconstructions. The reconstruction
was then cropped from the coverslip to be 1000 nm away from the coverslip
axially. The mean heights of the pillars is 884 nm with some pillar
heights being close to 1000 nm. However, the pillars rarely exceed
1000 nm in height. Thus, the analysis of data for up to 1000 nm probes
the number of molecules over the entire experimental distribution
of heights while also excluding many localizations above the pillars.
In addition, while AP-2 and actin can be located away from the membrane,
these molecules strongly associated with the curved membrane, further
highlighting that molecules not bound to the membrane were not significant
in the analysis. These cropped pillars were exported to custom written
python code where the localizations on the pillars were projected
onto the Z axis and binned in 50 nm bin widths. To
ensure similar Y axis scaling for comparison, histograms
were normalized such that each bin represents the probability of finding
a molecule at that location.
Nanopillar Simulations
To simulate the nanopillars
(Figure D), we treated
a tapering cylinder as a truncated hollow cone. The cylinder was capped
with a hemi-ellipsoid with a radius of 80 nm in XY and 100 nm axially. The total height of the pillar was sampled from
a Gaussian distribution where the mean was 884 nm and the standard
deviation was set at 71 nm, approximating the distribution of heights
extracted from the SEM images of the nanopillars. The bottom diameter
was set at 280 nm, and the top diameter was set at 160 nm. The bottom
and top diameters were chosen on the basis of averaged values of the
bottom and top diameters extracted from the top down SEM images as
shown in Figure A
(bottom right). The probability that a localization is found along
the pillar was determined by the surface area of the pillar at that
region. 700 localizations, roughly the average number of localizations
for each nanopillar, were scattered randomly along the pillar. A Gaussian
kick was added to the position of each localization laterally and
axially to reflect the localization precision in our experimental
measurements. The Gaussian kick distribution used a lateral σ
of 12 nm and an axial σ of 20 nm (median experimental precisions
in XY and Z). Sixteen total nanopillars
were simulated (equivalent to the number of nanopillars experimentally
analyzed). To calculate the number of molecules along the pillar,
the same analysis described in the preceding section was followed
for the simulated pillars.
Gaussian Curvature
3D localizations
of isolated nanopillars
from surface-labeled and membrane-labeled reconstructions were imported
to the freely available package MeshLab.[56] The screened Poisson surface reconstruction[55] was used to create a surface mesh of the imported data (see Figure S9A for a table of parameters for generating
the surface). The surface was then exported as a .STL file. This file
was imported to Matlab, and the Gaussian curvature was extracted along
the pillar using an approach previously described[46] using a mathematical formulation for Gaussian curvature
on unstructured triangulated surfaces.[57] The Gaussian curvature along the pillar was projected into Z-slices equidistant from one another. The average Gaussian
curvature over all pillars was then calculated. The previous analysis
was additionally used to calculate the Gaussian curvature along the
simulated nanopillars. The horizontal axis was scaled from 0 to 1
to reflect the variation in pillar heights.
Software
SEM height
and diameter measurements were
acquired using ImageJ. ImageJ was also used to process DL images and
the 2D SR reconstructions. 3D SR reconstructions were rendered using
Vutara SRX. Emitters were localized for 2D SR reconstructions using
ThunderSTORM, an ImageJ plugin. Emitters were localized for 3D SR
reconstructions using Easy-DHPSF in Matlab. Fitting immobilized beads
for objective comparison and correction collar adjustment was achieved
using custom written Matlab scripts. Custom written python scripts
were used for diameter and curvature measurements from SR reconstructions,
for simulating the nanopillars, and for quantifying the number of
molecules along the pillars. MeshLab was used to produce the surface
meshes from the 3D localization data. Matlab was used to extract the
Gaussian curvature along the surfaces.
Authors: Joshua Yoon; Colin J Comerci; Lucien E Weiss; Ljiljana Milenkovic; Tim Stearns; W E Moerner Journal: Biophys J Date: 2018-12-07 Impact factor: 4.033