Ilyaas Rehman1, Sarah Ali1, Conor O'Brien1, Christine Goodall1.
Abstract
Aim: The impact on physiological parameters and well-being from potential respiratory distress caused by FFP3 masks, particularly during extensive clinical sessions, has been widely speculated during the COVID-19 pandemic. This study aims to investigate the effect of FFP3 mask wear on clinicians' pulse rate and oxygen saturation. Material &Entities:
Keywords: Aerosol‐Generating Procedures; COVID‐19; FFP3; oxygen saturation; pulse rate
Year: 2021 PMID: 34548881 PMCID: PMC8447064 DOI: 10.1111/ors.12648
Source DB: PubMed Journal: Oral Surg ISSN: 1752-2471
FIGURE 1Clinical Staff Data Collection Form. Clinical staff within the Oral Surgery department were asked to record readings for SpO2 and pulse rate at three intervals during the clinical session: before donning, before doffing and after doffing, using a finger pulse oximeter while seated. A 30 s acclimitisation period was observed following application of the pulse oximeter before the reading was taken, for standardisation. Supplementary data was also collected on the duration of mask wear, the session (AM/PM), brand of mask and history of COVID‐19 symptoms
Raw data collected by each participant across the 10 specified parameters
| Participant | Baseline SpO2 (%) |
During SpO2 (%) |
End SpO2 (%) |
Baseline pulse rate |
During pulse rate |
End pulse rate |
Duration of wear (mins) | Session (AM/PM) | Brand of mask | Previous COVID symptoms |
|---|---|---|---|---|---|---|---|---|---|---|
| 1 | 97 | 96 | 98 | 65 | 53 | 53 | 125 | AM | Arco | No |
| 2 | 97 | 96 | 96 | 61 | 54 | 55 | 240 | PM | Arco | No |
| 3 | 99 | 98 | 99 | 77 | 80 | 70 | 90 | AM | 3 M | No |
| 4 | 98 | 100 | 100 | 95 | 108 | 94 | 155 | AM | Arco | No |
| 5 | 100 | 99 | 100 | 100 | 85 | 84 | 164 | AM | 3 M | No |
| 6 | 100 | 99 | 98 | 69 | 87 | 81 | 40 | AM | Arco | No |
| 7 | 100 | 98 | 99 | 54 | 63 | 62 | 240 | PM | Arco | Yes |
| 8 | 97 | 95 | 97 | 88 | 80 | 78 | 190 | PM | Arco | No |
| 9 | 94 | 98 | 98 | 103 | 90 | 87 | 170 | PM | Arco | No |
| 10 | 98 | 96 | 97 | 67 | 55 | 53 | 134 | AM | Arco | No |
| 11 | 98 | 97 | 99 | 90 | 88 | 86 | 75 | AM | Arco | No |
| 12 | 99 | 99 | 99 | 70 | 67 | 66 | 140 | AM | Arco | Yes |
| 13 | 98 | 97 | 96 | 66 | 58 | 66 | 100 | AM | Arco | No |
| 14 | 100 | 97 | 98 | 86 | 77 | 87 | 50 | PM | Arco | No |
| 15 | 99 | 97 | 97 | 51 | 66 | 65 | 180 | AM | Arco | No |
| 16 | 98 | 100 | 98 | 86 | 94 | 103 | 170 | PM | 3 M | No |
| 17 | 100 | 98 | 100 | 77 | 72 | 85 | 167 | AM | 3 M | No |
| 18 | 96 | 99 | 99 | 60 | 53 | 58 | 185 | PM | Arco | Yes |
| 19 | 98 | 97 | 97 | 69 | 56 | 53 | 155 | PM | Arco | No |
| 20 | 96 | 97 | 96 | 83 | 84 | 79 | 90 | AM | 3 M | No |
| 21 | 98 | 100 | 99 | 67 | 77 | 80 | 170 | PM | Arco | No |
| 22 | 99 | 95 | 98 | 59 | 52 | 57 | 135 | PM | Arco | No |
| 23 | 100 | 99 | 100 | 49 | 60 | 60 | 150 | AM | Arco | Yes |
| 24 | 100 | 99 | 98 | 97 | 82 | 70 | 190 | PM | 3 M | No |
| 25 | 100 | 97 | 96 | 56 | 63 | 61 | 180 | AM | Arco | No |
| 26 | 100 | 100 | 100 | 61 | 56 | 56 | 180 | PM | 3 M+ | Yes |
| 27 | 98 | 98 | 98 | 63 | 53 | 53 | 185 | PM | 3 M | No |
| 28 | 98 | 98 | 98 | 50 | 48 | 45 | 160 | PM | 3 M | No |
FIGURE 2Relationship of raw SpO2 and pulse rate readings for each participant across three intervals: before donning FFP3 mask, before doffing and after doffing. The figure demonstrates that FFP3 masks had minimal influence on SpO2 and pulse rate. The mean SpO2 readings across the three intervals were 98.39%, 97.92% and 98.14% respectively, demonstrating an overall mean reduction of 0.25%. The mean pulse rate readings across the three intervals were 72.11, 70.04 and 69.54, demonstrating an overall mean reduction of 3.56%
The statistical report of Oxygen saturation SpO2 for each participant across three intervals
| Parameter | Mean | Upper limit | Lower limit | Standard deviation |
|---|---|---|---|---|
| SpO2 before donning (%) | 98.39 | 100 | 94 | 1.52 |
| SpO2 before doffing (%) | 97.82 | 100 | 95 | 1.47 |
| SpO2 after doffing (%) | 98.14 | 100 | 96 | 1.30 |
The statistical report of pulse rate measurements for each participant across three intervals
| Parameter | Mean | Upper limit | Lower limit | Standard deviation |
|---|---|---|---|---|
| Pulse rate before donning | 72.11 | 103 | 49 | 15.95 |
| Pulse rate before doffing | 70.04 | 108 | 48 | 15.89 |
| Pulse rate after doffing | 69.54 | 103 | 45 | 15.02 |