Literature DB >> 34516005

Neutrophil dysfunction due to continuous mechanical shear exposure in mechanically assisted circulation in vitro.

Wenji Sun1, Jiafeng Zhang1, Aakash Shah1, Katherin Arias1,2, Zachary Berk1, Bartley P Griffith1, Zhongjun J Wu1,2.   

Abstract

OBJECTIVE: Leukocytes play an important role in the body's immune system. The aim of this study was to assess alterations in neutrophil phenotype and function in pump-assisted circulation in vitro.
METHODS: Human blood was circulated for four hours in three circulatory flow loops with a CentriMag blood pump operated at a flow of 4.5 L/min at three rotational speeds (2100, 2800, and 4000 rpm), against three pressure heads (75, 150, and 350 mm Hg), respectively. Blood samples were collected hourly for analyses of neutrophil activation state (Mac-1, CD62L, CD162), neutrophil reactive oxygen species (ROS) production, apoptosis, and neutrophil phagocytosis.
RESULTS: Activated neutrophils indicated by both Mac-1 expression and decreased surface expression of CD62L and CD162 receptors increased with time in three loops. The highest level of neutrophil activation was observed in the loop with the highest rotational speed. Platelet-neutrophil aggregates (PNAs) progressively increased in two loops with lower rotational speeds. PNAs peaked at one hour after circulation and decreased subsequently in the loop with the highest rotational speed. Neutrophil ROS production dramatically increased at one hour after circulation and decreased subsequently in all three loops with similar levels and trends. Apoptotic neutrophils increased with time in all three loops. Neutrophil phagocytosis capacity in three loops initially elevated at one hour after circulation and decreased subsequently. Apoptosis and altered phagocytosis were dependent on rotational speed.
CONCLUSIONS: Our study revealed that the pump-assisted circulation induced neutrophil activation, apoptosis, and functional impairment. The alterations were strongly associated with pump operating condition and duration.
© 2021 International Center for Artificial Organs and Transplantation and Wiley Periodicals LLC.

Entities:  

Keywords:  blood pump; mechanical shear stress; mechanically assisted circulation; neutrophil dysfunction

Mesh:

Substances:

Year:  2021        PMID: 34516005      PMCID: PMC8688241          DOI: 10.1111/aor.14068

Source DB:  PubMed          Journal:  Artif Organs        ISSN: 0160-564X            Impact factor:   3.094


  55 in total

1.  Neutrophil extracellular traps kill bacteria.

Authors:  Volker Brinkmann; Ulrike Reichard; Christian Goosmann; Beatrix Fauler; Yvonne Uhlemann; David S Weiss; Yvette Weinrauch; Arturo Zychlinsky
Journal:  Science       Date:  2004-03-05       Impact factor: 47.728

2.  Evaluation of von Willebrand factor with a fully magnetically levitated centrifugal continuous-flow left ventricular assist device in advanced heart failure.

Authors:  Ivan Netuka; Tomáš Kvasnička; Jan Kvasnička; Ingrid Hrachovinová; Peter Ivák; František Mareček; Jana Bílková; Ivana Malíková; Michaela Jančová; Jiří Malý; Poornima Sood; Kartik S Sundareswaran; Jean M Connors; Mandeep R Mehra
Journal:  J Heart Lung Transplant       Date:  2016-05-31       Impact factor: 10.247

3.  Neutrophil Mac-1 and MEL-14 adhesion proteins inversely regulated by chemotactic factors.

Authors:  T K Kishimoto; M A Jutila; E L Berg; E C Butcher
Journal:  Science       Date:  1989-09-15       Impact factor: 47.728

Review 4.  Left ventricular assist devices and their complications: A review for emergency clinicians.

Authors:  Brit Long; Jennifer Robertson; Alex Koyfman; William Brady
Journal:  Am J Emerg Med       Date:  2019-05-06       Impact factor: 2.469

5.  Activation of polymorphonuclear leukocytes by candidate biomaterials for an implantable glucose sensor.

Authors:  Andrey Sokolov; Bernt Christian Hellerud; John D Lambris; Erik A Johannessen; Tom Eirik Mollnes
Journal:  J Diabetes Sci Technol       Date:  2011-11-01

6.  Quantitative Characterization of Shear-Induced Platelet Receptor Shedding: Glycoprotein Ibα, Glycoprotein VI, and Glycoprotein IIb/IIIa.

Authors:  Zengsheng Chen; Steven C Koenig; Mark S Slaughter; Bartley P Griffith; Zhongjun J Wu
Journal:  ASAIO J       Date:  2018 Nov/Dec       Impact factor: 2.872

7.  Systemic Inflammatory Response Syndrome in End-Stage Heart Failure Patients Following Continuous-Flow Left Ventricular Assist Device Implantation: Differences in Plasma Redox Status and Leukocyte Activation.

Authors:  Nandan K Mondal; Erik N Sorensen; Si M Pham; Steven C Koenig; Bartley P Griffith; Mark S Slaughter; Zhongjun J Wu
Journal:  Artif Organs       Date:  2015-09-29       Impact factor: 3.094

Review 8.  A head-to-tail view of L-selectin and its impact on neutrophil behaviour.

Authors:  Aleksandar Ivetic
Journal:  Cell Tissue Res       Date:  2018-01-20       Impact factor: 5.249

Review 9.  The Inflammatory Response to Ventricular Assist Devices.

Authors:  Gemma Radley; Ina Laura Pieper; Sabrina Ali; Farah Bhatti; Catherine A Thornton
Journal:  Front Immunol       Date:  2018-11-15       Impact factor: 7.561

10.  The effect of ventricular assist device-associated biomaterials on human blood leukocytes.

Authors:  Gemma Radley; Ina Laura Pieper; Catherine A Thornton
Journal:  J Biomed Mater Res B Appl Biomater       Date:  2017-09-09       Impact factor: 3.368