Penetration and emanation of light into tissue are limited by the strong interaction of light with the tissue components, especially oxygenated hemoglobin and white adipose tissue. This limits the possibilities for all-optical minimal invasive sensing. In order to minimize the optical losses of light in and out of the tissue, only a narrow optical window between 630 and 900 nm is available. In this work, we realized for the first time all-optical temperature sensing within the narrow optical window for tissue by using the process of triplet-triplet annihilation photon energy upconversion (TTA-UC) as a sensing tool. For this, we apply the asymmetrical benzo-fused BODIPY dye as an optimal emitter and mixed palladium benzo-naphtho-porphyrins as an optimal sensitizer. The TTA-UC sensing system is excited with λ = 658 nm with an extremely low intensity of 1 mW × cm-2 and is factual-protected for a time period longer than 100 s against oxygen-stimulated damage, allowing a stable demonstration of this T-sensing system also in an oxygen-rich environment without losing sensitivity. The sensing dyes we embed in the natural wax/natural matrix, which is intrinsically biocompatible, are approved by the FDA as food additives. The demonstrated temperature sensitivity is higher than ΔT = 200 mK placed around the physiologically relevant temperature of T = 36 °C.
Penetration and emanation of light into tissue are limited by the strong interaction of light with the tissue components, especially oxygenated hemoglobin and white adipose tissue. This limits the possibilities for all-optical minimal invasive sensing. In order to minimize the optical losses of light in and out of the tissue, only a narrow optical window between 630 and 900 nm is available. In this work, we realized for the first time all-optical temperature sensing within the narrow optical window for tissue by using the process of triplet-triplet annihilation photon energy upconversion (TTA-UC) as a sensing tool. For this, we apply the asymmetrical benzo-fused BODIPY dye as an optimal emitter and mixed palladium benzo-naphtho-porphyrins as an optimal sensitizer. The TTA-UC sensing system is excited with λ = 658 nm with an extremely low intensity of 1 mW × cm-2 and is factual-protected for a time period longer than 100 s against oxygen-stimulated damage, allowing a stable demonstration of this T-sensing system also in an oxygen-rich environment without losing sensitivity. The sensing dyes we embed in the natural wax/natural matrix, which is intrinsically biocompatible, are approved by the FDA as food additives. The demonstrated temperature sensitivity is higher than ΔT = 200 mK placed around the physiologically relevant temperature of T = 36 °C.
The many biochemical
reactions responsible for cellular functions,
which are either exothermic or endothermic, are fundamentally co-regulated
by the intracellular temperature distribution. In addition, they are
exposed to different oxygen conditions depending on the particular
areas within cell organelles, at which they take place.[1,2] In an ideal case, the minimally invasive thermometry could be used
to probe many functional characteristics of biological specimens,
their physiological behavior under various conditions, and their responses
to external stimuli, such as chemical and environmental stress.[3] A series of compendious reviews have discussed
the progress in biocompatible temperature measurements. Optical methods[4] for temperature sensing are less invasive and
able to provide a time-resolved and two-dimensional spatial evolution
of the temperature distribution of a living cell.[5,6]Optically excited chromophores in the triplet state can be used
for applications in various fields, like bioimaging,[7] molecular sensing,[8] and photocatalytic
organic reactions.[9] The process of triplet–triplet
annihilation photon energy upconversion (TTA-UC) demonstrates good
prospects for temperature-sensing applications based on optically
excited triplet ensembles. This all-optical sensing technique, supported
by ratiometric-type signal registration, ensures relative independence
of the data obtained on small excitation intensity instabilities,
local molecular concentration variations, and field-of-view uncertainties
for the temperature region centered at the physiologically important
temperature of 36 °C.[10]Briefly,
the TTA-UC process is performed in a multi-chromophore
system built of energetically optimized pairs of sensitizers (metallated
macrocycles) and emitter molecules (aromatic hydrocarbons), as shown
in Figure .[10] Photon energy absorbed by the sensitizer (dark
red arrow, Figure ) is stored into the triplet state, created during the process of
intersystem crossing (ISC). As a next step, stored energy is transferred
to an emitter triplet state via the process of triplet–triplet
transfer. Furthermore, the excited triplet states of two emitter molecules
go through the triplet–triplet annihilation (TTA) process:
so, one emitter molecule relaxes to its singlet ground state, but
the other molecule gains the energy of both triplet states and populate
the excited emitter singlet state. After radiative relaxation of the
emitter singlet state to the ground state, a delayed emitter fluorescence
(red arrow, Figure , called shortly dF), bearing higher energy than that
of the excitation photon, is emitted. If triplet manifolds of the
emitter and sensitizer molecules are not optimally overlapped or if
the molecular rotational diffusion of the interacting sensitizer/emitter
triplet moieties is not high enough, complete depopulation of the
sensitizer triplet state does not happen: simultaneously, a residual
sensitizer phosphorescence (violet arrow, Figure , called shortly rPh) will be
observed.[11,12]
Figure 1
Simplified energetic scheme of the triplet–triplet
annihilation
upconversion process in an oxygen-rich environment. Inset: chemical
structures of the sensitizer-mixed palladium benzo-naphtho-porphyrins, n = 1,0 (PdBNP); emitter—MPh-MB-BODIPY.
Simplified energetic scheme of the triplet–triplet
annihilation
upconversion process in an oxygen-rich environment. Inset: chemical
structures of the sensitizer-mixed palladium benzo-naphtho-porphyrins, n = 1,0 (PdBNP); emitter—MPh-MB-BODIPY.The efficiency and sustainable operation of the TTA-UC process
depend drastically on the presence of oxygen molecules, known as effective
quenchers of the excited triplet states. Blends of natural waxes/oils
with pronounced singlet oxygen scavenging properties, containing TTA-UC
molecules, allow for almost complete chemically binding of the locally
dissolved molecular oxygen.[13] Thus, during
the excitation, the optically assessed spot is almost oxygen free,
and the temperature-sensing procedure can be performed in a sustainable
manner. Employing a matrix consisting of natural waxes/oils ensures
simultaneously the ability to tune the temperature-sensitivity range
toward the biologically relevant temperature window (centered at T = 36 °C) and to use natural biocompatible materials
(all used waxes/oils are approved from FDA as food additives).Despite the demonstrated experimental progress of the TTA-UC process
as an all-optical sensing tool, efficiently protected against the
influence of the local oxygen concentration on the provided temperature
data—there is a significant problem preventing straightforward
application of the TTA-UC sensing technology in vitro: the absorption and scattering properties of the human skin.There is a broad consensus[6−8] that the optical parameters, as
optical absorption and scattering of the living tissue of a particular
person, are subject to variations in the blood content, water content,
and collagen content, and the fiber development. In order to keep
the electromagnetic stress of the patient skin on an acceptable level
and to be minimally invasive, the targeted UC-sensing materials must
fulfill a chain of very specific requirements: (1) the living organisms
develop and accommodate to light intensities close to 1 Sun; therefore,
the excitation intensity of the TTA-UC process must be comparable
with it; (2) only excitation wavelengths, which coincide with the
transparency window of the different components of the human skin
penetrate optimally; and (3) simultaneously, in order to keep optical
losses low, the emission wavelengths of the optical signals must coincide
with the tissue transparency window. Figure demonstrates the optical properties of two
components of the human skin, for which absorption spectra are mostly
limiting the optical access: oxygenated hemoglobin (HbO2, the red curve) and purified white adipose tissue (WAT, the gray
curve).
Figure 2
Molar extinction coefficient for different breast tissue components
as follows: oxygenated hemoglobin (HbO2, dark red line,
in water) and purified WAT (gray line) compared with the emission
spectral range of the signals of delayed emitter fluorescence (dF, the red line) and residual sensitizer phosphorescence
(rPh, the violet line) excited in the upconversion regime,
using deep-red excitation light with an extremely low excitation intensity
of 1 mW × cm–2.
Molar extinction coefficient for different breast tissue components
as follows: oxygenated hemoglobin (HbO2, dark red line,
in water) and purified WAT (gray line) compared with the emission
spectral range of the signals of delayed emitter fluorescence (dF, the red line) and residual sensitizer phosphorescence
(rPh, the violet line) excited in the upconversion regime,
using deep-red excitation light with an extremely low excitation intensity
of 1 mW × cm–2.All these requirements predetermine a new, non-orthodox optimization
strategy for the process of TTA-UC: until now, all synthetic efforts[14] were directed toward as possible high anti-Stokes
shift of the UC-delayed fluorescence signal. The anti-Stokes shift
of the signal of delayed fluorescence is ΔEaS ∼ 0.55–0.7 eV. In this respect, in order
to squeeze the complete TTA-UC spectrum into the limited human skin
transparency window, it is essential to minimize the anti-Stokes shift
(the studied TTA-UC system demonstrates at least four times smaller
ΔEaS ∼ 0.08–0.15 eV).
The UC-fluorescence signal with central emission wavelength λ
≤ 620 nm is strongly absorbed (Figure , please refer to the HbO2—absorption).
Even, if such a delayed fluorescence signal is generated into the
studied tissue, only a small part of this emission will be able to
escape out. Similarly, if the central emission wavelength of the residual
sensitizer phosphorescence is λ ≥ 900 nm (please refer
to WAT-absorption/optical scattering, Figure ), the phosphorescence signal experiences
similar problems.
Results and Discussion
Efficient
TTA-UC was demonstrated with various sensitizer molecules;
in most cases, these were Pd-porphyrins, while simple octaethyl- and
tetraphenylporphyrins show Q-band absorption in the green region,[17] benzoannulatedporphyrins (benzo-,[18] naphtha-,[19] and anthra-[20]) have a Q-band absorption in red, deep-red,
and IR-A region, respectively. However, a symmetric benzoannulation
on all four positions of a porphyrin ring leads to a drastic bathochromic
shift of the absorption (80–100 nm), stepwise annulation of
one, two, or three benzene moieties allows small shifts of Q-band
absorption in order of 20–30 nm.[15] It was demonstrated that such asymmetric porphyrins act as efficient
sensitizers in TTA-UC.[18,19] A similar synthetic strategy
was applied for the UC emitters: for each sensitizer, a suitable emitter
with the highest UC efficiency could be prepared by modification of
the π-core of anthracene,[21] tetracene,[22] perylene,[23] or BODIPY[24] dyes.In the present paper, we combined
a mixed pyrrole condensation
strategy, previously known for porphyrins[15,19] with benzo-annulation on a pyrrole ring, for the synthesis of new
core-modified BODIPY dye having a high fluorescence quantum yield,
good photochemical stability, and acting as an efficient singlet emitter
in the TTA-UC process with mixed palladium benzo-naphtho-porphyrins
as a sensitizer. BODIPY was chosen on purpose because the energy position
of the triplet state[25] is laying relatively
high; thus, the anti-Stokes shift of the resulting UC-emission was
expected to be low.Modification of BODIPY dyes via π-extension
is a known method
to shift their absorption bathochromically.[25] Introduction of a phenyl ring is a common way to make monofunctional
dyes, whereas a substitution of pyrrole with aryl groups at the alpha
position shifts the absorption significantly.[26] Another way is the benzoannulation,[27] similar to porphyrins and perylene dyes.[28] For the double annulation, the same synthetic precursors as for
tetrabenzoporphyrins can be used. Recently, monobenzo-BODIPY was prepared
by a reaction with tetrahydroisoindole with formylpyrrole.[29]Here, we used two pyrroles and one aldehyde
to obtain a statistical
acid-mediated condensation with subsequent separation of the products
by column chromatography (Scheme ). In the first step, trifluoroacetic acid (TFA) was
used as a catalyst to afford dipyrromethane intermediates, which were
oxidized under mild conditions to the corresponding dipyrromethenes.
Then, the reaction with boron trifluorate etherate afforded BODIPY
dyes. The first one, 3,5-diphenyl-8-(3,5-di-tert-butylphenyl)BODIPY
(DPh-BODIPY) could be isolated directly after this step, but we used
the mixture for the final aromatization procedure, which was performed
with 2,3-dichloro-5,6-dicyano-1,4-benzoquinone (DDQ) in toluene under
reflux. Surprisingly, only DPh-BODIPY and 3-phenyl-5-ethoxycarbonyl-6:7-benzo-8-(3,5-di-tert-butylphenyl) BODIPY (MPh-MB-BODIPY) were separated
after column chromatography. No evidence for the formation of 3,5-bis(ethoxycarbonyl)-1,2,6,7-dibenzo-8-(3,5-di-tert-butylphenyl)BODIPY (DB-BODIPY) or its non-oxidized
precursors was found. For the synthesis of DB-BODIPY, only aldehyde
and tetrahydroisoindole ester were used. After similar steps, DB-BODIPY
was isolated as a blue solid (Scheme ). The obtained three BODIPY dyes are strongly colored
solids; their solutions show a ranging from deep-red to blue colors.
Strong fluorescence of first two dyes was visible even by the naked
eye. Absorption spectra revealed that every annulation step shifts
the absorption bathochromically by 40–45 nm; at the same time,
molar absorptivity is growing as well, due to enlargement of the π-system.
All dyes show bright fluorescence, with quantum yields of 0.56–0.74
and a Stokes shift of 608–1033 cm–1 (see Table ).
Scheme 1
Synthesis of DPh-BODIPY
and MPh-MB-BODIPY
Scheme 2
Synthesis of DB-BODIPY
Table 1
Spectral Properties of BODIPYs in
Toluene at Room Temperature
absorbance λmax [nm]
ε [M–1 cm–1]
emission λmax [nm]
Φfa
Stokes shift [cm–1]
DPh-BODIPY
557
30,600
591
0.68
1033
MPh-MB-BODIPY
597
52,650
630
0.74
877
DB-BODIPY
641
67,100
667
0.56
608
Fluorescence quantum yields for
all BODIPYs (λexc = 560 nm) were calculated using
Lumogen Red as a standard (Φf = 0.96 in chloroform).
Fluorescence quantum yields for
all BODIPYs (λexc = 560 nm) were calculated using
Lumogen Red as a standard (Φf = 0.96 in chloroform).The normalized absorption and
fluorescence spectra of the asymmetric
BODIPY, together with the normalized absorption spectrum of the family
of mixed benzo-naphtho-porphyrins are shown in Figure . The absorption and fluorescence spectra
of the symmetric BODIPYs—the DPh-BODIPY and the DB-BODIPY—are
shown in the Supporting Information, Figures
S4 and S5, respectively. As expected[24b] the studied BODIPY’s demonstrate efficient TTA-UC when the
UC-couples are combined, as shown in Table . As seen from the Table , the UC system with the smallest anti-Stokes
shift is the system obtained by the mixed-condensation strategy, both
for the sensitizer and for the emitter molecules. Despite this advantage,
there are three other optical parameters derived from Figure and Table making the UC dye-couple PdBNP/MPh-MB-BODIPY
an optimal system for under-cutaneous sensing applications: (1) comparing
the absorption coefficients of the HbO2 for the specific
excitation wavelengths λ = 635 nm (the UC-couple PdTBP/DPh-BODIPY)
and λ = 658 nm (the UC-couple PdBNP/MPh-MB-BODIPY), there is
a more than a 1.5 times higher absorption for the shorter excitation
wavelength; (2) the crucial advantage of the asymmetrical UC-couple
is the fact that the emission generated inside the tissue will be
more than 5.3 times less absorbed than the signal of the UC dye-couple
PdTBP/DPh-BODIPY; and (3) regarding the absorption coefficient for
the residual phosphorescence signal, the asymmetrical UC dye-couple
reveals more than 7.5 times lower optical losses than it is observed
for the strongly red-shifted UC dye-couple PdTNP/DB-BODIPY.
Figure 3
(a) Normalized
absorption spectrum of the mixed palladium benzo-naphtho-porphyrin
family (PdBNP); (b) normalized absorption (the green curve) and fluorescence
(the red curve) of MPh-MB-BODIPY in toluene.
Table 2
TTA—UC Parameters for Different
UC-Couples, in Toluene at Room Temperature, Glovebox Conditions
sensitizer
emitter
excitation [nm]
dF λmax [nm]
rPh λmax [nm]
anti-Stokes shift
[cm–1]
Q.Y. TTA-UC
PdTBP
DPh-BODIPY
635
591
795
1170 (0.145 eV)
0.02
PdBNP
MPh-MB-BODIPY
658
630
850
677 (0.084 eV)
0.021
PdTNP
DB-BODIPY
705
667
900
806 (0.100 eV)
0.018
(a) Normalized
absorption spectrum of the mixed palladium benzo-naphtho-porphyrin
family (PdBNP); (b) normalized absorption (the green curve) and fluorescence
(the red curve) of MPh-MB-BODIPY in toluene.Summarizing the data
presented in (1), (2), and (3), one can conclude
that the registered dF or rPh signals for
the UC dye-couple PdBNP/MPh-MB-BODIPY collected after the sequential
processes—excitation (tissue penetration), TTA-UC, that is,
generation of delayed fluorescence and residual phosphorescence, emanation
of the optical signal (escape from the tissue)—are more than
eight times higher, keeping all other experimental conditions the
same (namely, excitation photon flux, TTA-UC quantum yield, dye concentrations,
oxygen content, sample temperature, etc.) constant.In an oxygen-contaminated
environment, during the optical excitation,
singlet oxygen is generated continuously. The phytochemical compounds
of the vegetable oils (e.g., tocopherol, tocotrienol, and γ-oryzanol)
demonstrate a remarkable ability to bind chemically all existing amounts
of singlet oxygen. If the oxygen permeation rate through the sample
surface is much lower than the rate of chemical binding of singlet
oxygen across the optically assessed spot, after a short initial period
(around 4 s in this case, see Figure ), the entire oxygen content is chemically bound. This
fact is demonstrated by the truly stationary intensity of the signals
of dF and rPh, as verified in Figure .
Figure 4
Temporal evolution of
the signals of dF and rPh at sample temperature
of T = 22 °C.
The excitation intensity is kept constant at 1 mW × cm–2 for all measurements; cw—diode laser at λexc = 658 nm; air-saturated environment; excitation spot diameter d = 1.8 × 10–3 m; sample thickness b = 4 × 10–4 m. Material composition,
as follows, 1 × 10–5 M PdBNP/2 × 10–4 M MPh-MB-BODIPY/40 wt % carnauba wax/30 wt % squalene
oil/30 wt % peanut oil. The black lines are guide for the eye.
Temporal evolution of
the signals of dF and rPh at sample temperature
of T = 22 °C.
The excitation intensity is kept constant at 1 mW × cm–2 for all measurements; cw—diode laser at λexc = 658 nm; air-saturated environment; excitation spot diameter d = 1.8 × 10–3 m; sample thickness b = 4 × 10–4 m. Material composition,
as follows, 1 × 10–5 M PdBNP/2 × 10–4 M MPh-MB-BODIPY/40 wt % carnauba wax/30 wt % squalene
oil/30 wt % peanut oil. The black lines are guide for the eye.The signals of dF and rPh, even in an
oxygen saturated environment, demonstrate remarkable stability. This
allows us to study the temperature dependence of the TTA-UC process.
In Figure a, the luminescence
spectra of the studied material composition are demonstrated for two
boundary temperature values, namely, 18 and 42 °C. As expected,[10] a significant decrease in the residual sensitizer
phosphorescence, accompanied with a well-observable increase in the
emitter delayed fluorescence with increasing sample temperature, was
detected. The data presented in Figure a are summarized in Figure b, where the dependence of the dF and rPh signals for a stepwise increase in the sample
temperature is reported.
Figure 5
(a) Luminescence spectra of the UC systems for
different sample
temperatures; (b) temperature dependence of the signals of dF (at λmax = 630 nm, the blue dots) and rPh (at λmax = 850 nm, the red dots) on the sample
temperature. Experimental conditions for all measurements: material
composition, as follows, 1 × 10–5 M PdBNP/2
× 10–4 M MPh-MB-BODIPY/40 wt % carnauba wax/30
wt % squalene/30 wt % peanut oil. The spectra are obtained at the t = 4 s after starting the optical excitation. The excitation
intensity is kept constant, at 1 mW × cm–2 for
all measurements; cw—diode laser at λexc = 658 nm; air saturated environment.
(a) Luminescence spectra of the UC systems for
different sample
temperatures; (b) temperature dependence of the signals of dF (at λmax = 630 nm, the blue dots) and rPh (at λmax = 850 nm, the red dots) on the sample
temperature. Experimental conditions for all measurements: material
composition, as follows, 1 × 10–5 M PdBNP/2
× 10–4 M MPh-MB-BODIPY/40 wt % carnauba wax/30
wt % squalene/30 wt % peanut oil. The spectra are obtained at the t = 4 s after starting the optical excitation. The excitation
intensity is kept constant, at 1 mW × cm–2 for
all measurements; cw—diode laser at λexc = 658 nm; air saturated environment.As shown in Figure b, the signals of dF and rPh have comparable
intensity. Additionally, the dF signal increases monotonically
with increasing sample temperature; simultaneously, the rPh signal decreases monotonically with increasing sample temperature.
Thus, it allows us to achieve a non-ambiguous calibration curve, as
shown in Figure .
From this figure, it is evident that this biocompatible material composition
(1 × 10–5 M PdBNP/2 × 10–4 M MPh-MB-BODIPY/40 wt % carnauba wax/30 wt % squalene oil/30 wt
% peanut oil) demonstrates a high-temperature sensitivity since the
ratio dF/rPh is changed more than four times
within the physiologically relevant temperature window of interest
ΔT ∼ 18 – 42 °C.
Figure 6
Temperature
calibration curve—ratiometric response. Normalized
ratio of the signals of dF/rPh as a function of the sample
temperature, as demonstrated in Figure b.
Temperature
calibration curve—ratiometric response. Normalized
ratio of the signals of dF/rPh as a function of the sample
temperature, as demonstrated in Figure b.
Conclusions
In
this work, we demonstrated for the first time all-optical temperature
sensing with optimal excitation/emanation of the optical signals.
The synthesized asymmetrical benzo-fused BODIPY was identified as
an optimal emitter for the process of TTA-UC, performed with the mixed
palladium benzo-naphtho-porphyrins, used as a sensitizer. The sensing
technique is based on a ratiometric-type signal registration that
ensures significant independence of the obtained data on excitation
intensity instabilities, local molecular concentration fluctuations
and field-of-view variations. The identified that matrix materials
(natural wax/natural oils) are inherently biocompatible and FDA-approved
as food additives. The desired temperature sensitivity is better than
200 mK centered on the physiologically relevant temperature of 36
°C and is warranted by using the process of TTA-UC as a sensing
mechanism. The TTA-UC system is effectively protected for more than
100 s against oxygen-induced damages, allowing stable performance
of this temperature-sensing system even in the ambient environment
without losing sensitivity while applying the same calibration curve.
Experimental
Section
3,5-Di(tert-butyl)benzaldehyde (TCI
Chemicals),
DDQ, carnauba wax, squalene, peanut oil (Acros), triethylamine (Roth),
DIPEA (Roth), boron trifluoride etherate (Merck), anhydrous dichloromethane
(Aldrich), and 2-phenylpyrrole (Chempur) were used as received. Ethyl-4,5,6,7-tetrahydro-2H-isoindole-1-carboxylate was synthesized, as described
elsewhere.[15]1H and 13C NMR spectra were recorded on a Bruker Avance 250 and a Bruker Avance
500 spectrometers. Chemical shifts are denoted in d unit (ppm). Mass
spectra were recorded with an Advion Expression L spectrometer. UV/Vis
spectra were recorded at room temperature on a Shimadzu UV-1800 spectrophotometer.
Fluorescence spectra were recorded on a Spex Fluorolog 3 spectrometer.
Fluorescence quantum yields were determined using the relative method
using Lumogen Red as a ref (16).
Synthesis of DPh-BODIPY and MPh-MB-BODIPY
3,5-Di(tert-butyl)benzaldehyde (218 mg, 1 mmol), ethyl-4,5,6,7-tetrahydro-2H-isoindole-1-carboxylate (193 mg, 1 mmol), and 2-phenylpyrrole
(143 mg, 1 mmol) were dissolved in 100 mL of absolute CH2Cl2 under an Ar atmosphere. Three drops of TFA were added,
and the solution was stirred at room temperature overnight in the
darkness. Dry DDQ (250 mg) was added and stirring was continued for
2 h. Triethylamine (2 mL) was added, and the organic phase was washed
with aqueous sodium sulfite (3%, 2 × 100 mL). Organic layers
were separated, dried over anhydrous sodium sulfate, and evaporated
to dryness. N,N-diisopropylethylamine (DIEA) (3 mL)
and 100 mL of absolute CH2Cl2 were added under
an Ar atmosphere, and the solution was stirred at room temperature
for 10 min. BF3·OEt2 (3 mL) was added,
and stirring was continued for 2 h. The reaction mixture was washed
with NaHCO3 solution (5%, 2 × 100 mL) and water (100
mL). The combined organic extracts were dried over Na2SO4, filtered, and evaporated. Toluene (50 mL) and 1,4-dioxane
(100 mL) were added, stirred for 5 min, and then, DDQ (300 mg) was
added and stirred at 110 °C for 15 h. Solution was cooled and
washed with aqueous sodium sulfite (3%, 2 × 100 mL). Organic
layers were separated, dried over anhydrous sodium sulfate, and evaporated
to dryness. Column chromatography with silica gel (eluent–toluene)
afforded DPh-BODIPY as a first red fraction with yellow fluorescence,
which was evaporated and recrystallized from CH2Cl2/methanol to afford dark red crystals after drying under vacuum.
Yield 207 mg (39%).1H NMR (250 MHz, C2D2Cl4): δ 7.91–7.87 (m, 4H), 7.63
(s, 1H), 7.49–7.45 (m, 8H), 6.99 (d, J = 4.3
Hz, 1H), 6.67 (d, J = 4.2 Hz, 1H), 1.42 (s, 18H). 13C NMR (126 MHz, C2D2Cl4):
δ 158.12, 150.70, 146.04, 136.35, 133.22, 132.65, 125.25, 124.30,
120.91, 74.13, 34.83, 31.39; λmax (toluene)/nm 557
(ε/dm3 mol–1 cm–1 30,600); fluorescence (toluene): λmax = 591 nm
(ϕ = 68%); MS (FD, 8 kV): m/z (%) 532.5 (100), M+; the second violet fraction, possessing
red fluorescence, was evaporated, dissolved in cyclohexane (20 mL)
and freeze-dried for 24 h, to afford MPhMB-BODIPY as a violet powder
(133 mg, 23% yield).1H NMR (250 MHz, C2D2Cl4): δ 8.05 (d, J = 8.1 Hz, 1H), 7.99–7.96
(m, 2H), 7,68 (s, 1H), 7.53–7.51 (m, 3H), 7.38–7.36
(m, 2H), 7.31 (t, J = 7.6 Hz, 1H), 7.16 (t, J = 7.6 Hz, 1H), 6.84 (d, J = 4.3 Hz, 1H),
6.67 (d, J = 4.2 Hz, 1H), 6.40 (d, J = 8.4 Hz, 1H), 4.56 (q, J = 7.1 Hz, 2H), 1.51 (t, J = 7.1 Hz, 3H), 1.39 (s, 18H). 13C NMR (126
MHz, C2D2Cl4): δ 160.50, 158.10,
151.27, 144.28, 139.99, 137.60, 134.62, 132.53, 132.29, 130.68, 130.46,
129.38, 128.28, 123.84, 123.45, 121.87, 74.04, 62.25, 34.93, 31.31,
29.60, 14.13; λmax (toluene)/nm 597 (ε/dm3 mol–1 cm–1 52,650); fluorescence
(toluene): λmax = 630 nm (ϕ = 74%); MS (FD,
8 kV): m/z (%) 578.6 (100), M+; DPh-BODIPY was also synthesized directly from 3,5-di(tert-butyl)benzaldehyde and 2-phenylpyrrole following the
same procedure but without last aromatization step. The analytical
data are identical to those, obtained by mixed pyrrole condensation.
Yield 57%.
Synthesis of DB-BODIPY
3,5-Di(tert-butyl)benzaldehyde (218 mg, 1 mmol) and ethyl-4,5,6,7-tetrahydro-2H-isoindole-1-carboxylate
(386 mg, 2 mmol) were dissolved in 100 mL of absolute CH2Cl2 under an Ar atmosphere. Three drops of TFA were added,
and the solution was stirred at room temperature overnight in the
darkness. Dry DDQ (250 mg) was added, and stirring was continued for
2 h. Triethylamine (2 mL) was added, and the organic phase was washed
with aqueous sodium sulfite (3%, 2 × 100 mL). Organic layers
were separated, dried over anhydrous sodium sulfate, and evaporated
to dryness. DIEA (3 mL) and 100 mL of absolute CH2Cl2 were added under an Ar atmosphere, and the solution was stirred
at room temperature for 10 min. BF3·OEt2 (3 mL) was added, and stirring was continued for 4 h. The reaction
mixture was washed with NaHCO3 solution (5%, 2 × 100
mL) and water (100 mL). The combined organic extracts were dried over
Na2SO4, filtered, and evaporated. Toluene (50
mL) and 1,4-dioxane (100 mL) were added, stirred for 5 min, and then,
DDQ (500 mg) was added and stirred at 110 °C for 15 h. Solution
was cooled and washed with aqueous sodium sulfite (3%, 2 × 100
mL). Organic layers were separated, dried over anhydrous sodium sulfate,
and evaporated to dryness. Column chromatography with silica gel (eluent–toluene)
afforded blue fraction, which was evaporated, dissolved in cyclohexane,
and freeze-dried for 24 h to afford a blue powder. Yield 230 mg (37%).1H NMR (250 MHz, C2D2Cl4): δ 8.09 (d, J = 8.2 Hz, 2H), 7,76 (s, 1H),
7.36 (d, J = 1.7 Hz, 2H), 7.33–7.25 (m, 2H),
7.19–7.04 (m, 2H), 6.22 (d, J = 8.4 Hz, 2H),
4.61 (q, J = 7.1 Hz, 4H), 1.55 (t, J = 7.1 Hz, 6H), 1.38 (s, 18H). 13C NMR (126 MHz, C2D2Cl4): δ 160.45, 152.68, 142.82,
139.62, 134.58, 132.69, 130.78, 129.62, 129.16, 126.60, 123.76, 123.08,
122.57, 121.85, 74.04, 74.00, 73.78, 73.56, 62.27, 35.12, 31.27, 26.82,
14.18; λmax (toluene)/nm 641 (ε/dm3 mol–1 cm–1 67,100); fluorescence
(toluene): λmax = 667 nm (ϕ = 56%); MS (FD,
8 kV): m/z (%) 624.6 (100), M+.
Authors: Julia Bailey-Serres; Takeshi Fukao; Daniel J Gibbs; Michael J Holdsworth; Seung Cho Lee; Francesco Licausi; Pierdomenico Perata; Laurentius A C J Voesenek; Joost T van Dongen Journal: Trends Plant Sci Date: 2012-01-24 Impact factor: 18.313
Authors: Mikhail A Filatov; Stanislav Baluschev; Iliyana Z Ilieva; Volker Enkelmann; Tzenka Miteva; Katharina Landfester; Sergey E Aleshchenkov; Andrei V Cheprakov Journal: J Org Chem Date: 2012-12-12 Impact factor: 4.354
Authors: Joel A Spencer; Francesca Ferraro; Emmanuel Roussakis; Alyssa Klein; Juwell Wu; Judith M Runnels; Walid Zaher; Luke J Mortensen; Clemens Alt; Raphaël Turcotte; Rushdia Yusuf; Daniel Côté; Sergei A Vinogradov; David T Scadden; Charles P Lin Journal: Nature Date: 2014-03-02 Impact factor: 49.962