| Literature DB >> 34154558 |
Juan Li1, Hao Wang2, Lu Wang1, Ting Wei1, Minggang Wu1, Tingting Li1, Jifen Liao1, Bo Tan1, Man Lu3.
Abstract
BACKGROUND: The aim of this study was to investigate the concordance in lesion detection, between conventional Handhold Ultrasound (HHUS) and The Anatomical Intelligence for Breast ultrasound scan method. RESULT: The AI-breast showed the absolute agreement between the resident and an experienced breast radiologist. The ICC for the scan time, number, clockface location, distance to the nipple, largest diameter and mean diameter of the lesion obtained by a resident and an experienced breast radiologist were 0.7642, 0.7692, 0.8651, 0.8436, 0.7502, 0.8885, respectively. The ICC of the both practitioners of AI-breast were 0.7971, 0.7843, 0.9283, 0.8748, 0.7248, 0.8163, respectively. The k value of Anatomical Intelligence breast between experienced breast radiologist and resident in these image characteristics of boundary, morphology, aspect ratio, internal echo, and BI-RADS assessment were 0.7424, 0.7217, 0.6741, 0.6419, 0.6241, respectively. The k value of the two readers of AI-breast were 0.6531, 0.6762, 0.6439, 0.6137, 0.5981, respectively.Entities:
Keywords: Anatomical intelligence breast; Breast; Breast imaging reporting and data system; Ultrasound
Mesh:
Year: 2021 PMID: 34154558 PMCID: PMC8215794 DOI: 10.1186/s12880-021-00628-x
Source DB: PubMed Journal: BMC Med Imaging ISSN: 1471-2342 Impact factor: 1.930
Fig. 1Flowchart of the study
Fig. 2AI breast ultrasound scan method. A -patient maintained a supine position; B- and C- transducer marker verification on the on-screen breast graphic; D- image acquisition
Fig. 3A 45-year-old women presents with eight lesions in the left breast. The number of the breast markers in the left breast and the eight lesions of the left breast
Tabulated values for both scanning methods for the scan time, number, location, and size
| C-US | A-US | ICC (C-A) | ICC A1-2 | ||
|---|---|---|---|---|---|
| Reader 1 | Reader 2 | ||||
| Scan time | 0.7642 | 0.7971 | |||
| Nodule < 5 | 3.99 | 4.56 | 4.21 | ||
| Nodule5-10 | 6.86 | 6.25 | 6.33 | 0.7417 | 0.7996 |
| Nodule > 10 | 10.91 | 9.26 | 9.63 | 0.6664 | 0.6651 |
| Number | 73 | 73 | 73 | 0.7692 | 0.783 |
| < 5 | 22 | 19 | 20 | 0.7982 | |
| 6–10 | 36 | 33 | 34 | 0.7652 | 0.7825 |
| > 10 | 15 | 21 | 19 | 0.6913 | 0.8171 |
| Location (Clockface location; Distance to the nipple) | |||||
| Clockface location | |||||
| Distance to the nipple | |||||
| Size (Largest diameter; Mean diameter) | |||||
| Largest diameter | 6.7092 | 5.9415 | 6.2764 | 0.7502 | 0.7248 |
| 1–5 mm | 0.4241 | 0.4531 | |||
| 5–10 mm | 0.6538 | 0.5912 | |||
| > 10 mm | |||||
| Mean diameter | 5.7092 | 5.9477 | 6.2437 | ||
| 1–5 mm | 0.7241 | 0.7123 | |||
| 5–10 mm | 0.6538 | 0.6759 | |||
| > 10 mm | |||||
Data in bold are almost perfect agreement
CUS represents the conventional US scan method; AIUS represents the AI breast US scan method; C-A represents the ICC between the AI breast US scan method and AI breast US scan method; A1-2 represents the ICC between the two reader of AI breast US scan method
Tabulated values for the image characteristics for both investigated scanning methods
| No. of lesions value | k value | K value sub-feature | |||||
|---|---|---|---|---|---|---|---|
| CUS | AUS-1 | AUS-2 | C-A | A1-2 | C-A | A1-2 | |
| Boundary | 0.7424 | 0.6531 | |||||
| Morphology | 0.7217 | 0.6762 | |||||
| Aspect ratio | 0.6741 | 0.6439 | |||||
| Internal echo | 0.6419 | 0.6137 | |||||
| Cyst | 161 | 132 | 149 | ||||
| Mixed echogenic | 66 | 73 | 62 | 0.5132 | 0.4783 | ||
| Solid | 99 | 94 | 98 | 0.7425 | 0.7774 | ||
| BI-RADS assessment | 0.6241 | 0.5981 | |||||
| 2 | 171 | 141 | 148 | 0.5768 | 0.5174 | ||
| 3 | 128 | 122 | 130 | ||||
| 4A, 4B, 4C, 5 | 27 | 36 | 31 | 0.5132 | 0.4378 | ||
Data in bold are almost perfect agreement
CUS represents the conventional US scan method; AUS represents the AI breast US scan method; C-A represents the k Value between the AI breast US scan method and AI breast US scan method; A1-2 represents k Value between the two readers of AI breast US scan method