Pirbhat Shams1, Muhammad Mehdi1, Jamshed Ali1, Intisar Ahmed1, Sheema Saadia1, Sameen Iqbal1, Aamir Hameed Khan1, Yawer Saeed1.
Abstract
BACKGROUND: Smartphone-based applications to identify cardiac implantable electronic devices (CIED) are extremely useful in circumstances, where urgent device interrogation is needed, and a device identification card is not available. Few studies have provided insights regarding the utility of these applications. We have studied two widely available applications i.e., Pacemaker ID app (PMIDa) or Cardiac Rhythm Management Device-Finder (CRMD-f) to identify device manufacturers in CIEDs.Entities:
Keywords: Arrhythmias; CIED interrogation; Cardia-X; Device identity; Pacemaker
Year: 2021 PMID: 34484722 PMCID: PMC8408627 DOI: 10.1016/j.amsu.2021.102741
Source DB: PubMed Journal: Ann Med Surg (Lond) ISSN: 2049-0801
Fig. 1Pacemaker ID application user interface.
Fig. 2Cardiac Rhythm Management Device Finder application user interface.
Fig. 3Radiographic parts of the cardiac implantable electronic device.
Device models tested in the study.
| Device Model | Number (n) | Percentage (%) |
|---|---|---|
| Medtronic PPM Sensia | 89 | 14.9 |
| Medtronic Evera S MRI ICD | 30 | 5 |
| Medtronic Bravaquard CRT-D | 17 | 2.8 |
| Medtronic Attesta PPM MRI | 4 | 0.7 |
| Medtronic Ensura DR MRI PPM | 265 | 44.2 |
| Abbot Fortify Assura | 27 | 4.5 |
| Others | 115 | 21% |
Sensitivity, specificity, and accuracy of Pacemaker ID app and CRMD-finder app.
| Application | Sensitivity (%) | Specificity (%) | Accuracy (%) |
|---|---|---|---|
| PacemakerID | 55 | 82.5 | 60.5 |
| CRMD-Finder | 53.6 | 93.58 | 61.6 |
Fig. 4Sensitivity, specificity, and accuracy of Pacemaker ID app and CRMD-finder app.
- Comparison of our study with Chudow et al. demonstrating inter-study variability of accuracies of PMIDa and CRMD-f.
| Device characteristic | PMIDa was correct Shams et al. | PMIDa accuracy Chudow et al. | CRMD-Finder was correct | CaRDIA-X accuracy Chudow et al. | ||
|---|---|---|---|---|---|---|
| Medtronic | 57.9% | 96% | 56.6% | 72% | ||
| St. Jude Medical | 75.2% | 89% | 94.4% | 84% | ||
| Dual chamber PPM | 52.1% | 52.5% | 49.7% | 50.6% | 88% | |
| Single lead PPM | 71.4% | 100% | ||||
| CRT | 100% | 100% | ||||
| Single lead ICD | 75.4% | 95% (For all ICDs) | 93.4% | 88% (For all ICDs) | ||
| Anterior-posterior projection | 72.4% | 94% | ||||
| Posterior-anterior projection | 59.3% | 98.5% | ||||