Alexander Gangl1, Hannes Alexander Deutschmann2, Rupert Horst Portugaller1, Georg Stücklschweiger3.
Abstract
OBJECTIVE: By simulating a fluoroscopic-guided vascular intervention, two differently designed radiation safety glasses were compared. The impacts of changing viewing directions and body heights on the eye lens dose were evaluated. Additionally, the effect of variable magnification levels on the arising scattered radiation was determined.Entities:
Keywords: Cataract; Eye protective devices; Fluoroscopy; Phantoms imaging; Thermoluminescent dosimetry
Mesh:
Year: 2021 PMID: 34495352 PMCID: PMC8831265 DOI: 10.1007/s00330-021-08231-y
Source DB: PubMed Journal: Eur Radiol ISSN: 0938-7994 Impact factor: 5.315
Fig. 1Positioning of dosimeters (TLDs, RaySafe i3) at the phantom head
Fig. 2Safety glass models (left: Type BR 115, right: Type BR 126)
Fig. 3Experimental set-up: Generating eye dose values under protected conditions considering varying viewing directions (0°, 20°, 40°, 60°)
Dose reduction factors referring to changing viewing directions, eye levels and protective eyeglass models
| Eye level | Head rotation (°) | Dose reduction factor | Range | |||
|---|---|---|---|---|---|---|
| 0 | 20 | 40 | 60 | |||
| 100 | Type 115 | 6.9 | 8.0 | 6.0 | 3.9 | 4.0 |
| Type 126 | 3.3 | 4.0 | 2.8 | 2.4 | 1.6 | |
| 157 | Type 115 | 4.2 | 2.0 | 2.1 | 2.3 | 2.1 |
| Type 126 | 1.3 | 1.3 | 1.3 | 1.0 | 0.3 | |
| 176 | Type 115 | 2.0 | 1.3 | 1.4 | 1.2 | 0.9 |
| Type 126 | 1.5 | 1.1 | 1.3 | 1.1 | 0.4 | |
Absolute dose values considering changing eye levels, protective eyeglass models and viewing directions
| Eye level 100 cm | Eye level 157 cm | Eye level 176 cm | ||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Head rotation (°) | 0 | 20 | 40 | 60 | 0 | 20 | 40 | 60 | 0 | 20 | 40 | 60 |
| No protection (µGy) | 683 | 642 | 605 | 565 | 377 | 341 | 331 | 322 | 275 | 204 | 239 | 229 |
| Type 115 (µGy) | 98 | 81 | 100 | 143 | 89 | 85 | 155 | 142 | 136 | 159 | 166 | 196 |
| Type 126 (µGy) | 205 | 161 | 217 | 234 | 299 | 259 | 263 | 316 | 186 | 192 | 179 | 200 |
Fig. 4Radiation exposure to protected and unprotected eye lenses considering increasing eye levels (100, 157 and 176 cm) and head rotations (0°, 20°, 40°, 60°)
Fig. 5Linear regression between scattering dose and magnification level
X-ray parameters considering increasing magnification levels according to shrinking FOV sizes
| FOV size | Exposure parameters | |||||
|---|---|---|---|---|---|---|
| Tube voltage (kV) | Tube current (mA) | Radiation time per frame (ms) | Cumulative air kerma (mGy) | Dose rate (Gy/min) | Kerma-area product (µGym2) | |
| FOV 48 | 70 | 288 | 34.3 | 351.711.2 | 2.1 ± 0.1 | 14614 ± 320 |
| FOV 32 | 76 | 259 | 36.8 | 435 ± 4.0 | 2.4 ± 0.3 | 7966 ± 70.1 |
| FOV 11 | 83 | 478 | 38.8 | 1152 ± 17.3 | 6.3 ± 0.1 | 2683 ± 35.4 |