Jiali Yang1, Jiawu Li1, Guilin Ye1, Yan Luo1.
Abstract
PURPOSE: This study investigated the effectiveness and feasibility of shear wave elastography ((sound touch elastography) STE and (shear wave elastography) SWE) and visual transient elastography (ViTE) in the noninvasive quantitative diagnosis of liver fibrosis in chronic liver disease (CLD). PATIENTS AND METHODS: A total of 106 patients with CLD underwent STE, SWE and ViTE elastography evaluation. The Young's modulus of the three elastography was valuated and the diagnostic performances of the three techniques for liver fibrosis staging were compared. The area under the receiver operating curve (ROC) for the diagnosis of liver fibrosis was compared. The final diagnosis was based on the histological findings on the liver biopsy.Entities:
Keywords: chronic liver disease; liver fibrosis; liver stiffness measurement; shear wave elastography; visual transient elastography
Year: 2021 PMID: 34295181 PMCID: PMC8290848 DOI: 10.2147/IJGM.S319101
Source DB: PubMed Journal: Int J Gen Med ISSN: 1178-7074
Figure 1Elasticity maps of three different elastography (A) ViTE; (B) STE; (C) SWE.
Shows the Characteristics of Patients
| Age, mean ±SD, y | 51.22±12.91 (18–81) |
| Sex | n (%) |
| Male | 51 (48.11) |
| Female | 55 (51.89) |
| BMI (kg/m2) | 24.16±4.88 |
| Diagnosis | n (%) |
| HBV | 31 (29.25) |
| AIH | 35 (33.02) |
| Drug-induced liver injury | 8 (7.55) |
| NAFLD | 5 (4.72) |
| Others | 27 (25.47) |
| METAVIR | n (%) |
| F0 | 16 (15.09) |
| F1 | 21 (19.81) |
| F2 | 19 (17.92) |
| F3 | 22 (20.75) |
| F4 | 28 (26.42) |
LSMs of the ViTE, SWE and STE in Pathological Analysis of Liver Fibrosis in Each Group
| METAVIR | Cases | ViTE | STE | SWE |
|---|---|---|---|---|
| F0 | 16 | 6.62±2.07 | 6.74±1.15 | 5.7±1.56 |
| F1 | 21 | 8.90±2.79 | 8.96±2.42 | 8.53±3.15 |
| F2 | 19 | 12.27±5.40 | 12.03±5.24 | 10.1±2.84 |
| F3 | 22 | 13.77±5.93 | 12.65±5.00 | 13.5±7.80 |
| F4 | 28 | 17.56±7.46 | 13.10±4.86 | 14.3±6.47 |
| F values | 9.829 | 8.192 | 9.36 | |
| <0.001 | <0.001 | <0.001 |
Figure 2Box-plot of three elastography techniques in the stage of liver fibrosis (A) ViTE; (B) SWE; (C) STE.
Figure 3Comparison of the area under the curve (AUC) of ViTE, SWE and STE at different stages of fibrosis (A) comparison of ROC curves in the identification of F0 vs F1–4; (B) F0-1 vs F2-3; (C) F0-2 vs F3-4; (D) F0-3 vs F4.
Diagnostic Accuracy of ViTE, SWE and STE for Different Stages of Liver Fibrosis in CLD Patients
| Metavir Staging | AUC | Cut-Off (kPa) | Se (%) | Sp (%) |
|---|---|---|---|---|
| ViTE | ||||
| F1≤ | 0.88 | 6.3 | 75.0 | 84.4 |
| F2≤ | 0.84 | 8.8 | 70.3 | 81.2 |
| F3≤ | 0.80 | 10.0 | 64.3 | 80.0 |
| =F4 | 0.80 | 14.1 | 79.5 | 71.4 |
| SWE | ||||
| F1≤ | 0.92 | 5.7 | 75.0 | 95.6 |
| F2≤ | 0.84 | 7.0 | 70.3 | 84.1 |
| F3≤ | 0.79 | 8.1 | 62.6 | 84.0 |
| =F4 | 0.76 | 13.2 | 80.8 | 78.6 |
| STE | ||||
| F1≤ | 0.91 | 6.8 | 56.3 | 93.3 |
| F2≤ | 0.84 | 7.9 | 67.6 | 87.0 |
| F3≤ | 0.77 | 8.9 | 62.5 | 82.0 |
| =F4 | 0.71 | 12.5 | 73.1 | 78.6 |
Diagnostic Accuracy of ViTE and STE by Combining or Using Either Technique
| Metavir Staging | AUC | Se (%) | Sp (%) |
|---|---|---|---|
| F1≤ | |||
| ViTE≤6.3kPa and STE≤6.8kPa | 0.93 | 65.0 | 95.4 |
| ViTE≤6.3kPa or STE≤6.8kPa | 0.94 | 85.7 | 85.9 |
| F2≤ | |||
| ViTE≤8.8kPa and STE≤7.9kPa | 0.87 | 76.3 | 84.1 |
| ViTE≤8.8kPa or STE≤7.9kPa | 0.88 | 79.3 | 80.0 |
| F3≤ | |||
| ViTE≤10.0kPa and STE≤8.9kPa | 0.83 | 60.7 | 91.1 |
| ViTE≤10.0kPa or STE≤8.9kPa | 0.84 | 74.6 | 79.1 |
| =F4 | |||
| ViTE≤14.1kPa and STE≤12.5kPa | 0.87 | 70.5 | 93.1 |
| ViTE≤14.1kPa or STE≤12.5kPa | 0.88 | 82.0 | 79.3 |