| Literature DB >> 34191079 |
Miranda Morrison1, Athanasia Korda1, Ewa Zamaro1, Franca Wagner2, Marco D Caversaccio1, Thomas C Sauter3, Roger Kalla4, Georgios Mantokoudis5.
Abstract
OBJECTIVE: Cold and warm water ear irrigation, also known as bithermal caloric testing, has been considered for over 100 years the 'Gold Standard' for the detection of peripheral vestibular hypofunction. Its discovery was awarded a Nobel Prize. We aimed to investigate the diagnostic accuracy of Caloric Testing when compared to the video head impulse test (vHIT) in differentiating between vestibular neuritis and vestibular strokes in acute dizziness.Entities:
Keywords: Acute stroke; Caloric testing; Dizziness; Head-impulse test; Neuritis; Vertigo
Mesh:
Year: 2021 PMID: 34191079 PMCID: PMC8782777 DOI: 10.1007/s00415-021-10667-7
Source DB: PubMed Journal: J Neurol ISSN: 0340-5354 Impact factor: 4.849
Fig. 1Technical setup for the caloric exam compared to the Video-Head Impulse test. Diagram comparing the technical setup for the caloric exam with that of the vHIT; calorics are performed in the dark on a patient in a supine position and head rest positioned at 30° from horizontal. The outer ear canal on each side is irrigated sequentially for 30 s (at 30° C cold and 44° C warm water) and the resulting eye movements recorded for a duration of 3 min using VOG goggles. The whole procedures takes up to 30 min including waiting intervals of 5 min between irrigations. The vHIT is performed in a normal lit room on a upright sitting patient. The head is moved rapidly from side to side (20 times in an impulse-like motion) and eye movements are recorded using adapted vHIT-goggles. When done correctly, the vHIT takes under 5 min
Concordance vHIT versus calorics
Stroke risk estimation for vHIT and caloric asymmetry
| Increment steps | Regression coefficient | Standard error | Wald | Odd's ratio | 95% Confidence interval | ||||
|---|---|---|---|---|---|---|---|---|---|
| Lower | Upper | ||||||||
| vHIT gain | 0.1 | 1.041 | 0.308 | 11.387 | 1 | 0.001 | 2.832 | 1.547 | 5.183 |
| Caloric asymmetry | 1% | − 0.077 | 0.024 | 10.446 | 1 | 0.001 | 0.926 | 0.883 | 0.970 |
Fig. 2ROC curves. ROC curve demonstrating a higher sensitivity and specificity for vHIT for the detection of stroke compared to calorics. Black circles indicate the optimal test discrimination cut-off for each test. The dotted line illustrates a likelihood ratio of 1 with an area under the curve (AUC) at 0.5 indicating an unhelpful test
Sensitivity and specificity for vHIT and calorics
| vHIT(Gain) | Calorics (% Asymmetry) | |||
|---|---|---|---|---|
| Test cut-off | > 0.685 | > 0.805 | < 25.3% | < 30.9% |
| AUC (95% CI) | 0.926 (0.833–0.976) | 0.863 (0.755–0.936) | ||
| Sensitivity | 91.7% | 41.7% | 58.3% | 75% |
| Specificity | 88.7% | 96.2% | 96.2% | 86.8% |
| Negative test | 48 | 58 | 56 | 49 |
| Positive test | 17 | 7 | 9 | 16 |
| True positives | 11 | 5 | 7 | 9 |
| False positives | 6 | 2 | 2 | 7 |
| True negatives | 47 | 51 | 51 | 46 |
| False negatives | 1 | 7 | 5 | 3 |
| Likelihood ratio pos. test | 8.097 | 11.042 | 15.458 | 5.679 |
| Likelihood ratio neg. test | 0.094 | 0.606 | 0.433 | 0.288 |
| Accuracy | 89.3 | 86.2 | 89.2 | 84.6 |
Pre-test and post-test probabilities of stroke using calorics or vHIT to ‘rule out’ stroke
| Post-test probability of stroke | ||
|---|---|---|
| Test | Calorics (rule out stroke) | vHIT (rule out stroke) |
| Test cut-off | 30.9% asymmetry | 0.685 gain |
| Pre-test probability of stroke (based on risk stratification rules) | Sn 75%, Sp 86.8% NLR 0.29 | Sn 91.7%, Sp 88.7% NLR 0.094 |
| 10% (low) | 3.1% | 1.0% |
| 25% (average) | 8.8% | 3.0% |
| 50% (high) | 22.4% | 8.6% |
| 75% (very high) | 46.4% | 22.0% |
Sn sensitivity, SP specificity, NLR negative likelihood ratio