A diagnosis of lymphedema comes with a lifetime requirement for careful self-care and treatment to control skin deterioration and the consequences of excessive fluid and protein buildup leading to abnormal limb volume and an increased risk of infection. The burden of care and psychosocial aspects of physical disfiguration and loss of function are associated with compromised quality of life (QoL). The current standard therapeutic intervention is complex decongestive therapy with manual lymph drainage and frequent wearing of compression garments. With insurance limitations on therapy visits and the time and travel required, additional home treatment options are needed. Pneumatic compression pumps that mimic the manual massage pressure and pattern are sometimes prescribed, but these are bulky, difficult to apply, and require immobility during treatment. An open-label pilot study in 40 subjects was performed to evaluate the QoL and limb volume maintenance efficacy of a novel wearable compression system (Dayspring™) that is low profile, easy to use, and allows for mobility during treatment. After 28 days of use, subjects had a statistically significant 18% (p < 0.001) improvement in overall QoL as measured by the Lymphedema Quality-of-Life Questionnaire compared with baseline. Individual QoL domains, and limb volume improved with therapy. Adherence was 98% over the course of the study. Results of the clinical evaluation suggest the Dayspring wearable compression device is safe and effective and improves QoL and limb volume. The novel, low-profile device is easy to use and allows for mobility during treatment, addressing a potential barrier to adherence with pneumatic compression devices.Entities:
Keywords: advanced compression; lymphedema; mobility; quality of life; wearable
Mesh:
Year: 2021 PMID: 34227842 PMCID: PMC9081034 DOI: 10.1089/lrb.2020.0126
Source DB: PubMed Journal: Lymphat Res Biol ISSN: 1539-6851 Impact factor: 2.349
FIG. 1.Schematic of the Dayspring—a smart wearable device—shown with the controller (right) and garment (left).
Demographics
| Patients | 40 |
| Age (mean ± SD) | 65.7 ± 8.8 |
| Female (male) | 39 (1) |
| Race | |
| Caucasian | 32 |
| African American | 1 |
| Hispanic | 3 |
| Other | 4 |
| Affected arm | |
| Left | 22 |
| Right | 18 |
| Sites | 2 |
| Months since ALDN/RT | 26 ± 23 |
ALDN/RT, axillary lymph node dissection/radiotherapy; SD, standard deviation.
FIG. 2.Lymphedema overall quality-of-life (LYMQOL) outcomes from subjects before and after use of the Dayspring device over a 1-month evaluation period. Horizontal bars represent mean ± 1 standard deviation. LYMQOL, Lymphedema Quality-of-Life Questionnaire; QoL, quality of life.
FIG. 3.Lymphedema qualify-of-life (LYMQOL) functional outcomes from subjects before and after use of the Dayspring device over a 1-month evaluation period. Horizontal bars represent mean ± 1 standard deviation.
Lymphedema Quality-of-Life Questionnaire Results (n = 40)
| Subscale | Baseline (mean ± SD) | Posttest (mean ± SD) |
| Mean difference (from baseline) | % Difference (from baseline) |
|---|---|---|---|---|---|
| Function | 1.64 ± 0.43 | 1.42 ± 0.52 | <0.001 | 0.22 | 13 |
| Appearance | 2.26 ± 0.73 | 1.72 ± 0.65 | <0.001 | 0.54 | 24 |
| Mood | 1.73 ± 0.59 | 1.28 ± 0.35 | <0.001 | 0.45 | 22 |
| Symptoms | 2.01 ± 0.69 | 1.56 ± 0.52 | <0.001 | 0.45 | 26 |
Lymphedema Quality-of-Life Questionnaire Results (Prior Pneumatic Compression Device Use) (n = 15)
| Subscale | Baseline (mean | Posttest (mean ± SD) |
| Mean difference (from PCD users) | % Difference (from PCD users) |
|---|---|---|---|---|---|
| Function | 1.59 ± 0.40 | 1.40 ± 0.40 | <0.001 | 0.19 | 12 |
| Appearance | 2.23 ± 0.58 | 1.76 ± 0.62 | <0.001 | 0.47 | 21 |
| Mood | 1.55 ± 0.44 | 1.27 ± 0.30 | <0.001 | 0.28 | 18 |
| Symptoms | 1.99 ± 0.70 | 1.60 ± 0.58 | <0.001 | 0.39 | 19 |
PCD, pneumatic compression device.
FIG. 4.Volume reduction in subjects before and after use of the Dayspring device over a 1-month evaluation period. Horizontal bars represent mean ± 1 standard deviation.
FIG. 5.Dayspring companion app for digital interface and adherence monitoring.
FIG. 6.VAS score in preference of Dayspring over PCDs from subjects after the trial. PCDs, pneumatic compression devices; VAS, visual analog scale.