Blood Density Is Nearly Equal to Water Density: A Validation Study of the Gravimetric Method of Measuring Intraoperative Blood Loss.

Dominic J Vitello1, Richard M Ripper1, Michael R Fettiplace1, Guy L Weinberg1, Joseph M Vitello2.   

Abstract

Purpose. The gravimetric method of weighing surgical sponges is used to quantify intraoperative blood loss. The dry mass minus the wet mass of the gauze equals the volume of blood lost. This method assumes that the density of blood is equivalent to water (1 gm/mL). This study's purpose was to validate the assumption that the density of blood is equivalent to water and to correlate density with hematocrit. Methods. 50 µL of whole blood was weighed from eighteen rats. A distilled water control was weighed for each blood sample. The averages of the blood and water were compared utilizing a Student's unpaired, one-tailed t-test. The masses of the blood samples and the hematocrits were compared using a linear regression. Results. The average mass of the eighteen blood samples was 0.0489 g and that of the distilled water controls was 0.0492 g. The t-test showed P = 0.2269 and R (2) = 0.03154. The hematocrit values ranged from 24% to 48%. The linear regression R (2) value was 0.1767. Conclusions. The R (2) value comparing the blood and distilled water masses suggests high correlation between the two populations. Linear regression showed the hematocrit was not proportional to the mass of the blood. The study confirmed that the measured density of blood is similar to water.

Entities:  

Year:  2015        PMID: 26464949      PMCID: PMC4590883          DOI: 10.1155/2015/152730

Source DB:  PubMed          Journal:  J Vet Med        ISSN: 2314-6966


1. Introduction

The quantity of intraoperative blood loss is often determined using subjective visual estimation by the operating surgeon or anesthesiologist. Several studies document that this visual estimation may underestimate actual blood loss by as much as 89% [1-5]. Some studies also report that as actual blood loss increases, estimated blood loss is increasingly inaccurate [6]. Accurately measuring blood loss during an operation can assist with fluid resuscitation and the need for transfusion. Overestimation of blood loss can lead to unnecessary transfusion practices. For research purposes, precise measurement of intraoperative blood loss is critical to compare different operative techniques or the effects of medications on blood loss. During most operations, the direct measurement of the volume of blood collected in suction canisters is the most common clinical method used to determine intraoperative blood loss. When the volume of blood loss is small, the gravimetric method is a simple, accurate, and clinically relevant measurement technique [7]. With this method, surgical gauze sponges and laparotomy pads are weighed before and after use. The difference in their weight is generally believed to be an accurate measurement of blood loss. To convert the mass of blood to a more familiar volume statistic, knowledge of blood density is required. The density of blood is generally estimated to be one gram per milliliter [2, 7–9]. However, there is a paucity of documentation that the one milliliter of blood weighs one gram relationship is accurate or if it holds true for varying hematocrits. Existing data are conflicting and demonstrate blood density ranges between 1043 and 1060 kg/m3 (1.043–1.060 g/mL) [10]. The current study aims to determine the density of blood in an animal model and correlate it with the hematocrit. This will test the assumption that the specific gravity of blood is equal to water. If the relationship is valid and does not vary with hematocrit, it confirms the accuracy of weighing surgical sponges to determine intraoperative blood loss.

2. Materials and Methods

Rat blood used for this study was sampled from blood drawn as part of a protocol approved by the Animal Care Committee and Biologic Resources Laboratory at the University of Illinois at Chicago and the Institutional Animal Care and Utilization Committee of the Jesse Brown VA Medical Center (Chicago, Illinois). Male Sprague-Dawley rats, being utilized for another experiment and weighing between 385 and 435 grams, were sedated using atmospheric room air and 2% isoflurane. The rats were then placed on a heated surgical stage under a lightless heat lamp. The trachea was intubated and the rat ventilated with 1.7% isoflurane in oxygen. Cannulas were placed in both common carotid arteries and the left internal jugular vein. Heparinized microhematocrit tubes were filled with arterial whole blood from the cannulated left common carotid artery. The tubes were sealed at one end using two layers of white clay and centrifuged (International Equipment Company, Chattanooga, TN, Micro-Capillary Centrifuge, Model MB) at 10,000 revolutions per minute for 5 minutes. The hematocrit was read from the packed cell volume using a microcapillary reader (Damon/IEC Division Micro-Capillary Reader). A few drops (<0.2 mL) of arterial whole blood obtained from the carotid artery were placed on a plastic weigh dish. A 50 µL aliquot of this sample was collected using a disposable plastic micropipette (Fisher Scientific Fisherbrand, Loughborough, Leicestershire, United Kingdom, Finnpipette). This sample was placed on an electronic balance (Ohaus Corporation, Parsippany, NJ, Voyager) which had been previously zeroed, and the mass determined. Eighteen samples were collected and documented in this manner. Eighteen control trials using deionized water were also performed on the same electronic scale utilizing the identical technique as the blood samples. All measurements were determined at room temperature. Statistical methods included a linear regression to provide a model for testing the correlation between blood density and hematocrit. Additionally, a Student's unpaired, one-tailed t-test with a Welch's correction was used to compare the means of the blood and distilled water masses. Data are reported as the mean ± SD.

3. Results and Discussion

The masses of the 18 blood aliquots are recorded in Table 1. The average mass of the 50-microliter blood samples was 0.0489 ± 0.032 g with a range of 0.0445 g to 0.0512 g. Table 2 lists the masses of the 18 distilled water controls. The average mass of the 50-microliter water samples was 0.0492 g with a range from 0.0487 g to 0.0496 g. The average density of the blood samples was normalized to the average mass of the distilled water controls by dividing the two means to yield specific gravity and an adjusted mean blood density of 0.994 ± 0.032 g/mL. The average hematocrit of the blood samples was 40.3% (range 24%–48%). Figure 1 shows the mass of the 50 µL samples of blood as a function of hematocrit. The R 2 value was 0.1767. Therefore, there was poor correlation between the hematocrits and the density of the blood samples. Populations were deemed fit for parametric testing but their variances were unequal by ANOVA (P < 0.00001). To account for this a Student's t-test with a Welch's correction was utilized to compare the blood and distilled water samples. The results yielded a P = 0.2269 and R 2 = 0.03154. This analysis verifies the fact that the mass of the blood and water samples was not significantly different.
Table 1

Masses of the eighteen rat blood samples.

50 μL samples of blood
NumberHematocritMass (g)
145.00.0496
231.00.0479
344.50.0512
437.00.0506
548.00.0502
638.50.0499
746.50.0509
839.00.0497
947.00.0492
1045.00.0478
1142.50.0493
1232.00.0474
1346.00.0475
1439.00.0445
1545.00.0495
1631.00.0480
1724.00.0475
1844.00.0495

Average40.30.0489
Max48.00.0512
Min24.00.0445
Range24.00.0067
SD6.90.0016
Table 2

Masses of the eighteen deionized water controls.

50 μL deionized water controls
Control numberMass (g)
10.0495
20.0495
30.0494
40.0496
50.0496
60.0487
70.0494
80.0495
90.0490
100.0489
110.0491
120.0492
130.0489
140.0490
150.0492
160.0491
170.0493
180.0487

Average0.0492
Max0.0496
Min0.0487
Range0.0009
SD0.0003
Figure 1

Graph correlating rat blood mass to blood hematocrit.

The surgeon or anesthesiologist routinely estimates the volume of blood lost during each operation. This estimate can guide transfusion requirements or assist in the amount of intravenous resuscitation required to maintain euvolemia. Accurate measurement of intraoperative blood loss also allows the comparison of different surgical techniques and methods to minimize blood loss during surgery. Inaccurate determination of intraoperative blood loss can lead to unnecessary transfusions. There is ample evidence to suggest that blood is immunosuppressive [11, 12] and has been associated with worse outcomes in colorectal cancer [13]. Blood transfusion also increases the risk of surgical site infections [14] and is expensive. There are many methods to determine the volume of intraoperative blood loss. The most precise method utilizes spectrophotometry [15-17]. This method prepares a referenced standard of the patient's own blood. Blood lost into drapes, gowns, and sponges is extracted and filtered. The resulting sample is then read in a spectrophotometer along with the patient's reference standard. This method is considered highly reliable; however, it is expensive, labor intensive, and clinically impractical. During a typical operation, shed blood is aspirated via a suction system directly into canisters where the blood volume can be accurately measured. Blood lost into sponges and laparotomy pads is also visually estimated by the attending surgeon or anesthesiologist—a routine but subjective and highly inaccurate method [1-5]. The gravimetric method of determining blood loss requires weighing surgical sponges before and after use. The difference in weight is assumed to be the volume of blood lost as measured in milliliters. This assessment is based upon density which is defined as mass per unit volume. Water density is dependent upon temperature. At room temperature of 23°Centigrade, water has a density of 0.997538 g/mL [12]. This translates into the common metric equivalent that 1 gram of water is equal to 1 milliliter. The P value of 0.2269 for this study demonstrates that the two populations are not statistically significantly different. The R 2 for the two populations was 0.03154. This low value demonstrates that the means of the water and the blood samples are statistically similar. The relationship of the mass of blood and the hematocrit has not been clearly established [13, 14]. Our study confirmed that there was poor correlation between the mass of blood and hematocrits varying between 24 and 48 percent (Figure 1). The purpose of this study was to validate the commonly held assumption that the density of blood is one gram per milliliter. We found the average density of blood was 0.994 g/mL ± 0.032 g. Notably, changes in the hematocrit did not affect the density of the blood samples and analysis revealed no relationship between the two parameters. These data support the belief that the density of blood and water is very close and therefore the act of weighing gauze sponges before and after use in the operating room is a reliable guide to determine the volume of blood lost into the gauze. There were some limitations to this study. A 50 µL sample of water should have a mass of 0.05 grams although none of the samples were recorded with this value and probably reflects the error of pipetting and weighing. Aliquots were recorded immediately after dispensing. This minimized any errors due to evaporation. It was also noted that the blood samples adhered to the inside of the micropipette tip, possibly causing the dispensed volume to be less than actually measured. The quantity of blood or water retained on the inside of pipettes after dispensing the fluid was unknown and not measured, but presumably constant. Additional considerations included the fact that the density of water is a function of temperature, nearing a peak value of 999.9720 kg/m3 at 4°C [18]. At a temperature of 23°C, water has a density of 0.997538 g/mL [14] which offers a potential explanation for the control group's deviation from an expected density of 1 gm/mL. More sophisticated methods to determine mass could be used to reproduce and validate the results. A greater sample size may help to confirm the accuracy of the findings and repeating the study across a greater range of hematocrits would provide additional useful information. Finally, a direct comparison of rat and human blood densities is lacking in the literature. Based upon comparisons of other mammalian species, [19, 20] however, it is suggested that the comparative densities between human and rat blood are probably valid.

4. Conclusion

The measured mass of blood is nearly equal to distilled water. This confirms the assumption that the densities of blood and of distilled water are nearly equivalent. This helps verify the use of the gravimetric method of weighing sponges in the operating room to accurately determine blood loss and a one-gram increase in the weight of a blood soaked surgical gauze is equal to one milliliter of blood lost.
  19 in total

1.  Loss of blood at operation; a method for continuous measurement.

Authors:  F W BORDEN
Journal:  Calif Med       Date:  1957-08

2.  Quantification of surgical blood loss.

Authors:  Marcel H Lee; Britt T Ingvertsen; Jolle Kirpensteijn; Asger L Jensen; Annemarie T Kristensen
Journal:  Vet Surg       Date:  2006-06       Impact factor: 1.495

3.  Visually estimated and calculated blood loss in vaginal and cesarean delivery.

Authors:  Irene Stafford; Gary A Dildy; Steven L Clark; Michael A Belfort
Journal:  Am J Obstet Gynecol       Date:  2008-07-17       Impact factor: 8.661

Review 4.  Measurement of blood loss: review of the literature.

Authors:  Mavis N Schorn
Journal:  J Midwifery Womens Health       Date:  2010 Jan-Feb       Impact factor: 2.388

Review 5.  Immunomodulation by blood transfusion: an evolving scientific and clinical challenge.

Authors:  N Blumberg; J M Heal
Journal:  Am J Med       Date:  1996-09       Impact factor: 4.965

6.  Specific gravity of blood and plasma at 4 and 37 degrees C.

Authors:  R J Trudnowski; R C Rico
Journal:  Clin Chem       Date:  1974-05       Impact factor: 8.327

7.  A simple photo-electric method for the estimation of blood loss during surgery.

Authors:  J F Mainland
Journal:  Br J Anaesth       Date:  1966-01       Impact factor: 9.166

8.  Unreliability of intraoperative estimated blood loss in extended sagittal synostectomies.

Authors:  Mitchel Seruya; Albert K Oh; Michael J Boyajian; John S Myseros; Amanda L Yaun; Robert F Keating
Journal:  J Neurosurg Pediatr       Date:  2011-11       Impact factor: 2.375

9.  The association of perioperative blood transfusion with colorectal cancer recurrence.

Authors:  P I Tartter
Journal:  Ann Surg       Date:  1992-12       Impact factor: 12.969

10.  A comparison between visual estimation and laboratory determination of blood loss during the third stage of labour.

Authors:  K Razvi; S Chua; S Arulkumaran; S S Ratnam
Journal:  Aust N Z J Obstet Gynaecol       Date:  1996-05       Impact factor: 2.100

View more
  14 in total

1.  Evidence for exposure to selenium by breeding interior snowy plovers (Charadrius nivosus) in saline systems of the Southern Great Plains.

Authors:  H M Ashbaugh; W C Conway; D A Haukos; D P Collins; C E Comer; A D French
Journal:  Ecotoxicology       Date:  2018-05-29       Impact factor: 2.823

2.  Optical coherence tomography for evaluating capillary waves in blood and plasma.

Authors:  Hsiao-Chuan Liu; Piotr Kijanka; Matthew W Urban
Journal:  Biomed Opt Express       Date:  2020-01-24       Impact factor: 3.732

3.  Role of Prophylactic Tranexamic Acid in Reducing Blood Loss during Elective Caesarean Section: A Randomized Controlled Study.

Authors:  Sj Dhivya Lakshmi; Reena Abraham
Journal:  J Clin Diagn Res       Date:  2016-12-01

4.  Microbubble Size and Dose Effects on Pharmacokinetics.

Authors:  J Angel Navarro-Becerra; Kang-Ho Song; Payton Martinez; Mark A Borden
Journal:  ACS Biomater Sci Eng       Date:  2022-03-31

5.  A Centrifugal Microfluidic Platform That Separates Whole Blood Samples into Multiple Removable Fractions Due to Several Discrete but Continuous Density Gradient Sections.

Authors:  Scott T Moen; Christopher L Hatcher; Anup K Singh
Journal:  PLoS One       Date:  2016-04-07       Impact factor: 3.240

6.  Fluid surface tension evaluation using capillary wave measurement with optical coherence tomography.

Authors:  Hsiao-Chuan Liu; Piotr Kijanka; Matthew W Urban
Journal:  AIP Adv       Date:  2020-05-19       Impact factor: 1.548

7.  Comparing perioperative vaginal misoprostol with intraoperative pericervical hemostatic tourniquet in reducing blood loss during abdominal myomectomy: A randomized controlled trial

Authors:  Muhibat A. Afolabi; Grace G. Ezeoke; Rakiya Saidu; Munirdeen A. Ijaiya; Abiodun S. Adeniran
Journal:  J Turk Ger Gynecol Assoc       Date:  2018-11-30

8.  Exposure of Infants to Isoniazid via Breast Milk After Maternal Drug Intake of Recommended Doses Is Clinically Insignificant Irrespective of Metaboliser Status. A Physiologically-Based Pharmacokinetic (PBPK) Modelling Approach to Estimate Drug Exposure of Infants via Breast-Feeding.

Authors:  Estella Dora Germaine Garessus; Hans Mielke; Ursula Gundert-Remy
Journal:  Front Pharmacol       Date:  2019-01-22       Impact factor: 5.810

9.  Mid-Tibiofibular Amputation as a Method of Terminal Blood Collection in Xenopus Laevis.

Authors:  Benjamin C Dixon; Marilynn J Culbreth; David M Kumsher; Chance M Carbaugh; David P Fetterer; Cara P Reiter
Journal:  J Am Assoc Lab Anim Sci       Date:  2021-07-15       Impact factor: 1.232

10.  High levels of modified ceramides are a defining feature of murine and human cancer cachexia.

Authors:  Pauline Morigny; Julia Zuber; Mark Haid; Doris Kaltenecker; Fabien Riols; Joanna D C Lima; Estefania Simoes; José Pinhata Otoch; Sören Fisker Schmidt; Stephan Herzig; Jerzy Adamski; Marilia Seelaender; Mauricio Berriel Diaz; Maria Rohm
Journal:  J Cachexia Sarcopenia Muscle       Date:  2020-10-08       Impact factor: 12.910

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.