Changes in corrected carotid flow time induced by recruitment maneuver predict fluid responsiveness in patients undergoing general anesthesia.

Aya Kimura1, Koichi Suehiro2, Takashi Juri1, Katsuaki Tanaka1, Takashi Mori1.   

Abstract

Non-invasive methods to assess patients' fluid responsiveness during lung-protective ventilation are needed. We hypothesized changes in the corrected carotid flow time induced by the recruitment maneuver predict fluid responsiveness under general anesthesia. Thirty patients undergoing general anesthesia in the supine position were prospectively enrolled. The study protocol was conducted when the patient was hemodynamically stable during surgery. Flow time was measured on Doppler images of the common carotid artery. Carotid flow time, heart rate, stroke volume, stroke volume variation, and pulse pressure variation were recorded before and after a recruitment maneuver at a continuous airway pressure of 30 cmH2O for 30 s, and before and after volume expansion with 250 mL for 10 min. Patients were defined as fluid responders if the increase in stroke volume was > 10% after volume expansion. Twenty patients (67%) were fluid responders. All Doppler images for carotid flow time were obtained within 30 s. Changes in the corrected flow time accurately predicted fluid responsiveness (area under the curve: 0.82, 95% confidence interval [CI] 0.64-0.94, p = 0.002). The optimal threshold for changes in the corrected flow time was - 11.7% with a sensitivity of 95.0% (95% CI 75.1-99.9%) and a specificity of 80.0% (95% CI 44.4-97.5%). The gray-zone of changes in the corrected flow time was from - 25.1 to - 12.2% and included 12 patients (40%). Changes in the corrected carotid flow time were a useful, technically easy-to-perform, and non-invasive method to predict fluid responsiveness without a need for hemodynamic monitoring or arterial cannulation.
© 2021. The Author(s), under exclusive licence to Springer Nature B.V.

Entities:  

Keywords:  Carotid flow time; Fluid loading; Lung recruitment maneuver; Non-invasive monitoring; Stroke volume; Ultrasound

Mesh:

Year:  2021        PMID: 34191254     DOI: 10.1007/s10877-021-00736-7

Source DB:  PubMed          Journal:  J Clin Monit Comput        ISSN: 1387-1307            Impact factor:   1.977


  2 in total

1.  Changes in stroke volume induced by lung recruitment maneuver can predict fluid responsiveness during intraoperative lung-protective ventilation in prone position.

Authors:  Ryota Watanabe; Koichi Suehiro; Akira Mukai; Katsuaki Tanaka; Tokuhiro Yamada; Takashi Mori; Kiyonobu Nishikawa
Journal:  BMC Anesthesiol       Date:  2021-12-02       Impact factor: 2.217

2.  Correlation of carotid corrected flow time and respirophasic variation in blood flow peak velocity with stroke volume variation in elderly patients under general anaesthesia.

Authors:  Yu Chen; Ziyou Liu; Min Zhang; Jia Yang; Jun Fang; Yanhu Xie
Journal:  BMC Anesthesiol       Date:  2022-08-04       Impact factor: 2.376

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.