| Literature DB >> 34261481 |
Yaniv Shilo1, Jonathan Modai1, Dan Leibovici1, Ishai Dror2, Brian Berkowitz3.
Abstract
BACKGROUND: To compare the efficacy of different ureteral stents subject to extrinsic ureteral obstruction (EUO), in a controlled in vitro stented ureter experiment.Entities:
Keywords: Colloids; Endopyelotomy; Metal stents; Stent failure; Tandem stents
Mesh:
Year: 2021 PMID: 34261481 PMCID: PMC8281631 DOI: 10.1186/s12894-021-00865-w
Source DB: PubMed Journal: BMC Urol ISSN: 1471-2490 Impact factor: 2.264
Fig. 1Schematic showing the experimental setup, with details of a single ureter-stent unit, including of the angle of deformation, θ°
Fig. 2Photograph of part of the 12-unit experimental systems, showing two tandem 6F and two tandem 7F stent-ureter units, under deformation and compression
Time to stent failure (blockage), in hours, under deformation of 40° and compression of 2000 g in the region of the external ureteral obstruction
| Replicate | Stent | ||||
|---|---|---|---|---|---|
| Tandem 6F (time to blockage, hour) | Tandem 7F (time to blockage, hour) | Resonance® (time to blockage, hour) | Endopyelotomy 14F (time to blockage, hour) | 8F† (time to blockage, hour) | |
| 1 | 132 | 133.5 | 243.5 | 300.5 | –‡ |
| 2 | 274 | 103 | 124.5 | 235.5 | –‡ |
| 3 | 325 | 329 | 224 | ||
| 4 | –‡ | 303.5 | –‡ | ||
Entries with “–” indicate no failure. All failure times are rounded to the nearest half hour
†The same result (no blockage) for 4 similar replicates with the 8F stent was reported in [18]
‡Experiment duration of 348 h
Poiseuille flow calculations for flow in stents
| – | Stent | ||||||
|---|---|---|---|---|---|---|---|
| Single 4.8F | Tandem 4.8F | Single 6F | Tandem 6F | Single 7F | Tandem 7F | Single 8F | |
| Flow rate (mL/h) | 77 | 154 | 88 | 176 | 235 | 470 | 516 |
| Internal stent radius (mm) | 0.58 | – | 0.60 | – | 0.77 | – | 0.93 |
| Cross-sectional area (mm2) | 2.01 | 4.02 | 3.14 | 6.28 | 4.30 | 8.60 | 5.55 |