| Literature DB >> 34129644 |
Shunli Zhang1, Fei Cheng2, Hua Wang3, Jiangping Wen4, Jie Zeng5, Chuanbao Zhang5, Wensong Liu6, Ning Wang7, Tingting Jia1, Mo Wang1, Rui Zhang1, Yuhong Yue1, Jing Xu7, Zhanyong Wang7, Yilong Li6, Wenxiang Chen5, Qingtao Wang1.
Abstract
BACKGROUND: This study aimed to assess the comparability among assays using freshly frozen human sera and external quality assessment (EQA) data in China.Entities:
Year: 2021 PMID: 34129644 PMCID: PMC8205121 DOI: 10.1371/journal.pone.0253324
Source DB: PubMed Journal: PLoS One ISSN: 1932-6203 Impact factor: 3.240
Fig 3Commutability of National Center for Clinical Laboratories external quality assessment (EQA) materials among the eight assessed thyroid-stimulating hormone (TSH) instrument platforms based on CLSI approach.
Passing–Bablok regression analysis measured TSH values using NCCL EQA materials (2016~2019) after commutability-related biases were corrected.
| Platform | sample no. | Concentration, | CV, Median (%) | Regression parameters | ||
|---|---|---|---|---|---|---|
| Median (range) (μIU/mL) | Slope (95% CI) | Intercept, μIU/ml (95% CI) | r | |||
| ADVIA CentaurXP | 35 | 23.485 (0.610–55.670) | 4.16 | |||
| Immulite 2000 | 35 | 24.087 (0.650–56.519) | 6.14 | 0.996(0.960~1.024) | 0.064(0.030~0.327) | 0.997 |
| DXI800 | 35 | 23.392 (0.636–57.214) | 6.25 | 0.997(0.979~1.015) | 0.013(-0.020~0.105) | 0.999 |
| Autolumo A2000 plus | 35 | 20.693 (0.544–60.482) | 5.74 | 0.959(0.921~1.035) | 0.342(-0.041~0.784) | 0.992 |
| Maglumi2000 plus | 35 | 27.226 (0.902–65.788) | 5.71 | 1.211(1.172~1.285) | 0.164(0.063~0.324) | 0.995 |
| Cobas 601 | 35 | 29.988 (0825–60.973) | 3.75 | 1.254(1.162~1.277) | 0.216(0.163~0.682) | 0.995 |
| Architect i2000sr | 35 | 17.035 (0.518–44.247) | 4.67 | 0.735(0.707~0.757) | 0.093(0.048~0.225) | 0.997 |
| Liaison XL | 35 | 24.691 (0.736–59.929) | 5.93 | 1.034(1.022~1.047) | 0.136(0.102~0.221) | 0.999 |
a, included Siemens Advia Centaur CP/XP
b, included Siemens Immulite 2000/2000 XPi
c, included Beckman DXI 600,DXI 800
d, included AutoLumo A2000/A2000 plus
e, included Snibe Maglumi 600/800/1000/1000Plus/2000/2000plus
f, included Roche Cobas e601/e602
g, included Abbott Architect i2000SR/i2000/i1000srP
h, included DiaSorin S.p.A LIALSON/XL.
1, Median and range of TSH concentration of NCCL EQA samples were showed on eight platforms.
2, Median of coefficients of variation of TSH laboratory results for different NCCL EQA samples were used to describe imprecisions of eight platforms.
Regression parameters were estimated between ADVIA CentaurXP and other platform. Slope and intercept were used to reveal the proportional and constant deviations between platforms.
r, denote pearson correlation coefficient. CI, denote confidence interval.
Passing–Bablok regression analysis measured TSH values using commutable EQA materials (2017~2019).
| Platform | sample no. | Concentration | CV, Median (%) | Regression parameters | ||
|---|---|---|---|---|---|---|
| Median (range) (μIU/mL) | Slope (95% CI) | Intercept, μIU/ml (95% CI) | r | |||
| ADVIA CentaurXP | 25 | 2.98(1.32–16.80) | 5.44 | |||
| Immulite 2000 | 15 | 3.26(1.34–15.45) | 5.55 | 1.065(0.976~1.141) | 0.139(-0.111~0.572) | 0.997 |
| DXI800 | 25 | 3.15(1.31–17.58) | 6.33 | 1.027(1.010~1.054) | 0.003(-0.075~0.089) | 1.000 |
| Autolumo A2000 plus | 15 | 3.59(1.65–16.18) | 4.17 | 1.056(0.955~1.121) | 0.284(0.053~0.596) | 0.997 |
| Maglumi2000 plus | 20 | 3.50(1.46–15.88) | 4.35 | 1.115(1.044~1.181) | -0.149(-0.363~0.081) | 0.995 |
| Cobas 601 | 25 | 3.42(1.51–18.35) | 4.24 | 1.105(1.083~1.161) | 0.072(-0.102~0.135) | 1.000 |
| Architect i2000sr | 25 | 2.36(1.04–12.25) | 6.19 | 0.729(0.697~0.755) | 0.089(0.022~0.190) | 0.999 |
| Liaison XL | 10 | 3.19(1.53–17.01) | 2.82 | 1.111(1.083~1.151) | 0.096(0.004~0.275) | 1.000 |
a, included Siemens Advia Centaur CP/XP
b, included Siemens Siemens Immulite/One/1000/2000
c, included Beckman DXI 600,DXI 800
d, included AutoLumo A2000/A2000plus
e, included Snibe Maglumi 600/800/1000/1000Plus/2000/2000plus
f, includedRoche Cobas e601/e602
g, included Abbott Architect i2000SR/i2000/i1000sr
h, included DiaSorin S.p.A LIALSON
1, Median and range of TSH concentration of commutable EQA samples were showed on eight platforms.
2, Median of coefficients of variation of TSH laboratory results for different commutable EQA samples were used to describe imprecisions of eight platforms.
Regression parameters were estimated between ADVIA CentaurXP and other platform. Slope and intercept were used to reveal the proportional and constant deviations between platforms.
r, denote pearson correlation coefficient. CI, denote confidence interval.
Passing–Bablok regression analysis measured TSH values for patient samples after systematic biases were corrected.
| Platform | sample no. | Concentration, | CV, Median(%) | Regression parameters | r | |
|---|---|---|---|---|---|---|
| Median(range) (μIU/mL) | Slope (95% CI) | Intercept, μIU/ml (95% CI) | ||||
| ADVIA CentaurXP | 29 | 4.707(0.076–82.625) | 1.82 | |||
| Immulite 2000 | 28 | 4.421(0.091–53.395) | 3.07 | 1.033(0.956~1.107) | 0.135(0.052~0.367) | 0.996 |
| DXI800 | 29 | 4.762(0.116–78.601) | 2.28 | 1.024(0.986~1.089) | 0.093(-0.005~0.226) | 0.998 |
| Autolumo A2000 plus | 29 | 5.438(0.071–93.576) | 1.36 | 1.049(1.026~1.110) | 0.337(0.145~0.528) | 0.990 |
| Maglumi2000plus | 29 | 5.197(0.087–80.081) | 3.82 | 1.172(1.063~1.242) | 0.093(-0.001~0.295) | 0.992 |
| Cobas 601 | 29 | 6.193(0.100–92.081) | 4.09 | 1.299(1.222~1.371) | 0.147(0.043~0.372) | 0.995 |
| Architect i2000sr | 29 | 4.202(0.081–54.140) | 1.57 | 0.801(0.727~0.861) | 0.077(0.021~0.258) | 0.990 |
| Liaison XL | 29 | 4.811(0.097–88.712) | 1.17 | 1.069(1.015~1.087) | 0.093(0.022~0.166) | 0.999 |
1, Median and range of TSH concentration of patient samples were showed on eight platforms.
2, Median of coefficients of variation of TSH triplicate measurements for different patient samples were used to describe imprecisions of eight platforms.
Regression parameters were estimated between ADVIA CentaurXP and other platform. Slope and intercept were used to reveal the proportional and constant deviations between platforms.
r, denote pearson correlation coefficient. CI, denote confidence interval.