Sergey V Nesterov1,2, Roberto Sciagrà3, Luis Eduardo Juarez Orozco4, John O Prior5, Leonardo Settimo3, Chunlei Han4, Emmanuel Deshayes5,6, Robert A deKemp7, Darja V Ryzhkova8, Kilem L Gwet9, Juhani M Knuuti4.
Abstract
PURPOSE: To cross-compare three software packages (SPs)-Carimas, FlowQuant, and PMOD-to quantify myocardial perfusion at global, regional, and segmental levels.Entities:
Keywords: Imaging software; Myocardial perfusion imaging; N-13 ammonia; absolute quantification; agreement; positron emission tomography; quantitative imaging; reproducibility; standardization of PET
Mesh:
Substances:
Year: 2021 PMID: 34409572 PMCID: PMC9553813 DOI: 10.1007/s12350-021-02741-4
Source DB: PubMed Journal: J Nucl Cardiol ISSN: 1071-3581 Impact factor: 3.872
Population characteristics
| 48 | |
| No. of males (% of total) | 30 (63%) |
| Age, years. (range) | 44 ± 15.4 (16–79) |
| Weight, kg (range) | 70 ± 11.6 (4–98) |
| Body mass index, kg/m2 (range) | 24.1 ± 3.43 (18.3–32.9) |
| Symptoms | |
| Angina | 4 (8%) |
| Dyspnea | 20 (42%) |
| Syncope | 5 (10%) |
Values are n (%), or arithmetic means ± SD
Stress MBF values
| Carimas-1TCM | FlowQuant-1TCM | PMOD-1TCM | Max/Min ratio | Median | |
|---|---|---|---|---|---|
| Global | 1.47 ± 0.52 | 1.18 | 1.55 | ||
| LAD | 1.32 ± 0.49 | 1.15 | 1.38 | ||
| LCx | 1.81 ± 0.63 | 1.27 | 1.77 | ||
| RCA | 1.57 ± 0.59 | 1.15 | 1.59 | ||
| Seg1 | 1.64 ± 0.58 | 1.17 | 1.66 | ||
| Seg2 | 1.38 ± 0.49 | 1.23 | 1.38 | ||
| Seg3 | 1.52 ± 0.59 | 1.15 | 1.47 | ||
| Seg4 | 1.74 ± 0.69 | 1.16 | 1.81 | ||
| Seg5 | 1.88 ± 0.70 | 1.36 | 1.78 | ||
| Seg6 | 1.90 ± 0.68 | 1.28 | 1.82 | ||
| Seg7 | 1.66 ± 0.61 | 1.19 | 1.58 | ||
| Seg8 | 1.43 ± 0.53 | 1.16 | 1.33 | ||
| Seg9 | 1.52 ± 0.58 | 1.17 | 1.54 | ||
| Seg10 | 1.78 ± 0.74 | 1.10 | 1.68 | ||
| Seg11 | 1.93 ± 0.73 | 1.30 | 1.79 | ||
| Seg12 | 1.98 ± 0.70 | 1.21 | 1.86 | ||
| Seg13 | 1.40 ± 0.53 | 1.04 | 1.38 | ||
| Seg14 | 1.28 ± 0.52 | 1.10 | 1.19 | ||
| Seg15 | 1.43 ± 0.55 | 1.11 | 1.31 | ||
| Seg16 | 1.59 ± 0.64 | 1.09 | 1.51 | ||
| Seg17 | 1.20 ± 0.53 | 1.06 | 1.12 |
Values are mean ± SD (n = 48); median values are calculated for n = 144 in 1TCM. Minimum (Min) and maximum (Max) values in each row for 1TCM are in bold
Figure 1Comparison of stress MBF values on global and regional levels. The X-axis is the difference in MBF values expressed in percent of corresponding medians; Y-axis is 1-ICC. The shaded iris area limits the acceptable agreement: ± 20% of the median value on the X-axis and ICCs over 0.75 on the Y-axis. The chart element corresponds to the comparison in the left ventricle area of the same name. C2, Carimas; FQ, FlowQuant; PM, PMOD
Figure 2Comparison of stress MBF values on the segmental level. The X-axis is the difference in MBF values expressed in percent of corresponding medians; Y-axis is 1-ICC. The shaded iris area limits the acceptable agreement: ± 20% of the median value on the X-axis and ICCs over 0.75 on the Y-axis. The chart element corresponds to the comparison in the same name’s left ventricle area (e.g., ‘15’ stands for comparison in segment 15). C2, Carimas; FQ, FlowQuant; PM, PMOD
Figure 3Polar maps demonstrating the differences in estimated MBF values across SPs depicted across the levels of segmentation of the myocardium.