| Literature DB >> 34458192 |
Jalal Tabesh1, Maziyar Mahdavi2,3, Gholamhasan Haddadi2,3, Rezvan Ravanfar Haghighi4, Reza Jalli5.
Abstract
BACKGROUND: The diagnostic reference level (DRL) is measured with different methods in the common Computed tomography (CT) exams, but it has not been measured through the size-specific dose estimate (SSDE) method in Iran, yet.Entities:
Keywords: Body Mass Index; Diagnostic Reference Levels; Multidetector Computed Tomography; Quality Control-Based Dose Survey Method; Size-Specific Dose Estimate
Year: 2021 PMID: 34458192 PMCID: PMC8385214 DOI: 10.31661/jbpe.v0i0.2105-1322
Source DB: PubMed Journal: J Biomed Phys Eng ISSN: 2251-7200
Parameters of computed tomography (CT) scans performed with 8 scanners in Shiraz University CT centers. Computed tomography dose index (CTDIvol), dose length product (DLP), Abd Pel, b and c stand for, the volume computed tomography dose index, dose length product, abdomen/pelvis examination, skull base and cerebrum, respectively.
| CT scan type | CT Examination/ scanner mode | kVp | Average mAs (Range) | Pitch/ Interval (mm) | Scan Length (cm) (Range) | Beam Collimation (mm) | Average CTDIvol (mGy) (Range) | Average DLP (mGy.cm) (Range) |
|---|---|---|---|---|---|---|---|---|
| Siemens 16 Emotion | Head/Axial | 130 | 90 | 22.1 | 15.29 (12-17.5) | 19.2 | 19.9 (18.3-20.9) | 303 (249-332) |
| Chest/Spiral | 110 | 89.8 (58-100) | 1.395 (1.35-1.5) | 41.14 (36-49.1) | 19.2 | 6.7 (6.5-7.2) | 276 (252-354) | |
| Abd,Pel/Spiral | 130 | 113.2 (92-120) | 1.46 (1.4-1.5) | 53.7 (49.13-6) | 19.2 | 12.7 (10.4-13.4 | 691 (539-767) | |
| Neusoft 2 | Head/Axial | 120 | 189 (140-210) | 6 | 13.86 (11-15.7) | 5 | 20 (16.4-21.5) | 275 (229-338) |
| GE 8 Bright Speed | Head/Axial | 120 | 192 (180-240) | 10.43 b 15.65 c | 10.08 (8-12.8) | 10 b15 c | 35.2 (32.7-42.9) | 469 (438-536) |
| Sinus/Spiral | 120 | 123 (87-161) | 1.35 | 13.4 (11-15) | 10 | 12.4 (5.7-19.2) | 166 (68-306) | |
| Chest/Spiral | 120 | 149 (105-239) | 1.35 | 33.67 (29-41) | 20 | 9.4 (6.1-15.9) | 314 (189-519) | |
| Abd,Pel/Spiral | 120 | 173 (143-224) | 1.35 | 48.2 (40-55) | 20 | 10.6 (7.5-14.7) | 346 (334-747) | |
| GE 16 Bright Speed | Head/Axial | 120 | 265 | 22.09 | 14.13 (12-18) | 20 | 41.7 (38.3-42.5) | 595 (542-747) |
| Sinus/Spiral | 120 | 71 (40-200) | 1.75 | 14.23 (12.4-17.7) | 10 | 7.6 (4.8-27.1) | 125 (62-480) | |
| Chest/Spiral | 120 | 172 (117-294) | 1.75 | 31.69 (27-35) | 20 | 9.4 (6.3-15.6) | 296 (196-517) | |
| Abd,Pel/Spiral | 120 | 206 (123-294 | 1.75 | 45.58 (36-53) | 20 | 9.5 (5.3-15.9) | 456 (241-715) | |
| Phillips 16 Brilliance | Head/Axial | 120 | 265 (250,400) | 18.9 | 13.85 (11.7-16.5) | 18 | 31 (28.4-48.6) | 439 (380-585) |
| Sinus/Spiral | 120 | 79 (24-200) | 1.75 | 12.88 (11-16) | 12 | 9.9 (11.6-27.1) | 134 (18-301) | |
| Chest/Spiral | 120 | 129 (85-227) | 1.188 | 35 (27.5-42.6) | 24 | 8 (5.2-14.1) | 283 (176-486) | |
| Abd,Pel/Spiral | 120 | 179 (133-234) | 1.188 | 49.64 (46-56.4) | 24 | 11.2 (8.3-14.5) | 378 (396-821) | |
| GE 2 Hi Speed | Head/Axial | 120 | 225 b195 c | 10.17 b 14.24 c | 13.22 (12.1-15.2) | 10 b14 c | 34.6 (33.6-35.1) | 358 (412-512) |
| Phillips 128 Ingenuity | Head/Spiral | 120 | 166 (140-224) | 0.328 | 20.87 (18.3-26.4) | 40 | 21.4(18.8-28.9) | 449 (333-597) |
| Sinus/Spiral | 120 | 26 (23-32) | 0.399 | 17.37 (15.6-18.6) | 40 | 3.4 (3-4.1) | 58 (48-72) | |
| Chest/Spiral | 120 | 139 (102-230) | 1.2 | 34.3 (26-38.2) | 40 | 9.1 (6.7-15.1) | 313 (229-556) | |
| Abd,Pel/Spiral | 120 | 173 (78-339) | 0.797 | 51.2 (44.1-56.6) | 40 | 8.3 (5.1-13.4) | 431 (246-696) | |
| Siemens 2 Sensation | Head/Axial | 130 | 275 (260-290) | 6 b 10 c | 14.9 (13.6-17) | 3 b10 c | 38.5 (36.1-39.4) | 573 (493-652) |
| Sinus/Axial | 130 | 90 | 5 | 10.08 (7.7-13.2) | 5 | 16.5 (16.2,17.8) | 167 (125-214) |
CT: Computed tomography, CTDIvol: Computed tomography dose index, DLP: Dose length product
The local diagnostic reference level (LDRL) based on the volume computed tomography dose index (CTDIvol), and dose length product (DLP) obtained for the four common computed tomography (CT) examinations in four University centers in Shiraz, using data collection (DC), quality control based (QC) and modified quality control based (MQC) methods, respectively.
| CT examination | Dose Index | LDRL (Method) | ||
|---|---|---|---|---|
| DC | QC | MQC | ||
|
| CTDIvol mGy | 37.4 | 54.5 | 50.7 |
| DLP mGy.cm | 543 | 718 | 735 | |
|
| CTDIvol mGy | 16.2 | 16 | 18.1 |
| DLP mGy.cm | 173 | 216 | 232 | |
|
| CTDIvol mGy | 9.3 | 11.2 | 15.4 |
| DLP mGy.cm | 399 | 387 | 519 | |
|
| CTDIvol mGy | 12.5 | 13.1 | 18.9 |
| DLP mGy.cm | 672 | 718 | 984 | |
CT: Computed tomography, LDRL: Local diagnostic reference level, DC: Data collection, QC: Quality control based, MQC: Modified quality control based, CTDIvol Computed tomography dose index, DLP: Dose length product
Figure 1The diagnostic reference level (DRL) based on the volume computed tomography dose index (CTDIvol) for the four common CT scan examinations in four university centers in Shiraz obtained using the data collection (DC), quality control based (QC), and modified quality control based (MQC) methods, respectively. Abd/Pel, represents the abdomen/pelvis examination.
Figure 2The diagnostic reference level (DRL) based on dose length product (DLP) for the four common computed tomography (CT) scan examinations in four University centers in Shiraz obtained using the data collection (DC), quality control based (QC), and modified quality control based (MQC) methods, respectively. Abd/Pel, represents the abdomen/pelvis examination.
The mean effective diameter (Eff Diameter (cm)) and conversion factor f with the standard deviation (SD) for the computed tomography (CT) examinations of the head, sinus, chest, abdomen and pelvis based on the American Association of Physicist in Medicine (AAPM) report 204 [ 13 ].
| CT Exam | Eff. Diameter (cm) | Mean f Factor | SD |
|---|---|---|---|
|
| 16-18 | 0.99 | ± 0.03 |
|
| 13-16 | 1.06 | ± 0.04 |
|
| 26-30 | 1.3 | ± 0.10 |
|
| 23-29 | 1.43 | ± 0.14 |
CT: Computed tomography, SD: Standard deviation
Figure 3Measurements of the anteroposterior and lateral dimensions on the chest computed tomography (CT) scan images with contrast for the calculation of the patient effective diameter.
The comparison of the local diagnostic reference level (LDRL) values obtained based on the volume computed tomography dose index (CTDIvol) and dose length product (DLP) in this study with the national and local values: Parsi et al. [ 11 ], Najafi et al. [ 18 ], Sohrabi et al. [ 21 ], Deevband et al. [ 22 ] and Sohrabi et al. [ 23 ].
| CT Exam | Studies Dose Index | This Study Shiraz DC 2020 (8) | This Study Shiraz QC 2020 (8) | This Study Shiraz MQC 2020 (8) | Sohrabi Shiraz (DPQC) 2018 (5) | Parsi Tehran (QC) 2015 (70) | Najafi Iran (DC) 2014 (22) | Deevband Iran (Di) 2016 (120) |
|---|---|---|---|---|---|---|---|---|
| Head | CTDIvol (mGy) | 37 | 55 | 51 | 57 | 59 | 43 | 44 |
| DLP (mGy.cm) | 543 | 718 | 735 | 771 | 834 | 700 | 647 | |
| Sinus | CTDIvol (mGy) | 16 | 16 | 18 | 19 | 29 | 22 | 9 |
| DLP (mGy.cm) | 173 | 216 | 232 | 193 | 235 | 290 | 142 | |
| Chest | CTDIvol (mGy) | 9 | 11 | 15 | 10 | 10 | 10 | 9 |
| DLP (mGy.cm) | 399 | 387 | 519 | 280 | 233 | 330 | 289 | |
| Abd/Pel | CTDIvol(mGy) | 13 | 13 | 19 | 13 | 13 | 10 | 11 |
| DLP (mGy.cm) | 672 | 718 | 984 | 552 | 522 | 550 | 513 |
CT: Computed tomography, DC: Data collection, QC: Quality control based, MQC: Modified quality control based, DPQC: Dual purpose quality control, CTDIvol Computed tomography dose index, DLP: Dose length product
Figure 4The diagram shows diagnostic reference levels (DRL) of four routine examination based on the volume computed tomography dose index (CTDIvol) calculated with different methods in this study and the national and local studies conducted by Parsi et al. [ 11 ], Najafi et al. [ 18 ], Sohrabi et al. [ 21 , 23 ], Deevband et al. [ 22 ], respectively. The MQC, QC, DPQC, DC and Di stand for: modified quality control based, quality control based, dual purpose quality control based, data collection and direct methods, respectively. The American national study carried out by Kanal et al. [ 20 ] that used size specific dose estimate (SSDE) method with two water effective diameters in the chest, abdomen/pelvis, and the lateral diameter of the skull also included for comparison. The sinus examination is not included in the American study. Ab/p and Ch, represent the abdomen/pelvis and chest examinations, respectively.
Comparison of diagnostic reference levels (DRL) based on the volume computed tomography dose index method (CTDIvol) and dose length product (DLP) in the chest, abdomen and pelvis examinations in the American (USA) national study conducted by Kanal et al. [20] using the size specific dose estimate (SSDE) and data collection (DC) methods, respectively, with the present study using the modified quality control based (MQC), quality control based (QC), and DC methods, respectively. The Ab, Ch and Abd/Pel stand for abdomen and chest effective diameter in present study, and the abdomen/pelvis examination in both studies, respectively.
| DRL [CTDIvol mGy, (DLP mGy.cm)] | |||||||
|---|---|---|---|---|---|---|---|
| Shiraz (Effective Diameter) cm | USA (Water Equivalent Diameter) cm | ||||||
|
| MQC 26-30 Ch 23-29 Ab | QC 26-30 Ch 23-29 Ab | DC 26-30 ch 23-29Ab | SSDE 25-29 | SSDE 29-33 | DC 25-29 | DC 29-33 |
|
| 15 (519) | 11 (387) | 9 (399) | 13 (366) | 15 (469) | 10 (238) | 13 (353) |
|
| 19 (984) | 13 (718) | 13 (672) | 15 (524) | 18 (755) | 11 (409) | 15 (608) |
DRL: Diagnostic reference level, CTDIvol Computed tomography dose index, DLP: Dose length product, CT: Computed tomography, MQC: Modified quality control based, QC: Quality control based, SSDE: Size specific dose estimate, DC: Data collection