Vanja Harsaker1, Kristin Jensen2, Hilde Kjernlie Andersen2, Anne Catrine Martinsen3,4.
Abstract
BACKGROUND: The aim of this study was to quantitatively benchmark iodine imaging across specific virtual monoenergetic energy levels, iodine maps and virtual non-contrast images with different phantom sizes and iodine concentrations, using a rapid switching dual-energy CT (DECT) and a dual source DECT, in order to investigate accuracy and potential differences between the technologies.Entities:
Keywords: Contrast media; Iodine; Phantoms (imaging); Tomography (x-ray; computed)
Mesh:
Substances:
Year: 2021 PMID: 34159477 PMCID: PMC8219825 DOI: 10.1186/s41747-021-00224-2
Source DB: PubMed Journal: Eur Radiol Exp ISSN: 2509-9280
Scan parameter settings used for phantom scans for all phantom sizes and iodine concentrations
| Tube voltage (kV) | Rotation time (s) | Tube current (mA) | Pitch | Collimation (mm) | Slice width (mm) | Kernel | (DFOV) (mm) | CTDI | |
|---|---|---|---|---|---|---|---|---|---|
| General Electric Revolution | 80/140 | 0.5 | 320 | 0.516 | 40 | 2.5 plus | Standard | 400 | 15 |
| Siemens Somatom Drive | 80/Sn140 | 0.5 | 400/155 | 0.5 | 12 | 3 | I30f | 210 | 15.1 |
DFOV Display field of view, CTDI Volumetric computed tomography dose index, GE General Electric, Sn Tin filter
Fig. 1The Catphan CTP682 phantom was positioned in the isocentre of the gantry. The regions of interest were placed in the centre of the iodine solution in the phantom, here visualised as a red ring in the image
Iodine concentration measurements for all concentrations (mg/mL) and phantom sizes
| General Electric Revolution | Siemens Somatom Drive | |||||
|---|---|---|---|---|---|---|
| Solution | Small | Medium | Large | Small | Medium | Large |
| Water | -0.2 | -0.1 | -0.9 | 0.1 | 0.2 | 0.3 |
| Saline | -0.2 | -0.6 | -0.7 | 0.4 | 0.2 | 0.3 |
| Iodine 10 mg/mL | 10.3 (3.0) | 10.7 (7.0) | 10.9 (9.0) | 11.2 (12.0) | 11.1 (11.0) | 11.2 (12.0) |
| Iodine 20 mg/mL | 20.6 (3.0) | 21.0 (5.0) | 21.7 (8.5) | 21.8 (9.0) | 21.6 (8.0) | 22.2 (11.0) |
| Iodine 30 mg/mL | 30.8 (2.7) | 31.5 (5.0) | 32.4 (8.0) | 31.7 (5.7) | 31.8 (6.0) | 32.2 (7.3) |
| Iodine 50 mg/mL | 51.4 (2.8) | 52.8 (5.0) | 52.2 (4.4) | 51.0 (2.0) | 51.0 (2.0) | 52.0 (4.0) |
| Iodine 100 mg/mL | 101.8 (1.8) | 104.0 (4.0) | 103.1 (3.1) | 66.2 (-33.8) | 73.2 (-26.8) | 76.6 (-23.4) |
Percentage difference between known and measured iodine concentration in parentheses
CTDI Volumetric computed tomography dose index
HU and standard deviation (SD) in virtual non-contrast enhanced images for all concentrations and different phantom sizes
| General Electric Revolution | Siemens Somatom Drive | |||||
|---|---|---|---|---|---|---|
| Solution | Small | Medium | Large | Small | Medium | Large |
| Water | -1.5±5.0 | -2.7±13.8 | -8.4±16.0 | -3.7±3.0 | -6.4±5.6 | -4.6±8.5 |
| Saline | 7.1±5.8 | 2.4±16.2 | 1.3±16.9 | 1.9±3.1 | 2.4±7.2 | 1.5±8.0 |
| Iodine 10 mg/mL | 5.4±3.0 | -6.7±5.8 | -4.3±9.6 | -7.2±2.7 | -7.1±5.8 | -5.2±8.3 |
| Iodine 20 mg/mL | -1.3±3.1 | -24.5±15.2 | -30.3±10.4 | -2.3±4.4 | -0.8±6.1 | -3.7±9.8 |
| Iodine 30 mg/mL | -13.1±3.1 | -30.3±5.9 | -48.2±9.5 | 9.5±4.8 | 8.4±9.6 | 3.1±9.5 |
| Iodine 50 mg/mL | -26.9±4.2 | -63.5±8.8 | -75.4±11.7 | 39.7±11.6 | 35.0±12.0 | 21.5±16.6 |
| Iodine 100 mg/mL | -64.6±4.3 | -121.6±8.5 | -127.8±14.5 | 539.1±39.2 | 420.1±34.6 | 367.4±22.4 |
HU Hounsfield units
Measured HU values in monoenergetic images with General Electric Revolution CT
| Phantom size | Energy level (kV) | Water | Saline | 10 mg/mL | 20 mg/mL | 30 mg/mL | 50 mg/mL | 100 mg/mL |
|---|---|---|---|---|---|---|---|---|
| Small | 70 | -1.6 | 6.8 | 261.1 | 520.6 | 777.9 | 1292.6 | 2556.5 |
| 80 | 0.1 | 8.5 | 189.3 | 375.8 | 562.2 | 937.4 | 1850.8 | |
| 90 | 1.1 | 9.4 | 142.0 | 281.2 | 420.9 | 702.2 | 1388.6 | |
| 100 | 1.7 | 10.1 | 109.1 | 215.7 | 322.4 | 538.0 | 1065.1 | |
| 110 | 2.3 | 10.4 | 85.7 | 169.5 | 254.2 | 422.6 | 834.4 | |
| 120 | 2.7 | 10.7 | 69.5 | 137.1 | 205.4 | 341.7 | 673.9 | |
| 130 | 2.9 | 11.2 | 56.6 | 111.6 | 166.2 | 278.9 | 551.9 | |
| 140 | 3.0 | 11.3 | 49.0 | 92.2 | 138.4 | 229.2 | 457.8 | |
| Medium | 70 | -2.6 | 3.3 | 258.0 | 520.7 | 775.2 | 1300.5 | 2565.4 |
| 80 | 1.7 | 7.7 | 183.4 | 370.6 | 555.9 | 930.6 | 1843.2 | |
| 90 | 5.3 | 10.6 | 134.8 | 280.3 | 411.7 | 689.4 | 1370.0 | |
| 100 | 8.0 | 12.3 | 101.6 | 203.3 | 310.6 | 522.1 | 1038.4 | |
| 110 | 9.6 | 14.2 | 77.0 | 154.4 | 240.5 | 401.6 | 802.4 | |
| 120 | 10.9 | 16.5 | 60.7 | 132.5 | 189.5 | 316.6 | 638.3 | |
| 130 | 12.5 | 15.5 | 47.8 | 95.9 | 151.8 | 254.4 | 511.6 | |
| 140 | 13.2 | 17.4 | 37.6 | 87.7 | 121.2 | 203.4 | 413.4 | |
| Large | 70 | -7.6 | 1.8 | 266.8 | 524.4 | 785.4 | 1305.1 | 2556.4 |
| 80 | -2.1 | 7.1 | 190.4 | 375.3 | 559.9 | 932.6 | 1836.1 | |
| 90 | 2.2 | 11.2 | 141.0 | 276.3 | 410.7 | 685.0 | 1360.5 | |
| 100 | 6.6 | 12.7 | 105.2 | 206.8 | 303.3 | 513.6 | 1029.1 | |
| 110 | 8.6 | 13.4 | 80.4 | 158.1 | 233.0 | 392.8 | 794.6 | |
| 120 | 9.6 | 14.8 | 63.2 | 123.1 | 181.8 | 305.7 | 630.9 | |
| 130 | 11.0 | 15.3 | 49.5 | 96.8 | 141.7 | 242.6 | 504.0 | |
| 140 | 12.2 | 16.1 | 40.4 | 76.3 | 110.8 | 192.0 | 401.3 |
HU Hounsfield units
Measured HU values in monoenergetic images with Siemens Somatom Drive
| Phantom size | Energy level (kV) | Water | Saline | 10 mg/mL | 20 mg/mL | 30 mg/mL | 50 mg/mL | 100 mg/mL |
|---|---|---|---|---|---|---|---|---|
| Small | 70 | -0.1 | 11.4 | 286.7 | 566.3 | 835.9 | 1362.2 | 2263.3 |
| 80 | 0.6 | 10.1 | 212 | 419.9 | 621.4 | 1017.3 | 1795.6 | |
| 90 | 1.0 | 9.3 | 162.4 | 322.6 | 478.9 | 788.2 | 1483.9 | |
| 100 | 1.3 | 8.7 | 128.1 | 255.3 | 380.5 | 629.4 | 1268.4 | |
| 110 | 1.5 | 8.3 | 103.7 | 207.5 | 310.4 | 516.7 | 1115.0 | |
| 120 | 1.7 | 8.0 | 85.9 | 172.6 | 259.3 | 434.4 | 1003.0 | |
| 130 | 1.8 | 7.7 | 72.6 | 146.5 | 221.1 | 373 | 919.5 | |
| 140 | 1.9 | 7.6 | 62.5 | 126.7 | 192 | 326.2 | 856.0 | |
| Medium | 70 | -2.5 | 8.0 | 280.6 | 559.2 | 832.1 | 1357.2 | 2324.5 |
| 80 | -1.9 | 7.7 | 207.5 | 414.9 | 618.6 | 1013 | 1812.7 | |
| 90 | -1.6 | 7.6 | 158.8 | 318.7 | 476.3 | 783.5 | 1471.5 | |
| 100 | -1.3 | 7.5 | 125.0 | 252.4 | 378.0 | 624.9 | 1236.1 | |
| 110 | -1.1 | 7.4 | 101.1 | 205.0 | 308.0 | 512.3 | 1068.2 | |
| 120 | -1.0 | 7.3 | 83.6 | 170.5 | 257.0 | 430.0 | 945.8 | |
| 130 | -0.9 | 7.3 | 70.6 | 144.7 | 218.9 | 368.7 | 854.5 | |
| 140 | -0.8 | 7.3 | 60.7 | 125.1 | 189.9 | 321.9 | 785.0 | |
| Large | 70 | 0.6 | 8.8 | 287.8 | 565.2 | 841.7 | 1367.6 | 2350.1 |
| 80 | 0.9 | 8.2 | 213.4 | 418.8 | 625.0 | 1020.2 | 1823.5 | |
| 90 | 1.2 | 7.6 | 163.8 | 321.1 | 480.5 | 788.5 | 1472.5 | |
| 100 | 1.3 | 7.2 | 129.4 | 253.5 | 380.2 | 628.6 | 1230.3 | |
| 110 | 1.4 | 6.9 | 105.2 | 205.4 | 308.9 | 514.8 | 1057.8 | |
| 120 | 1.5 | 6.6 | 87.5 | 170.4 | 256.8 | 431.8 | 931.3 | |
| 130 | 1.6 | 6.5 | 74.3 | 144.3 | 218.0 | 369.8 | 836.8 | |
| 140 | 1.6 | 6.3 | 64.2 | 124.4 | 188.5 | 322.6 | 765.1 |
HU Hounsfield units
Mean difference in HU values between phantom sizes for all monoenergetic images and iodine concentrations with General Electric Revolution CT
| Phantom sizes | 10 mg/mL | 20 mg/mL | 30 mg/mL | 50 mg/mL | 100 mg/mL |
|---|---|---|---|---|---|
| M vs S | -7.68 (0.036) | -7.30 (0.294) | -11.42 (0.281) | -15.51 (0.368) | -24.54 (0.457) |
| L vs S | -3.17 (0.365) | -8.33 (0.233) | -15.13 (0.158) | -21.65 (0.213) | -33.26 (0.316) |
| L vs M | 4.52 (0.201) | -1.030 (0.880) | -3.71 (0.719) | -6.14 (0.719) | -8.72 (0.790) |
HU Hounsfield units, M vs S Medium versus small phantom size, L vs M Large versus medium phantom size, L vs S Large versus small phantom size
p-value in parentheses
Mean difference in HU values between phantom sizes for all monoenergetic images and iodine concentrations with Siemens Somatom Drive
| Phantom sizes | 10 mg/mL | 20 mg/mL | 30 mg/mL | 50 mg/mL | 100 mg/mL |
|---|---|---|---|---|---|
| M vs S | -3.25 (0.334) | -3.36 (0.607) | -2.59 (0.788) | -4.49 (0.772) | -25.80 (0.333) |
| L vs S | 1.46 (0.661) | -1.79 (0.784) | 0.01 (0.999) | -0.44 (0.997) | -29.66 (0.268) |
| L vs M | 4.71 (0.167) | 1.58 (0.809) | 2.6 (0.787) | 4.05 (0.793) | -3.86 (0.883) |
HU Hounsfield units, M vs S medium versus small phantom size, L vs M large versus medium phantom size, L vs S large versus small phantom size
p-value in parentheses
Mean difference in HU values between the vendors for all monoenergetic images and iodine concentrations
| 10 mg/mL | 20 mg/mL | 30 mg/mL | 50 mg/mL | 100 mg/mL | |
|---|---|---|---|---|---|
| Mean HU difference | 21.98 (< 0.001) | 42.69 (< 0.001) | 64.47 (< 0.001) | 98.85 (< 0.001) | -222.65–358.41 (0.032) |
HU Hounsfield units
p-value in parentheses
Fig. 2The mean HU in reconstructed monoenergetic images for each concentration of iodine. The virtual monoenergetic energy levels have the unit keV. HU Hounsfield units