MRI chemical shift artifact produced by center-out radial sampling of k-space: a potential pitfall in clinical diagnosis.

Mark Bydder1, Michael Carl2, Graeme M Bydder3,4, Jiang Du3.   

Abstract

BACKGROUND: Center-out radial sampling of k-space in magnetic resonance imaging employs a different direction for each readout. Off-resonance artifacts (including those produced by chemical shift between water and fat) found with this type of sampling are usually described as blurring, however more specific characterization of these artifacts can be ascertained from the fact that their point spread function is ring-shaped. This produces effects that differ from those seen with Cartesian sampling of k-space. Experiments were designed to demonstrate the origin of these artifacts and a volunteer was imaged to show them.
METHODS: Two phantoms containing oil in a syringe and an annulus of oil surrounded by water were scanned with a range of bandwidths from 62.5 down to 4 kHz. In a human volunteer, head, pelvis and spine images were obtained with bandwidths of 62.5 and 4 kHz.
RESULTS: The two phantoms showed displacement of the oil signal away from the center into the region of the surrounding water. The effect increased as the bandwidth was decreased. In the head of the volunteer, signal from fat in red bone marrow in the skull was displaced centrally and peripherally relative to water within the marrow, and appeared in the region between the skull and the brain, as well as in the surrounding scalp. Displacements of the former type simulated subdural hematomas. Displacement of perivesical fat signal centrally over the wall of the bladder simulated bladder tumor, and displacement of fat signal from red bone marrow in the lumbar spine to the intervertebral discs simulated their cartilaginous endplates.
CONCLUSIONS: Center-out radial artifacts are important to recognize on clinical images since they may mimic anatomy and simulate pathology. The article shows how these artifacts originate, includes examples, and describes how the artifacts differ from Cartesian chemical shift artifacts. 2021 Quantitative Imaging in Medicine and Surgery. All rights reserved.

Entities:  

Keywords:  Radial center-out sampling of k-space; chemical shift artifact; fat imaging; magnetic resonance imaging

Year:  2021        PMID: 34341741      PMCID: PMC8245944          DOI: 10.21037/qims-21-115

Source DB:  PubMed          Journal:  Quant Imaging Med Surg        ISSN: 2223-4306


  10 in total

Review 1.  Chemical shift: the artifact and clinical tool revisited.

Authors:  M N Hood; V B Ho; J G Smirniotopoulos; J Szumowski
Journal:  Radiographics       Date:  1999 Mar-Apr       Impact factor: 5.333

2.  Chemical-shift imaging utilizing the positional shifts along the readout gradient direction.

Authors:  M I Altbach; T P Trouard; R Van de Walle; R J Theilmann; E Clarkson; H H Barrett; A F Gmitro
Journal:  IEEE Trans Med Imaging       Date:  2001-11       Impact factor: 10.048

Review 3.  Magnetic resonance: an introduction to ultrashort TE (UTE) imaging.

Authors:  Matthew D Robson; Peter D Gatehouse; Mark Bydder; Graeme M Bydder
Journal:  J Comput Assist Tomogr       Date:  2003 Nov-Dec       Impact factor: 1.826

4.  Lung parenchyma: projection reconstruction MR imaging.

Authors:  C J Bergin; J M Pauly; A Macovski
Journal:  Radiology       Date:  1991-06       Impact factor: 11.105

5.  Signal decay correction in 2D ultra-short echo time imaging.

Authors:  Detlef Mentrup; Holger Eggers
Journal:  MAGMA       Date:  2006-05-09       Impact factor: 2.310

6.  Three-dimensional radial ultrashort echo-time imaging with T2 adapted sampling.

Authors:  Jürgen Rahmer; Peter Börnert; Jan Groen; Clemens Bos
Journal:  Magn Reson Med       Date:  2006-05       Impact factor: 4.668

7.  Highly localized positive contrast of small paramagnetic objects using 3D center-out radial sampling with off-resonance reception.

Authors:  Peter R Seevinck; Hendrik de Leeuw; Clemens Bos; Chris J G Bakker
Journal:  Magn Reson Med       Date:  2011-01       Impact factor: 4.668

8.  Multifrequency interpolation for fast off-resonance correction.

Authors:  L C Man; J M Pauly; A Macovski
Journal:  Magn Reson Med       Date:  1997-05       Impact factor: 4.668

Review 9.  Short-T2 MRI: Principles and recent advances.

Authors:  Markus Weiger; Klaas P Pruessmann
Journal:  Prog Nucl Magn Reson Spectrosc       Date:  2019-07-30       Impact factor: 9.795

10.  Depiction of the Periosteum Using Ultrashort Echo Time Pulse Sequence with Three-Dimensional Cone Trajectory and Histologic Correlation in a Porcine Model.

Authors:  Dae Joong Kim; Kun Hwang; Hun Kim; Jang Gyu Cha; Hyungseok Jang; Ju Yong Park; Yeo Ju Kim
Journal:  Korean J Radiol       Date:  2021-02-09       Impact factor: 3.500

  10 in total
  1 in total

Review 1.  Ultrashort Echo Time Magnetic Resonance Imaging Techniques: Met and Unmet Needs in Musculoskeletal Imaging.

Authors:  Amir Masoud Afsahi; Yajun Ma; Hyungseok Jang; Saeed Jerban; Christine B Chung; Eric Y Chang; Jiang Du
Journal:  J Magn Reson Imaging       Date:  2021-12-28       Impact factor: 5.119

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.