Visual Performance of Two Diffractive Trifocal Intraocular Lenses: A Randomized Trial.

Julius Hienert, Kristina Stjepanek, Nino Hirnschall, Manuel Ruiss, Hannah Zwickl, Oliver Findl.   

Abstract

PURPOSE: To compare two trifocal intraocular lenses (IOLs), the RayOne Trifocal RAO603F IOL (closed-loop haptic IOL; Rayner Intraocular Lenses Limited) and the AT Lisa tri 839 MP IOL (plate-haptic IOL; Carl Zeiss Meditec AG), concerning optical and capsular bag performance.
METHODS: Patients scheduled for cataract surgery received either a closed-loop haptic IOL or a plate-haptic IOL in the first eye and the other IOL in the second eye. Three months postoperatively, autorefraction and subjective refraction, uncorrected and corrected distance visual acuity at 4 m, 80 cm, and 40 cm, an objective reading test (Salzburg Reading Desk; SRD Vision), a defocus curve, IOL tilt and decentration, a questionnaire about dysphotopsia, and grading of halos with a halometer were performed.
RESULTS: Eighty-eight eyes of 44 patients were included. Visual acuity was comparable between both IOLs. The closed-loop haptic IOL performed better in the defocus curve at -1.50 diopters (D) (0.08 ± 0.10 vs 0.12 ± 0.09 logMAR; P < .01). The plate-haptic IOL had better contrast sensitivity without glare under mesopic and photopic conditions in miosis (P = .0018 and .002, respectively) and mydriasis (P = .017 and .003, respectively). Significant differences were found for less overall subjective disturbance (P = .047) and starbursts (P = .039) for the plate-haptic IOL, but not for the other positive dysphotopsia symptoms.
CONCLUSIONS: Both trifocal IOLs delivered good and comparable visual function with low degrees of disturbing dysphotopsia. The closed-loop haptic IOL was slightly superior in the defocus curve, whereas the plate-haptic IOL was slightly superior concerning contrast sensitivity and positive dysphotopsia. [J Refract Surg. 2021;37(7):460-465.].

Entities:  

Year:  2021        PMID: 34236906     DOI: 10.3928/1081597X-20210420-01

Source DB:  PubMed          Journal:  J Refract Surg        ISSN: 1081-597X            Impact factor:   3.573


  2 in total

1.  Efficacy of a Secondary Trifocal Sulcus IOL in Providing Near and Intermediate Vision in Patients with Prior Myopic Laser Vision Correction and Cataract Surgery.

Authors:  Kjell Gunnar Gundersen; Bjørn Gjerdrum; Richard Potvin
Journal:  Clin Ophthalmol       Date:  2022-07-08

2.  Active-fluidics versus gravity-fluidics system in phacoemulsification for age-related cataract (AGSPC): study protocol for a prospective, randomised, double-blind, controlled clinical trial.

Authors:  Yu Luo; Hongyu Li; Wenqian Chen; Yi Gao; Tianju Ma; Zi Ye; Zhaohui Li
Journal:  BMJ Open       Date:  2022-01-20       Impact factor: 2.692

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.