Nadia Sabbah1,2, Peter Wolf1,3, Céline Piedvache4, Séverine Trabado5, Tristan Verdelet1, Catherine Cornu6, Jean-Claude Souberbielle7, Philippe Chanson1.
Abstract
OBJECTIVE: Measurement of IGF-I is important in the management of patients with growth hormone disorders. Here we aim to establish normative data for two new IGF-I assay kits based on a large random sample of the French general adult population. SUBJECTS AND METHODS: We measured IGF-I in 911 healthy adults (18-90 years) with two immunoassays (ROCHE Elecsys® and IMMULITE-2000 calibrated against the new IS 02/2547). We compared the data with those of the six immunoassays (iSYS, LIAISON XL, IMMULITE-2000 calibrated against the first IS 87/518, IGF-I RIACT, Mediagnost ELISA, and Mediagnost RIA) that we reported previously. The pairwise concordance among the eight assays was assessed with Bland-Altman plots for both the IGF-1 raw data and the standard deviation scores (SDS), as well as with the percentage of observed agreement and the weighted Kappa coefficient for categorizing IGF-I SDS (ClinicalTrials.gov Identifier: NCT01831648).Entities:
Keywords: IGF-I; SD score; Z-score; acromegaly; growth hormone deficiency; normal healthy population; normative data; reference range
Year: 2021 PMID: 34343107 PMCID: PMC8428081 DOI: 10.1530/EC-21-0175
Source DB: PubMed Journal: Endocr Connect ISSN: 2049-3614 Impact factor: 3.335
Characteristics of the tested IGF-I assays as provided by the manufacturers. The limit of quantification (LOQ) is the lowest amount of IGF-I that can be accurately quantified with an allowable error of ≤20%. The limit of detection (LOD) is the IGF-I concentration corresponding to the 95th percentile value from a number of determinations of IGF-I concentration in free serum samples.
| Name of the assay | Manufacturer | Automated | Tracer | International standard against which the assay is calibrated | Intra-assay CV | Inter-assay CV | LOQ or LOD in ng/mL | Highest measurable value without dilution (ng/mL) | Reference adult population recruited by the manufacturer |
|---|---|---|---|---|---|---|---|---|---|
| Elecsys® IGF-1 | Roche Diagnostics | Yes | Ruthenium | WHO international standard 02/254 | Human samples between 1.4 and 3.5% | Human samples between 2.5 and 4.6% | LoQ 15 ng/mL | 1600 ng/mL | 3652 men between 3 months and 80 years |
| IMMULITE® 20000 IGF-I | Siemens | Yes | Phosphatase alcaline | WHO international standard 02/254 | Human samples between 2.4 and 6.3% | Human samples between 3 and 7.6% | LoQ 24.9 ng/mL | 90,000 ng/mL | 701 pediatric and 1104 adult samples |
Normative reference intervals (95% CI) of IGF-1 measured by eight assay methods according to age range and sex in a cohort of 899 healthy subjects.
| Age range | iSYS IGF-1 (ng/mL) 95% | LIAISON XL IGF-1 (ng/mL) 95% | IMMULITE 2000 IGF-1 (ng/mL) 95% | IGF-I-RIACT IGF-1 (ng/mL) 95% | Mediagnost ELISA IGF-1 (ng/mL) 95% | Mediagnost RIA IGF-1 (ng/mL) 95% | ROCHE Elecsys® IGF-1 (ng/mL) 95% | IMMULITE new standard IGF-1 (ng/mL) 95% | |
|---|---|---|---|---|---|---|---|---|---|
| Males (years) | |||||||||
| 18–20 | 56 | 168–391 | 186–453 | 195–537 | 197–486 | 177–400 | 168–430 | 173–395 | 117–295 |
| 21–23 | 61 | 147–346 | 168–411 | 171–477 | 173–430 | 159–388 | 150–337 | 152–350 | 105–266 |
| 24–26 | 53 | 132–313 | 153–377 | 152–430 | 155–389 | 144–355 | 135–308 | 136–317 | 96–243 |
| 27–29 | 49 | 122–292 | 142–351 | 138–396 | 143–363 | 133–331 | 126–289 | 125–294 | 89–227 |
| 30–39 | 56 | 108–265 | 124–310 | 118–348 | 127–329 | 115–295 | 112–265 | 110–263 | 80–206 |
| 40–49 | 51 | 91–233 | 106–271 | 98–301 | 107–286 | 98–261 | 97–237 | 92–229 | 70–184 |
| 50–59 | 54 | 81–214 | 97–252 | 85–273 | 94–262 | 88–245 | 86–218 | 82–210 | 61–164 |
| 60–69 | 49 | 75–208 | 92–245 | 77–260 | 87–250 | 80–237 | 82–214 | 76–202 | 56–154 |
| 70–89 | 34 | 64–192 | 80–220 | 66–242 | 75–231 | 71–233 | 72–200 | 65–182 | 52–145 |
| Females (years) | |||||||||
| 18–20 | 41 | 155–421 | 191–483 | 180–586 | 169–517 | 169–487 | 161–412 | 162–419 | 102–324 |
| 21–23 | 54 | 144–383 | 176–448 | 166–541 | 159–476 | 156–446 | 149–379 | 149–384 | 101–304 |
| 24–26 | 45 | 134–353 | 163–418 | 153–501 | 150–440 | 144–412 | 139–353 | 138–354 | 98–286 |
| 27–29 | 48 | 126–330 | 152–391 | 142–467 | 142–410 | 134–385 | 131–332 | 128–329 | 95–272 |
| 30–39 | 47 | 113–294 | 131–345 | 121–403 | 126–356 | 118–341 | 118–298 | 112–286 | 86–243 |
| 40–49 | 50 | 97–253 | 109–296 | 98–331 | 107–297 | 100–296 | 103–258 | 94–242 | 72–203 |
| 50–59 | 54 | 80–209 | 93–253 | 80–271 | 90–247 | 82–248 | 97–220 | 77–203 | 59–165 |
| 60–69 | 47 | 64–170 | 84–222 | 68–227 | 76–209 | 68–208 | 75–190 | 66–174 | 50–140 |
| 70–89 | 50 | 56–154 | 81–204 | 60–188 | 67–189 | 60–187 | 68–175 | 59–159 | 50–134 |
Figure 1Reference intervals (upper panel, males; lower panel, females) according to the age intervals of the eight immunoassays tested. Lower limits (2.5th percentile) and upper limits (97.5th percentile) of the normal range are drawn as full lines and medians as dotted lines. Each assay has a different color.
Figure 2Comparisons between iSYS and ROCHE Elecsys® expressed as scatter plots (A) or Bland-Altman plots (B) for raw data, or scatter plots (C) and Bland-Altman plots (D) for SDS showing a good overall agreement between both immunoassays, with no significant bias. Comparisons between iSYS and Immulite new calibration expressed as scatter plots (E) or Bland-Altman plots (F) for raw data, or scatter plots (G) and Bland-Altman plots (H) for SDS showing a bad overall agreement between these two immunoassays.
Agreement of each IGF-1 assay method against each of the others, with the two new kits IMMULITE (new standard) and ROCHE expressed as weighted Kappa.
| LIAISON XL | iSYS | IMMULITE 2000 | Mediagnost ELISA | Mediagnost RIA | IGF1-RIACT | IMMULITE New standard | ROCHE Elecsys® | |
|---|---|---|---|---|---|---|---|---|
| Liaison XL | – | 0.49 | 0.50 | 0.47 | 0.38 | 0.48 | 0.38 | 0.50 |
| iSYS | 0.49 | – | 0.64 | 0.61 | 0.70 | 0.64 | 0.55 | 0.75 |
| IMMULITE 2000 | 0.50 | 0.64 | – | 0.61 | 0.58 | 0.64 | 0.65 | 0.64 |
| Mediagnost ELISA | 0.47 | 0.61 | 0.61 | – | 0.59 | 0.53 | 0.55 | 0.61 |
| Mediagnost RIA | 0.38 | 0.70 | 0.58 | 0.59 | – | 0.48 | 0.49 | 0.68 |
| IGF1-RIACT | 0.48 | 0.64 | 0.64 | 0.53 | 0.48 | – | 0.55 | 0.62 |
| IMMULITE New standard | 0.38 | 0.55 | 0.65 | 0.55 | 0.49 | 0.55 | – | 0.56 |
| ROCHE Elecsys® | 0.50 | 0.75 | 0.64 | 0.61 | 0.68 | 0.62 | 0.56 | – |