Literature DB >> 34377380

Estimation of Blood Loss in Oral and Maxillofacial Surgery by Measurements of Low Haemoglobin Levels in Mixtures of Blood, Saliva and Saline: a Laboratory Study.

Krister Johansson1,2, Martin Lindström1, Manaf Alhabshi1, Marianne Ahmad1, Peter J Svensson3, Jonas P Becktor1.   

Abstract

OBJECTIVES: Estimating blood loss is an important factor in several surgical procedures. The accuracy of blood loss measurements in situations where blood is mixed with saliva and saline is however uncertain. The purpose of this laboratory study was to ascertain if blood loss measurements in mixtures of blood, saline, and saliva are reliable and could be applicable in a clinical setting.
MATERIAL AND METHODS: Venous blood and resting saliva were collected from six volunteers. Saliva, saline, and combinations thereof were mixed with blood to obtain different concentrations. A portable spectrophotometer was first used to measure the haemoglobin concentration in undiluted venous blood followed by measurements of the haemoglobin concentration after each dilution. To examine the strength of linear relationships, linear regression and Pearson correlations were used.
RESULTS: The measurements of haemoglobin concentrations in mixtures of blood, saline, and saliva were proven to be accurate for haemoglobin measurements > 0.3 g/dl (correlation = 0.986 to 1). For haemoglobin measurements < 0.3 g/dl, a small increase in haemoglobin values were reported, which was directly associated to the saliva concentration in the solution (correlation = 0.983 to 1). This interference of saliva was significantly eliminated by diluting the samples with saline, mimicking the clinical situation.
CONCLUSIONS: The results suggest that a portable spectrophotometer can be used clinically to preoperatively measure the haemoglobin value of a venous blood sample and postoperatively measure the haemoglobin value of the collected liquids, including shed blood, thereby achieving a highly accurate method of measuring blood loss during oral and maxillofacial surgery.
Copyright © Johansson K, Lindström M, Alhabshi M, Ahmad M, Svensson PJ, Becktor JP. Published in the JOURNAL OF ORAL & MAXILLOFACIAL RESEARCH (http://www.ejomr.org), 30 June 2021.

Entities:  

Keywords:  oral surgery; postoperative hemorrhage; saliva; surgical blood loss

Year:  2021        PMID: 34377380      PMCID: PMC8326882          DOI: 10.5037/jomr.2021.12203

Source DB:  PubMed          Journal:  J Oral Maxillofac Res        ISSN: 2029-283X


INTRODUCTION

Peri- and postoperative blood loss in oral and maxillofacial surgery (OMFS) is seldom a clinical problem but under certain circumstances major haemorrhage may occur. Multiple measurements like pulse, cutaneous vasoconstriction, urinary output, and estimated blood loss are important factors to detect hypovolemia and initiate fluid management such as transfusion [1]. Various methods are available to estimate the amount of blood loss during different surgeries [2,3]. However, the data on monitoring blood loss during OMFS is rather limited. There are published studies on relatively rough estimations of blood loss in OMFS performed by weighing surgical gauze swabs and measuring the contents after deducting the saline from the total volume of the suction bottle [4,5]. Another method for monitoring blood loss that may be useful during OMFS is to calculate the blood loss by multiplying the total volume of the liquid collected in the suction bottle by its haemoglobin value, and then divide the product by the haemoglobin value from a preoperatively collected venous blood sample of the patient [6,7]. To facilitate the procedure, a portable device for haemoglobin measurements may be used. The HemoCue® system (HemoCue AB; Ängelholm, Sweden), is a compact device that provides an immediate haemoglobin measurement from a small blood volume deposited to a disposable cuvette [8]. The HemoCue® system has been compared with other spectrophotometric devices and has shown a high accuracy regarding haemoglobin concentration measurements [8,9]. Spectrophotometry in general has been proven quite accurate in measurements of the haemoglobin concentration in blood but in OMFS, where blood is often diluted due to the saline irrigation and saliva, it could be difficult with the current systems to accurately measure haemoglobin concentrations [10]. To overcome these difficulties, the HemoCue® Plasma/Low Hb system has been developed and has been proven to accurately measure haemoglobin levels in low blood concentrations [11-13]. How ever, to the best of our knowledge, neither the Hemocue® system nor other devices intended to measure haemoglobin levels have been tested on blood diluted with saliva. Hence, the aim of the present study was to, in a laboratory model, validate haemoglobin measurements in diverse mixtures of blood, saliva, and saline using a Hemocue® haemoglobin system and thereby test the hypothesis that saline and saliva would not influence the results of the measurements.

MATERIAL AND METHODS

Population Blood and saliva were collected from six healthy volunteers (donor 1 - 6), two males and four females, aged from 26 to 64. The subjects were divided into groups. Five participants (donor 1, 2, 4 - 6) were included in the preparatory test were different concentrations of saliva were analyzed with both devices. Three subjects (donor 1 - 3) underwent measurements of different combinations of blood and saliva with the HemoCue® Hb 201+ system and three subjects (donor 2 - 4) underwent the analyses of various concentrations of blood and saliva with the HemoCue® Plasma/Low Hb system. Recruitment was done among the staff at the department were the study was conducted and informed consent was obtained from all individual participants. The study was conducted from 5 May, 2017 to 28 June, 2017 at the Department of Oral and Maxillofacial Surgery and Oral Medicine, Faculty of Odontology, Malmö University, Malmö, Sweden. The study was approved by the by the Ethical Review Board in Lund, Sweden on 12 May, 2016 (Dnr. 2016/286) and has been performed according to the general principles of the Declaration of Helsinki (General Assembly, Helsinki, Finland, June 1964), on ethical principles for medical research involving human subjects. Blood and saliva collection Blood and saliva samples were collected 1 to 2 hours before implementation of the experiments and stored at room temperature (22 to 25 °C). Blood was collected through venepuncture into evacuated tubes containing K2 EDTA (BD Vacutainer® Plus Plastic K2EDTA tubes - Becton, Dickinson and Co.; Franklin Lakes, New Jersey, USA) [14]. Resting saliva was collected in sterile tubes by spitting. Before collection of saliva the donors were asked to thoroughly rinse the mouth. The tubes were stored on a tube rocker after collection and during the experiments in order to achieve homogeneity of the samples. Before utilisation of saliva, foam was removed from the tubes to enable correct volume measurements. One blood sample and one tube of saliva were collected from each participant. Assay methods HemoCue® Hb 201+ system and HemoCue® Plasma/Low Hb system and microcuvettes recommended by the manufacturer were used for analysing haemoglobin concentrations in diverse mixtures of blood, saliva, and saline. Both devices are factory-calibrated against the International Council for Standardization in Haematology (ICSH) reference method. The reaction in the microcuvettes is a modified azidmethaemoglobin reaction. The erythrocyte membranes are disintegrated by sodium deoxycholate and haemoglobin is released. Sodium nitrite transforms haemoglobin into methaemoglobin, which combines with sodium azide to form azidemethaemoglobin. To correct for turbidity, two wavelengths (570 nm and 880 nm) are used for the absorbance measurements. All analyses were performed in an accredited laboratory (Skåne County Council, Office for Medical Services, Division of Laboratory Medicine, Department of Clinical Chemistry, Section of Special Chemistry, Unit of Malmö) with validated instruments at room temperature. To avoid potential immunological reactions, blood and saliva from different individuals were not mixed. As a preparatory test, saliva from five subjects (donor 1, 2, 4 - 6) was diluted with saline to saliva concentrations of 100%, 75%, 50%, 25% and 12.5% and analysed to evaluate any impact on the HemoCue® Hb 201+ and HemoCue® Plasma/Low Hb systems of saliva only. According to the manufacturer, the intended use for HemoCue® Hb 201+ system is quantitative determination of haemoglobin in whole blood within the measuring range from 0 to 25.6 g/dl. Three different liquids were prepared: one reference/golden standard solution containing saline only; one containing a mixture of equal amounts of saliva and saline (50/50) and one solution containing saliva. The prepared liquids were poured into several test tubes and, when applicable, mixed with blood to final blood concentrations of 0%, 25%, 50% and 75%. These samples and samples containing 100% blood were analysed. According to the manufacturer, HemoCue® Plasma/Low Hb system is designed for quantitative determination of low levels of haemoglobin in plasma, serum, and aqueous solutions. Based on calculations of normal salivary flow, estimated duration of OMFS interventions and average volume of saline used during dento-alveolar surgery three different liquids were prepared: one reference/golden standard solution containing saline only, one containing saliva 12.5%/saline 87.5% and one solution containing saliva only [15-17]. The HemoCue® Plasma/Low Hb system is linear and calibrated for measurements between 0.03 and 3 g/dl. Consequently, to achieve haemoglobin levels within this range, each one of the 3 mixtures was poured into several test tubes and mixed with blood to final blood concentrations of 0.4%, 0.8%, 2.4%, 7.6% and 15.2%. To analyse the samples obtained from the experiments, each liquid was thoroughly mixed and then a drop of the solution was placed on a hydrophobic clean surface. A microcuvette was then filled according to the instruction for use and analysed within a minute. This was repeated five times for each sample. Previous to any series of measurement each system was checked using quality control solutions obtained from the manufacturer. Statistical analysis To examine the strength of linear relationships of the samples containing blood in different concentrations, linear regression and Pearson and Spearman´s correlations were used. For both the HemoCue® Hb 201+ system and the HemoCue® Plasma/Low Hb system the haemoglobin values retained from the measurements of the series of blood samples diluted with saline only, saliva only or a mixture of saline and saliva were compared. The samples from different donors were not compared with each other due to variances in blood and saliva composition. For the HemoCue® Plasma/Low Hb system the limits of detection (LOD) were calculated. Analyses were performed using Microsoft Office Excel 2013 program (Microsoft Corporation, Redmond, Washington, USA) and figures were produced in SPSS Statistics version 25 (IBM®, Chicago, Illinois, USA).

RESULTS

Whole blood system (HemoCue For every single donor, the values from the different measurements correlated well, hence the HemoCue® Hb 201+ system did not react either on saliva from five subjects (donor 1, 2, 4 - 6) in different concentrations diluted with saline (Table 1) or on series of samples from donor 1 - 3 containing blood mixed with saliva and/or saline, except for isolated outliers when blood was diluted with saliva only (Table 2, Figure 1).
Table 1

Analyses of saliva diluted in saline only

Donor 1 2 4 5 6
Whole blood system Results not elevated Results not elevated Results not elevated Results not elevated Results not elevated

Low haemoglobin system 25% 75% 25% 50% 50%

Description of the lowest saliva concentration required (0 to 100%) to get elevated results (> 0 g/dL) from the analyses. Saliva 0 to 100% did not affect the whole blood measurement device. Saliva 25 to 75% or more affected the low haemoglobin device.

Table 2

Analyses made with the whole blood system of blood diluted with saliva and/or saline

Donor 1 (n = 25) 2 (n = 25) 3 (n = 25)




Diluent Saline Saline 50%, saliva 50% Saliva Saline Saline 50%, saliva 50% Saliva Saline Saline 50%, saliva 50% Saliva
Spearman´s rank correlation 0.991 0.99 0.989 0.989 0.991 0.99 0.989 0.988 0.986

Pearson correlation 1 1 1 1 1 0.998 1 0.999 1

Slope 0.157 0.157 0.157 0.13 0,13 0.13 0.137 0.138 0.137

Y-intercept -0.12 -0.076 -0.084 -0.076 -0.052 0.028 -0.036 -0.024 -0.096

Blood was diluted with either saline only (reference/golden standard), saline 50%/saliva 50% or saliva only.

For every single donor, the values from the different measurements showed a high degree of correlation, reflecting that saliva did not have any impact on the measurements.

Figure 1

Retained haemoglobin values at different blood concentrations from the measurements with the whole blood system: A = donor 1; B = donor 2; C = donor 3.

Twenty-five measurements were done for saline 100%, saliva 50% and saliva 100% each.

Not all results are distinguishable in the figure due to overlapping, indicating high correlation of the measurements.

Analyses of saliva diluted in saline only Description of the lowest saliva concentration required (0 to 100%) to get elevated results (> 0 g/dL) from the analyses. Saliva 0 to 100% did not affect the whole blood measurement device. Saliva 25 to 75% or more affected the low haemoglobin device. Analyses made with the whole blood system of blood diluted with saliva and/or saline Blood was diluted with either saline only (reference/golden standard), saline 50%/saliva 50% or saliva only. For every single donor, the values from the different measurements showed a high degree of correlation, reflecting that saliva did not have any impact on the measurements. Retained haemoglobin values at different blood concentrations from the measurements with the whole blood system: A = donor 1; B = donor 2; C = donor 3. Twenty-five measurements were done for saline 100%, saliva 50% and saliva 100% each. Not all results are distinguishable in the figure due to overlapping, indicating high correlation of the measurements. Retained haemoglobin values at different blood concentrations from the measurements with the low haemoglobin system: A = donor 2; B = donor 3; C = donor 4. Twenty-five measurements were done for saline 100%, saliva 12.5% and saliva 100% each. Not all results are distinguishable due to overlapping, indicating high correlation of the measurements. The more saliva a sample consisted of, the higher were the haemoglobin levels reported from the low haemoglobin system, given that the blood concentration was unaltered and that the blood came from the same donor. Low haemoglobin system (HemoCue Saliva from five subjects (donor 1, 2, 4 - 6) diluted with saline only was analysed, showing that the HemoCue® Plasma/Low Hb system reacted on saliva concentrations of more or equal to 25 - 75%, depending on saliva donor (Table 1). Results from the series of samples from donor 2 to 4 containing blood mixed with saliva and/or saline revealed that the correlations were overall close to one and the slopes for measurements within a single donor showed high consistency. Because the intended measure range of the HemoCue® Plasma/Low Hb system is close to zero (0.03 to 3 g/dl), the trend of differences in Y-intercept observed within the measurements from every donor could therefore be used as a measurement of the impact of saliva (Table 3). The more saliva a blood sample contained the higher were the haemoglobin levels reported from the HemoCue® Plasma/Low Hb system, given that the blood concentration was unaltered and that the blood came from the same donor (Figure 2). Limits of detection for the HemoCue® Plasma/Low Hb system were in the range of 0.03 to 0.1 g/dl, and thus quite concordant with the information stated by the manufacturer (the system should be linear between 0.03 and 3 g/dl).
Table 3

Analyses made with the low haemoglobin system of blood diluted with saliva and/or saline

Donor 2 (n = 25) 3 (n = 25) 4 (n = 25)




Diluent Saline Saline 87.5%, saliva 12.5% Saliva Saline Saline 87.5%, saliva 12.5% Saliva Saline Saline 87.5%, saliva 12.5% Saliva
Spearman´s rank correlation 0.99 0.983 0.984 0.987 0.988 0.988 0.987 0.986 0.984

Pearson correlation 1 1 1 1 1 1 1 1 0.999

Slope 13.261 13.445 13.051 12.568 12.437 12.469 16.271 16.39 16.118

Y-intercept -0.013 0.007 0.099 -0.006 0.009 0.05 -0.01 0.011 0.092

Blood was diluted with either saline only (reference/golden standard), saline 87.5%/saliva 12.5% or saliva only. The correlations were overall close to one and the slopes for measurements within a single donor showed high consistency.

The trend of differences in Y-intercept that could be observed within the measurements from every donor reflects the impact of saliva on the measurements.

Figure 2

Retained haemoglobin values at different blood concentrations from the measurements with the low haemoglobin system: A = donor 2; B = donor 3; C = donor 4.

Twenty-five measurements were done for saline 100%, saliva 12.5% and saliva 100% each.

Not all results are distinguishable due to overlapping, indicating high correlation of the measurements. The more saliva a sample consisted of, the higher were the haemoglobin levels reported from the low haemoglobin system, given that the blood concentration was unaltered and that the blood came from the same donor.

Analyses made with the low haemoglobin system of blood diluted with saliva and/or saline Blood was diluted with either saline only (reference/golden standard), saline 87.5%/saliva 12.5% or saliva only. The correlations were overall close to one and the slopes for measurements within a single donor showed high consistency. The trend of differences in Y-intercept that could be observed within the measurements from every donor reflects the impact of saliva on the measurements.

DISCUSSION

The present study investigated how haemoglobin measurements were affected by various mixtures of blood, saliva, and saline. The results suggested that haemoglobin analyses made with a whole blood measurement device were not influenced by saliva when compared to reference samples diluted with saline. When measuring haemoglobin with a spectrophotometer developed for low haemoglobin measurements, a slight increase in the reported haemoglobin levels was observed compared to reference samples diluted with saline, especially for the combination of low blood concentrations and high saliva concentrations. These false high haemoglobin results could possibly be related to the protein phase of the saliva and the settings of the absorbance measurement device. Because the correlations for all measurements, including those of the saliva-diluted samples, were close to one, one could expect that the small impact of saliva could be foreseen, and thereby adjusted for. One other explanation for the fact that saliva only seemed to affect measurements performed with the device for low haemoglobin measurements could be that the HemoCue® Plasma/Low Hb spectrophotometer used in the study presents data with one more decimal than the instrument used for whole blood measurements. Both devices are factory-calibrated against the International Council for Standardization in Haematology (ICSH) reference method and consequently the blood samples that were diluted with saline only to different haemoglobin concentrations (Table 2 and 3), may be used as reference values. The descriptive statistics that were used to generate the results (Table 2 and 3, Figure 1 and 2) displayed obvious linear relationships. Consequently, there was no need for any interferential statistical analysis and no P-values are presented in this study. The results from the present study could be judged as general even though the study had few participants because the correlation for all measurements were close to one. To test if the protein phase of saliva would interfere with the measurements, resting saliva, containing more proteins than stimulated saliva, was used [15]. However, during OMFS stimulated saliva might be produced and normal stimulated salivary flow is reported 0.63 to 2.94 ml/min [16,17]. To diminish the saliva impact on haemoglobin measurements in solutions with low blood proportions (haemoglobin levels ≤ 0.3 g/dl), one could suggest adding saline before measurements with a HemoCue® Plasma/Low Hb spectrophotometer are performed. Clinically assessment of blood loss is often based on visual estimation and subtraction of saline from the liquid collected. However, the results from the present study suggest that a portable spectrophotometer could measure haemoglobin levels with high precision even when blood is mixed with saline and saliva. This finding supports the possibility of performing accurate blood loss measurements in OMFS. The blood loss may be calculated by multiplying the total volume of the liquid collected in the suction bottle by its haemoglobin value, and then divide the product by the haemoglobin value from a preoperatively collected venous blood sample of the patient [6,7]. The total volume of the liquid in the suction canister could be visually estimated by using the reference levels marked on the canister. Another, perhaps more accurate way, of establishing the total volume of the liquid in the suction bottle is to measure the mass of the collected fluid by weighting the filled bottle after the treatment and subtracting the weight of an empty suction bottle [8]. The haemoglobin value of the liquid in the suction canister may be analysed using a portable spectrophotometer. The patient´s venous haemoglobin value will exhibit minor variations during surgery due to infusion of fluids into the blood stream. These small haemoglobin changes are in most cases not expected to influence the calculation of blood loss described, and thus a preoperative venous haemoglobin value is often sufficient when calculating the blood loss [19]. But if significant blood loss has occurred and the surgery has lasted for several hours, utilisation of the mean haemoglobin value from a venous blood sample collected preoperatively and another sample collected postoperatively could be suggested in order to increase the accuracy of the calculation. Other possible applications of the method described in this study could be measurements of bleeding tendency caused by different drugs or comparison of different haemostatic agents. It might be appropriate to perform such measurements intraorally because bleeding tendency is related to temperature [20] and the temperature of the oral cavity could be expected to show less variation than the temperature of the skin [21,22]. Factors influencing the measurements of the spectrophotometer such as clot formation and debris accumulation could easily be counteracted by application of anticoagulants in the solution and usage of filtering traps. A significant limitation of the clinical utility of the method is however the operator´s ability to collect the entire blood loss during surgery. Making the operator aware of the need to properly collect the blood loss and to avoid the interference of blood accumulation within different sites such as surgical sponges may reduce this problem.

CONCLUSIONS

Measurements of haemoglobin concentrations using a device for whole blood measurements in mixtures of blood, saline, and saliva are reliable and comparable to measurements in mixtures of blood and saline and could be recommended for haemoglobin levels ≥ 0.3 g/dl. For haemoglobin levels ≤ 0.3 g/dl in mixtures of blood, saline, and saliva utilisation of a spectrophotometer intended for low haemoglobin measurements could be recommended if the mixture is well diluted with saline prior to any measurements in order to diminish the saliva impact. Given these differences, it is suggested that both systems can be applied in the field of oral and maxillofacial surgery, which could improve blood loss measurements and support the surgeon and anaesthesiologist in avoiding hypovolemia and determine fluid management.
  22 in total

1.  The novel HemoCu plasma/low hemoglobin system accurately measures small concentrations of three different hemoglobin-based oxygen carriers in plasma: hemoglobin glutamer-200 (bovine) (Oxyglobin), hemoglobin glutamer-250 (bovine) (Hemopure), and hemoglobin-Raffimer (Hemolink).

Authors:  Fedor Lurie; Jonathan S Jahr; Bernd Driessen
Journal:  Anesth Analg       Date:  2002-10       Impact factor: 5.108

Review 2.  Saliva as research material: biochemical, physicochemical and practical aspects.

Authors:  Raymond G Schipper; Erika Silletti; Monique H Vingerhoeds
Journal:  Arch Oral Biol       Date:  2007-08-10       Impact factor: 2.633

3.  Intraoperative blood loss and blood transfusion requirements in patients undergoing orthognathic surgery.

Authors:  Leonardo Perez Faverani; Gabriel Ramalho-Ferreira; André Luis Silva Fabris; Tárik Ocon Braga Polo; Guilherme Henrique Souza Poli; Cláudio Maldonado Pastori; Clóvis Marzola; Wirley Gonçalves Assunção; Idelmo Rangel Garcia-Júnior
Journal:  Oral Maxillofac Surg       Date:  2013-04-26

4.  Comparison of the accuracy of noninvasive hemoglobin monitoring by spectrophotometry (SpHb) and HemoCue® with automated laboratory hemoglobin measurement.

Authors:  Lionel Lamhaut; Roxana Apriotesei; Xavier Combes; Marc Lejay; Pierre Carli; Benoît Vivien
Journal:  Anesthesiology       Date:  2011-09       Impact factor: 7.892

5.  Use of a new HemoCue system for measuring haemoglobin at low concentrations.

Authors:  L D Morris; A Pont; S M Lewis
Journal:  Clin Lab Haematol       Date:  2001-04

6.  Operative time, blood loss, hemoglobin drop, blood transfusion, and hospital stay in orthognathic surgery.

Authors:  Ra'ed Ghaleb Salma; Fahad Mohammed Al-Shammari; Bishi Abdullah Al-Garni; Mohammed Abdullah Al-Qarzaee
Journal:  Oral Maxillofac Surg       Date:  2017-05-02

Review 7.  The role of ethylenediamine tetraacetic acid (EDTA) as in vitro anticoagulant for diagnostic purposes.

Authors:  Giuseppe Banfi; Gian Luca Salvagno; Giuseppe Lippi
Journal:  Clin Chem Lab Med       Date:  2007       Impact factor: 3.694

8.  Skin temperature measurement using an infrared thermometer on patients who have been exposed to cold.

Authors:  Ufuk Erenberk; Emel Torun; Emin Ozkaya; Selcuk Uzuner; Aysegul Dogan Demir; Rusen Dundaroz
Journal:  Pediatr Int       Date:  2013-12       Impact factor: 1.524

9.  Blood Density Is Nearly Equal to Water Density: A Validation Study of the Gravimetric Method of Measuring Intraoperative Blood Loss.

Authors:  Dominic J Vitello; Richard M Ripper; Michael R Fettiplace; Guy L Weinberg; Joseph M Vitello
Journal:  J Vet Med       Date:  2015-01-29

10.  Measurement of Intraoperative Blood Loss in Pediatric Orthopaedic Patients: Evaluation of a New Method.

Authors:  Philip D Nowicki; Akunne Ndika; John Kemppainen; Jeffrey Cassidy; Michael Forness; Siddarth Satish; Nabil Hassan
Journal:  J Am Acad Orthop Surg Glob Res Rev       Date:  2018-05-08
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.