Analytical Bias Exceeding Desirable Quality Goal in 4 out of 5 Common Immunoassays: Results of a Native Single Serum Sample External Quality Assessment Program for Cobalamin, Folate, Ferritin, Thyroid-Stimulating Hormone, and Free T4 Analyses.

Gunn B B Kristensen1, Pål Rustad2, Jens P Berg3, Kristin M Aakre4.   

Abstract

BACKGROUND: We undertook this study to evaluate method differences for 5 components analyzed by immunoassays, to explore whether the use of method-dependent reference intervals may compensate for method differences, and to investigate commutability of external quality assessment (EQA) materials.
METHODS: Twenty fresh native single serum samples, a fresh native serum pool, Nordic Federation of Clinical Chemistry Reference Serum X (serum X) (serum pool), and 2 EQA materials were sent to 38 laboratories for measurement of cobalamin, folate, ferritin, free T4, and thyroid-stimulating hormone (TSH) by 5 different measurement procedures [Roche Cobas (n = 15), Roche Modular (n = 4), Abbott Architect (n = 8), Beckman Coulter Unicel (n = 2), and Siemens ADVIA Centaur (n = 9)]. The target value for each component was calculated based on the mean of method means or measured by a reference measurement procedure (free T4). Quality specifications were based on biological variation. Local reference intervals were reported from all laboratories.
RESULTS: Method differences that exceeded acceptable bias were found for all components except folate. Free T4 differences from the uncommonly used reference measurement procedure were large. Reference intervals differed between measurement procedures but also within 1 measurement procedure. The serum X material was commutable for all components and measurement procedures, whereas the EQA materials were noncommutable in 13 of 50 occasions (5 components, 5 methods, 2 EQA materials).
CONCLUSIONS: The bias between the measurement procedures was unacceptably large in 4/5 tested components. Traceability to reference materials as claimed by the manufacturers did not lead to acceptable harmonization. Adjustment of reference intervals in accordance with method differences and use of commutable EQA samples are not implemented commonly.
© 2016 American Association for Clinical Chemistry.

Entities:  

Mesh:

Substances:

Year:  2016        PMID: 27384539     DOI: 10.1373/clinchem.2016.258962

Source DB:  PubMed          Journal:  Clin Chem        ISSN: 0009-9147            Impact factor:   8.327


  8 in total

Review 1.  Lot-to-Lot Variation.

Authors:  Simon Thompson; Douglas Chesher
Journal:  Clin Biochem Rev       Date:  2018-05

2.  How to Really Understand and Improve the System of Internal Quality Control and External Quality Assessment in the Accreditation Process of the Medical Laboratory?

Authors:  Milan Skitek; Flávia Martinello; Aleš Jerin
Journal:  EJIFCC       Date:  2022-04-11

Review 3.  Interpretation of EQA results and EQA-based trouble shooting.

Authors:  Gunn Berit Berge Kristensen; Piet Meijer
Journal:  Biochem Med (Zagreb)       Date:  2017-02-15       Impact factor: 2.313

4.  Comparability of thyroid-stimulating hormone immunoassays using fresh frozen human sera and external quality assessment data.

Authors:  Shunli Zhang; Fei Cheng; Hua Wang; Jiangping Wen; Jie Zeng; Chuanbao Zhang; Wensong Liu; Ning Wang; Tingting Jia; Mo Wang; Rui Zhang; Yuhong Yue; Jing Xu; Zhanyong Wang; Yilong Li; Wenxiang Chen; Qingtao Wang
Journal:  PLoS One       Date:  2021-06-15       Impact factor: 3.240

Review 5.  Performance and comparability of laboratory methods for measuring ferritin concentrations in human serum or plasma: A systematic review and meta-analysis.

Authors:  Maria N Garcia-Casal; Juan P Peña-Rosas; Eloisa Urrechaga; Jesus F Escanero; Junsheng Huo; Ricardo X Martinez; Lucero Lopez-Perez
Journal:  PLoS One       Date:  2018-05-03       Impact factor: 3.240

6.  Management of Amiodarone-Induced Thyrotoxicosis at a Cardiac Transplantation Centre.

Authors:  Michelle Isaacs; Monique Costin; Ron Bova; Helen L Barrett; Drew Heffernan; Katherine Samaras; Jerry R Greenfield
Journal:  Front Endocrinol (Lausanne)       Date:  2018-08-22       Impact factor: 5.555

7.  Prevalence of low folate deficiency after wheat flour supplementation - should we still measure serum folate?

Authors:  Alejandro Godoy; Aldo Hugo Tabares
Journal:  Hematol Transfus Cell Ther       Date:  2018-06-13

8.  Comparability of four clinical laboratory measurement methods for GGT and commutability of candidate reference materials.

Authors:  Rui Zhang; Qingtao Wang
Journal:  J Clin Lab Anal       Date:  2020-09-11       Impact factor: 3.124

  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.