Literature DB >> 34124283

Development of a fluoroscopically guided robotic assistant for instrument placement in pelvic trauma surgery.

Rohan C Vijayan1, Runze Han1, Pengwei Wu1, Niral M Sheth1, Michael D Ketcha1, Prasad Vagdargi2, Sebastian Vogt3, Gerhard Kleinszig3, Greg M Osgood4, Jeffrey H Siewerdsen1,2, Ali Uneri1.   

Abstract

Purpose: A method for fluoroscopic guidance of a robotic assistant is presented for instrument placement in pelvic trauma surgery. The solution uses fluoroscopic images acquired in standard clinical workflow and helps avoid repeat fluoroscopy commonly performed during implant guidance. Approach: Images acquired from a mobile C-arm are used to perform 3D-2D registration of both the patient (via patient CT) and the robot (via CAD model of a surgical instrument attached to its end effector, e.g; a drill guide), guiding the robot to target trajectories defined in the patient CT. The proposed approach avoids C-arm gantry motion, instead manipulating the robot to acquire disparate views of the instrument. Phantom and cadaver studies were performed to determine operating parameters and assess the accuracy of the proposed approach in aligning a standard drill guide instrument.
Results: The proposed approach achieved average drill guide tip placement accuracy of 1.57 ± 0.47    mm and angular alignment of 0.35 ± 0.32    deg in phantom studies. The errors remained within 2 mm and 1 deg in cadaver experiments, comparable to the margins of errors provided by surgical trackers (but operating without the need for external tracking). Conclusions: By operating at a fixed fluoroscopic perspective and eliminating the need for encoded C-arm gantry movement, the proposed approach simplifies and expedites the registration of image-guided robotic assistants and can be used with simple, non-calibrated, non-encoded, and non-isocentric C-arm systems to accurately guide a robotic device in a manner that is compatible with the surgical workflow.
© 2021 Society of Photo-Optical Instrumentation Engineers (SPIE).

Entities:  

Keywords:  3D–2D image registration; image-guided surgery; orthopaedic surgery; pelvic trauma; surgical robotics

Year:  2021        PMID: 34124283      PMCID: PMC8189698          DOI: 10.1117/1.JMI.8.3.035001

Source DB:  PubMed          Journal:  J Med Imaging (Bellingham)        ISSN: 2329-4302


  34 in total

Review 1.  Completely derandomized self-adaptation in evolution strategies.

Authors:  N Hansen; A Ostermeier
Journal:  Evol Comput       Date:  2001       Impact factor: 3.277

2.  Robotic systems in orthopaedic surgery.

Authors:  J E Lang; S Mannava; A J Floyd; M S Goddard; B P Smith; A Mofidi; T M Seyler; R H Jinnah
Journal:  J Bone Joint Surg Br       Date:  2011-10

3.  Anatomy-based registration of CT-scan and intraoperative X-ray images for guiding a surgical robot.

Authors:  A Guéziec; P Kazanzides; B Williamson; R H Taylor
Journal:  IEEE Trans Med Imaging       Date:  1998-10       Impact factor: 10.048

4.  Robotic drill guide positioning using known-component 3D-2D image registration.

Authors:  Thomas Yi; Vignesh Ramchandran; Jeffrey H Siewerdsen; Ali Uneri
Journal:  J Med Imaging (Bellingham)       Date:  2018-02-06

5.  Robotic assisted laparoscopic radical prostatectomy versus retropubic radical prostatectomy: a prospective assessment of postoperative pain.

Authors:  Todd M Webster; S Duke Herrell; Sam S Chang; Michael S Cookson; Roxelyn G Baumgartner; Laura W Anderson; Joseph A Smith
Journal:  J Urol       Date:  2005-09       Impact factor: 7.450

6.  Self-calibration of cone-beam CT geometry using 3D-2D image registration.

Authors:  S Ouadah; J W Stayman; G J Gang; T Ehtiati; J H Siewerdsen
Journal:  Phys Med Biol       Date:  2016-03-10       Impact factor: 3.609

7.  First spine surgery utilizing real-time image-guided robotic assistance.

Authors:  A Karim Ahmed; Corinna C Zygourakis; Samuel Kalb; Alex M Zhu; Camilo A Molina; Bowen Jiang; Ari M Blitz; Ali Bydon; Neil R Crawford; Nicholas Theodore
Journal:  Comput Assist Surg (Abingdon)       Date:  2019-03-01       Impact factor: 1.787

8.  Pedicle screw accuracy assessment in ExcelsiusGPS® robotic spine surgery: evaluation of deviation from pre-planned trajectory.

Authors:  Bowen Jiang; A Karim Ahmed; Corinna C Zygourakis; Samuel Kalb; Alex M Zhu; Jakub Godzik; Camilo A Molina; Ari M Blitz; Ali Bydon; Neil Crawford; Nicholas Theodore
Journal:  Chin Neurosurg J       Date:  2018-09-03

9.  2D-fluoroscopic navigated percutaneous screw fixation of pelvic ring injuries--a case series.

Authors:  Florian Gras; Ivan Marintschev; Arne Wilharm; Kajetan Klos; Thomas Mückley; Gunther O Hofmann
Journal:  BMC Musculoskelet Disord       Date:  2010-07-07       Impact factor: 2.362

10.  Minimally invasive surgery under fluoro-navigation for anterior pelvic ring fractures.

Authors:  Kai-Hua Zhou; Cong-Feng Luo; Nong Chen; Cheng-Fang Hu; Fu-Gen Pan
Journal:  Indian J Orthop       Date:  2016 May-Jun       Impact factor: 1.251

View more
  1 in total

1.  Preclinical evaluation of a prototype freehand drill video guidance system for orthopedic surgery.

Authors:  Niral Sheth; Prasad Vagdargi; Alejandro Sisniega; Ali Uneri; Gregory Osgood; Jeffrey H Siewerdsen
Journal:  J Med Imaging (Bellingham)       Date:  2022-08-26
  1 in total