Objective.Compared to biomedical devices with implanted batteries, wirelessly powered technologies can be longer-lasting, less invasive, safer, and can be miniaturized to access difficult-to-reach areas of the body. Magnetic fields are an attractive wireless power transfer modality for such bioelectronic applications because they suffer negligible absorption and reflection in biological tissues. However, current solutions using magnetic fields for mm sized implants either operate at high frequencies (>500 kHz) or require high magnetic field strengths (>10 mT), which restricts the amount of power that can be transferred safely through tissue and limits the development of wearable power transmitter systems. Magnetoelectric (ME) materials have recently been shown to provide a wireless power solution for mm-sized neural stimulators. These ME transducers convert low magnitude (<1 mT) and low-frequency (∼300 kHz) magnetic fields into electric fields that can power custom integrated circuits or stimulate nearby tissue.Approach.Here we demonstrate a battery-powered wearable magnetic field generator that can power a miniaturized MagnetoElectric-powered Bio ImplanT 'ME-BIT' that functions as a neural stimulator. The wearable transmitter weighs less than 0.5 lbs and has an approximate battery life of 37 h.Main results.We demonstrate the ability to power a millimeter-sized prototype 'ME-BIT' at a distance of 4 cm with enough energy to electrically stimulate a rat sciatic nerve. We also find that the system performs well under translational misalignment and identify safe operating ranges according to the specific absorption rate limits set by the IEEE Std 95.1-2019.Significance.These results validate the feasibility of a wearable system that can power miniaturized ME implants that can be used for different neuromodulation applications. Creative Commons Attribution license.
Objective.Compared to biomedical devices with implanted batteries, wirelessly powered technologies can be longer-lasting, less invasive, safer, and can be miniaturized to access difficult-to-reach areas of the body. Magnetic fields are an attractive wireless power transfer modality for such bioelectronic applications because they suffer negligible absorption and reflection in biological tissues. However, current solutions using magnetic fields for mm sized implants either operate at high frequencies (>500 kHz) or require high magnetic field strengths (>10 mT), which restricts the amount of power that can be transferred safely through tissue and limits the development of wearable power transmitter systems. Magnetoelectric (ME) materials have recently been shown to provide a wireless power solution for mm-sized neural stimulators. These ME transducers convert low magnitude (<1 mT) and low-frequency (∼300 kHz) magnetic fields into electric fields that can power custom integrated circuits or stimulate nearby tissue.Approach.Here we demonstrate a battery-powered wearable magnetic field generator that can power a miniaturized MagnetoElectric-powered Bio ImplanT 'ME-BIT' that functions as a neural stimulator. The wearable transmitter weighs less than 0.5 lbs and has an approximate battery life of 37 h.Main results.We demonstrate the ability to power a millimeter-sized prototype 'ME-BIT' at a distance of 4 cm with enough energy to electrically stimulate a rat sciatic nerve. We also find that the system performs well under translational misalignment and identify safe operating ranges according to the specific absorption rate limits set by the IEEE Std 95.1-2019.Significance.These results validate the feasibility of a wearable system that can power miniaturized ME implants that can be used for different neuromodulation applications. Creative Commons Attribution license.
Entities:
Keywords:
bioelectronics; implantable device; magnetoelectric effect; wearable system; wireless power transfer
Bioelectronics is transforming the medical field through engineered implants that can deliver precise, personalized, and long term therapies that are difficult to achieve using traditional pharmaceuticals [1]–[7]. One example application for bioelectronics is neural stimulation. Neural stimulators modulate the electrical activity of the nervous system providing effective treatment for a wide range of neurological and psychiatric disorders [8]–[12].Miniaturization is an important goal for bioelectronic devices like neural stimulators. By making devices smaller, one can reduce invasiveness and improve accessibility to difficult-to-reach areas of the body. However, on-board batteries used to power a device are often a bottleneck in reducing the device footprint. Moreover, the infections, movement, and breakage associated with wires that connect the battery and stimulator impose serious risks to the patient’s health and require periodic battery replacement surgeries [13]. To make extremely miniature bioelectronics, it is necessary to eliminate batteries in favor of wireless power and data transfer.Several methods are proposed in the literature including acoustic [14], electromagnetic [15], [16], and magnetic technologies with the latter comprising magnetothermal [17], [18], magnetoelectric(ME) [19]–[21], and inductive coupling [22]. While there are advantages to each approach, the physics of wireless data and power transfer imposes performance trade-offs [23]. Using ultrasound waves to transmit both power and data has been demonstrated for miniaturized ‘Stimdust’ motes [14]; however, ultrasound impedance mismatches between bone and tissue typically limit ultrasonic technology to soft tissue applications. In addition, because the ultrasound waves are typically focused by the transmitter, power and data transfer efficiency is sensitive to translational misalignment with the external transceiver. Likewise, the misalignment between the transmitter and the receiver coils limits the power received by the inductively coupled sub-millimeter neurostimulator reported in [22]. While misalignment has minor effects on the wireless power link of the RF implantable rectenna proposed in [16], a high operating frequency limits the amount of power that can be delivered safely to the implant. Although the magnetothermal techniques used to heat magnetic nanoparticles operate in the frequency range of a few hundred kHz where tissue absorption is low, the requirement for a relatively large magnetic field typically more than 10 mT at the location of the nanoparticles requires benchtop field generators that are not easily moved. The large size is due to the fact that these generators typically operate at kW power levels, and are often water-cooled [17], [18].Magnetoelectrics is a promising solution for millimeter-sized bioelectronic devices due to the large power density and ability to operate with low amplitude magnetic fields (<10 mT) at frequencies in the 100–500 kHz band where tissue absorption is low [24]. This technology utilizes a bi-layer thin film of mechanically coupled magnetostrictive and piezoelectric materials that transduce the magnetic field energy to electric energy that can be used to power implanted electronics or stimulate nearby tissues. For example, a mm-sized ME film activated using a low amplitude and low-frequency magnetic field has been used to successfully provide therapeutic deep brain stimulation in a freely moving rodent model for Parkinson’s disease [19]. Exploiting the CMOS technology, the first programmable neural implant leveraging the ME effect (MagNI) has been proposed in [20], [21], [25]. MagNI has a miniaturized volume of 8.2mm3 and can deliver programmable bi-phasic current stimulation for different neurostimulation applications. ME micro-resonators operating at GHz frequencies can also be used to power ultra-small low-power devices [26], [27].While proof-of-principle ME technologies show promise for wireless bioelectronics, many applications require a portable system that includes a wearable solution for data and power transfer. For instance, the spinal cord stimulator used for chemotherapy-induced pain could need up to 6–12 hours of use per day [28]. Likewise, patients must use an auricular vagus nerve stimulator used for depression and epilepsy treatment [29] and chronic pain [30] daily for at least 3–4 sessions.To design a wearable system for ME implants, it’s important to understand the working principles of both the transmitter and receiver and the design challenges associated with the coupling between them. Also, one should take into account the misalignment issues that are inevitable due to relative body motion, the movement of the device inside the body, and the difficulty that comes with patients wearing the transmitter belt to align the system properly. Moreover, since ME technology utilizes magnetic fields for power transfer, it’s important to ensure that the field exposure has no adverse effects on the user’s health.In our previous work [20], [21], we have proposed a conceptual design for a transmitter system that can be assembled in a wearable belt and power ME implants for spinal cord stimulation applications. In this paper, we demonstrate a prototype of a battery-powered wearable system capable of generating a low-amplitude (~1 mT) and low-frequency (~100s kHz) magnetic field sufficient to generate several mW of power in millimeter-sized magnetoelectric implant at a distance up to 4 cm. We first present the receiver and transmitter design theory and compare the simulation results to the experimental results. We then show the experimental results of the overall system performance at different distances, and translational misalignments. To validate the efficacy of our system in an in vivo environment, we demonstrate its ability to electrically stimulate the sciatic nerve of a rat with a transmitter-to-receiver distance of 4 cm. We further verify the system’s compliance with the IEEE C95.1–2019 safety standard by using COMSOL multiphysics software to model the brain and study the effects of magnetic field exposure on the implant’s surrounding tissue.
System overview
Here we describe a proof-of-principle wearable wireless power transfer (WPT) system for Magneto-Electric-powered Bio ImplanT or “ME-BIT” (Fig. 1). We define these general ME-powered implants as having a ME material to harvest data and power, and a rectifier circuit (Fig. 1.b,d). To power these millimeter-sized bioelectronic implants we designed a light-weight transmitter that combines custom drive electronics, wire coil, batteries, and a Bluetooth module integrated into a wearable belt as shown in Fig. 1.c. This transmitter system weighs less than 0.5 lbs and can generate a sufficient magnetic field to power the ME-BIT at an implantation depth of up to 4 cm.
Fig. 1.
Conceptual illustration of the wearable system, (a)applications including DBS,VNS, and SCS by adjusting the transmitter coil position in the head, neck, and back respectively, (b)ME receiver working principle, (c)prototype of the wearable transmitter system, (d)block diagram of the overall system.
In addition to providing power to the implant, this system can modulate the amplitude and frequency of the field to send data to the digitally programmable ME-BIT for neural interfaces “MagNI” [20], [21] or to control the stimulation waveform of analog ME-BIT neural simulators [19]. Moreover, the same system can be used to power the ME-BITs used in different neurostimulation applications including the deep brain (DBS), vagus nerve (VNS), and spinal cord stimulation (SCS) by adjusting the transmitter coil position in the head, neck, and back respectively (Fig. 1.a). We will discuss the receiver and transmitter characteristics and design specifications in the following sub-sections.
Receiver
Our ME-BIT utilizes a thin bilayer film of two smart materials, Metglas (magnetostrictive) and PZT-5 (piezoelectric), with the two layers bonded using epoxy as shown in Fig. 2.a. The film length, width, Metglas layer thickness, and PZT-5 layer thickness are denoted by l, w, t, and t respectively. The working principle depends on the material behavior of changing its properties in response to external stimuli like a magnetic field, electric field, or mechanical stress.
Fig. 2.
(a)ME film structure, (b)working principle of ME effect, (c) Equivalent circuit model of the ME film, (d) Equivalent impedance of the ME film.
Applying an alternating magnetic field H along the longitudinal direction of the Metglas layer excites a longitudinal mechanical strain. Due to the nonlinear characteristics of the magnetostrictive material, a significant mechanical strain can only develop when an additional bias field His applied [31]. An optimal H would shift the operating point of the Metglas layer to the inflection point of the magnetic field-strain curve at which the derivative of mechanical strain with respect to the magnetic field is maximum as shown in Fig. 2.b [32]. Mechanical coupling between the layers leads to the transfer of strain from the Metglas layer to the PZT-5 layer through the interfacial epoxy layer. This coupling induces an alternating electric voltage across the laminate’s thickness. The transduction of the magnetic energy to electric energy through the elastic coupling between the two laminates is what we refer to as the magnetoelectric effect.For wireless power transfer applications, it is important to quantify the electric energy that can be harvested by the receiver in response to a certain magnetic field as well as evaluating the transduction efficiency and the parameters affecting it. The electric energy harvested by the ME transducer depends on the ME coupling between the two laminates which is usually evaluated using the ME voltage coefficient α. α is defined as the change in the film’s output electric voltage V in response to a change in the applied magnetic field such that:According to the equivalent circuit model developed in [33], [34] and depicted in Fig. 2.c, the open circuit ME voltage coefficient is expressed as:Where , and are the magneto-elastic and electro-elastic coupling factors, C0 is the clamping capacitance of the PZT-5 laminate, and Z represents the mechanical impedance.The mechanical impedance is mainly a function of the mechanical quality factor Q, the resonant frequency and Z = Avρ given that the cross-sectional area A = A1 + A2 where A1, A2 represents the cross-sectional area of the PZT-5 and Metglas laminates respectively, the geometric ratio , the average density and the sound velocity . All material constants used to calculate the ME coupling coefficient are defined and listed in Table. I.
Table 1.
Material properties of Metglas and PZT-5 laminates.
We study the ME-transducer geometric and fabrication properties that affect the ME coupling hence the transduction efficiency of the ME receiver. Specifically we find that the ME voltage coefficient is independent of the area of the film and increases as the interface coupling factor k, the thickness ratio a, or the mechanical quality factor Q increases. Moreover, the resonance frequency of the film is shifted to the right as the length, interface coupling factor, or thickness ratio decreased whereas it’s independent of the mechanical quality factor.Our modeling shows that the ME voltage coefficient is a function of the frequency, material properties, and geometry. The ME coupling is also affected by the interfacial epoxy layer characteristics. As reported in [35], the ME coupling coefficient decreases as the thickness of the bonding layer increases, or Young’s modulus of the bonding material decreases. To account for this imperfect coupling, authors in [34] introduced an interface coupling factor k, 0 < k < 1, defined as the percentage of the mechanical stress transferred from the magnetostrictive layer to the piezoelectric layer.Fig. 3 depicts the general trends of the calculated ME voltage coefficient considering the variation of the area, quality factor, interface coupling factor, and thickness ratio a = t/t. To begin with, the film is assumed to have a 6mm×3mm perfectly coupled Metglas and PZT-5 layers with a thickness ratio of 5.2, and a quality factor of 35.
Fig. 3.
Calculated ME voltage coefficient with a variation of (a)film area where aspect ratio l/w is fixed to 3, (b)interface coupling factor, (c)thickness ratio, (d)quality factor.
The ME voltage coefficient is:
independent of area:
In Fig. 3.a, the aspect ratio of the film l/w is fixed to 3, and the length is varied from 4mm to 8mm. While the ME voltage coefficient is not affected by changes in the film area, the resonant frequency decreases as the length increases. Although both length and width are changed, the resonant frequency of the fundamental mode is only determined by the length as defined earlier .
increased with larger interface coupling factors :
The bonding layer characteristics significantly affect the film performance as shown in Fig. 3.b, where a stronger coupling between the magnetostrictive and piezoelectric layers results in a higher ME voltage coefficient. However, achieving perfect coupling is limited by the available bonding materials and the fabrication process.
increased with larger piezoelectric to magnetostrictive thickness ratios:
The ME voltage coefficient is also significantly affected by the ratio between the two laminate thicknesses as shown in Fig. 3.c. A thicker piezoelectric layer compared to the magnetostrictive layer is more desirable.
increased with larger mechanical quality factors:
A larger mechanical quality factor results in stronger ME coupling as depicted in Fig. 3.d. This factor is usually used to quantify the strain amplification in the ME laminate and is defined as the ratio between the resonant frequency and the 3 dB frequency bandwidth [36]. As can be seen for films with the same resonant frequency, the ones with higher Q have narrower bandwidth.
Factors affecting the ME receiver impedance:
In addition to the voltage profile of the ME film, it’s important to study the impedance characteristics to optimize the output power and better match the film with the load impedance. Here we find that, at the resonance frequency, the film impedanceis dominated by the resistive part and it’s magnitude mainly depends on the material properties and film’s geometry.The equivalent impedance of the film Z can be defined as the parallel combination of the mechanical impedance Z reflected at the electric part and the clamping capacitor C0 impedance as shown in Fig. 2.d.:Fig. 4 shows the calculated real and imaginary parts of the equivalent impedance of the film defined as the base case previously which has a resonant frequency of 242 kHz. As can be seen, the film impedance is dominated by the real component of the impedance at the resonant frequency, while the imaginary part is more significant off-resonance. Since our system will operate at the resonant frequency to ensure efficient performance, we focused on the characteristics of the real part of the equivalent impedance.
Fig. 4.
Calculated real and imaginary parts of a 6*3 mm ME film impedance.
Our model shows that the real impedance is also a function of the frequency, material properties, and geometry. The real impedance magnitude:
decreases with width and increases with aspect ratio:
In Fig. 5.a the length of the film is fixed to 6 mm whereas the width is changed such that the aspect ratio varies from 1 to 5. Clearly, the resonant frequency does not significantly change since the length is fixed (Fig. 5.a), however, the magnitude of the real impedance decreases as the width increases. Also, the impedance is linearly proportional to the aspect ratio.
Fig. 5.
Calculated real impedance with variation of (a)aspect ratio, (b)interface coupling factor, (c)thickness ratio, (d)quality factor.
increases with larger interface coupling factors :
The interface coupling factor affects both the impedance amplitude and the resonance frequency as shown in Fig. 5.b, a stronger coupling results in larger impedance and a lower resonance frequency.
increases with with larger piezoelectric to magnetostrictive thickness ratios:
As shown in Fig. 5.c, the thickness ratio can noticeably change the impedance and frequency such that a thicker PZT layer with respect to the Metglas layer results in a larger impedance magnitude and lower resonance frequency.
increases with larger mechanical Q factor:
as can be seen in Fig. 5.d, the magnitude of the real impedance is linearly proportional to the magnitude of the mechanical quality factor, whereas the resonance frequency is independent of Q.
Transmitter
To generate sufficient magnetic fields to power the implanted ME-BIT, we designed a battery-powered transmitter system as shown in Fig. 6.a. A rechargeable lithium-ion battery connected to an H-bridge converter drives a pulsed alternating current through a resonant circuit of an inductive coil and capacitor bank in series. Three parameters characterize the coil’s pulsed current: the frequency, duty, and pulse frequency as shown in Fig. 6.b. These parameters determine the magnetic field generated by the coil which directly affects the voltage induced at the ME transducer terminal. The user can set these parameters using a mobile application that communicates them through Bluetooth to the driver circuit that combines an HM-19 Bluetooth module, Teensy LC microcontroller, and custom electronics to generate the H-bridge switching signals accordingly.
Fig. 6.
(a)Transmitter system, (b) Coil’s pulsed current parameters.
To power implanted devices using a wearable system, one needs to efficiently generate a sufficient magnetic field at a distance from the coil surface, therefore, the design of the transmitter coil is specially important. For the transmitter coil, we tested using a circular coil that can be wrapped around the patient’s waist or a spiral coil that can be placed parallel to the implantation site. We built a 3D model of both coils using COMSOL multiphysics and computed the magnetic field norm values using the AC/DC module as shown in Fig. 7. The circular coil has a major radius of 15 cm whereas the spiral coil outer radius is 2.8 cm. For the same excitation current of 10 A, the spiral coil has higher magnetic field at a distance compared to the circular coil. In addition, the spiral coil has a more compact size and superior misalignment tolerance [37]–[39].
Fig. 7.
(a) Spiral coil that can be placed parallel to the implantation site and its corresponding magnetic field with 10 A excitation current. (b) Circular coil that can be wrapped around the patient waist and its corresponding magnetic field with 10 A excitation current.
The inductance of the spiral coil can be calculated as [40]:
Where N is the turns number, Din is the inner diameter, Dout is the outer diameter, and μ0 is the free space permeability= 4π × 10−7H/m.TAs mentioned in the previous sub-section, the implanted ME-BIT is sensitive to the direction of the magnetic field and reaches a maximum response when the field lines are parallel to its long axis therefore it is critical to determine the magnetic field spatial distribution and its components to place the mote at the optimal position/orientation.Using the model of the spiral coil shown in Fig. 7.a (15 turns, 1.5 cm inner diameter, and 5.6 cm outer diameter) we studied the spatial distribution of the generated magnetic field as can be seen in Fig. 8, the grey arrows represent the pattern of the magnetic field lines and are proportional to the field magnitude. As shown here, the magnetic field lines are parallel to the z-direction at the center of the coil and bend toward the coil edges as the radial distance from the center increases.
Fig. 8.
(a) 3D model of the circular spiral coil used as a transmitter coil, (b)Magnetic field distribution at the coil surface, and magnetic field lines pattern(grey arrows).
We evaluated the field distribution of the magnitude of the z-, x-, and y-components at different vertical distances from the coil surface as shown in Fig. 9.
Fig. 9.
z-component, x-component, and y-component of the magnetic field generated by the transmitter coil at 1 cm vertical distance from the coil surface, (b) the maximum z-component, x-component, and y-component of the magnetic field as a function of the vertical distance from the coil surface.
As can be seen, the z-component has a maximum value at the center of the coil and decays with an increase in the distance from the coil surface and the radial distance from the center. On the other hand, the x-component has a symmetric distribution around the y-axis and zero along the y-axis itself whereas the y-component has a symmetric distribution around the x-axis and zero along the x-axis itself. As expected, since we have a circular coil, the x and y components are of equal magnitude and the y-component distribution is shifted by 90° compared to the x-component distribution. Moreover, the x- and y-components have a lower magnitude compared to the z-component and their maximum is shifted from the center.
System Performance
To demonstrate that we can create a wearable system capable of powering our ME-BIT implants we built a prototype transmitter as shown in Fig. 10.a. This system is powered by a set of four 3.7 V, 2000 mAh rechargeable Lithium-ion batteries as depicted in Fig. 10.a. The batteries are connected to the resonant circuit through an H-bridge that turns the input DC current into pulsed AC current based on the switching signals provided by the driver circuit. To generate the switching signals, the HM-19 module connected to the driver circuit receives the stimulation parameters set by the user through a mobile application and communicated via Bluetooth. Based on the received parameters, the Teensy LC microcontroller and custom electronics determine the timing and voltage levels to generate the switching signals accordingly. A resonant circuit of a circular spiral coil connected in series to a capacitor is used to generate the alternating magnetic field required to activate the ME film, moreover, a permanent magnet is used to provide a bias field to ensure efficient transduction.
Fig. 10.
(a)System prototype, (b)physical realization of the system, (c)ME-BIT used as a receiver.
For the proof-of-concept ME receiver, we fabricated a 7 mm*3 mm bilayer film of 30 μm-thick layer of Metglas 2605SA1 (Metglas Inc) attached using HARDMAN® epoxy to a 270 μm-thick layer of PZT-5A4E (Piezo Inc) as shown in Fig. 10.c.To measure the film ME voltage coefficient, we placed the film at the center of the spiral coil such that it’s long axis is parallel to the z-axis then we measured the open-circuit voltage and magnetic field strength at different frequencies to compute the ME voltage coefficient as shown in Fig. 11.a. As can be seen, the film’s mechanical resonant frequency is around 222 KHz at which it has a maximum ME voltage coefficient of 0.5 V/Oe. Fitting the measured data to the equivalent circuit model mentioned in section II.B suggests that the film has an interfacial coupling factor of 0.43 and a mechanical quality factor of 41. Fig. 11.b shows the calculated and measured real impedance of the film, at the resonance frequency, the film has a maximum resistive impedance of 1700 Ω.
Fig. 11.
(a) Measured and calculated ME voltage coefficient of the ME film, (b) Measured and calculated real impedance of the film.
To generate a magnetic field that resonates with the ME film, the pulsed current parameters are set via Bluetooth as follows: frequency = 100 Hz, pulse frequency = 222 KHz, and duty = 0.012. The pulse frequency is set based on the film resonant frequency, while the frequency and the duty values can be changed based on the application. For the resonant circuit, a coil with the same features mentioned in section II.A (D=1.5 cm, D=5.6 cm,N=15) is made using 1 mm thick Litz wires to minimize the skin effect. The measured inductance of the coil is 7.3 μH, hence, a 70 nF capacitor is used. This setting results in 54 mA DC current being drawn from the batteries, therefore, the system can operate continuously for up to 37 hours.As mentioned in previous sections, the dominant component of the magnetic field at the coil’s center is the z-component, so we placed the film at the center of the coil such that its longer axis is parallel to the z-axis. In addition to the AC magnetic field, a permanent magnet is used to provide a DC bias field of 3mT. We note that for some surgical approaches it may be preferred to place the long axis of the film parallel to the skin rather than perpendicular as described here. In that case, the film can still be powered by simply shifting the coil so that the film is not directly under the coil center. In that case the fringing x or y components of the magnetic field can be used to power devices that lie parallel to the skin’s surface Fig. 9.a.To measure the maximum power, the film is connected to a resistive load of 1700 Ω, the load voltage is measured and the load power is computed as as shown in Fig. 12a. As the distance from the coil increases, the magnetic field decays resulting in lower output power from the ME film. Nevertheless, we can achieve an output power of 1.8 mW at a 3 cm distance where the applied magnetic field at that distance is equal to approximately 0.5 mT. Although for some applications the load impedance can not match the optimal value, an impedance matching network can be used to deliver the maximum available power.
Fig. 12.
(a) Film’s output power and output voltage as a function of the distance from the coil surface, (b) Normalized values of the film output power and output voltage, and the applied magnetic field and it’s square.
The normalized value of the film’s output voltage, power, applied magnetic field and the square of the field as a function of the vertical distance are shown in Fig. 12.b. As mentioned in section II.B, the voltage is a linear function of the applied magnetic field whereas the power is proportional to the square of the magnetic field.The efficiency of the wireless power transfer is shown in Fig. 13.a. As can be seen, the efficiency has a maximum value of 1% at the coil surface and decays as the distance between the transmitter coil and ME film increases.
Fig. 13.
(a) Measured power transfer efficiency over various coil- film distance.(b) Normalized value of the film output voltage and applied magnetic field as a function of radial distance shift from the coil center.
To measure how robust the power transfer efficiency is to positional misalignments we measured the film’s output voltage for different displacements relative to the optimal coupling position. Fig. 13.b shows the normalized value of the z component of the magnetic field and the output voltage of the film as it is shifted radially from the coil center at different vertical distances from the coil. The long axis of the film is maintained parallel to the z-axis, in this case, the output voltage of the film is directly related to the z component of the magnetic field which decays radially as mentioned previously. We find that 80% and 50% of the maximum voltage can still be maintained at the film terminal with 10 and 20mm shifts, respectively (Fig. 13.b). This misalignment tolerance is due to the fact that the voltage of the ME film is directly proportional to the applied ac magnetic field as mentioned in section II-A, and that the spiral coil has a high magnetic field uniformity as shown in Fig. 7.Moreover, due to the flux concentration properties of the Metglas layer,the ME power link shows improved angular misalignment tolerance compared to inductive coupling as reported in [21].
In vivo Electrical Stimulation for Sciatic Nerve
As a proof-of-concept, we demonstrate that our wireless transmitter can power an ME film at a distance of 4 cm in Phosphate Buffer Solution (PBS) with enough energy to stimulate rat sciatic nerve in-vivo. Demonstrating the ability to power an implant at a depth of 4 cm in a rodetn model is challenging due to the fact that most nerve targets are with 1 cm of the skin’s surface. Therefore, to characterize the in vivo transmitter performance at an implant depth of 4 cm, which is relevant for human applications, we used the experimental setup shown in Fig. 14.a. The ME film encapsulated with 20 μm of Parylene-C is submerged in PBS at a distance of 4 cm from a transmitter coil. The film is connected to platinum electrodes that are placed into contact with the sciatic nerve of the rat. Since neurons are unresponsive to the high frequency ME voltages, we incorporated a commercial diode in parallel with the ME film to rectify the signal and modulated the field to a lower frequency (as shown previously in [19]). By pulsing the applied alternating magnetic field we were able to deliver a rectified voltage envelop of 1 Hz over the 222 kHz carrier frequency that was tuned to the film resonance (Fig. 14b–c). When we tuned the carrier frequency to the ME resonance we observed leg kick, which conformed that the voltage generated by the ME film was able to stimulate the nerve. Furthermore, we recorded the resulting compound muscle action potentials through EMG recording needles placed in the plantar muscles of the rat foot as shown in Fig. 14.d, blue line). To confirm that the stimulation was the result of the voltage produced by the ME film, we removed the DC bias magnet, which reduces the amplitude of the ME voltage. In the absence of the DC bias we observed no EMG response of leg kick (Fig. 14.d, red line).
Fig. 14.
(a)Experimental setup for rat sciatic nerve stimulation and EMG recording.(b)Transmitter coil’s current.(c)Rectified ME voltage.(d)EMG elect rod’s voltage.
Safety Analysis
Comparing the magnetic and electric fields generated by our transmitter to the IEEE Std C95.1–2019 exposure limits we find that our transmitter is well within the range of safe exposure levels. Transmitting a time-varying magnetic field through the conductive biological tissues will induce an electric field according to Maxwell’s equations. The induced electric field drives an electrical current through the tissue resulting in heat dissipation that could be detrimental to the tissue’s health. The generated heat depends on two frequency-dependent properties of the tissue: the dielectric permittivity and the electrical conductance.Given the potential of magnetic and electric fields to heat tissue, the IEEE Std C95.1–2019 set the limits of the electric and magnetic field exposure in the 0 Hz - 300 GHz that would ensure safe operation and protect against adverse health effects in humans [41]. The compliance with the standard implies meeting the electrostimulation limits when working below 100 kHz, the thermal limits above 5 MHz, and both limits when working in the transition region between 100 kHz and 5 MHz. The electrostimulation limit is set by the dosimetric reference limit (DRL) that is expressed as the electric field in the transition region, whereas the thermal limit is expressed by the specific absorption rate (SAR) that is defined as [41]:
Where σ is the conductivity of the tissue (S/m), ρ is the density of the tissue (kg/m3), and E is the RMS electric field strength (V/m).Since our system works in the 100s kHz range, both the electric field and the SAR limits should be met to ensure safe operation. To calculate these values, we modeled the seven-layers brain model reported in [42] using COMSOL. The model includes seven distinct layers of the brain tissues: skin, fat, cortical bone, dura, cerebrospinal fluid (CSF), grey matter, and white matter as shown in Fig. 15.a.
Fig. 15.
(a)Seven-layer brain tissue model,(b)RMS values of the magnetic field for a vertical cross-section at x=0 such that 0.5mT peak value is generated at 3 cm, (c)Max RMS electric field associated with magnetic field in (b) across the different brain tissues at 100, 250, 500 kHz(solid lines), IEEE Std C95.1–2019 electric field safety limits(dashed lines).
For a ME film with a length in the range of 3–10 mm, the resonance frequency is expected to be within the 100–500 kHz range as shown in Fig. 3, therefore, we analyzed the system’s compliance with the safety limits at this range. The properties of the tissues at these frequencies are listed in Table. IV (Appendix. A) [43]. As reported in the IEEE Std C95.1–2019, the limit of the SAR in unrestricted environment is 2W/kg, whereas the limit of the electric field for the head tissues for persons in unrestricted environment is given by:
Table 4.
Brain tissues properties at 100, 250, 500 khz for each layer.
Frequency = 100kHz
Frequency = 250kHz
Frequency = 500kHz
Tissue
Density kg/m−3
εr
σ
εr
σ
εr
σ
Skin
1109
1.12 × 103
4.5 × 10−4
1.1 × 103
1.46 × 10−3
1.06 × 103
4.36 × 10−3
Fat
911
1.01 × 102
4.34 × 10−2
6.78 × 101
4.36 × 10−2
5.68 × 101
4.38 × 10−2
Cortical bone
1908
2.28 × 102
2.08 × 10−2
1.97 × 102
2.12 × 10−2
1.75 × 102
2.22 × 10−2
Dura
1174
3.26 × 102
5.02 × 10−1
2.82 × 102
5.02 × 10−1
2.64 × 102
5.03 × 10−1
CSF
1007
1.09 × 102
2
1.09 × 102
2
1.09 × 102
2
Grey matter
1045
3.22 × 103
1.34 × 10−1
1.75 × 103
1.43 × 10−1
1.19 × 103
1.52 × 10−1
White matter
1041
2.11 × 103
8.18 × 10−2
1.11 × 103
8.85 × 10−2
7.12 × 102
9.47 × 10−2
Where f is the frequency of the applied magnetic field.Fig. 15.a shows the model of the transmitter coil and brain tissues with a transverse area of 100mm x 100mm. The magnetic fields, electric fields, and the average SAR are computed for the different layers using the magnetic and electric field modules available in COMSOL. Based on this model, we computed the magnetic field distribution through the different tissues such that a~ 0.4mT peak value is generated at 4 cm. Fig. 15.b depicts the distribution of the RMS values of the magnetic field for a vertical cross-section at x=0, at which we expect maximum field strength, moreover, the maximum values at each layer are listed.The generated electric field associated with this magnetic field strength depends on the operating frequency. Assuming the worst case scenario which is continuous exposure of magnetic field, the maximum value at each layer of the RMS electric field is shown in Fig. 15.c, the dashed lines represent the safety limit at each frequency, as can be seen, the electric field values are lower than the safety limits for all frequencies. The values of the SAR averaged over 10g for each layer are listed in Table. II. All values are below the safety limit of 2 W/Kg. It can be noted that the largest SAR value is for the CSF layer due to its high conductivity across the different frequencies whereas the largest electric field is generated at the skin layer as it is the closest to the coil. Further, to determine the maximum magnetic field values that can be transferred safely through the tissue in this frequency range we computed the electric field value at the skin layer and the SAR values at the CSF layer for different magnetic field exposure. As can be seen in Fig. 16, a maximum value of 8 mT can be generated at the coil surface and transmitted through tissue safely.
Table 2.
Computed average SAR (W/Kg−1) values over 10 g at 100, 250, 500 for each layer.
100k
250k
500k
Skin
3.4 × 10−5
6.8 × 10−4
8.1 × 10−3
Fat
3.0 × 10−3
1.9 × 10−2
7.5×10−2
Bone
5.0 × 10−5
3.2 × 10−4
1.3 × 10−2
Dura
7.9 × 10−3
4.9 × 10−2
19.4 × 10−2
CSF
2.6 × 10−2
17.2 × 10−2
68.8 × 10−2
Grey matter
7.2 × 10−4
4.8 × 10−3
2.0 × 10−2
White matter
2.5 × 10−6
1.7 × 10−5
7.2 × 10−5
Fig. 16.
(a)Maximum electric field generated at the skin layer for different coil’s magnetic field, (b)Maximum SAR generated at the CSF layer for different coil’s magnetic field.
Conclusion
This work demonstrates a proof of principle light-weight, battery-powered wearable transmitter system that can power a miniaturized ME neural stimulator and has a battery life of 37 hr. The working principle, theory, and design of the transmitter and receiver are reported. The proposed system is capable of wirelessly powering a miniaturized 7×3 mm Metglas/PZT-5 ME receiver at a distance of 4 cm with enough engergy to stimulate the sciatic nerve of a rat in vivo. Moreover, the experimental results show that the system performs well under translational misalignment. We also provide some general guidelines for how the size, composition, and quality factors of ME films affect energy harvesting performance. Compared to the state-of-the-art wearable solutions for wirelessly powered implants (Table. III), our system utilizes rechargeable batteries to power the transmitter coil that generates sufficient magnetic fields to power deeply implanted miniaturized ME stimulators. Moreover, our system operates at a low-frequency range where tissue absorption of the magnetic field is negligible. Finally, using a COMSOL model of brain tissue we computed the magnetic field, electric field, and SAR values to verify the system compliance with the IEEE C95.1–2019 safety standard across the operating frequency range.
Table 3.
Comparison with state of the art wearable systems for wireless powered implants.
Authors: Amanda Singer; Shayok Dutta; Eric Lewis; Ziying Chen; Joshua C Chen; Nishant Verma; Benjamin Avants; Ariel K Feldman; John O'Malley; Michael Beierlein; Caleb Kemere; Jacob T Robinson Journal: Neuron Date: 2020-06-08 Impact factor: 17.173
Authors: Zhanghao Yu; Joshua C Chen; Fatima T Alrashdan; Benjamin W Avants; Yan He; Amanda Singer; Jacob T Robinson; Kaiyuan Yang Journal: IEEE Trans Biomed Circuits Syst Date: 2020-12-31 Impact factor: 3.833
Authors: Daniel K Freeman; Jonathan M O'Brien; Parshant Kumar; Brian Daniels; Reed A Irion; Louis Shraytah; Brett K Ingersoll; Andrew P Magyar; Andrew Czarnecki; Jesse Wheeler; Jonathan R Coppeta; Michael P Abban; Ronald Gatzke; Shelley I Fried; Seung Woo Lee; Amy E Duwel; Jonathan J Bernstein; Alik S Widge; Ana Hernandez-Reynoso; Aswini Kanneganti; Mario I Romero-Ortega; Stuart F Cogan Journal: Front Neurosci Date: 2017-11-27 Impact factor: 4.677
Authors: Joshua C Chen; Peter Kan; Zhanghao Yu; Fatima Alrashdan; Roberto Garcia; Amanda Singer; C S Edwin Lai; Ben Avants; Scott Crosby; Zhongxi Li; Boshuo Wang; Michelle M Felicella; Ariadna Robledo; Angel V Peterchev; Stefan M Goetz; Jeffrey D Hartgerink; Sunil A Sheth; Kaiyuan Yang; Jacob T Robinson Journal: Nat Biomed Eng Date: 2022-03-31 Impact factor: 29.234