| Literature DB >> 34189251 |
Yui O Nozaki1, Shinichiro Fujimoto1, Chihiro Aoshima1, Yuki Kamo1, Yuko O Kawaguchi1, Kazuhisa Takamura1, Ayako Kudo1, Daigo Takahashi1, Makoto Hiki1, Yoshiteru Kato1, Iwao Okai1, Tomotaka Dohi1, Shinya Okazaki1, Nobuo Tomizawa2, Kanako K Kumamaru2, Shigeki Aoki2, Tohru Minamino1,3.
Abstract
BACKGROUND: Computed tomography fractional flow reserve (CT-FFR), which can be acquired on-site workstation using fluid structure interaction during the multiple optimal diastolic phase, has an incremental diagnostic value over conventional coronary computed tomography angiography (CCTA). However, the appropriate location for CT-FFR measurement remains to be clarified.Entities:
Keywords: Coronary CT angiography; Fluid structure interaction; Fractional flow reserve; Lesion-specific ischemia
Year: 2021 PMID: 34189251 PMCID: PMC8215214 DOI: 10.1016/j.ijcha.2021.100815
Source DB: PubMed Journal: Int J Cardiol Heart Vasc ISSN: 2352-9067
Fig. 1Examples of a target lesion and each measurement: (a) the CT-FFR values at 1 and 2 cm distal to the end of the target lesion, and at the target vessel terminus were evaluated. (b) When the target lesion was sequential, we defined the target as shown by the black line arrow. (c) When the target lesion with ≥30% stenosis was divided into parts, we defined the target as shown the black line arrow. (d) When there was a stenosis <30% distal to a target lesion with ≥30% stenosis, we did not consider the small stenosis as a target.
Comparison between various locations of measurements for CT-FFR to identify invasive FFR ≤ 0.80 on per vessel.
| CT-FFR | Sensitivity (%) | Specificity (%) | PPV (%) | NPV (%) | Accuracy (%) |
|---|---|---|---|---|---|
| 1 cm | 70.9 | 87.2 | 76.5 | 83.7 | 81.2 |
| 2 cm | 80.0 | 79.8 | 69.8 | 87.2 | 79.9 |
| lowest | 100 | 36.2 | 47.8 | 100 | 59.7 |
PPV; positive predictive value NPV; negative predictive value CI; confidence intervals.
Fig. 2AUC for the characteristics in CT-FFR measured at each point (CT-FFR1cm, 2cm and lowest) in predicting invasive FFR ≤ 0.80 per vessel. The AUCs to predict invasive FFR ≤ 0.80 on a per-vessel basis were significantly higher at CT-FFR1cm and 2cm than CT-FFRlowest (p < 0.05 in each case.). AUC; area under the curve. Values are mean with 95% confidence intervals.