Literature DB >> 34307611

Hemoglobin Fukuoka caused unexpected hemoglobin A1c results: A case report.

Xue-Ping Lin1, Qiu-Rong Yuan1, Shi-Qiong Niu1, Xi Jiang1, Zhi-Kun Wu1, Zhao-Fan Luo2.   

Abstract

BACKGROUND: Glycated hemoglobin (Hb) (HbA1c) is an indicator that is used to diagnose and monitor the treatment of diabetes. Many factors can affect the detection of HbA1c. One of the most important of these factors is the Hb variant. Here, we report a rare Hb variant and evaluate its effect on HbA1c. CASE
SUMMARY: A 35-year-old man was suspected of harboring an Hb variant following the measurement of HbA1c with the Variant II Turbo 2.0 Hb detection system during a routine examination. Subsequently, we used the Arkray HA-8160 and ARCHITECT c4000 system to reanalyze HbA1c. Finally, the Hb variant was detected with a Capillary2FP analyzer that operates on the principle of capillary electrophoresis. We also used gene sequencing to investigate the mutation site. The value of HbA1c detected with the Variant II Turbo 2.0 system was 52.7%. However, the Arkray HA-8160 system did not display a result while the ARCHITECT c16000 system showed a result of 5.4%. The Capillary2FP analyzer did not reveal any abnormal Hb zones. However, gene sequencing identified the presence of a mutation in the Hb β2 chain [CD2(CAC>TAC), His>Tyr, HBB: c.7C>T]; the genotype was Hb Fukuoka.
CONCLUSION: Hb variants could cause abnormal HbA1c results. For patients with Hb variants, different methods should be used to detect HbA1c. ©The Author(s) 2021. Published by Baishideng Publishing Group Inc. All rights reserved.

Entities:  

Keywords:  Case report; Enzymatic method; Hemoglobin A1c; Hemoglobin Fukuoka; Hemoglobin variant; High-performance liquid chromatography

Year:  2021        PMID: 34307611      PMCID: PMC8281434          DOI: 10.12998/wjcc.v9.i20.5568

Source DB:  PubMed          Journal:  World J Clin Cases        ISSN: 2307-8960            Impact factor:   1.337


Core Tip: Hemoglobin A1c (HbA1c) is an indicator of diabetes diagnosis and blood glucose monitoring. Therefore, the accuracy of HbA1c results is of great significance to clinical diagnosis and treatment. This case had an abnormal HbA1c result and traced back to a rare hemoglobin variant. Hemoglobin variants are one of the important factors affecting the accuracy of HbA1c results. In this case, different methods were used to detect HbA1c, which can provide reference evidence for subsequent cases and reduce false results reports.

INTRODUCTION

Hemoglobin A1c (HbA1c) is the main indicator recommended by the American Diabetes Association and the World Health Organization for the diagnosis of diabetes[1]. HbA1c is a stable compound that covalently bonds glucose and hemoglobin (Hb) β-chain N-terminal valine residues in human blood. It reflects the average blood glucose concentration over the previous 2 mo to 3 mo. However, the results of the HbA1c are affected by a wide range of factors, including Hb variants, severe hemolytic anemia, and severe liver disease, etc. The most common factors that interfere with HbA1c tests are Hb variants[2,3]. In this paper, we report a case with abnormal HbA1c values, which was eventually diagnosed as a rare Hb mutation.

CASE PRESENTATION

Chief complaints

A 35-year-old man arrived at our hospital for a physical examination in November 2018.

History of past illness

The patient had no special past history.

Personal and family history

The patient had no special personal and family history.

Physical examination

Normal physical examination without any obvious signs of abnormality.

Laboratory examinations

The HbA1c result of 52.7% was out of the expected detection range (Figure 1A). The HbA1c of the patient was remarkably elevated in comparison to the reference range (4.0%-6.4%). Data arising from routine blood analyses are shown in Table 1. The patient had normal liver function, normal fasting glucose, and no previous history of diabetes. The HbA1c test was performed using the Biorad Variant II Turbo 2.0 automatic glycated Hb analyzer (Biorad, United States); the HbA1c value was abnormally elevated, although the patient had no history of diabetes. Therefore, we suspected that the patient possessed Hb variants that interfered with the results of the HbA1c test.
Figure 1

Detection of hemoglobin AA: The peak of hemoglobin A1c (HbA1c) was abnormally elevated (orange arrow); B: The HbA1c peak did not appear, and there was an abnormal protruding peak on the left of peak A0 (orange arrow). HbA1c: Hemoglobin A1c.

Table 1

Routine blood data


Results
Reference range
Red blood cells (× 1012/L)5.64.3-5.8
Hemoglobin (g/L)161130-175
Mean erythrocyte volume (fL)9182-100
Mean erythrocyte hemoglobin volume (pg)2927-34
Mean erythrocyte hemoglobin concen-tration (g/L)315316-354
Erythrocyte distribution width (%)1312-15
Total protein (g/L)79.365.0-85.0
Albumin (g/L)46.132.0-45.0
Alanine aminotransferase (U/L)190-55
Aspartate transferase (U/L)2215-40
Blood creatinine (µmol/L)7664-104
Blood urea nitrogen (mmol/L)4.83.2-7.4
Fasting glucose (mmol/L)4.713.90-6.10
Detection of hemoglobin AA: The peak of hemoglobin A1c (HbA1c) was abnormally elevated (orange arrow); B: The HbA1c peak did not appear, and there was an abnormal protruding peak on the left of peak A0 (orange arrow). HbA1c: Hemoglobin A1c. Routine blood data

Further investigations for Hb variants

We attempted to confirm the HbA1c value by using the Arkray HA-8160 automatic glycated Hb analyzer (Arcolai, Japan), which features high-performance liquid chromatography (HPLC), and the ARCHITECT c4000 system (Abbott, United States), which operates by enzymatic methods. The Arkray HA-8160 system returned no result; the instrument showed that the HbA1c peak did not appear, and there was an abnormal protruding peak on the left of peak A0 (Figure 1B). The ARCHITECT c16000 enzyme assay returned a result of 5.4%. Next, we used the Capillary2FP automated Hb analyzer (Sebia, France) to find the presence of abnormal Hb. Capillary electrophoresis showed no abnormal Hb bands (Figure 2).
Figure 2

A Capillary2FP system was used to find the hemoglobin variant. Hb: Hemoglobin.

A Capillary2FP system was used to find the hemoglobin variant. Hb: Hemoglobin. In addition, the patient's genomic DNA was extracted using a Genomic DNA Isolation Kit (Kaipu, China) and the HBA1, HBA2, and HBB genes were amplified with specific forward/reverse primers and sequenced on an ABI3500 sequencer. HBA1 forward primer was 5´-CGCGCCAGCCAATGAGC-3´ and HBA1 reverse primer was 5´-ACACACA TGGCTAGAACCTCTCTG-3´. HBA2 forward primer was 5´-GGGCT CCGCGCCAGCCAA-3´ and HBA2 reverse primer was 5´-CAAGGACCTCTCT GCAGCT-3´. HBB forward primer was 5´-TACGGCTGTCATCACTTAG-3´and HBB reverse primer was 5´-GCCACACTGAGTGAGCTGCACT-3´. We extracted a sample of whole blood from the patient and extracted genomic DNA. This was used as a template to amplify the Hb gene. Polymerase chain reaction amplicons were then sequenced. No thalassemia mutation types were found in the sequencing range for α1 and α2. However, the β-globin gene possessed a histidine > tyrosine (CAC >TAC) mutation in the 7th amino acid, C>T (Figure 3). Next, we searched the HbA1c variant library website (http://globin.bx.psu.edu/hbvar/menu.html), which revealed that the Hb genotype was Hb Fukuoka (HBB:c.7C>T). Finally, the patient was confirmed to possess a heterozygous Hb Fukuoka mutation.
Figure 3

Gene sequencing results revealed a β chain mutation in the gene [CD2(CAC>TAC), His>Tyr, HBB:c.7C>T] (orange arrow).

Gene sequencing results revealed a β chain mutation in the gene [CD2(CAC>TAC), His>Tyr, HBB:c.7C>T] (orange arrow).

FINAL DIAGNOSIS

Finally, the patient was confirmed to possess a heterozygous Hb Fukuoka mutation.

TREATMENT

The patient was not treated.

OUTCOME AND FOLLOW-UP

For patients with Hb variants, we recommend the use of an assay that is not affected by interference caused by Hb variants.

DISCUSSION

The guidelines relating to diabetes mellitus that were published by the American Chemical Society in 2011 suggest that a patient should receive further research/investigation if HbA1c > 15% or if the laboratory data are inconsistent with clinical manifestations. The guidelines also recommend the use of different test principles and methods so that the patient can be re-tested[4]. In this case, the patient’s HbA1c was 52.7%, significantly higher than 15%. The patient had no history of diabetes. Laboratory data were all within the normal quantitative range, including Hb, kidney and liver function parameters, and fasting blood glucose. Therefore, it was necessary to retest this patient and identify reasons that might underlie the abnormal HbA1c result. Biorad Variant II Turbo 2.0 and Arkray HA-8160 Glycation Hemoglobin analyzers are based on HPLC technology. Using this technology, it is possible to separate Hb fractions by differential affinity to the internal surface of the column or to the buffer. The greater the affinity to the buffer, the faster the elution, and the shorter the retention time. The greater the affinity to the analytical column, the slower the elution, and the longer the retention time. In this case, the Biorad Variant II Turbo 2.0 returned an HbA1c result of 52.7%, and no result was detected by Arkray HA-8160. Therefore, we considered the possibility of interference caused by Hb variants. Hb variants can affect the separation of variants from HbA1c or HbA by changing their charge, co-eluting with HbA1c, and replacing glycosylation sites[5,6]. The NGSP website has declared that the common variants including S, C, D, and E have no effect on Biorad Variant II Turbo 2.0 (http://ngsp.org/factors.asp). However, some rare Hb variants can affect HbA1c with HPLC[7,8]. The ARCHITECT c16000 uses enzymatic methods to detect HbA1c values. By lysing the whole blood sample and performing extensive proteolytic digestion, the Hb β chain releases amino acids, especially the glycated N-terminal valine. The signal generated by the glycosylated valine in the subsequent color-based reaction can then be used to calculate the HbA1c. At the analytical level, these methods are unaffected by the presence of Hb variants, such as HbC, HbD, HbE, and HbS. However, this method is unable to detect Hb variants[9]. Therefore, an HbA1c result of 5.4% can accurately reflect the glycated Hb of the patient. The Capillary 2FP system operates on the principle of capillary electrophoresis and was used to investigate for the presence of an Hb variant. The Capillary 2FP system showed no abnormal Hb bands. Gene sequencing identified the variant as an Hb Fukuoka heterozygote. This variant was first reported in Japan in 1985, and subsequently reported by both Harano et al[10] and Farah et al[11]. Harano et al[10] found abnormal Hb peak in a diabetic patient's HbA1c test in 1990, and identified that the patient had Hb Fukuoka. Furthermore, Farah et al[11] identified the presence of Hb Fukuoka by performing electrophoresis examinations in a patient with anemia. As far as we know, this case was the first case to be discovered in China. Several methods can be used to detect glycosylated Hb, including electrophoresis, immunoturbidimetric assay (immunoassay), enzymatic assay, boronate affinity, ion-exchange HPLC, or capillary electrophoresis. Laboratories should be aware of the limitations of their methods with respect to interference from other Hb variants and suggest alternative HbA1c quantification methods[12,13]. For patients with Hb variants, we recommend the use of an assay that is not affected by interference caused by Hb variants; this strategy will help avoid erroneous results that can mislead data in clinical practice[14]. Glycated albumin and fructosamine are recommended as alternative indicators to monitor glycaemia when HbA1c is disturbed[15].

CONCLUSION

The chromatographic pattern of abnormal Hb variants differs from that of normal samples, making it difficult for some instruments to determine accurate HbA1c values. HbA1c is an important indicator for the diagnosis, treatment, and monitoring of diabetes mellitus, and the accuracy of HbA1c results is of paramount importance. But meanwhile, it is important for laboratory staff to be familiar with the factors that may cause interference, and to be able to communicate with clinicians in a timely manner to clarify the situation and make clinical suggestions. The ultimate goal is to ensure accurate and realistic test results so as to help clinical diagnosis.

ACKNOWLEDGEMENTS

We are grateful for instructive discussion with Professor Luo ZF.
  15 in total

1.  Hb Fukuoka [beta 2(NA2)His----Tyr]: a new mutation at the 2,3-diphosphoglycerate binding site.

Authors:  T Harano; K Harano; S Ueda; K Imai; A Ohkuma; Y Koya; H Takahashi
Journal:  Hemoglobin       Date:  1990       Impact factor: 0.849

Review 2.  Pathology consultation on HbA(1c) methods and interferences.

Authors:  Jeanne M Rhea; Ross Molinaro
Journal:  Am J Clin Pathol       Date:  2014-01       Impact factor: 2.493

Review 3.  A review of variant hemoglobins interfering with hemoglobin A1c measurement.

Authors:  Randie R Little; William L Roberts
Journal:  J Diabetes Sci Technol       Date:  2009-05-01

4.  Effects of 49 Different Rare Hb Variants on HbA1c Measurement in Eight Methods.

Authors:  Randie R Little; Sonia L La'ulu; Steven E Hanson; Curt L Rohlfing; Robert L Schmidt
Journal:  J Diabetes Sci Technol       Date:  2015-02-17

5.  Guidelines and recommendations for laboratory analysis in the diagnosis and management of diabetes mellitus.

Authors:  David B Sacks; Mark Arnold; George L Bakris; David E Bruns; Andrea Rita Horvath; M Sue Kirkman; Ake Lernmark; Boyd E Metzger; David M Nathan
Journal:  Clin Chem       Date:  2011-05-26       Impact factor: 8.327

6.  Effects of common hemoglobin variants on HbA1c measurements in China: results for α- and β-globin variants measured by six methods.

Authors:  Anping Xu; Weidong Chen; Yong Xia; Yu Zhou; Ling Ji
Journal:  Clin Chem Lab Med       Date:  2018-07-26       Impact factor: 3.694

7.  Recognition of rare hemoglobin variants by hemoglobin A1c measurement procedures.

Authors:  Sydney W Strickland; Sean T Campbell; Randie R Little; David E Bruns; Lindsay A L Bazydlo
Journal:  Clin Chim Acta       Date:  2017-11-14       Impact factor: 3.786

Review 8.  HbA1c, Fructosamine, and Glycated Albumin in the Detection of Dysglycaemic Conditions.

Authors:  Rogério Tavares Ribeiro; Maria Paula Macedo; João Filipe Raposo
Journal:  Curr Diabetes Rev       Date:  2016

9.  Hemoglobin variants detected by hemoglobin A1c (HbA1c) analysis and the effects on HbA1c measurements.

Authors:  Nadzimah Mohd Nasir; M Thevarajah; Chew Yee Yean
Journal:  Int J Diabetes Dev Ctries       Date:  2010-04

Review 10.  2. Classification and Diagnosis of Diabetes: Standards of Medical Care in Diabetes-2020.

Authors: 
Journal:  Diabetes Care       Date:  2020-01       Impact factor: 19.112

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.