Literature DB >> 34394740

Bicortical versus unicortical fixation of plated clavicular fractures: A biomechanical study.

Ezequiel E Zaidenberg1,2, Michael Voor3, Enrique Pereira4, Luciano A Rossi1, Carlos R Zaidenberg2.   

Abstract

PURPOSE: To compare the biomechanical properties of three plate stabilization techniques for midshaft clavicle fractures: anatomical bicortical locking construct, anatomical unicortical locking construct, and reconstruction bicortical locking construct.
METHODS: We analyzed superior plating of the clavicle using an anatomical clavicle plate (Acumed) with both bicortical or unicortical screw fixation and a locking reconstruction plate (DePuy-Synthes). Twenty-one fourth generation composite clavicles were used for non-destructive stiffness testing in axial loading, bending, and torsion. Fifteen composite clavicles and 17 foam clavicles were used for cyclic failure testing using a combined loading method that included all three loading modes.
RESULTS: No significant differences were found between the three constructs in torsional stiffness nor in vertical bending loading. In axial loading, the anatomical bicortical locking construct was significantly stiffer than either anatomical unicortical locking construct or the reconstruction bicortical locking construct. The unicortical fixation was also significantly stiffer than the reconstruction bicortical locking construct. Regarding failure testing, there was not a significant difference between the bicortical and unicortical anatomical locking constructs; however, both were significantly stronger than bicortical screw fixation in the reconstruction plate.
CONCLUSION: Based on the biomechanical performance of these constructs, unicortical locked plate fixation may be a reasonable option in the treatment of displaced midshaft clavicle fracture fixation.
© 2020 The British Elbow & Shoulder Society.

Entities:  

Keywords:  biomechanical; clavicle; fracture; unicortical

Year:  2020        PMID: 34394740      PMCID: PMC8355658          DOI: 10.1177/1758573220914217

Source DB:  PubMed          Journal:  Shoulder Elbow        ISSN: 1758-5732


  22 in total

Review 1.  Plate Fixation Versus Nonoperative Treatment for Displaced Midshaft Clavicular Fractures: A Meta-Analysis of Randomized Controlled Trials.

Authors:  Sarah Woltz; Pieta Krijnen; Inger B Schipper
Journal:  J Bone Joint Surg Am       Date:  2017-06-21       Impact factor: 5.284

Review 2.  Evolution of plate design and material composition.

Authors:  David J Hak; Rodrigo Banegas; Kyros Ipaktchi; Cyril Mauffrey
Journal:  Injury       Date:  2018-06       Impact factor: 2.586

3.  Biomechanical comparison of bicortical locking versus unicortical far-cortex-abutting locking screw-plate fixation for comminuted radial shaft fractures.

Authors:  Steven J Overturf; Randal P Morris; Zbigniew Gugala; Ronald W Lindsey
Journal:  J Hand Surg Am       Date:  2014-08-20       Impact factor: 2.230

Review 4.  Locked Plating and Advanced Augmentation Techniques in Osteoporotic Fractures.

Authors:  Graham J DeKeyser; Patrick J Kellam; Justin M Haller
Journal:  Orthop Clin North Am       Date:  2019-02-12       Impact factor: 2.472

5.  Biomechanical analysis of bicortical versus unicortical locked plating of mid-clavicular fractures.

Authors:  Daniel Hamman; Derek Lindsey; Jason Dragoo
Journal:  Arch Orthop Trauma Surg       Date:  2010-11-11       Impact factor: 3.067

6.  Biomechanical comparison of contemporary clavicle fixation devices.

Authors:  Timothy Renfree; Bryan Conrad; Thomas Wright
Journal:  J Hand Surg Am       Date:  2010-04       Impact factor: 2.230

7.  Forces across the middle of the intact clavicle during shoulder motion.

Authors:  Maria Iannolo; Frederick W Werner; Levi G Sutton; Sean M Serell; Scott M VanValkenburg
Journal:  J Shoulder Elbow Surg       Date:  2010-07-16       Impact factor: 3.019

8.  Plate fixation compared with nonoperative treatment of displaced midshaft clavicular fractures: a randomized clinical trial.

Authors:  A H Qvist; M T Væsel; C M Jensen; S L Jensen
Journal:  Bone Joint J       Date:  2018-10       Impact factor: 5.082

9.  Upper-extremity deep-vein thrombosis following a clavicular fracture.

Authors:  Deepthi Nandan Adla; A Ali; S A Shahane
Journal:  Eur J Orthop Surg Traumatol       Date:  2004-06-18

10.  Clavicle fixation screw impingement causing subclavian artery pseudoaneurysm.

Authors:  Steven D Lewis; Felix S Chew
Journal:  Radiol Case Rep       Date:  2019-07-29
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.