Diagnostic abdominal MR imaging on a prototype low-field 0.55 T scanner operating at two different gradient strengths.

Hersh Chandarana1, Barun Bagga2, Chenchan Huang2, Bari Dane2, Robert Petrocelli2, Mary Bruno2, Mahesh Keerthivasan3, David Grodzki4, Kai Tobias Block2, David Stoffel2, Daniel K Sodickson2.   

Abstract

PURPOSE: To develop a protocol for abdominal imaging on a prototype 0.55 T scanner and to benchmark the image quality against conventional 1.5 T exam.
METHODS: In this prospective IRB-approved HIPAA-compliant study, 10 healthy volunteers were recruited and imaged. A commercial MRI system was modified to operate at 0.55 T (LF) with two different gradient performance levels. Each subject underwent non-contrast abdominal examinations on the 0.55 T scanner utilizing higher gradients (LF-High), lower adjusted gradients (LF-Adjusted), and a conventional 1.5 T scanner. The following pulse sequences were optimized: fat-saturated T2-weighted imaging (T2WI), diffusion-weighted imaging (DWI), and Dixon T1-weighted imaging (T1WI). Three readers independently evaluated image quality in a blinded fashion on a 5-point Likert scale, with a score of 1 being non-diagnostic and 5 being excellent. An exact paired sample Wilcoxon signed-rank test was used to compare the image quality.
RESULTS: Diagnostic image quality (overall image quality score ≥ 3) was achieved at LF in all subjects for T2WI, DWI, and T1WI with no more than one unit lower score than 1.5 T. The mean difference in overall image quality score was not significantly different between LF-High and LF-Adjusted for T2WI (95% CI - 0.44 to 0.44; p = 0.98), DWI (95% CI - 0.43 to 0.36; p = 0.92), and for T1 in- and out-of-phase imaging (95%C I - 0.36 to 0.27; p = 0.91) or T1 fat-sat (water only) images (95% CI - 0.24 to 0.18; p = 1.0).
CONCLUSION: Diagnostic abdominal MRI can be performed on a prototype 0.55 T scanner, either with conventional or with reduced gradient performance, within an acquisition time of 10 min or less.
© 2021. The Author(s), under exclusive licence to Springer Science+Business Media, LLC, part of Springer Nature.

Entities:  

Keywords:  Abdominal MRI; Accessible MRI; Low-cost MRI; Low-field MRI

Mesh:

Year:  2021        PMID: 34415411      PMCID: PMC8720489          DOI: 10.1007/s00261-021-03234-1

Source DB:  PubMed          Journal:  Abdom Radiol (NY)


  2 in total

1.  New-Generation Low-Field Magnetic Resonance Imaging of Hip Arthroplasty Implants Using Slice Encoding for Metal Artifact Correction: First In Vitro Experience at 0.55 T and Comparison With 1.5 T.

Authors:  Iman Khodarahmi; Inge M Brinkmann; Dana J Lin; Mary Bruno; Patricia M Johnson; Florian Knoll; Mahesh B Keerthivasan; Hersh Chandarana; Jan Fritz
Journal:  Invest Radiol       Date:  2022-03-04       Impact factor: 10.065

2.  Simulated diagnostic performance of low-field MRI: Harnessing open-access datasets to evaluate novel devices.

Authors:  T Campbell Arnold; Steven N Baldassano; Brian Litt; Joel M Stein
Journal:  Magn Reson Imaging       Date:  2021-12-28       Impact factor: 3.130

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.