| Literature DB >> 35054391 |
Andreas S Brendlin1, David Plajer1, Maryanna Chaika1, Robin Wrazidlo1, Arne Estler1, Ilias Tsiflikas1, Christoph P Artzner1, Saif Afat1, Malte N Bongers1.
Abstract
(1) Background: To evaluate the effects of an AI-based denoising post-processing software solution in low-dose whole-body computer tomography (WBCT) stagings; (2)Entities:
Keywords: AI (artificial intelligence); computed tomography; image quality enhancement; protection; radiation; tumor staging
Year: 2022 PMID: 35054391 PMCID: PMC8774552 DOI: 10.3390/diagnostics12010225
Source DB: PubMed Journal: Diagnostics (Basel) ISSN: 2075-4418
Patient characteristics and radiation dose.
| Female | Male | Overall | ||||
|---|---|---|---|---|---|---|
| SOMATOM Definition AS+ | Patient characteristics | |||||
| N | 30 | 30 | 60 | |||
| Age | 63 ± 12 | 59 ± 16 | 61 ± 14 | |||
| BMI | 26 ± 3 | 27 ± 2 | 26 ± 3 | |||
| Radiation dose | ||||||
| CTDIvol (mean ± SD) | 4.31 ± 0.95 | 4.46 ± 0.94 | 4.38 ± 0.95 | |||
| DLP (mean ± SD) | 746.63 ± 165.58 | 759.54 ± 168.45 | 753.08 ± 167.15 | |||
| ED (mean ± SD) | ||||||
| 100% mAs | 10.72 ± 2.53 | 11.39 ± 2.53 | 11.05 ± 2.55 | |||
| 50% mAs | 5.36 ± 1.27 | 5.70 ± 1.26 | 5.53 ± 1.28 | |||
| 40% mAs | 4.29 ± 1.01 | 4.56 ± 1.01 | 4.43 ± 1.02 | |||
| 30% mAs | 3.22 ± 0.76 | 3.42 ± 0.76 | 3.32 ± 0.77 | |||
| SOMATOM Force | Patient characteristics | |||||
| N | 30 | 30 | 60 | |||
| Age | 63 ± 12 | 59 ± 16 | 61 ± 14 | |||
| BMI | 26 ± 3 | 27 ± 2 | 26 ± 3 | |||
| Radiation dose | ||||||
| CTDIvol (mean ± SD) | 4.15 ± 0.97 | 4.32 ± 0.97 | 4.22 ± 0.97 | |||
| DLP (mean ± SD) | 714.42 ± 168.85 | 726.78 ± 171.77 | 720.60 ± 170.43 | |||
| ED (mean ± SD) | ||||||
| 100% mAs | 9.17 ± 2.17 | 9.59 ± 2.13 | 9.38 ± 2.16 | |||
| 50% mAs | 4.59 ± 1.08 | 4.79 ± 1.06 | 4.69 ± 1.08 | |||
| 40% mAs | 3.67 ± 0.87 | 3.83 ± 0.85 | 3.75 ± 0.86 | |||
| 30% mAs | 2.75 ± 0.65 | 2.88 ± 0.64 | 2.81 ± 0.65 | |||
N = number; SD = standard deviation; BMI = body mass index (kg/m2); CTDIvol = Computer Tomography Dose Index (mGy); DLP = dose length product (mGy x cm); ED = effective radiation dose (mSv).
Figure 1Study flowchart and patient enrollment. wFBP = weighted filtered back-projection; ADMIRE 2 = Advanced Modeled Iterative Reconstruction strength 2.
Subjective image quality ratings, inter-rater-agreement, and intra-rater reliability.
| Timepoint 1 | Timepoint 1 | Timepoint 1 vs. 2 | ||||||||
|---|---|---|---|---|---|---|---|---|---|---|
| Rating | Inter-Rater Agreement | Rating | Inter-Rater Agreement | Intra-Rater Reliability | ||||||
| ED | Mode | Median (IQR) | ICC (95% CI) |
| Median (IQR) | ICC (95% CI) |
| ICC (95% CI) |
| |
| SOMATOM Definition AS+ | 100% | wFBP | 4 (4–4) | 0.97 (0.96–0.98) | <0.001 | 4 (3–4) | 0.95 (0.93–0.97) | <0.001 | 0.98 (0.97–0.99) | <0.001 |
| ADMIRE 2 | 4 (4–5) | 0.96 (0.94–0.98) | <0.001 | 4 (4–5) | 0.96 (0.95–0.98) | <0.001 | 0.98 (0.97–0.98) | <0.001 | ||
| wFBP + PS | 4 (4–5) | 0.97 (0.96–0.98) | <0.001 | 4 (4–5) | 0.86 (0.80–0.91) | <0.001 | 0.98 (0.97–0.98) | <0.001 | ||
| ADMIRE 2 + PS | 4 (4–5) | 0.97 (0.95–0.98) | <0.001 | 4 (4–5) | 0.93 (0.90–0.96) | <0.001 | 0.98 (0.98–0.99) | <0.001 | ||
| 50% | wFBP | 3 (3–3) | 0.98 (0.98–0.99) | <0.001 | 3 (2–3) | 0.98 (0.97–0.99) | <0.001 | 0.98 (0.98–0.99) | <0.001 | |
| ADMIRE 2 | 3 (3–4) | 0.98 (0.97–0.99) | <0.001 | 3 (3–4) | 0.97 (0.95–0.98) | <0.001 | 0.98 (0.97–0.99) | <0.001 | ||
| wFBP + PS | 4 (4–5) | 0.94 (0.92–0.96) | <0.001 | 4 (4–5) | 0.89 (0.83–0.93) | <0.001 | 0.94 (0.92–0.96) | <0.001 | ||
| ADMIRE 2 + PS | 4 (4–5) | 0.95 (0.93–0.97) | <0.001 | 4 (4–5) | 0.93 (0.90–0.96) | <0.001 | 0.95 (0.93–0.97) | <0.001 | ||
| 40% | wFBP | 2 (2–2) | 0.98 (0.97–0.99) | <0.001 | 2 (1–2) | 0.97 (0.95–0.98) | <0.001 | 0.99 (0.98–0.99) | <0.001 | |
| ADMIRE 2 | 2 (2–2) | 0.99 (0.98–0.99) | <0.001 | 2 (2–3) | 0.96 (0.94–0.97) | <0.001 | 0.98 (0.97–0.99) | <0.001 | ||
| wFBP + PS | 4 (4–5) | 0.92 (0.89–0.95) | <0.001 | 4 (4–5) | 0.91 (0.87–0.94) | <0.001 | 0.92 (0.88–0.95) | <0.001 | ||
| ADMIRE 2 + PS | 4 (4–5) | 0.95 (0.93–0.97) | <0.001 | 4 (4–5) | 0.95 (0.92–0.96) | <0.001 | 0.95 (0.92–0.96) | <0.001 | ||
| 30% | wFBP | 1 (1–1) | 0.99 (0.98–0.99) | <0.001 | 1 (1-1) | 0.96 (0.94–0.97) | <0.001 | 0.98 (0.97–0.99) | <0.001 | |
| ADMIRE 2 | 1 (1–2) | 0.97 (0.96–0.98) | <0.001 | 1 (1-2) | 0.96 (0.94–0.97) | <0.001 | 0.98 (0.97–0.99) | <0.001 | ||
| wFBP + PS | 4 (4–5) | 0.91 (0.87–0.94) | <0.001 | 4 (4-5) | 0.96 (0.94–0.98) | <0.001 | 0.91 (0.87–0.94) | <0.001 | ||
| ADMIRE 2 + PS | 4 (4–5) | 0.95 (0.92–0.96) | <0.001 | 4 (4-5) | 0.98 (0.97–0.98) | <0.001 | 0.94 (0.92–0.96) | <0.001 | ||
| SOMATOM Force | 100% | wFBP | 4 (4–5) | 0.98 (0.97–0.99) | <0.001 | 4 (4–5) | 0.98 (0.97–0.99) | <0.001 | 0.99 (0.98–0.99) | <0.001 |
| ADMIRE 2 | 5 (4–5) | 0.95 (0.92–0.97) | <0.001 | 5 (4–5) | 0.96 (0.95–0.98) | <0.001 | 0.96 (0.95–0.98) | <0.001 | ||
| wFBP + PS | 5 (4–5) | 0.82 (0.74–0.88) | <0.001 | 5 (4–5) | 0.89 (0.84–0.93) | <0.001 | 0.84 (0.78–0.90) | <0.001 | ||
| ADMIRE 2 + PS | 5 (5–5) | 0.91 (0.86–0.94) | <0.001 | 5 (4–5) | 0.91 (0.87–0.94) | <0.001 | 0.85 (0.79–0.90) | <0.001 | ||
| 50% | wFBP | 3 (3–4) | 0.99 (0.98–0.99) | <0.001 | 3 (3–4) | 0.98 (0.97–0.99) | <0.001 | 0.99 (0.98–0.99) | <0.001 | |
| ADMIRE 2 | 4 (3–4) | 0.96 (0.94–0.97) | <0.001 | 4 (3–4) | 0.95 (0.93–0.97) | <0.001 | 0.97 (0.96–0.98) | <0.001 | ||
| wFBP + PS | 4 (4–5) | 0.92 (0.88–0.95) | <0.001 | 4 (4–5) | 0.91 (0.88–0.94) | <0.001 | 0.84 (0.77–0.89) | <0.001 | ||
| ADMIRE 2 + PS | 4 (4–5) | 0.94 (0.92–0.96) | <0.001 | 4 (4–5) | 0.83 (0.75–0.89) | <0.001 | 0.88 (0.83–0.92) | <0.001 | ||
| 40% | wFBP | 2 (2–3) | 0.99 (0.99–1.00) | <0.001 | 2 (2–3) | 0.98 (0.96–0.98) | <0.001 | 0.99 (0.98–0.99) | <0.001 | |
| ADMIRE 2 | 3 (2–3) | 0.98 (0.97–0.99) | <0.001 | 3 (2–3) | 0.95 (0.92–0.97) | <0.001 | 0.97 (0.96–0.98) | <0.001 | ||
| wFBP + PS | 4 (4–5) | 0.92 (0.88–0.94) | <0.001 | 4 (4–5) | 0.86 (0.80–0.91) | <0.001 | 0.86 (0.80–0.91) | <0.001 | ||
| ADMIRE 2 + PS | 4 (4–5) | 0.93 (0.89–0.95) | <0.001 | 4 (4–5) | 0.9 (0.85–0.93) | <0.001 | 0.85 (0.79–0.90) | <0.001 | ||
| 30% | wFBP | 1 (1–2) | 0.98 (0.97–0.99) | <0.001 | 1 (1–2) | 0.97 (0.96–0.98) | <0.001 | 0.99 (0.99–0.99) | <0.001 | |
| ADMIRE 2 | 2 (1–2) | 0.96 (0.94–0.97) | <0.001 | 2 (1–2) | 0.93 (0.89–0.95) | <0.001 | 0.98 (0.97–0.99) | <0.001 | ||
| wFBP + PS | 4 (4–5) | 0.86 (0.80–0.91) | <0.001 | 4 (4–5) | 0.92 (0.88–0.95) | <0.001 | 0.87 (0.82–0.91) | <0.001 | ||
| ADMIRE 2 + PS | 4 (4–5) | 0.92 (0.88–0.95) | <0.001 | 4 (4–5) | 0.95 (0.92–0.96) | <0.001 | 0.93 (0.90–0.95) | <0.001 | ||
ED = effective radiation dose; wFBP = weighted filtered back-projection; ADMIRE 2 = Advanced Modeled Iterative Reconstruction strength 2; PS = PixelShine; ICC = intraclass correlation coefficient; 95% CI = 95% confidence interval; p = significance level.
Linear regression metrics: subjective image quality.
| Variable | B | SE | 95% CI | |t| |
|
|---|---|---|---|---|---|
| Intercept | 1.21 | 0.06 | 1.09–1.33 | 20.1 | <0.001 |
| ED (reference: 30%) | |||||
| 40% | 0.78 | 0.08 | 0.63–0.93 | 10.1 | <0.001 |
| 50% | 1.78 | 0.08 | 1.63–1.93 | 23 | <0.001 |
| 100% | 2.78 | 0.08 | 2.63–2.93 | 35.9 | <0.001 |
| Scanner (reference: SOMATOM Definition AS+) | |||||
| SOMATOM Force | 0.22 | 0.07 | 0.09–0.35 | 3.26 | 0.053 |
| Mode (reference: wFBP) | |||||
| ADMIRE 2 | 0.04 | 0.08 | −0.11–0.2 | 0.58 | 0.564 |
| ADMIRE 2 + PS | 2.88 | 0.08 | 2.72–3.03 | 37.2 | <0.001 |
| wFBP + PS | 3.06 | 0.08 | 2.91–3.21 | 39.6 | <0.001 |
| Rater (reference: Rater 1) | |||||
| Rater 2 | 0.01 | 0.08 | –0.16–0.16 | 0 | >0.999 |
| Rater 3 | 0.01 | 0.08 | –0.16–0.16 | 0 | >0.999 |
| Rater 4 | 0.01 | 0.08 | –0.14–0.17 | 0.15 | 0.882 |
| Rater 5 | 0.01 | 0.08 | –0.16–0.16 | 0 | >0.999 |
| Timepoint (reference: Timepoint 1) | |||||
| Timepoint 2 | 0.03 | 0.07 | –0.1–0.16 | 0.46 | 0.648 |
B = estimate; SE = standard error; 95% CI = 95% confidence interval; |t| = absolute value of t statistics, p = significance level; ED = effective radiation dose; Mode = reconstruction/postprocessing mode; wFBP = weighted filtered back-projection; ADMIRE 2 = Advanced Modeled Iterative Reconstruction strength 2; PS = PixelShine; Timepoint = time of subjective image quality rating.
Figure 2Pooled subjective image quality scores. wFBP = weighted Filtered Back Projection, ADMIRE 2 = Advanced Modeled Iterative Reconstruction strength 2.
Figure 3Comparison of image quality for all scanner types, radiation dose levels, and reconstruction methods. wFBP = weighted Filtered Back Projection, ADMIRE 2 = Advanced Modeled Iterative Reconstruction strength 2.
Mean noise levels and pairwise noise comparisons.
| Noise | |||||||
|---|---|---|---|---|---|---|---|
| ED | Mode | Mean ± SD | wFBP | ADMIRE 2 | wFBP + PS | ADMIRE 2 + PS | |
| SOMATOM Definition AS+ | 100% | wFBP | 12.85 ± 1.63 | <0.001 | <0.001 | <0.001 | |
| ADMIRE 2 | 10.75 ± 1.36 | <0.001 | <0.001 | <0.001 | |||
| wFBP + PS | 7.8 ± 0.13 | <0.001 | <0.001 | 0.423 | |||
| ADMIRE 2 + PS | 7.8 ± 0.13 | <0.001 | <0.001 | 0.423 | |||
| 50% | wFBP | 17.99 ± 2.51 | <0.001 | <0.001 | <0.001 | <0.001 | |
| ADMIRE 2 | 15.05 ± 2.11 | <0.001 | <0.001 | <0.001 | <0.001 | ||
| wFBP + PS | 7.8 ± 0.15 | <0.001 | <0.001 | 0.936 | 0.943 | ||
| ADMIRE 2 + PS | 7.8 ± 0.15 | <0.001 | <0.001 | 0.860 | 0.991 | ||
| 40% | wFBP | 20.06 ± 2.89 | <0.001 | <0.001 | <0.001 | <0.001 | |
| ADMIRE 2 | 16.78 ± 2.42 | <0.001 | <0.001 | <0.001 | <0.001 | ||
| wFBP + PS | 7.8 ± 0.18 | <0.001 | <0.001 | 0.947 | 0.941 | ||
| ADMIRE 2 + PS | 7.8 ± 0.17 | <0.001 | <0.001 | 0.963 | 0.969 | ||
| 30% | wFBP | 23.07 ± 3.46 | <0.001 | <0.001 | <0.001 | <0.001 | |
| ADMIRE 2 | 19.29 ± 2.89 | <0.001 | <0.001 | <0.001 | <0.001 | ||
| wFBP + PS | 7.8 ± 0.2 | <0.001 | <0.001 | 0.980 | 0.981 | ||
| ADMIRE 2 + PS | 7.8 ± 0.2 | <0.001 | <0.001 | 0.979 | 0.969 | ||
| SOMATOM Force | 100% | wFBP | 12.35 ± 1.57 | <0.001 | <0.001 | <0.001 | <0.001 |
| ADMIRE 2 | 10.73 ± 1.36 | <0.001 | <0.001 | <0.001 | |||
| wFBP + PS | 7.8 ± 0.13 | <0.001 | <0.001 | 0.422 | |||
| ADMIRE 2 + PS | 7.8 ± 0.12 | <0.001 | <0.001 | 0.422 | |||
| 50% | wFBP | 17.29 ± 2.42 | <0.001 | <0.001 | <0.001 | <0.001 | |
| ADMIRE 2 | 15.02 ± 2.11 | <0.001 | <0.001 | <0.001 | <0.001 | ||
| wFBP + PS | 7.8 ± 0.15 | <0.001 | <0.001 | 0.927 | 0.807 | ||
| ADMIRE 2 + PS | 7.8 ± 0.14 | <0.001 | <0.001 | 0.920 | 0.927 | ||
| 40% | wFBP | 19.27 ± 2.78 | <0.001 | <0.001 | <0.001 | <0.001 | |
| ADMIRE 2 | 16.74 ± 2.41 | <0.001 | <0.001 | <0.001 | <0.001 | ||
| wFBP + PS | 7.8 ± 0.17 | <0.001 | <0.001 | 0.930 | 0.868 | ||
| ADMIRE 2 + PS | 7.8 ± 0.17 | <0.001 | <0.001 | 0.936 | 0.895 | ||
| 30% | wFBP | 22.17 ± 3.32 | <0.001 | <0.001 | <0.001 | <0.001 | |
| ADMIRE 2 | 19.26 ± 2.89 | <0.001 | <0.001 | <0.001 | <0.001 | ||
| wFBP + PS | 7.8 ± 0.19 | <0.001 | <0.001 | 0.936 | 0.897 | ||
| ADMIRE 2 + PS | 7.8 ± 0.19 | <0.001 | <0.001 | 0.978 | 0.978 | ||
ED = effective radiation dose; SD = standard deviation; mAs = tube current; wFBP = weighted Filtered Back Projection; ADMIRE 2 = Advanced Modeled Iterative Reconstruction strength 2; PS = PixelShine.
Figure 4Visualization of mean noise levels. wFBP = weighted Filtered Back Projection, ADMIRE 2 = Advanced Modeled Iterative Reconstruction strength 2.
Figure 5Image quality aspects in the setting of a hepatic melanoma metastasis at the different radiation dose levels, conventional reconstruction (top row) vs. postprocessing (bottom row). wFBP = weighted Filtered Back Projection, PS = PixelShine.