Christian Gerges1, Helmut Neumann5, Sauid Ishaq3, Visvakanth Sivanathan4, Peter R Galle4, Horst Neuhaus1, Helmut Neumann5. 1. Department of Gastroenterology, Evangelisches Krankenhaus Düsseldorf, Düsseldorf, Germany. 2. GastroZentrum Lippe, Germany. 3. Department of Gastroenterology, Dudley Group NHS, Foundation Trust and Birmingham City University, Birmingham, UK. 4. Department of Interdisciplinary Endoscopy, I. Medizinische Klinik und Poliklinik, University Hospital, Mainz, Germany. 5. Interdisciplinary Endoscopy, First Medical Department, University Medical Center Mainz, Mainz, Germany.
Abstract
BACKGROUND: Although colonoscopy is the gold standard for colorectal cancer screening, colonic looping may make complete colonoscopy challenging. Commonly available stiffening device colonoscopy has been described as helpful but not effective enough to prevent looping. In this context the effect on cecal intubation time and rate was described differently in various studies and in some studies had no impact on cecal intubation time at all. The aim of this study was to evaluate whether a novel colonoscope with gradual stiffness (Fujifilm EC760R-V/I- flexibility adjuster, Tokyo, Japan) using four significantly different grades of stiffness can be an alternative to established devices in terms of loop prevention, cecal intubation rate and time, adverse events, and patient/examiner satisfaction. METHODS: Consecutive patients without previous colorectal surgery were analyzed retrospectively. Colonoscopy was performed with the new colonoscope and performance characteristics, including time to cecum, withdrawal time, total examination time, and patient and endoscopist satisfaction were recorded. RESULTS: Among 180 consecutive procedures, 98.3% of examinations were complete to the cecum. The endoscopic flexibility adjuster was used in 150 of 180 cases (83.3%). Overall, the device was scored by the examiner as helpful to prevent looping in 146 of the 150 cases (97.7%). Mean cecal intubation time was 6.5 min, with 35% of examination performed in under 5 min with a mean withdrawal time of 7 min. Mean total examination time was 18 min. Patient satisfaction was rated as high in all examinations performed. CONCLUSION: The new flexibility adjuster colonoscope was shown to be helpful in loop prevention, allowed for fast and successful cecal intubation, and led to a high rate of patients satisfaction.
BACKGROUND: Although colonoscopy is the gold standard for colorectal cancer screening, colonic looping may make complete colonoscopy challenging. Commonly available stiffening device colonoscopy has been described as helpful but not effective enough to prevent looping. In this context the effect on cecal intubation time and rate was described differently in various studies and in some studies had no impact on cecal intubation time at all. The aim of this study was to evaluate whether a novel colonoscope with gradual stiffness (Fujifilm EC760R-V/I- flexibility adjuster, Tokyo, Japan) using four significantly different grades of stiffness can be an alternative to established devices in terms of loop prevention, cecal intubation rate and time, adverse events, and patient/examiner satisfaction. METHODS: Consecutive patients without previous colorectal surgery were analyzed retrospectively. Colonoscopy was performed with the new colonoscope and performance characteristics, including time to cecum, withdrawal time, total examination time, and patient and endoscopist satisfaction were recorded. RESULTS: Among 180 consecutive procedures, 98.3% of examinations were complete to the cecum. The endoscopic flexibility adjuster was used in 150 of 180 cases (83.3%). Overall, the device was scored by the examiner as helpful to prevent looping in 146 of the 150 cases (97.7%). Mean cecal intubation time was 6.5 min, with 35% of examination performed in under 5 min with a mean withdrawal time of 7 min. Mean total examination time was 18 min. Patient satisfaction was rated as high in all examinations performed. CONCLUSION: The new flexibility adjuster colonoscope was shown to be helpful in loop prevention, allowed for fast and successful cecal intubation, and led to a high rate of patients satisfaction.
One of the most important quality criteria for colonoscopy is a cecal intubation rate
greater than 90%.[1-6] However, achieving complete
colonoscopy can be challenging due to loops preventing advancement of the endoscope
to the cecum. In order to overcome loop formation, several maneuvers and principles
have been established, including the push and retract-technique, minimal use of
CO2/air-insufflation, or external abdominal pressure. In selected
cases, a more flexible and smaller diameter pediatric colonoscope can be used
successfully to facilitate the passage of a tortuous and/or narrowed sigmoid
colon.[7,8]
However, the advantage of the flexible and thinner endoscope can become a drawback
once the sigmoid colon is passed, resulting in more frequent looping, making a cecal
intubation even more difficult. In this context, variable stiffness colonoscopes
allow for passage of the rectosigmoid colon in a flexible mode followed by
stiffening of the device to prevent looping, thus potentially accelerating the
endoscopic procedure. The idea of a variable stiffening has been evaluated in the
past, mainly using a variable stiffening device with three grades of
stiffness.[9-12] Even though the stiffening
device has been judged as “helpful” by examiners, and despite being ranked highest
of the three available settings, it was, however, not stiff enough to prevent
looping.[9-12] In this context, the range of
the three settings was identified as not wide enough to prevent looping and may also
be the reason why contrary effects on cecal intubation time and rate have been
described in various studies, with some studies illustrating no impact on cecal
intubation time at all.[9,12,13] A novel colonoscope with gradual stiffness has been developed
by Fujifilm (EC760R-V/I Tokyo, Japan) using four significantly different grades of
stiffness (Figure 1).
Gradual stiffness means that the stiffness is changing over the length of the scope,
being softest at the distal end and hardest at the proximal end.
It remains unclear if this new colonoscope can be an alternative to the
established devices in terms of loop prevention, cecal intubation rate and time,
adverse events, and patient and examiners satisfaction.
Figure 1.
Flexibility adjuster.
Flexibility adjuster.The aim of this study was to evaluate the above mentioned performance characteristics
of a novel colonoscope offering a new type of gradual stiffness (flexibility
adjuster).
Materials and methods
A retrospective observational study of prospectively collected data was performed.
The retrospective design was chosen in order to collect statistically valid data as
a basis for a prospective randomized trial. The study population included patients
aged >18 years with an indication for screening or surveillance colonoscopy and
without previous colorectal surgery. All patient data since 2016 were consecutively
included in the study, if they were eligible, based on inclusion and exclusion
criteria. Written informed consent was obtained from all patients. Patients were
unaware whether the flexibility adjuster was applied for their procedure.
High-definition video endoscopes with a fully integrated stiffing device allowing
for manual adjustment of the stiffing in four different grades were used
(EC760R-V/I, Fujifilm, Tokyo, Japan) (Figure 2). The gradual stiffness insertion
section is designed to optimize force and torque transmission during colonoscope
insertion, and thereby enhance insertion capability without the need for an external
imaging device. The colonoscopes have a multi LED-Technology with a 170° field view,
a 12-mm diameter and a 3.8-mm working channel. The stiffness can be adjusted by
turning a dedicated wheel at the shaft of the colonoscope (Figure 1). All colonoscopies were performed
under conscious sedation.
Figure 2.
Fully integrated stiffing device allowing for manual adjustment.
Fully integrated stiffing device allowing for manual adjustment.
Outcome measures
The following outcome measures were assessed: (1) successful cecum intubation;
(2) time to intubate cecum; (3) withdrawal time; (4) endoscopist use and
satisfaction of the stiffening device; (5) patients satisfaction; (6) adverse
events.
Procedure
The colonoscopies were performed by an expert endoscopist with more than 10 years
experience in screening and interventional colonoscopy to avoid bias. The
insertion protocol was based on previous described standards in evaluating a
variable stiffness colonoscope.
The sigmoid colon was passed using the most flexible mode of the
colonoscope. Advancement was then continued in the flexible mode until any
looping started, at which point the colonoscope was straightened and immediately
placed in the maximal stiffness mode, followed by additional attempts at
reinsertion. Successful cecal intubation was documented by the endoscopist for
each examination in the report. The time to intubate the cecum was measured by
the endoscopy technician using a stopwatch. The stopwatch was started when the
first endoscopic view of rectal mucosa was attained and stopped when the scope
fully entered the base of the caecum. The stopwatch was not stopped while the
patient was being rotated, while abdominal pressure was initiated, or while the
flexibility adjuster device was being activated or deactivated. As instrument
advancement progressed, insertion was at times facilitated by changing back to
the flexible mode. The endoscopist scored the value of the stiffening device as
helpful or not helpful, depending on the extent to which it appeared to counter
the tendency to form loops or whether they used the stiffening option at all.
This was documented in a questionnaire at the end of each procedure.
Patients satisfaction was assessed before discharge and documented in
simple questionnaire ranging from not satisfied to highly satisfied.
Results
A total of 180 patients were included; 81 of the patients were male and 99 female.
The mean age of patients was 59.9 years.Colonoscopy was completed to the cecum in 177 of 180 cases (98.3%). Of the three
incomplete colonoscopies, one case was due to incomplete bowel preparation with
hard, lumen filling stool. In one case, a malignant stricture in the hepatic flexure
could not be passed, and in one case, in a patient with a history of hysterectomy
and stenosing diverticulosis, the endoscope could not be passed past over the
sigmoid colon. Mean cecal intubation time was 6.5 min (range 2–15 min) and mean
withdrawal time was 7 min (range 6–25 min). In 63 (35%) of the procedures intubation
time was under 5 min. The mean total examination time was 18 min (range 8–32 min)
and patient satisfaction was rated high in all examinations (100%). The endoscopic
flexibility adjuster was used in 150 of 180 cases (83.3%). Overall, the device was
scored by the examiner as helpful to prevent looping in 146 of the 150 cases
(97.7%). In 4 of the 11 first cases, the device was scored as not helpful. After the
11 first cases the device was always scored as helpful. No adverse events occurred
(0%) (Table 1).
Table 1.
Results obtained with Fujifilm EC760R-V/I colonoscope.
Total procedures (N)
180
Completed procedures (N)
177
Mean cecal intubation time (min)
6.5 (2–15)
Mean withdrawal time (min)
7 (6–25)
Mean total exam time (min)
18 (8–32)
Patients satisfactory
Highly satisfied
100%
Examiner score
Helpful
97.7%
Not helpful
2.3%
Adverse events
0%
Results obtained with Fujifilm EC760R-V/I colonoscope.
Discussion
The cecal intubation rate in this study evaluating a Fuji colonoscope with a novel
stiffening device was 98.3% and therefore much higher than the generally recommended
threshold of 90% and at the upper end of previously described results.[2,4,5] Previous studies evaluating
stiffening devices, mainly in Olympus endoscopes, showed a cecal intubation rate of
87–99.2% and examiners judged these variable stiffness colonoscopes frequently as
useful.[9-11] Time to reach
the cecum in this study was below 5 min in about one-third of procedures, which is
an additional quality indicator for this device as there have been differing
previous reports on whether a stiffening device reduces the time to cecum.[9,12] However, this result might be
affected by the experience of the endoscopist. Short cecal insertion time below
6 min has been associated with increased rates of detection of small adenomas, and a
prolonged cecal insertion time leads to a reduced adenoma detection rate.[15,16] The
stiffening was used in 83.3% of procedures and the endoscopist described the
stiffening option as “helpful” in almost 98% of the cases. While previous studies
described the range of the stiffness as not wide enough to have an impact on
preventing looping and cecal intubation time, this four-grade gradual stiffening
technique seems to assist loop prevention and a faster cecal intubation time.
However, all studies mentioned, including ours, were performed by experienced
examiners, which might reduce the generalizability of the results; further
evaluation in a randomized controlled trial is required. The results obtained here
indicate that this novel stiffening device is at least equally effective as the
former evaluated devices.The fact that the endoscopist did not score the device as “helpful” only within the
first couple of procedures suggests a distinct learning curve. It remains unclear if
unexperienced endoscopists who usually have more difficulties in passing the sigmoid
would derive greater benefit from using this device.Patient comfort is a key parameter in screening and surveillance colonoscopy as it
helps increase screening rates and prevent recurrence of adenomas. It is unclear if
a stiffening device can help increase patient comfort. In this study, patients
consistently described a high satisfaction rate after the procedure, which is very
likely the result of reduced looping during colonoscopy. However, the stiffening
device could not be used in all cases as, in some cases, the endoscopist felt the
tension on the colon would become too strong or simply because it was not needed.
There are natural concerns regarding whether a stiffening device might increase the
perforation rate. This large cohort, as well as previous studies with other devices,
observed no adverse events as the endoscopist feels how much tension is applied
during pushing.[9,10,12] The study had a few limitations. The lack of a control group
and the retrospective design may have introduced a bias. Strengths of this study
were a standardized evaluation of the endoscopist and patients, number of cases, and
patients being blinded as to whether the stiffening option was used or not.In conclusion, the new flexibility adjuster colonoscope allowed for fast and
successful cecal intubation within a short period of time, was frequently judged as
helpful by the examiner in loop prevention, and led to a high rate of patient
satisfaction. These findings still have to be reproduced in a randomized, controlled
trial and future studies should now also focus on the effect of the new colonoscope
in unsedated patients and for endoscopy trainees.
Authors: U Denzer; U Beilenhoff; A Eickhoff; S Faiss; P Hüttl; S In der Smitten; R Jakobs; C Jenssen; M Keuchel; F Langer; M M Lerch; P Lynen Jansen; A May; R Menningen; G Moog; T Rösch; U Rosien; T Vowinkel; T Wehrmann; U Weickert Journal: Z Gastroenterol Date: 2015-12-14 Impact factor: 2.000
Authors: B Rembacken; C Hassan; J F Riemann; A Chilton; M Rutter; J-M Dumonceau; M Omar; T Ponchon Journal: Endoscopy Date: 2012-09-17 Impact factor: 10.093
Authors: Michal F Kaminski; Jaroslaw Regula; Ewa Kraszewska; Marcin Polkowski; Urszula Wojciechowska; Joanna Didkowska; Maria Zwierko; Maciej Rupinski; Marek P Nowacki; Eugeniusz Butruk Journal: N Engl J Med Date: 2010-05-13 Impact factor: 91.245
Authors: Syed G Shah; Jim C Brooker; Christopher B Williams; Catherine Thapar; Noriko Suzuki; Brian P Saunders Journal: Gastrointest Endosc Date: 2002-08 Impact factor: 9.427
Authors: Douglas A Corley; Christopher D Jensen; Amy R Marks; Wei K Zhao; Jeffrey K Lee; Chyke A Doubeni; Ann G Zauber; Jolanda de Boer; Bruce H Fireman; Joanne E Schottinger; Virginia P Quinn; Nirupa R Ghai; Theodore R Levin; Charles P Quesenberry Journal: N Engl J Med Date: 2014-04-03 Impact factor: 91.245
Authors: Colin J Rees; Siwan Thomas Gibson; Matt D Rutter; Phil Baragwanath; Rupert Pullan; Mark Feeney; Neil Haslam Journal: Gut Date: 2016-08-16 Impact factor: 23.059