CT in Patients With External Fixation for Complex Lower Extremity Fractures: Impact of Iterative Metal Artifact Reduction Techniques on Metal Artifact Burden and Subjective Quality.

Andreas Stefan Brendlin1, Christian Philipp Reinert1, Heiko Baumgartner2, Malte Niklas Bongers1, Christoph Thomas3, Saif Afat1, Fabian Springer1,4, Haidara Almansour1.   

Abstract

BACKGROUND. Lower extremity external fixators have complex geometries that induce pronounced metal artifact on CT. Iterative metal artifact reduction (iMAR) algorithms help reduce such artifact, although no dedicated iMAR preset exists for external fixators. OBJECTIVE. The purpose of our study was to compare iMAR presets for CT examinations in terms of quantitative metal artifact burden and subjective image quality in patients with external fixators for complex lower extremity fractures. METHODS. This retrospective study included 72 CT examinations in 56 patients (20 women, 36 men; mean age, 56 ± 18 [SD] years) with lower extremity external fixators (regular, hybrid, or monotube). Examinations were reconstructed without iMAR (hereafter referred to as "noMAR") and with three iMAR presets (iMARspine, iMARhip, iMARextremity). A radiology resident quantified metal artifact burden using software. Two radiology residents independently assessed overall image quality and diagnostic confidence using 4-point scales (4 = excellent [highest quality or highest confidence]). Techniques were compared using Bonferroni-corrected post hoc tests. Interreader agreement was assessed by intraclass correlation coefficients (ICCs). A post hoc multinomial regression model was used for predicting overall image quality. RESULTS. Mean quantitative metal artifact burden was 100,816 ± 45,558 for noMAR, 88,889 ± 44,028 for iMARspine, 82,295 ± 41,983 for iMARhip, and 81,956 ± 41,890 for iMARextremity. Overall image quality yielded an ICC of 0.94 or greater. Using pooled reader data, median overall image quality score for the regular fixator was 2 (noMAR), 3 (iMARspine and iMARhip), and 4 (iMARextremity); for the hybrid fixator, 1 (noMAR), 2 (iMARspine), and 3 (iMARhip and iMARextremity); and for the monotube fixator, 2 (noMAR), 3 (iMARspine and iMARhip), and 4 (iMARextremity). Metal artifact burden was lower and overall image quality was higher (p < .05) for iMARhip and iMARextremity than noMAR and iMARspine for all fixators (aside from image quality of iMARhip and iMARextremity vs iMARspine for regular fixators) but were not different (all, p > .05) between iMARhip and iMARextremity. Median diagnostic confidence was 4 for all fixators and reconstructions. Independent predictors of overall quality relative to noMAR were iMARspine (odds ratio [OR] = 1.92-5.51), iMARhip (OR = 5.56-31.10), and iMARextremity (OR = 7.07-38.21). All iMAR presets introduced new reconstruction artifacts for all examinations for both readers. CONCLUSION. For the three fixator types, iMARhip and iMARextremity achieved greatest metal artifact burden reduction and highest subjective image quality, although both introduced new reconstruction artifacts. CLINICAL IMPACT. CT using the two identified iMAR presets may facilitate perioperative management of external fixators.

Entities:  

Keywords:  CT; artifacts; external fixation; fracture; image reconstruction; monotube

Mesh:

Substances:

Year:  2021        PMID: 34523951     DOI: 10.2214/AJR.21.26442

Source DB:  PubMed          Journal:  AJR Am J Roentgenol        ISSN: 0361-803X            Impact factor:   3.959


  2 in total

1.  Clinical Evaluation of an Innovative Metal-Artifact-Reduction Algorithm in FD-CT Angiography in Cerebral Aneurysms Treated by Endovascular Coiling or Surgical Clipping.

Authors:  Felix Eisenhut; Manuel Alexander Schmidt; Alexander Kalik; Tobias Struffert; Julian Feulner; Sven-Martin Schlaffer; Michael Manhart; Arnd Doerfler; Stefan Lang
Journal:  Diagnostics (Basel)       Date:  2022-05-04

2.  AI Denoising Significantly Improves Image Quality in Whole-Body Low-Dose Computed Tomography Staging.

Authors:  Andreas S Brendlin; David Plajer; Maryanna Chaika; Robin Wrazidlo; Arne Estler; Ilias Tsiflikas; Christoph P Artzner; Saif Afat; Malte N Bongers
Journal:  Diagnostics (Basel)       Date:  2022-01-17
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.