Dynamic analysis of flutter in disk type mechanical heart valve prostheses.

A A Prabhu1, N H Hwang.   

Abstract

Parametric study of the low frequency oscillations occasionally observed in certain types of disc type prosthetic heart valves (PHV) are carried out using a finite element technique. The analysis is performed to determine the frequencies of the dynamic fluttering with the help of the 'ANSYS' computer program. The results show that the frequencies of the dynamic fluttering for both the circular occluders and the semi-circular occluders are at least two orders of magnitude higher than that observed in vivo. It is thus concluded that the clinically observed leaflet oscillations should not be a dynamic flutter phenomenon. Rather, the vortex shedding has been assumed to be the cause of these oscillations.

Mesh:

Year:  1988        PMID: 3410861     DOI: 10.1016/0021-9290(88)90221-7

Source DB:  PubMed          Journal:  J Biomech        ISSN: 0021-9290            Impact factor:   2.712