Manuela Lualdi1, Adalberto Cavalleri2, Andrea Bianco3, Mara Biasin4, Claudia Cavatorta5, Mario Clerici6,7, Paola Galli3, Giovanni Pareschi3, Emanuele Pignoli5.
Abstract
BACKGROUND: UltraViolet-C (UV-C) lamps may be used to supplement current hospital cleaning and disinfection of surfaces contaminated by SARS-CoV-2. Our aim is to provide some practical indications for the correct use of UV-C lamps.Entities:
Keywords: Environmental contamination; SARS-CoV-2; UV-C disinfection; UV-C dosimeters; UV-C lamps
Mesh:
Year: 2021 PMID: 34157967 PMCID: PMC8218289 DOI: 10.1186/s12879-021-06310-5
Source DB: PubMed Journal: BMC Infect Dis ISSN: 1471-2334 Impact factor: 3.090
Technical features of the three UV-C lamps used in this study
| Model | Deluxe110 | Sterilight-S72-UV-C | AirZing™ PRO5040 |
|---|---|---|---|
| Nominal wavelength [nm] | 253.7 | 253.7 | 253.7 |
| Nominal power [W] | 2 × 55 | 2 × 36 | 1 × 36 |
| Dimension [mm] | 600x240x170 | 540x210x170 | 1363x54x78 |
| Weight [Kg] | 6.5 | 4.5 | 1.3 |
| Nominal irradiance @1 m [μW/cm2] | 600 | 220 | 140 |
| Life time [h] | 9000 | 5000 | 9000 |
| Rear reflector | Yes | Yes | No |
| Side reflector | Yesa | No | No |
aTwo side reflectors parallel to the luminous elements
Fig. 1Experimental set up for the evaluation of the impact of the choice of the lamps and their positioning. A and B show the scheme of the UV-C irradiation obtained in the cubicle for swabs with two AirZing™ PRO5040 lamps mounted on the mobile unit or on the ceiling. C and D show the scheme of the UV-C irradiation obtained in the waiting room for COVID-19 triage with two Deluxe 110 lamps mounted on the ceiling or on two opposite walls. E and F show the scheme of the UV-C irradiation of the CT bed obtained with the mobile unit equipped with AirZing™ PRO5040 lamp or with Sterilight-S72-UV-C lamp. In the simplify representation of UV-C radiation only the direct component emerging from the luminous elements was considered
Fig. 2Spectral irradiance at 1 m of the three lamps evaluated in the study. The measurements were made 2 min after switching on
Fig. 3Temporal variation of the normalized integrated irradiance of the three lamps in the first 2 min after being switching on
Fig. 4Integrated irradiance (mean and standard deviation of five measurements) of the three lamps at 1 m distance from the center of the luminous elements in three different positions: on the axis orthogonal to the luminous elements, on an axis inclined at 60° to the orthogonal axis in the plane orthogonal to the major axis of the luminous elements and on an axis inclined at 60° to the orthogonal axis in the plane containing the major axis of the luminous elements
Irradiation times calculated with and without direct irradiance measurements
| Position | Distance from the lamp | Angle | Expected irradiance from nominal data | Irradiation time corresponding to expected irradiance | Measured irradiance | Irradiation time corresponding to measured irradiance | Time required to compensate for the emission not immediately fully operational | Total irradiation time corresponding to measured irradiance | Difference between measured and expected time |
|---|---|---|---|---|---|---|---|---|---|
| [m] | [°] | [μWatt/cm2] | [s] | [μWatt/cm2] | [s] | [s] | [s] | ||
| Side edge of the bed | 0.90 | 20 | 165 | 224 | 160 | 231 | 19 | 250 | 10.4% |
| Headrest | 1.40 | 45 | 71 | 521 | 46 | 804 | 17 | 821 | 36.6% |
| Medical chair | 0.85 | 20 | 199 | 187 | 191 | 194 | 17 | 211 | 11.4% |
| Side edge of the bed | 1.30 | 30 | 83 | 446 | 74 | 500 | 18 | 518 | 13.9% |
| First chair | 1.90 | 0 | 166 | 223 | 196 | 189 | 10 | 199 | −12.1% |
| Second chair | 2.60 | 45 | 89 | 416 | 89 | 416 | 9 | 425 | 2.1% |
| First chair | 1.90 | 0 | 39 | 949 | 42 | 881 | 18 | 899 | −5.6% |
| Second chair | 2.60 | 45 | 21 | 1762 | 13 | 2846 | 17 | 2863 | 38.5% |
| First chair | 1.90 | 0 | 61 | 607 | 109 | 339 | 13 | 352 | −72.4% |
| Second chair | 2.60 | 45 | 33 | 1121 | 30 | 1233 | 13 | 1246 | 10.0% |
| First chair | 2.30 | 0 | 113 | 327 | 139 | 266 | 10 | 276 | −18.5% |
| Second chair | 4.60 | 45 | 28 | 1321 | 28 | 1321 | 9 | 1330 | 0.7% |
aThe distance is calculated from the reference position to the center of the luminous elements
bAngle between the axis orthogonal to the luminous elements and the line joining the reference position and the center of the luminous elements
cIrradiance measurements were made two minutes after switching on
dIrradiance measurements were made by switching on only one lamp at a time
Comparison between the dose calculated from direct measurements of irradiance and the dose indicated by the UVC-100 TRI photocromic dosimeters
| Position | Measured irradiance | Expected dose | UVC-100 TRI color | Is the target dose of 37 mJ/cm |
|---|---|---|---|---|
| Reference position: side edge of the bed | 160 | 37 | LO-DO | Probably yes |
| Reference position: headrest | 46 | 37 | LO-DO | Probably yes |
| Test position: lateral edge of the headrest | 44 | 35 | LO-DO | Probably yes |
| Reference position: medical chair (armrest of the chair) | 191 | 37 | LO-DO | Probably yes |
| Reference position: side edge of the bed | 74 | 37 | LO-DO | Probably yes |
| Test position: medical chair (lateral edge of the seat) | 69 | 35 | LO-DO | Probably yes |
| Reference position: first chair (armrest of the chair) | 196 | 37 | DO | Yes |
| Test position: first chair (lateral edge of the seat) | 190 | 36 | DO | Yes |
| Reference position: second chair (armrest of the chair) | 89 | 37 | LO-DO | Yes |
| Test position: second chair (lateral edge of the seat) | 84 | 35 | LO-DO | Probably yes |
| Reference position: first chair (armrest of the chair) | 42 | 37 | LO-DO | Probably yes |
| Test position: first chair (lateral edge of the seat) | 39 | 34 | LO-DO | Probably yes |
| Reference position: second chair (armrest of the chair) | 13 | 37 | LO-DO | Probably yes |
| Test position: second chair (lateral edge of the seat) | 9 | 26 | LO | Probably no |
| Reference position: first chair (armrest of the chair) | 109 | 37 | LO-DO | Probably yes |
| Test position: first chair (lateral edge of the seat) | 100 | 34 | LO-DO | Probably yes |
| Reference position: second chair (armrest of the chair) | 30 | 37 | LO-DO | Probably yes |
| Test position: second chair (lateral edge of the seat) | 23 | 28 | LO | Probably no |
| Reference position: first chair (armrest of the chair) | 139 | 37 | LO-DO | Probably yes |
| Test position: first chair (lateral edge of the seat) | 130 | 35 | LO-DO | Probably yes |
| Reference position: second chair (armrest of the chair) | 28 | 37 | LO-DO | Probably yes |
| Test position: second chair (lateral edge of the seat) | 23 | 30 | LO | Probably no |
aLO-DO intermediate between light orange and dark orange, LO very similar to light orange, DO very similar to dark orange