| Literature DB >> 34216435 |
Gonca E Karahan1, Yvonne de Vaal1, Kim Bakker1, Dave Roelen1, Frans H J Claas1, Sebastiaan Heidt1.
Abstract
Detection of HLA-specific memory B cells can provide additional information on sensitization of alloantigen-exposed individuals and refine immunological risk assessment. We have recently developed an assay enabling profiling of memory B cell-derived HLA antibodies using luminex single antigen bead (SAB) assay. Here, we compared the performance of the SAB kits from two vendors for memory B cell-derived HLA antibody detection. IgG was isolated from culture supernatants of polyclonally activated B cells from alloantigen-exposed (n = 7) or nonexposed (n = 5) individuals, using our previously established method. Eluates containing isolated IgG from culture supernatants were tested for the presence of HLA antibodies using luminex SAB analysis from both One Lambda and Lifecodes (Immucor). In contrast to Lifecodes, high mean fluorescence intensity (MFI) signals were found for negative control beads in One Lambda (median MFI for class I:1730 and for class II:728), accompanied by high MFI values for self HLA-coated beads, especially for HLA-C. Despite high background in the One Lambda assays, 91% concordance for HLA class I and 85% concordance for HLA class II were found between the specificities detected using SAB kits from the two vendors. Our results show that HLA-specific memory B cells can be profiled using kits from both vendors. However, when analyzing One Lambda results one should be aware of the restrictions related to nonspecific binding particularly in HLA-C-coated beads, and pay attention to self HLA-coated beads in order to accurately identify the reactivities leading to the definition of the actual HLA antibody specificities.Entities:
Keywords: bead-based assays; donor-specific antibody; humoral alloimmune response; memory B cells
Mesh:
Substances:
Year: 2021 PMID: 34216435 PMCID: PMC8456970 DOI: 10.1111/tan.14356
Source DB: PubMed Journal: HLA ISSN: 2059-2302 Impact factor: 4.513
FIGURE 1MFI values for negative control beads and self HLA‐coated beads of the study cohort (n = 12) eluates tested using Lifecodes and One Lambda single antigen bead kits. BC, background corrected; MFI, mean fluorescence intensity
FIGURE 2Bead reactivity pattern of all Lifecodes and One Lambda class I beads in eluates from alloantigen nonexposed individuals (n = 5). Dashed lines are set at MFI of 750 and MFI/LRA ratio of 4.62 (highest bead‐specific cut‐off). Gray dots indicate HLA‐C‐coated beads. NBG, normalized background
Memory B cell‐derived HLA antibody profiles as detected by SAB kits from two different vendors in samples from alloantigen exposed individuals
| Case | HLA class | Mismatched HLA (partner and/or child) | HLA antibody profiles by Lifecodes | HLA antibody profiles by One Lambda |
|---|---|---|---|---|
|
|
| B | B | |
|
| n.a | Negative | Negative | |
|
|
| A |
| |
|
| DRB1 | Negative | Negative | |
|
|
| B | Negative |
|
|
| DRB1 | DRB1 | DRB1 | |
|
|
| A | A | A |
|
| DRB1 | |||
|
|
| A | ||
|
| n.a | DRB1 | DRB1 | |
| DPB1 | ||||
| DPB1 | ||||
|
|
| B | ||
|
| n.a | DRB1 | DRB1 | |
|
|
| A1, A3, B7, B27, Cw2 | ||
|
| n.a | DRB1 | DRB1 | |
Note: Antibodies directed to a mismatched HLA of an immunizing event are written in bold and italic. Specificities (within the common beads) that were found positive by only one vendor's kits are marked with red. Beads exclusive for one vendor are underlined.
Abbreviation: n.a, not available.
Bead positivity based on raw MFI >750 and bead‐specific MFI/LRA cut‐off by Lifecodes MATCHIT! antibody analysis software.
Bead positivity based on score 8 according to baseline (normalized) MFI values by One Lambda HLA FUSION antibody analysis software.