Age-related changes in diffuse optical tomography sensitivity profiles in infancy.

Xiaoxue Fu1, John E Richards1.   

Abstract

Diffuse optical tomography uses near-infrared light spectroscopy to measure changes in cerebral hemoglobin concentration. Anatomical interpretations of the location that generates the hemodynamic signal requires accurate descriptions of diffuse optical tomography sensitivity to the underlying cortical structures. Such information is limited for pediatric populations because they undergo rapid head and brain development. The present study used photon propagation simulation methods to examine diffuse optical tomography sensitivity profiles in realistic head models among infants ranging from 2 weeks to 24 months with narrow age bins, children (4 and 12 years) and adults (20 to 24 years). The sensitivity profiles changed systematically with the source-detector separation distance. The peak of the sensitivity function in the head was largest at the smallest separation distance and decreased as separation distance increased. The fluence value dissipated more quickly with sampling depth at the shorter source-detector separations than the longer separation distances. There were age-related differences in the shape and variance of sensitivity profiles across a wide range of source-detector separation distances. Our findings have important implications in the design of sensor placement and diffuse optical tomography image reconstruction in (functional) near-infrared light spectroscopy research. Age-appropriate realistic head models should be used to provide anatomical guidance for standalone near-infrared light spectroscopy data in infants.

Entities:  

Year:  2021        PMID: 34101747      PMCID: PMC8186805          DOI: 10.1371/journal.pone.0252036

Source DB:  PubMed          Journal:  PLoS One        ISSN: 1932-6203            Impact factor:   3.240


Introduction

Diffuse optical tomography (DOT) uses near-infrared light spectroscopy (NIRS) to measure changes in cerebral hemoglobin concentration [1]. Multi-channel NIRS instruments measures hemoglobin changes in the scalp channel space. DOT instruments use overlapping and channels with multiple separation distances to enhance spatial resolution [2]. However, both types of the scalp-based measurement do not directly provide anatomical information about the brain regions of the hemodynamic signal. Localizing the channel-wise signals to specific brain regions requires comprehensive understanding of the DOT sensitivity profile. It is difficult to experimentally measure DOT sensitivity inside human head [3]. Instead, Monte Carlo simulations are used to simulate the propagation of photons through head and brain tissues [4]. The forward model can then guide DOT image reconstruction to recover the brain locations of hemoglobin concentration changes [5, 6]. DOT sensitivity profiles have been studied in adults. However, there are considerable brain structural changes during infancy through childhood and adulthood [7, 8]. Accurate forward models depend on the use of realistic head models that account for age-related changes in the head [9, 10]. The present study examined age-related changes in DOT sensitivity profiles. We used age-appropriate, realistic head models with extensive coverage of the infancy period (2 weeks through 2 years), two child ages (4, 12 years), and young adults (20 to 24 years).

The use of Monte Carlo simulations to estimate DOT sensitivity profile

A goal of DOT is to model how hemoglobin changes at the channel space (what is measured) correspond to changes in optical properties within the brain (to be determined). This requires modeling the DOT sensitivity. One approach to solve the forward model is using finite element method (FEM) quantitative algorithms to describe photon propagations through heterogeneous tissue structures such as the human head (NIRFAST; [11]; TOAST++; [12]). Alternatively, Monte Carlo simulations provide numerical solutions to the radiative transport equation that models the photon migration through the head tissues based on fewer assumptions [1, 13]. The Monte Carlo method is computationally demanding but provides more accurate solutions for the forward model [14]. Details of the input parameters and procedures for Monte Carlo simulations are described in [4]. Briefly, the simulations require 1) segmented MRI(s) that define tissue types; 2) optical properties of the tissue types are assigned to each voxel; 3) an injection point and the direction of the photon are specified; 4) an algorithm is applied to describe the path of the photon through the tissues. The steps 3 and 4 are repeated to compute the accumulative photon fluence and the total path length traveled through the head tissues [3]. A computer program is used that takes the input information and simulates the path of injected photons through the head media (e.g., tMCimg; [4]; MCX; [15]; MMC; [16]; MCML; [13, 17]). DOT sensitivity can be quantified from Monte Carlo simulation outputs. The outputs are fluence distributions of photons travel across the head. The Monte Carlo simulations commonly model fluence distributions for the continuous-wave systems as they are more widely implemented in commercial NIRS instruments than the time-domain and frequency-domain systems [1, 14]. The continuous-wave systems measure fluence at the detector location but do not separately quantify the effects of light absorption and scattering. The DOT sensitivity is conceptualized as the extent to which the detected signal is influenced by changes in optical properties in the given region of tissue [18]. One representation of DOT sensitivity is the spatial sensitivity profile. The profile depicts the fluence distribution of detected photons [4]. It can be quantified by the probability of a photon travelling from a source-optode location to a voxel inside the head (“2-point Green’s function”; [3]). A second approach to quantify DOT sensitivity is to use the photon measurement density function (PMDF). The PMDF quantifies optical sensitivity in NIRS source-detector recording by calculating the product of the fluence distribution at the source location and the detector location (PMDF; [9, 19]; or “3-point Green’s function; [3]). We will call this the “Source-Detector Channel DOT” (S-D Channel DOT). S-D Channel DOT represents the probability that a photon injected from the source and measured at the detector location has propagated through a given tissue. DOT sensitivity is quantified as each location at which a photon traverses through a given tissue region. “Sampling depth” represents the depth of the photon measured in the head medium.

DOT sensitivity profile by source-detector separation distances

The DOT sensitivity changes as a function of source-detector separation distances [3, 18, 20, 21]. As the source-detector separation distance increases, the fluence distribution covers wider tissue regions, extends deeper into the brain, and decreases in DOT sensitivity. For example, it has been found that for 20 mm source-detector separation distance the fluence distribution was largely confined to shallow layers of the head model and about 6% of the total sensitivity was attributed to the brain (gray and white matters) [3, 20]. Sensitivity to the brain increased linearly from 20 mm up to 45 mm separation distance [3, 18, 20, 21]. The increase in sensitivity to the brain was less steep at larger separation distances [3, 20]. There was a trade-off between penetration depth and signal intensity. Signal intensity decreased monotonically and exponentially with increased depth at all separation distances [3]. Wang et al. [21] showed that the signal strength was reduced as source-detector separation distances increased. Hence, the optimal separation distances for measuring brain activity need to be determined based on the consideration of DOT sensitivity, the desired spatial resolution of the measurement regions, and the achievable signal strength for the measurement device. It is important to use age-specific realistic head models for estimating DOT sensitivity. There is substantial increase in skull thickness [22], CSF volume [23], and global gray matter / white matter growth [7, 8, 23, 24]. There are developmental effects in DOT sensitivity. Age was positively associated with the DOT sensitivity in the frontal and occipital regions in head models of children between 5 to 11 years old [25]. Fukui et al. [10] compared DOT sensitivity by separation distances in neonates (20–24 weeks) and adults. They found that the sensitivity to brain tissues was greater in neonates than adults across all separation distances (10 mm to 60 mm). Additionally, there was an age-by-separation-distance effect. Larger source-detector separation distances (from 40 mm to 50 mm) extended the penetration of the fluence distribution to the gray matter in neonates but did not increase the penetration depth in the adult head model. Age differences in DOT sensitivity may be attributed to anatomical variations across ages. Increased scalp-to-brain distance was related to reduced path length through the brain regions in children [25]. Moreover, greater scalp and skull thickness was associated with reduced sensitivity to the gray matter across source-detector separation distances from 20 mm to 50 mm in adults [20]. These studies concluded that age-related changes in head/brain anatomy and composition affect the DOT sensitivity function. However, these studies are based on a limited range of ages (e.g., neonate; 5 to 11 years) and do not cover the age span across infancy when dramatic development of the head and brain occurs.

The applications of S-D channel DOT sensitivity profile

The DOT sensitivity profile provides a measure of the scalp-location-to-cortex correspondence. Fu and Richards (under review) estimated the distance between the scalp and the maximum fluence for the S-D Channel DOT fluence. They found that scalp-to-cortex distance increased from infancy to childhood and from childhood to adulthoods. This confirmed that longer source-detector separation distances are required for adults to achieve comparable S-D Channel DOT sensitivity profile. The S-D Channel fluence distribution can also be used to localize the regions of interest (ROIs) that would be sampled with NIRS. Zimeo Morais et al. [26] used the S-D Channel DOT fluence to compute the specificity of each channel (10–10 electrode pairings) in measuring the corresponding ROIs and non-brain tissues. Fu and Richards (under review) [27] also used the S-D Channel DOT measure to map 10–10 channel locations with underlying ROIs from macro- and micro-structural atlases. Such channel-to-ROI look-up tables facilitate the optimization of channel array design to maximize DOT sensitivity to user-specified ROIs [26, 28] and help to localize the possible ROIs that generated the fNIRS signals [29-31]. Existing tools for designing channel placements based on ROIs estimates DOT sensitivity using adult head models. Hence, it is important to evaluate whether the estimations from adult head models can be applied to developmental samples.

The present study

The current study examined age-related variations in DOT sensitivity profiles. Our primary contribution was to use existing methods for DOT sensitivity across the period of infancy (2 weeks through 24 months) with narrow age bands, and compare these to older children (4, 12 years) and adults (20–24 years). We used the MCX photon migration simulation program to model photon propagation through the segmented tissues in the individual head MRIs (Fang & Boas [15]; see Supporting Information for further information). We examined S-D Channel DOT fluence as a function of sampling depth. We also depicted how the DOT sensitivity profile changes across source-detector separation distances. We expect to replicate existing findings in adults. That is, the fluence strength will decrease with increased depth at all source-detector separation distances. The shape of the DOT sensitivity profile will change from short to longer separation distances. The study will provide novel evidence revealing developmental changes in how the DOT sensitivity profiles vary across separation distances. The present study has important implications in the design of age-specific channel placements for NIRS recordings and make anatomical interpretations of stand-alone NIR data across a wide range of age groups.

Materials and methods

Participants

The participants were 1058 typically developing participants ranging from 2 weeks to 24 years of age. The MRIs were obtained from open-access databases and a local scanning facility. Details of the sample are described in [27]. Table 1 presents the number of MRIs for the open access databases, separately for age and gender. The sample ages were narrowest in the infancy period (1-, 1.5-, 3-, or 6-month intervals from 2 weeks through 2 years) and included exemplar ages in children and adolescent ages (4, 12 years) and adult (20–24 years). All studies had institutional review board approval and informed consent. All data were collected specifically for research purposes. Data were completely anonymized and deidentified for processing and analysis. The University of South Carolina Institutional Review Board approved the current study (including data collection at the MacCausland Center for Brain Imaging (MCBI) and the use of data from all open-access databases).
Table 1

Demographical information of study participants by age group, sex, and data source.

Participant InformationData Source
Age GroupTotal NFemale NABIDE NBCP NEBDS NIBIS NMCBI and Collaboration Sites NPING N
2 Weeks41240338000
1 Month964001779000
2 Months68400860000
3 Months382102400140
4.5 Months542904100130
6 Months743500060140
7.5 Months611700049120
9 Months60350480390
10.5 Months42210400020
12 Months1093600089200
15 Months78410630870
18 Months763103608320
2 Years66220006510
4 Years24900001014
12 Years37149000280
20–24 Years1347700001340

Note. ABIDE = Autism Brain Imaging Data Exchange; BCP = Baby Connectome Project; EBDS = Early Brain Development Study; IBIS = Infant Brain Imaging Study; MCBI = McCausland Center for Brain Imaging; PING = Pediatric Imaging, Neurocognition, and Genetics Data Repository

Note. ABIDE = Autism Brain Imaging Data Exchange; BCP = Baby Connectome Project; EBDS = Early Brain Development Study; IBIS = Infant Brain Imaging Study; MCBI = McCausland Center for Brain Imaging; PING = Pediatric Imaging, Neurocognition, and Genetics Data Repository

MRI sequences

The present study utilized T1-weighted (T1W) and T2-weighted (T2W) scans from each collection site. Details of the MRI acquisition protocols have been described in literatures on the Neurodevelopmental MRI Database [32-37]. All MRIs were converted to NIFTI compressed format with 32-bit floating point resolution. Bias-field inhomogeneity correction (N4 algorithm) was performed on the extracted T1-weighted images [38, 39].

MRI preprocessing and segmentation

First, the brains were extracted from the whole-head MRI volume in a procedure adapted from the FSL VBM pipeline [40]. The T1W volume for each participant was registered to an age-appropriate average MRI template. The MRI templates came from the Neurodevelopmental MRI database [35-37]. The brain from the average template was transformed into the participant MRI space and used a mask on the head volume. The extracted masked data was then used with the FSL brain extraction tool program [41, 42]. Each brain was visually inspected and manually modified if necessary. Second, each head MRI volume was segmented into 9 or 10 media types: gray matter (GM), white matter (WM), cerebrospinal fluid (CSF), non-myelinated axons (NMA), other brain matter, skin, skull, air, eyes, and other inside skull material. Details of the segmentation methods are presented in the Supporting Information. The segmented regions were assembled into a single MRI volume we will refer to as the “segmented head MRI volume”. Fig 1A shows a 3D rending of the T1W volume from a 3-month-old infant with a cutout revealing the segmented MRI volume. The realistic head model represents the geometry of the head and allow us to differentiate optical properties of different tissue types.
Fig 1

Segmented head MRI volumes.

The examples were taken from the same 3-month-old infant MRI. A. The segmented head model. Aqua is the scalp, green is the non-bone structure (muscle, sinew, fat), gold is the nasal and mouth air cavities, turquoise is the skull, light blue is the dura, orange is the cerebrospinal fluid, dark blue is the gray matter, and purple is the white matter. B. The segmented head model with dense finite element (FE) mesh.

Segmented head MRI volumes.

The examples were taken from the same 3-month-old infant MRI. A. The segmented head model. Aqua is the scalp, green is the non-bone structure (muscle, sinew, fat), gold is the nasal and mouth air cavities, turquoise is the skull, light blue is the dura, orange is the cerebrospinal fluid, dark blue is the gray matter, and purple is the white matter. B. The segmented head model with dense finite element (FE) mesh.

Mesh generation

A “finite element” (FE) tetrahedral mesh was constructed from the segmented head MRI volume. Fig 1B displayed the dense meshes. They were produced using the iso2mesh toolbox with CGAL 3.6 mesh program (“v2m”; [43]). The FE volumetric meshes have nodes that represent the voxel locations for the tetrahedra, a 4-element matrix representing the corners of each tetrahedron, and a vector representing the media type from the segmented head MRI volume. A mesh was generated for each segmented head MRI volume. The number of nodes, elements, and tetra volume were calculated for the infants (2 weeks through 2 years), children (4 and 12 years), and adults (Fig 1 in S1 File). We will refer to this as the “segmented FE mesh”. The dense “segmented FE mesh” was used for MCX [15] to find a segment element that is closest to an optode position.

Scalp locations

Virtual optodes placement

The locations for the 10–10 and 10–5 electrode systems [44] were constructed on each head MRI volume. These standard locations were used for virtual optodes placement. First, we manually marked cranial fiducial points using MRIcron [45, 46]: nasion (Nz), vertex (Vz), inion (Iz), left preauricular point (LPA), right preauricular point (RPA), left mastoid (LMa), and right mastoid (RMa) [47]. The coordinates of the fiducials were transferred on to the scalp mesh. Next, we calculated 81 virtual positions based on the “unambiguously illustrated 10–10 system [44]. Details for constructing the 10–10 locations are described in [47] and the Supporting Information. We simulated a total of 358 optodes on 10–5 locations by calculating center points between 10–10 positions. The optode positions were also computed for the average MRI templates. Fig 2 illustrates the virtual optodes placements at 10–10 and 10–5 electrode positions.
Fig 2

Virtual optodes placement.

A. Virtual optodes placement on the standard 10–10 electrode system. From left to right: a three-month individual head model, an average template for three-month-old’s, and a two-dimensional layout of the 10–10 system). Ten-ten electrodes were divided into six groups for visualization purposes. The six groups were color-coded on the two-dimensional schematic of the 10–10 system. Group 1: electrodes on the central curve (Nz-Cz-Iz). Group 2: electrodes on the left curve between Nz and Iz (N1-LPA/T9-I1). Group 3: electrodes on the right curve between Nz and Iz (N2-RPA/T10-I2). Group 4: electrodes on the left curve between Fpz and Oz (Fp1-T7-O1). Group 5: electrodes on the right curve between Fpz and Oz (Fp2-T8-O2). Group 6: the remaining electrodes enclosed by the central curve, the left and right curves between Fpz and Oz. B. Virtual optodes placement on the 10–5 electrode system. From left to right: the same three-month individual head model and the three-month average template.

Virtual optodes placement.

A. Virtual optodes placement on the standard 10–10 electrode system. From left to right: a three-month individual head model, an average template for three-month-old’s, and a two-dimensional layout of the 10–10 system). Ten-ten electrodes were divided into six groups for visualization purposes. The six groups were color-coded on the two-dimensional schematic of the 10–10 system. Group 1: electrodes on the central curve (Nz-Cz-Iz). Group 2: electrodes on the left curve between Nz and Iz (N1-LPA/T9-I1). Group 3: electrodes on the right curve between Nz and Iz (N2-RPA/T10-I2). Group 4: electrodes on the left curve between Fpz and Oz (Fp1-T7-O1). Group 5: electrodes on the right curve between Fpz and Oz (Fp2-T8-O2). Group 6: the remaining electrodes enclosed by the central curve, the left and right curves between Fpz and Oz. B. Virtual optodes placement on the 10–5 electrode system. From left to right: the same three-month individual head model and the three-month average template.

Source-detector channels, all optode pairs

Source-detector channels were defined in two ways for the analyses. : Source-detector channel locations were defined using optode combinations centered on each 10–10 electrode location. The 10–10 electrode locations were centered between surrounding adjacent pairs of 10–10 or 10–5 electrode locations. There were 251 source-detector channels formed with adjacent 10–10 electrodes, MeanSeparation = 58.0 mm; SD = 14.2, and 251 channels formed with adjacent 10–5 electrodes, MeanSeparation = 28.9 mm; SD = 7.1. The channel locations were used to estimate “S-D Channel DOT fluence” described below. : Some channel locations were defined from the combination of all 10–5 electrode positions for the source-detector separation distance analyses. The path from each optode to all other optodes was traced on the scalp and the distance of the path recorded (63,903 paths). The channel location was defined at the half-way point in this path. The optode pairs, scalp path distances, and channel locations were recorded. Source-detector separation distances were calculated as half of the scalp path distances. The separation distances were used for the S-D Channel DOT sensitivity profile analysis described below.

Photon migration simulations

The DOT fluence values were estimated with photon migration simulations. We modeled photon migration through brain tissues using the Monte Carlo eXtreme package (MCX; [15, 48]). The package implements an GPU-accelerated, voxel-based Monte Carlo simulation algorithm. The inputs of Monte Carlo simulations include a segmented head model that defines the tissue types. Five tissue types are typically identified: the scalp, skull, cerebrospinal fluid (CSF), GM, and WM [3, 10, 18]. The voxel-based MCX simulations do not require “segmented FE mesh” [15]. The default voxel size 1×1×1mm3 was used. We launched 108 photons for the time window of 0 to 5 nanoseconds. The fluence resolution was 50 time-gates [26]. The wavelength was set at 690 nanometers. Photons were sent from the 358 10–5 positions. The optical properties of the 10 tissue types that are specified include the absorption coefficient, the scattering coefficient, the anisotropy coefficient, and the index of refraction. There is no consensus on the optical properties in the literature [26]. The values differ considerably across studies [2–4, 19, 25, 26, 49]. Different values may be needed for adult and infant-child participants (cf. [10, 19, 25]). S1 Table summarizes optical properties used in several adult and infant-child studies. The table shows that studies commonly set optical properties for scalp and skull, CSF, GM, and WM. Our input values for the optical properties of the head media and the sources of these values were shown in Table 2. They were based on values used in [50], which primarily came from [16, 49, 51].
Table 2

Optical properties used in Monte Carlo photon migration simulations.

Tissue Typeμa (mm-1)μs (mm-1)gNSources and Rationales
White matter0.0740.10.851.37[50, 51]
Gray matter0.028.40.901.37[50, 51]
CSF0.000410.991.37[16, 49, 50, 52]; mimic the optical properties of water.
Dura0.0101800.991.37[50] for skin and non-brain tissues
Skull0.01011000.991.37[2, 49, 50]
Skin0.0101800.991.37[50] for skin and non-brain tissues
Muscle0.0101800.991.37[50] for skin and non-brain tissues
Eyes0.000410.991.37Mimic the optical properties of water.
Nasal Cavity0.0101800.991.37[50] for skin and non-brain tissues
Non-myelinated Axons0.0740.10.851.37Mimic the optical properties of WM for infants at 12 months of age or younger.

μ is the optical absorption coefficient; μ is the scattering coefficient, g is the anisotropy coefficient, and N is the index of refraction.

The wavelength was set at 690 nanometers.

μ is the optical absorption coefficient; μ is the scattering coefficient, g is the anisotropy coefficient, and N is the index of refraction. The wavelength was set at 690 nanometers. The Monte Carlo simulations estimate fluence distribution [3, 4]. The photons were injected at the optode position. The “flux” strength of the photons, or “fluence” of the sum photon flux, decreased monotonically due to absorption and scattering as photons traveled through brain tissues. The simulations accounted for absorption, scattering and reflection based on the specified optical properties [4, 15]. Each simulation continued until the photon exited the media, was absorbed by the media, or after 5 nanoseconds. The output of the simulations included the photon fluence at each nanosecond for each voxel from the injected photons. The number ranged from zero to the total number of injected photons. We summed this value across the 5 nanoseconds of the simulation. This value represents the fluence of the simulation. We took the log (fluence) as our number representing the DOT sensitivity for each voxel.

DOT sensitivity analyses

The output from the Monte Carlo simulation contained the fluence across the entire MRI volume separately for each optode. We simulated the path from a source-optode to a detector-optode. This was done by multiplying the source-optode fluence distribution by the detector-optode fluence distribution. We will refer to this as “S-D Channel DOT fluence”. It is a unitless measure that represents the sensitivity of the DOT measure for detecting changes in optical properties at a given point inside the head medium (PMDF; [9, 19]; or “3-point Green’s function; [3]). An ad-hoc method has been used in some studies to divide the product by the estimated value of the photons from the source in the vicinity of the detector [50]. This step was not taken in the current study. Fig 3 shows the S-D Channel DOT fluence plotted on an MRI for a single source-detector channel. This figure shows a similar monotonic decrease in fluence strength as photons traveled deeper into the head tissues.
Fig 3

Source-detector (S-D) channel DOT fluence distribution.

Monte Carlo photon migration simulations were used to estimate fluence distributions. The red area represents greater fluence.

Source-detector (S-D) channel DOT fluence distribution.

Monte Carlo photon migration simulations were used to estimate fluence distributions. The red area represents greater fluence.

S-D channel DOT sensitivity profile by source-detector separation distances

We examined how the of S-D Channel DOT sensitivity profile changed as a function of source-detector separation distance. The S-D Channel DOT fluence was extracted voxel by voxel. We recorded the fluence value and the distance (depth) from the channel location to the voxel with the fluence estimation. We refer the distance from the channel to the location of the fluence value inside the head model as “sampling depth” for short. The S-D Channel DOT sensitivity profile was defined as the fluence value at the voxel as a function of the sampling depth. The target source-detector separation distances for testing were from 10 mm to 60 mm by 5 mm increments. They were selected from “all optode pairs”. S2 Table lists the minimum and maximum separation distances for each target separation distance for all age group. Fig 2 in S1 File displays the mean number the source-detector channels with the target separation distances by age group. For each participant and each channel with the target separation distance (e.g. 400 channels with 10mm separation distance), we computed the S-D Channel DOT fluence value at each sampling depth. The fluence values were then averaged across channels by sampling depth (1 to 100), target separation distance, and participant. We additionally assessed the variance of the sensitivity profile by examining the standard error of the mean fluence value as a function of sampling depth. The shape of the S-D Channel DOT fluence distribution was quantified by the half-width half-maximum (HWHM; [53]). The HWHM identified the sampling depth at which the fluence first dropped below half of the maximum value. We referred to this as the HWHM location.

Results

S-D channel DOT sensitivity profiles by source-detector separation distances

We examined the S-D Channel DOT fluence as a function of the depth between channel location and the voxel sampled in the fluence distribution (“sampling depth”). This was done by selecting 10–5 location optode pairs with separation distances from 10 mm to 60 mm by 5 mm increments and calculating the S-D Channel DOT fluence for each pair and averaging over pairs of the same separation distance. Fig 4 displays examples of the S-D Channel DOT sensitivity profiles plotted by channels (separate lines) and separation distances (panels) for one individual from four different ages at 20 mm and 50 mm separation distances (more examples are displayed in Fig 3A and 3B in S1 File). The sensitivity profiles across the four individuals showed greater peaks and steeper slopes for channels at 20 mm than 50 mm separation distance. The between-channel variance showed a systematic decline from the infants to the adults. The variance increased from 20 mm to 50 mm separation distances for the infant groups whereas it remained stable for the adult group.
Fig 4

Source-detector (S-D) channel DOT fluence sensitivity profile by channels.

Four example individuals were selected, one from each age category. The S-D Channel DOT fluence value was plotted as a function of the sampling depth separately for channels at the source-detector separation distances for 20 mm and 50 mm for everyone. For each participant and each channel with the target separation distance, we computed the S-D Channel DOT fluence and the distance from the channel location to the voxel with the fluence estimation (“sampling depth”).

Source-detector (S-D) channel DOT fluence sensitivity profile by channels.

Four example individuals were selected, one from each age category. The S-D Channel DOT fluence value was plotted as a function of the sampling depth separately for channels at the source-detector separation distances for 20 mm and 50 mm for everyone. For each participant and each channel with the target separation distance, we computed the S-D Channel DOT fluence and the distance from the channel location to the voxel with the fluence estimation (“sampling depth”). The sensitivity profiles changed with separation distances across age groups. Fig 5A shows the S-D Channel DOT fluence as a function of sampling depth separately for the target source-detector separation distances, averaged across channel locations, and separately for four age bins. We grouped the samples to four age bins for averages based on preliminary data visualization: 2 weeks through 4.5 months; 6 months through 12 months; 15 months through 2 years; 4, 12, and 20–24 years. The sensitivity profile changed systematically with the source-detector separation distance. The peak of the sensitivity function was largest at the smallest separation distance and decreased as source-detector separation distance increased. However, the fluence value dissipated more quickly at the shorter source-detector separation distances than the longer separation distances, with the distributions becoming flatter across increasing source-detector separation distances. Compared to the sensitivity profile at shorter separation distances, the same fluence value was carried to greater sampling depth at larger separation distances.
Fig 5

Source-detector (S-D) channel DOT fluence sensitivity profile by age groups.

The S-D Channel DOT fluence value was plotted as a function of the sampling depth, defined as the distance (depth) from the channel location to the voxel location in the head model where the S-D channel DOT fluence was measured. A. S-D Channel DOT fluence sensitivity function at each target source-detector separation distances. For visualization, we created four age categories: 2 weeks to 4.5 months, 6 months to 12 months, 15 months to 2 years, and 4 years to 20–24 years. Mean values by each age category are displayed. B. S-D Channel DOT fluence sensitivity function by individual age categories at 20 mm source-detector separation distance. C. S-D Channel DOT fluence sensitivity function by individual age categories at 30 mm separation distance. D. S-D Channel DOT fluence sensitivity function by individual age categories at 50 mm separation distance.

Source-detector (S-D) channel DOT fluence sensitivity profile by age groups.

The S-D Channel DOT fluence value was plotted as a function of the sampling depth, defined as the distance (depth) from the channel location to the voxel location in the head model where the S-D channel DOT fluence was measured. A. S-D Channel DOT fluence sensitivity function at each target source-detector separation distances. For visualization, we created four age categories: 2 weeks to 4.5 months, 6 months to 12 months, 15 months to 2 years, and 4 years to 20–24 years. Mean values by each age category are displayed. B. S-D Channel DOT fluence sensitivity function by individual age categories at 20 mm source-detector separation distance. C. S-D Channel DOT fluence sensitivity function by individual age categories at 30 mm separation distance. D. S-D Channel DOT fluence sensitivity function by individual age categories at 50 mm separation distance. The patterns of age-related differences in the sensitivity profiles were relatively consistent across separation distances. Fig 5B–5D, and S1 File in Fig 4 show the S-D Channel DOT fluence sensitivity functions for the 20 mm, 30 mm, and 50 mm source-detector separation distances, respectively, and separately for the testing ages. There were consistent age-related changes in the shape of the sensitivity function over the source-detector separation distances. These are most easily seen in the profile for the 50 mm separation distance (Fig 5D and S1 File in Fig 4C). The line graph in the Main Text shows that the 2-week-to-4.5-month age groups (Fig 5A solid line, aqua color) had a relatively high peak and rapid decline, the 6-month-to-12-month (Fig 5A solid line blue color) and the 15-month-to-2-year age groups (Fig 5A dashed line red color) had a smaller peak value and less rapid decline, i.e., flatter shape. The bar graph in the Supporting Information additionally shows that the sensitivity function at the 50 mm separation distance had a more flattened shape across all age groups. Age-related differences were greater at between 0 to 40 mm sampling depth. Together, the figures indicate that the age differences were similar across all source-detector separation distances but became greater with increasing source-detector separation distances.

Summary

Our findings showed that the DOT sensitivity profiles display similar characteristics across age groups but also important age-related differences. The S-D Channel DOT fluence decreased exponentially in all age groups as the light traveled deeper into the head tissues. The sensitivity profiles for the 20-to-24-year-olds showed that as the separation distances increased, the peak of the fluence value decreased, but the fluence strength declined less rapidly as the light traveled through the tissues. These patterns confirmed the DOT sensitivity profiles found in previous studies using adult head models [3, 18, 20, 21]. Older infants (6 months to 2 years) displayed different DOT sensitivity profiles than young infants (2 weeks to 4.5 months) and adults, and the age-group differences are more discernible in larger source-detector separation distances.

The variance of S-D channel DOT sensitivity profiles

We also examined the between-channel variance of the S-D Channel DOT fluence as a function of sampling depth. Fig 4 shows the average between-channel variance of the sensitivity profiles across separation distances, which apparently differed across ages and showed a larger change with source-detector separation distance for infants than adults. We further examined the variance by computing the average standard error of the mean S-D Channel DOT fluence value. Fig 6A displays the average of the standard error as a function of sampling depth for each separation distance. The average variance as a function of sampling depth increased at separation distances greater than 35 mm for the younger infants (2 weeks to 12 months). The average variance became larger with increased sampling depth for the infants between 15 months to 2 years of age at separation distances greater than 20 mm. The average variance across sampling depth significantly increased at 55 mm and 60 mm separation distances for the 15-month-to-2-year group. The average variance across sampling depth remained relatively stable for the child and adult group. Fig 6B and S1 File in Fig 5 show the standard error of the mean S-D Channel DOT fluence value as a function of sampling depth separately by the testing age groups at 20 mm separation distance. The variance across sampling depth was the greatest for infants aged between 3 months to 2 years relative to the children and adults. The youngest infants (2 weeks, 1 month, 2 months) showed relatively small variance across separation distances. We present similar figures for 30 mm and 50 mm separation distances in Fig 6A and 6B in S1 File. These figures show that the variance as a function of sampling depth was greater at larger separation distances for all age groups. The age-related differences in variance were enhanced at 50 mm separation distances. The increase in variance at larger separation distance is more discernible for infants between 3 months to 2 years. The patterns of age-group differences in variance of the sensitivity profile across separation distances were consistent with the individual-subject plots displayed in Fig 4 and S1 File in Fig 3.
Fig 6

Variance of source-detector (S-D) channel DOT fluence sensitivity profile.

This is the standard error of the S-D Channel DOT fluence value as a function of the sampling depth, defined as the distance (depth) from the channel location to the voxel location in the head model where the S-D channel DOT fluence was measured. A. Standard error of the S-D Channel DOT fluence sensitivity function averaged by the age categories. B. Standard error of the S-D Channel DOT fluence sensitivity function by individual age groups at 20 mm source-detector separation distance.

Variance of source-detector (S-D) channel DOT fluence sensitivity profile.

This is the standard error of the S-D Channel DOT fluence value as a function of the sampling depth, defined as the distance (depth) from the channel location to the voxel location in the head model where the S-D channel DOT fluence was measured. A. Standard error of the S-D Channel DOT fluence sensitivity function averaged by the age categories. B. Standard error of the S-D Channel DOT fluence sensitivity function by individual age groups at 20 mm source-detector separation distance. Between-channel variance was relatively stable for the children and adults. The variance increased as a function of source-detector separation distance in infant groups. The increase was most observable in older infants (6 months to 2 years).

The shape of S-D channel DOT sensitivity profiles

The last analysis examined the half-width half-maximum (HWHM) location for each S-D Channel DOT fluence distribution. Fig 7 shows the HWHM location as a function of the source-detector separation distance separately for the testing ages (Fig 7A) and for four separate age groupings (Fig 7B). Fig 7 in S1 File displays the HWMH location as a function of age groups, separately for separation distances. The HWHM increased as a function of the source-detector separation distance for all age groups. This increase quantifies the apparent change in steepness in the S-D Channel DOT fluence sensitivity functions over source-detector separation distances (Fig 5A). The change in HWHM across source-detector separation distance implies that there was a gain in sensitivity to deeper cortex over the source-detector separation distance, though the fluence strength systematically decreased with increasing source-detector separation distance (e.g., Fig 5A). The patterns of the HWHM location over age parallel the age differences found in the sensitivity profile (e.g., Fig 5A). The mean HWHM locations across all separation distances were greater in the youngest infants (2-week-to-4.5-month group, Fig 7A and 7B solid line, aqua color), children and adults (Fig 7A and 7B green dashed line), and at smaller levels in the 6-month-to-12-month group (Fig 7A and 7B solid line blue color) and the 15-month-to-2-year group (Fig 7A and 7B dashed line red color). These differences were consistent with the age differences found in the sensitivity profile (e.g., Fig 5A). The larger HWHM location values imply the DOT sensitivity to deeper cortical areas for the youngest infants, children, and adults.
Fig 7

Half-width half-maximum (HWHM) locations of source-detector (S-D) channel fluence as a function of source-detector separation distances.

The figure shows age-related differences in how the shape of the S-D Channel DOT sensitivity profile changes with separation distances. The HWHM location was defined as the point in the fluence distribution at which the fluence first dropped below half of the maximum value. A. HWHM locations by source-detector separation distances for individual age groups. B. Mean HWHM locations by source-detector separation distances for the four age categories: 2 weeks to 4.5 months, 6 months to 12 months, 15 months to 2 years, and 4 years to 24 years.

Half-width half-maximum (HWHM) locations of source-detector (S-D) channel fluence as a function of source-detector separation distances.

The figure shows age-related differences in how the shape of the S-D Channel DOT sensitivity profile changes with separation distances. The HWHM location was defined as the point in the fluence distribution at which the fluence first dropped below half of the maximum value. A. HWHM locations by source-detector separation distances for individual age groups. B. Mean HWHM locations by source-detector separation distances for the four age categories: 2 weeks to 4.5 months, 6 months to 12 months, 15 months to 2 years, and 4 years to 24 years. The shape of DOT sensitivity profiles corroborated existing evidence that the fluence distribution extends deeper into the head tissues as the source-detector distance increases [3, 18, 20, 21]. This pattern was found in all age groups. The important age-related difference was that the same proportion of fluence strength carried deeper into the head tissues for the young infant, child, and adult groups across all separation distances.

Discussion

The present study examined age differences in DOT sensitivity profiles. We extended existing photon migration simulation methods to realistic head models from infants ranging from 2 weeks to 2 years, children (4, 12 years) and adults. The DOT sensitivity profiles had a common shape across source-detector separation and ages. There were expected differences in the S-D Channel DOT sensitivity profile across source-detector separation distances, with larger separation distances being more sensitivity to greater cortical depths but showing decreasing signal strength and larger variance in the fluence measure. A noticeable age-group difference was that the fluence distribution penetrated greater depth with the same level of signal loss in the younger infant (2 weeks to 4.5months) than the rest of the infant age groups. The shape of S-D Channel DOT sensitivity profile was similar between the younger infant groups and child and adult groups. The present study showed that the S-D Channel DOT sensitivity profiles changed systematically as a function of source-detector separation distance for all age groups. The study enhanced the DOT sensitivity estimation accuracy by using realistic FEM head models with refined segmentation of tissue types. The S-D Channel DOT sensitivity profiles depict S-D Channel DOT fluence as a function of sampling depth in the fluence distribution. The peak value of the S-D Channel DOT fluence was larger at shorter separation distances but showed rapid decay as sampling depth increased. Alternatively, at larger separation distances there was a smaller peak value but a less rapid decline with increased sampling depth. The shape of the sensitivity profile was characterized by measuring the point in the fluence distribution where the fluence value first dissipated below half of the maximum value (i.e., “HWHM location”). The HWHM location increased with source-detector separation distance for all age groups. This indicates that there is a gain in sensitivity to deeper cortical structures with increased separations at the expense of decreased signal strength. We replicated the robust characteristics of the DOT sensitivity profiles found in existing studies that modeled photon migrations in head models from infants [10] and adults [3, 18, 20, 21]. The trade-off between the gain in sensitivity to deeper cortical regions and loss in signal strength and spatial resolution needs to be considered in the design of source-detector separation distances. The sensitivity profiles showed relatively consistent age-related differences across source-detector separation distances. Young infant groups (2 weeks to 4.5 months) displayed a sensitivity profile that was significantly different from the older infant groups (6 months to 2 years). The sensitivity profiles for the young infant groups at all separation distances were characterized by higher peaks and steeper decreases of S-D Channel DOT fluence values with increased sampling depth. Their sensitivity profiles displayed relatively stable variance across separation distances. They had greater HWHM locations than older infant group at all separation distances. This means that the same fluence level extended deeper into the cortex in the younger infant groups. In contrast, the sensitivity profiles for the older infant groups were characterized by smaller peaks and more gradual declines with increased sampling depth. Their sensitivity profiles showed greater variance at larger separation distances. The slopes of increase in HWHM location across source-detector separation distance were similar across age groups. This suggests that the gain in S-D Channel DOT sensitivity by increasing separation distances is comparable across age groups. Developmental changes in the morphology of the head and the brain may contribute to the age-related differences in S-D Channel DOT sensitivity profiles. Infants have thinner extracerebral layers (scalp and skull) and thicker CSF than adults [19]. The skull of younger infants has greater curvature and is more flexible. Younger infants might have a thicker CSF layer in the back of their heads due to the head positioning during the MRI scan [54]. The low scattering CSF may confine the S-D Channel DOT fluence distribution to shallower layers of the cortex especially at shorter source-detector separation distances. Fu and Richards (under review) [27] showed that the distance between the scalp channel location and the maximum fluence in the S-D Channel DOT fluence distribution was relatively shorter for the 2-week to 2-month groups. A direction for future research is to characterize morphological changes in specific tissue layers that are linked to age-related differences in S-D Channel DOT sensitivity profiles across separation distances.

Applications of age-specific DOT sensitivity profiles

The age differences highlight the importance of making age-specific decisions on the appropriate source-detector separation distances for the target age group [25]. NIRS studies conventionally use 20mm to 30mm separation distances for infants [55, 56] and 30mm to 35mm for children (e.g., [57, 58]) and adults [59]. Young infants and adults displayed comparable DOT sensitivity profiles at 20mm-to-30mm separation distance. However, Fu and Richards (under review) [27] showed that young infants had closer scalp-to-cortex distances than adults (3 months: 8mm; 20–24 years: 13mm). Hence, young infants may have greater DOT sensitivity to the cortex than adults (cf. [10]). The age difference was more discernible at larger separation distances. Increasing separation distances to 35mm or greater may cause more concern in terms of reduction of signal strength and increased variance with relatively less gain in improving sensitivity to the cortex for older infants compared to adults. Detailed descriptions of changes in DOT sensitivity profiles by separation distances across developmental and adult samples can better inform age-specific designs of channel/optode placement. Several studies have used the S-D Channel DOT fluence measure to estimate the mapping between channel locations and ROIs [26-31]. For example, Zimeo Morais et al. [26] provided a toolbox that displays the channels that measure the user-specified ROI with a specificity value (1%-100%) for each channel. For the selected candidate channels, the users can also obtain their specificity to measure other (non-selected) brain regions. The channel(s) that show good discrimination to the user-selected ROI (i.e., high specificity to the ROI and low specificity to other brain and non-brain regions) are desirable. However, the toolbox uses a pool of 130 10–10 channels with a small range of separation distances for the adult head model (median = 3.60cm, lower quartile = 3.13cm, upper quartile = 3.98cm; [26]). The toolbox and existing look-up tables do not consider that the separation distance for a given channel changes with age as head sizes get bigger. Additionally, the current findings indicate that the separation-related changes in sensitivity profile also vary by age groups. Increasing source-detector separation distances beyond 30mm for infants may reduce the channel-to-ROI discrimination. This is problematic especially when researchers use ROIs from micro-structure atlases and are interested in comparing results between age groups. Future tools for designing channel placements should allow users to optimize the constellation of optode locations based on ROIs, age group, and separation distances. The age-related differences in sensitivity profiles reinforce the importance of using age-appropriate head models for DOT image reconstruction. The sensitivity profile estimated from Monte Carlo photon migration simulations can be recorded into a forward matrix. This is then used for solving the inverse problem in DOT image reconstruction that localizes the ROIs that showed hemoglobin concentration changes [5, 6]. Contemporary toolboxes for DOT image source reconstruction (AtlasViewer: [60]; TOAST++: [12]) use adult head models as the default option. The present study provides data that show the importance of using age-appropriate templates as an anatomical prior to improve accuracy in channel location co-registration and forward model estimation. Delgado Reyes et al. [61] and Wijeakumar et al. [62] show examples of this. They selected head models with atlas parcellations from the Neurodevelopmental MRI database [34-37] that were closely matched in age of the participants (4-month- to 2-year-olds: [61]; 4- to 48-month-olds: [62]). The head models were used for estimating the forward model and image reconstruction using the AtlasViewer [60]. The resultant voxel-wise hemodynamic concentration estimates were then entered in group-level analyses.

Limitations

Different optical properties might need to be applied to infant and adult head models for photon migration simulations [10, 19]. There is a paucity of reference for appropriate optical property values for different tissue types in adult and infant realistic head models, given existing studies performed less fined-grained segmentation of tissue layers (scalp, skull, CSF, GM, and WM). It is also computationally challenging to systematically quantify how variations in optical properties of different media types may impact S-D Channel DOT sensitivity estimations and whether the impact may differ between infant and adult realistic head models. In addition, the present study did not differentiate optical property values for some extracerebral tissue layers including the skin, dura, muscle, and nasal cavity. The impact of optical property settings for extracerebral tissues on photon migration might differ between infant and adult head models due to the age-related differences in the thickness of these tissue media [10, 19, 27, 54]. However, Brigadoi and Cooper [19]’s validation analyses showed that variations in the CSF scattering coefficients in the infant head model and scalp absorption coefficients in the adult head model within the range of commonly adopted values did not considerately impact the DOT sensitivity estimations. Future validation studies are needed to ensure that the age differences in DOT sensitivity profiles are over and above the age-related variations in optical properties of extracerebral and brain tissue layers. The current study did not compute DOT sensitivity by tissue types separately. Existing evidence indicated that the change of relative sensitivity as a function of source-detector separation distance displayed different shapes for GM, WM, CSF, and extracerebral tissues (scalp and skull) in infants and adults [3, 10, 18, 20]. Our data could be used to evaluate the fluence strength to the cortex based on the S-D Channel DOT scalp-to-cortex distance and the sensitivity profiles. We also could directly compare the relative sensitive to the brain versus extracerebral tissues at different separation distances for infants and adults [3, 19]. Future analyses that compare sensitivity profiles by tissue types could more precisely inform age-related differences in the optimal separation distances for sampling GM signal changes.

Conclusions

The current study examined S-D Channel DOT sensitivity profiles in infants, children, and adults. The profiles changed systematically as a function of source-detector separation distance for all age groups. There were also important age-related differences in the S-D Channel DOT sensitivity profiles. These are particularly discernible between younger (2 weeks to 4.5 months) and older infants at larger separation distances. Our findings of age-related differences in S-D Channel DOT sensitivity profiles are useful for determining age-appropriate source-detector separation distances and placement that match sensitivity, signal strength, and penetration depth for different ages. Our findings imply that the optimization of channel configurations and accurate anatomical interpretations of NIRS/fNIRS data are dependent on developmentally sensitive estimations of S-D Channel DOT sensitivity that account for the head and cortical development. The present study has demonstrated that age-appropriate realistic head models can be used with photon migration simulations to provide anatomical guidance for standalone DOT data. (DOCX) Click here for additional data file.

References of optical properties used in published studies.

(XLSX) Click here for additional data file.

Source-detector separation information for “all optode pairs” constructed from 10–5 optode locations.

It provides the minimum and maximum distances specified for each target separation distance (10mm to 60mm in 5mm increment) for all age groups. (XLSX) Click here for additional data file. 22 Feb 2021 PONE-D-20-39767 Age-related Changes in Diffuse Optical Tomography Sensitivity by Source-Detector Separation in Infancy PLOS ONE Dear Dr. Fu, Thank you for submitting your manuscript to PLOS ONE. After careful consideration, we feel that it has merit but does not fully meet PLOS ONE’s publication criteria as it currently stands. Therefore, we invite you to submit a revised version of the manuscript that addresses the points raised during the review process. Please submit your revised manuscript by Apr 02 2021 11:59PM. If you will need more time than this to complete your revisions, please reply to this message or contact the journal office at plosone@plos.org. When you're ready to submit your revision, log on to https://www.editorialmanager.com/pone/ and select the 'Submissions Needing Revision' folder to locate your manuscript file. Please include the following items when submitting your revised manuscript: A rebuttal letter that responds to each point raised by the academic editor and reviewer(s). You should upload this letter as a separate file labeled 'Response to Reviewers'. A marked-up copy of your manuscript that highlights changes made to the original version. You should upload this as a separate file labeled 'Revised Manuscript with Track Changes'. An unmarked version of your revised paper without tracked changes. You should upload this as a separate file labeled 'Manuscript'. If you would like to make changes to your financial disclosure, please include your updated statement in your cover letter. Guidelines for resubmitting your figure files are available below the reviewer comments at the end of this letter. If applicable, we recommend that you deposit your laboratory protocols in protocols.io to enhance the reproducibility of your results. Protocols.io assigns your protocol its own identifier (DOI) so that it can be cited independently in the future. For instructions see: http://journals.plos.org/plosone/s/submission-guidelines#loc-laboratory-protocols We look forward to receiving your revised manuscript. Kind regards, Andrea Farina Academic Editor PLOS ONE Journal Requirements: When submitting your revision, we need you to address these additional requirements. 1. Please ensure that your manuscript meets PLOS ONE's style requirements, including those for file naming. The PLOS ONE style templates can be found at https://journals.plos.org/plosone/s/file?id=wjVg/PLOSOne_formatting_sample_main_body.pdf and https://journals.plos.org/plosone/s/file?id=ba62/PLOSOne_formatting_sample_title_authors_affiliations.pdf 2. We note that the grant information you provided in the ‘Funding Information’ and ‘Financial Disclosure’ sections do not match. When you resubmit, please ensure that you provide the correct grant numbers for the awards you received for your study in the ‘Funding Information’ section. 3. We note that you have indicated that data from this study are available upon request. PLOS only allows data to be available upon request if there are legal or ethical restrictions on sharing data publicly. For information on unacceptable data access restrictions, please see http://journals.plos.org/plosone/s/data-availability#loc-unacceptable-data-access-restrictions. In your revised cover letter, please address the following prompts: a) If there are ethical or legal restrictions on sharing a de-identified data set, please explain them in detail (e.g., data contain potentially identifying or sensitive patient information) and who has imposed them (e.g., an ethics committee). Please also provide contact information for a data access committee, ethics committee, or other institutional body to which data requests may be sent. b) If there are no restrictions, please upload the minimal anonymized data set necessary to replicate your study findings as either Supporting Information files or to a stable, public repository and provide us with the relevant URLs, DOIs, or accession numbers. Please see http://www.bmj.com/content/340/bmj.c181.long for guidelines on how to de-identify and prepare clinical data for publication. For a list of acceptable repositories, please see http://journals.plos.org/plosone/s/data-availability#loc-recommended-repositories. We will update your Data Availability statement on your behalf to reflect the information you provide. [Note: HTML markup is below. Please do not edit.] Reviewers' comments: Reviewer's Responses to Questions Comments to the Author 1. Is the manuscript technically sound, and do the data support the conclusions? The manuscript must describe a technically sound piece of scientific research with data that supports the conclusions. Experiments must have been conducted rigorously, with appropriate controls, replication, and sample sizes. The conclusions must be drawn appropriately based on the data presented. Reviewer #1: Partly Reviewer #2: Yes ********** 2. Has the statistical analysis been performed appropriately and rigorously? Reviewer #1: N/A Reviewer #2: N/A ********** 3. Have the authors made all data underlying the findings in their manuscript fully available? The PLOS Data policy requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception (please refer to the Data Availability Statement in the manuscript PDF file). The data should be provided as part of the manuscript or its supporting information, or deposited to a public repository. For example, in addition to summary statistics, the data points behind means, medians and variance measures should be available. If there are restrictions on publicly sharing data—e.g. participant privacy or use of data from a third party—those must be specified. Reviewer #1: Yes Reviewer #2: Yes ********** 4. Is the manuscript presented in an intelligible fashion and written in standard English? PLOS ONE does not copyedit accepted manuscripts, so the language in submitted articles must be clear, correct, and unambiguous. Any typographical or grammatical errors should be corrected at revision, so please note any specific errors here. Reviewer #1: Yes Reviewer #2: Yes ********** 5. Review Comments to the Author Please use the space provided to explain your answers to the questions above. You may also include additional comments for the author, including concerns about dual publication, research ethics, or publication ethics. (Please upload your review as an attachment if it exceeds 20,000 characters) Reviewer #1: The work reported in interesting and there is potential for novel findings. However, the work in the current form requires some modifications and improvements. More specifically: - Sentence in Introduction “DOT does not provide anatomical information about the location of the hemodynamic signal.” is unclear. DOT can be used to provide spatial information where the contrast is generated by light absorption and scattering. What would be the anatomical information of the location that the authors mean? - Manuscript lacks an explanation on Monte Carlo simulation. Basic principles and parameters should be explained. - Three different experimental systems and datatypes are used in DOT: time-domain, frequency-domain and continuous-wave. It seems that the authors are considering continuous-wave simulations. However, that is not mentioned. - In the manuscript, acronym FEM is used for finite element model. However, FEM typically refers to finite element method that is a numerical method for solving partial differential equations. Further, when the authors refer to their discretisation, it should referred as finite element (FE) mesh. Not FEM mesh, if finite element method is not used. - The authors form tetrahedral discretisation from segmented MR images. However, they use voxel-based MCX to simulate light propagation. This is inconsistent. What discretization was used in simulations? Why use two different meshes when one could also use tetrahedral mesh in Monte Carlo? If two different meshes were used, how was the interpolation between the meshes implemented? - The authors use work “electrodes” to describe their light sources and detectors. Electrodes are not used in optical measurements. One can call those “optodes”, or just sources and detectors. - The results show a large number of graphs. However, no statistical analysis of the results is presented. One would be keen to get some more insight or conclusions of the work. If not statistical, then maybe more spatial information. After all, Monte Carlo provides fluence in the whole volume, not only the measurement signal. Reviewer #2: This manuscript studied the fluence change during infancy by using accurate head models with small age intervals and Monte Carlo simulations. This research topic is valuable for current DOT and fNIRS communities since the brain anatomy changes drastically from infancy to adulthood, which will have a significant impact on the fluence distribution in the human brain. This manuscript can be accepted if following issues can be properly addressed. Major: 1. It is believed that the focus of this manuscript is how the age will impact the fluence distribution in the human brain and the importance to have age-appropriate realistic head models. However, the manuscript spends too much time discussing other topics such as separation distance. The mixed discussions on age and separation makes readers confused about the message this manuscript wants to convey. The manuscript should be organized in a way the emphasis can be easily caught by the readers. If the message is the necessity of using accurate infant brain models, the accuracy of the brain models should be discussed. 2. The terminologies used in the manuscript are inconsistent and the definitions of these terminologies are problematic. There is a mixed usage between PPL, penetration depth, separation distance, electrode, optode, etc. The author should stick with one naming convention throughout the manuscript. Also, if possible, the formula should be given to define the terminologies. 3. It is very difficult to extract information by reading the figures. The focus of this paper is about the impact of age on the fluence. However, many figures are plotting the relationship between PPL and fluence. And it is not clear why this information is important to the DOT community since earlier studies have similar analyses. The figures should be arranged in a way showing why it is important to use realistic brain models for infancy. In such a case, age should be the x-axis. Minor: 1. The title is ambiguous. What is the focus of this title? Is it age-related fluence changes or source-detector separation. 2. In the abstract, in statement ‘the peak of the sensitivity function was largest at the smallest separation distance and decreased as source-detector distance increased’, which brain region does the sensitivity indicate? Sensitivity in skin, skull or brain? 3. In the 3rd paragraph, last sentence ‘PPL represent the penetration depths of the photon measured in the medium’. It sounds to be an inaccurate description for PPL. PPL should be the 'photon trajectory length in a specific tissue'. Can you clarify how you computed PPL in this manuscript. 4. In Fig. 1B and C, a brain model with dense and sparse densities is shown. Which mesh did you use in the manuscript? The dense or the sparse one? Why are these two meshes shown? 5. In Fig. 1A, can the author add the legend showing the mapping between color->tissue type? Also, it looks to me the dura is pretty thick and the CSF (dark orange) is not filling between brain and skull? What is the dark blue color in the figure, which appears both outside and inside of the brain? 6. In section ‘Scalp Location’, 1st paragraph, can you add citations for 10-10 and 10-5 electrode systems? 7. In section ‘Source-Detector Channels, All Electrode Pairs’, it is confusing to use electrodes here while you are placing light source or optode in the position. Please just call it source or detector and use the same names throughout the manuscript. 8. In section ‘Photon Migration Simulations’, while the author admitted the optical properties differ considerably across studies and it is likely different values are needed for adults versus development samples, the author did not give a reason or method how these optical properties are compiled in Table 2. This needs to be more clear and discussed in limitations. 9. In Table 2, the wavelength of these optical properties is not specified. In addition, citations should be added for the optical properties, showing from which papers these values are derived. Also, the absorption of muscle seems to be low because muscle is highly absorbing. The nasal cavity, which should be air, has an absorption coefficient of 0.0101 mm^-1. What is the reason for this? Please refer to the paper: a) Selective photobiomodulation for emotion regulation: model-based dosimetry study. 10. In section ‘DOT Sensitivity Analyses’, last sentence, there is a statement ‘This figure shows a similar monotonic decrease in fluence strength as a function of partial path length’. However, which PPL are you referring to? And PPL and penetration depth should not be the same thing. The author should stick with one naming in the manuscript. 11. In Fig. 3, the S-D Channel DOT fluence is a 3-D distribution. But In Fig. 4 and 5, it becomes a scalar that changes depending on PPL. How did you convert it to a value? Also, int Fig. 4 and 5, the S-D Channel DOT fluence does not have a unit. 12. As mentioned in the major issues, the figures from Fig. 4 to 7 are very difficult to read. It is difficult to extract effective information from these figures. These should be re-organized. 13. In section ‘The Shape of S-D Channel DOT Fluence Sensitivity Profiles’, last sentence ‘The larger HWHM location values for the youngest age group implies the DOT sensitivity to deeper cortical areas for the youngest ages.’ However, in Fig. 7B, the HWHM for age 4 to 24 years old is the highest, which conflicts with the claim that the youngest age group has DOT sensitivity in deeper cortical areas. The same statement was made in the 5th sentence in Discussion. Please clarify. 14. In Discussion, while the findings about sensitivity change and separations for infancy are given, the insights on what may cause these changes are not given. While these may not be included in this manuscript, they can be future directions. ********** 6. PLOS authors have the option to publish the peer review history of their article (what does this mean?). If published, this will include your full peer review and any attached files. If you choose “no”, your identity will remain anonymous but your review may still be made public. Do you want your identity to be public for this peer review? For information about this choice, including consent withdrawal, please see our Privacy Policy. Reviewer #1: No Reviewer #2: No [NOTE: If reviewer comments were submitted as an attachment file, they will be attached to this email and accessible via the submission site. Please log into your account, locate the manuscript record, and check for the action link "View Attachments". If this link does not appear, there are no attachment files.] While revising your submission, please upload your figure files to the Preflight Analysis and Conversion Engine (PACE) digital diagnostic tool, https://pacev2.apexcovantage.com/. PACE helps ensure that figures meet PLOS requirements. To use PACE, you must first register as a user. Registration is free. Then, login and navigate to the UPLOAD tab, where you will find detailed instructions on how to use the tool. If you encounter any issues or have any questions when using PACE, please email PLOS at figures@plos.org. Please note that Supporting Information files do not need this step. 25 Mar 2021 Journal Requirements: 1. Please ensure that your manuscript meets PLOS ONE's style requirements, including those for file naming. We have made sure that our manuscript meets PLOS ONE’s style requirements. 2. We note that the grant information you provided in the ‘Funding Information’ and ‘Financial Disclosure’ sections do not match. When you resubmit, please ensure that you provide the correct grant numbers for the awards you received for your study in the ‘Funding Information’ section. We have made sure that the grant information is accurate in the “Funding Information” and “Financial Disclosure” sections. In addition, we edited the “Acknowledgement” section to remove the grant information. 3. We note that you have indicated that data from this study are available upon request. PLOS only allows data to be available upon request if there are legal or ethical restrictions on sharing data publicly. In your revised cover letter, please address the following prompts: a) If there are ethical or legal restrictions on sharing a de-identified data set, please explain them in detail (e.g., data contain potentially identifying or sensitive patient information) and who has imposed them (e.g., an ethics committee). Please also provide contact information for a data access committee, ethics committee, or other institutional body to which data requests may be sent. b) If there are no restrictions, please upload the minimal anonymized data set necessary to replicate your study findings as either Supporting Information files or to a stable, public repository and provide us with the relevant URLs, DOIs, or accession numbers. We do not have legal or ethical restrictions on data sharing. We have uploaded the sensitivity profile dataset and the SAS scripts for data visualization in the Open Science Framework. We have edited the “Data and Code Availability Statement” to reflect this change. Reviewer 1’s Comments: 1. Sentence in Introduction “DOT does not provide anatomical information about the location of the hemodynamic signal.” is unclear. DOT can be used to provide spatial information where the contrast is generated by light absorption and scattering. What would be the anatomical information of the location that the authors mean? Thanks for pointing this out. We agree that DOT instruments with dense, multi-distance channels have improved spatial resolutions. We acknowledge this in the first paragraph: “Multi-channel NIRS instruments measures hemoglobin changes in the scalp channel space. DOT instruments use overlapping and channels with multiple separation distances to enhance spatial resolution (Custo et al., 2010)”. We also specified: “However, both types of the scalp-based measurement do not directly provide anatomical information about the brain regions of the hemodynamic signal. Localizing the channel-wise signals to specific brain regions requires comprehensive understanding of the DOT sensitivity profile.” We noted later in the paragraph: “The forward model can then guide DOT image reconstruction to recover the brain locations of hemoglobin concentration changes (Cooper et al., 2012; Culver, Siegel, Stott, & Boas, 2003). 2. Manuscript lacks an explanation on Monte Carlo simulation. Basic principles and parameters should be explained. We agree the Monte Carlo simulation technique should be explained in more details. We have now highlighted the basic principle in “The Use of Monte Carlo Simulations to Estimate DOT Sensitivity Profile” section in the Introduction: “A goal of DOT is to model how hemoglobin changes at the channel space (what is measured) correspond to changes in optical properties within the brain (to be determined). This requires modeling the DOT sensitivity. One approach to solve the forward model is using finite element method (FEM) quantitative algorithms to describe photon propagations through heterogeneous tissue structures such as the human head (NIRFAST: Dehghani et al., 2009; TOAST++: Schweiger & Arridge, 2014). Alternatively, Monte Carlo simulations provide numerical solutions to the radiative transport equation that models the photon migration through the head tissues based on fewer assumptions (Huppert, Franceschini, & Boas, 2009; Jacques & Wang, 1995). The Monte Carlo method is computationally demanding but provide more accurate solutions for the forward model (Wheelock, Culver, & Eggebrecht, 2019).” We have also explained in more details the input parameters and basic steps for the simulations in the same paragraph. In addition, we moved details on the inputs of the simulations used in the present study from the Supporting Information to the main text Methods section (“Photon Migration Stimulations”). 3. Three different experimental systems and datatypes are used in DOT: time-domain, frequency-domain and continuous-wave. It seems that the authors are considering continuous-wave simulations. However, that is not mentioned. Thank you for pointing out the important detail. Our Monte Carlo simulations are for the continuous-wave systems indeed. We have stated the information in the “The Use of Monte Carlo Simulations to Estimate DOT Sensitivity Profile” section (second paragraph) in the Introduction: “The Monte Carlo simulations commonly model fluence distributions for the continuous-wave systems as they are more widely implemented in commercial NIRS instruments than the time-domain and frequency-domain systems (Huppert et al., 2009; Wheelock et al., 2019). The continuous-wave systems measure fluence at the detector location but do not separately quantify the effects of light absorption and scattering.” 4. In the manuscript, acronym FEM is used for finite element model. However, FEM typically refers to finite element method that is a numerical method for solving partial differential equations. Further, when the authors refer to their discretisation, it should referred as finite element (FE) mesh. Not FEM mesh, if finite element method is not used. Thank you very much for the suggestion. We have changed “FEM mesh” to “finite element (FE) mesh”. 5. The authors form tetrahedral discretisation from segmented MR images. However, they use voxel-based MCX to simulate light propagation. This is inconsistent. What discretization was used in simulations? Why use two different meshes when one could also use tetrahedral mesh in Monte Carlo? If two different meshes were used, how was the interpolation between the meshes implemented? We apologize for the confusion. We used dense tetrahedral discretization. We stated in the “Mesh Generation” section: “The dense “segmented FE mesh” was used for MCX (Fang & Boas, 2009) to find a segment element that is closest to an optode position. In addition, we added in the “Photon Migration Simulations” section: “The voxel-based MCX simulations do not require “segmented FE mesh” (Fang & Boas, 2009). The default voxel size 1×1×1mm3 was used.” We removed information about the “sparse meshes” from the main text and the Supporting Information. Sparse meshes were not used in the procedures implemented for the current paper. 6. The authors use work “electrodes” to describe their light sources and detectors. Electrodes are not used in optical measurements. One can call those “optodes”, or just sources and detectors. Thank you for pointing out the potential confusion. The 10-10 and 10-5 systems are designed for EEG electrode positions (Jurcak, Tsuzuki, & Dan, 2007). We are using “electrode(s)” when referring specifically to the 10-10 or 10-5 electrode systems. We used “optode(s)” and channel(s)” when referring specifically to the NIRS sensors and the positioning of these sensors. We have gone through the main text and Supporting Information to ensure that this rule is consistently applied. 7. The results show a large number of graphs. However, no statistical analysis of the results is presented. One would be keen to get some more insight or conclusions of the work. If not statistical, then maybe more spatial information. After all, Monte Carlo provides fluence in the whole volume, not only the measurement signal. We have now added summary paragraphs in the Results section to highlight the key insights that can be drawn from each finding. For the “S-D Channel DOT Sensitivity Profiles by Source-Detector Separation Distances” results, we added: “Summary: Our findings showed that the DOT sensitivity profiles show robust characteristics across age groups but also important age-related differences. In all age groups, the S-D Channel DOT fluence decreased exponentially as the light traveled deeper into the head tissues. The sensitivity profiles for the 20-to-24-year-olds showed that as the separation distances increased, the peak of the fluence value decreased, but the fluence strength declined less rapidly as the light traveled through the tissues. These patters confirmed the robust characters of DOT sensitivity profiles found in previous studies using adult head models (Mansouri, L'Huillier, Kashou, & Humeau, 2010; Strangman, Li, & Zhang, 2013; Strangman, Zhang, & Li, 2014; Wang, Ayaz, & Izzetoglu, 2019). Older infants (6 months to 2 years) displayed different DOT sensitivity profiles than young infants (2 weeks to 4.5 months) and adults, and the age-group differences are more discernible in larger source-detector separation distances. For “The Variance of S-D Channel DOT Sensitivity Profiles”, we summarized that “Between-channel variance was relatively stable for the child and adult. The variance increased as a function of source-detector separation distance in infant groups. The increase was most observable in older infants (6 months to 2 years).” For “The Shape of S-D Channel DOT Sensitivity Profiles”, we highlighted that “the shape of DOT sensitivity profiles corroborated existing evidence that the fluence distribution extends deeper into the head tissues as the source-detector distance increases (Mansouri et al., 2010; Strangman et al., 2013; Strangman et al., 2014; Wang et al., 2019). This pattern was found in all age groups. The important age-related difference is that the same proportion of fluence strength carries deeper into the head tissues for the young infant, child, and adult groups across all separation distances.” Our current analyses have focused on presenting spatial information of the S-D Channel DOT sensitivity profile. The sensitivity profile is presented as changes in S-D Channel DOT as the function of sampling depth. We also examined changes in half width half maximum locations of the S-D Channel DOT fluence distribution as a function of source-detector separation distances. Reviewer 2’s Comments: 1. It is believed that the focus of this manuscript is how the age will impact the fluence distribution in the human brain and the importance to have age-appropriate realistic head models. However, the manuscript spends too much time discussing other topics such as separation distance. The mixed discussions on age and separation makes readers confused about the message this manuscript wants to convey. The manuscript should be organized in a way the emphasis can be easily caught by the readers. If the message is the necessity of using accurate infant brain models, the accuracy of the brain models should be discussed. Thank you for your suggestion. The two main points that our findings convey are that 1) the basic characteristics of the DOT sensitivity profiles are present for all age groups, and 2) there are also age differences in DOT sensitivity profiles as a function of source-detector separation distances. In response to Reviewer 1 Comment 7, we have now added summaries of the key findings in the Results section. We also edited the Discussion to highlight the central messages. Specificity, in the “DOT sensitivity Profile by Source-Detector Separation Distances” section, we mirrored the organization in the Introduction. The first paragraph stated the common characteristics of the DOT sensitivity profiles across all age groups. The second paragraph highlighted the age-group differences, especially between the young infant groups (2 weeks to 4.5 months) and the older infant groups (6 months to 2 years). This paragraph commented on age-related differences in S-D channel DOT fluence as a function of sampling depth across separation distances, the variance of the sensitivity profiles, and the shape of the sensitivity profiles across separation distances. In the “Applications of Age-specific DOT Sensitivity Profiles” section of the Discussion, we argued that our findings implied that optimal design of channel configurations that maximizes DOT sensitivity to the underlying region(s) of interest (ROIs) and accurate DOT image reconstruction rely on the use of age-specific head models. This point is also underscored in the Conclusions. 2. The terminologies used in the manuscript are inconsistent and the definitions of these terminologies are problematic. There is a mixed usage between PPL, penetration depth, separation distance, electrode, optode, etc. The author should stick with one naming convention throughout the manuscript. Also, if possible, the formula should be given to define the terminologies. Thank you for pointing out the inconsistencies. We have changed “PPL” to “sampling depth” in response to Comment 6. We have described the computation of the sampling depth in our response. We used “penetration depth” to refer to the depth that the S-D channel DOT fluence distribution (an example shown in Figure 3) extends to the head model (e.g. Huppert et al., 2009; Mansouri et al., 2010). We agree that “penetration depth” is not equivalent to the “sampling depth” we computed. We have changed “penetration depth” to “sampling depth” when describing our findings. For example, in the second paragraph of the Results: “Compared to the sensitivity profile at shorter separation distances, the same fluence value was carried to greater sampling depth at larger separation.” We have now specified the “source-detector separation distance” in the “Source-Detector Channels, All Optode Pairs” section in the Materials and Methods. To summarize, we recorded the distance between each of the optode pairs. Source-detector separation distance was half of the distance between the optode pair. We have changed the phrases “source-detector distance(s)”, “source-detector separation(s)”, “distance(s)”, “separation(s)”, and “channel separation” to “source-detector separation distance(s)” or “separation distance(s)” in the main text and Supporting Information. In response to Reviewer 1 Comment 6, we have changed “electrode(s)” to “optode(s) and “electrode paring” to “channel” in appropriate places. In addition, we changed “depth sensitivity profile” to “S-D channel DOT sensitivity profile” to keep consistency. We defined the S-D channel DOT sensitivity profile as “S-D Channel DOT fluence as a function of the depth between channel location and the voxel sampled in the fluence distribution (“sampling depth”)” (the first paragraph of the Results). 3. It is very difficult to extract information by reading the figures. The focus of this paper is about the impact of age on the fluence. However, many figures are plotting the relationship between PPL and fluence. And it is not clear why this information is important to the DOT community since earlier studies have similar analyses. The figures should be arranged in a way showing why it is important to use realistic brain models for infancy. In such a case, age should be the x-axis. Thank you for the suggestion. We plotted S-D Channel DOT fluence as a function of sampling depth to ensure that the current study can replicate the findings of existing studies using adult head models. That is, the fluence value decreased monotonically and exponentially with increased depth at all separation distances (Strangman et al., 2013). In addition, the shape of the DOT sensitivity profile (S-D Channel DOT fluence as a function of sampling depth fluence) will change from short to longer separation distances. We have also highlighted in “The Present Study” section of the Introduction: “The study will provide novel evidence revealing developmental changes in how the DOT sensitivity profiles vary across separation distances.” All the current figures present visualization of age-related differences. Separate lines were used to display age group as a categorical variable. However, we have taken the reviewer’s suggestion to highlight age-group differences in our visualizations. These additional figures are presented in the Supporting Information. For example, Figure 4A, 4B, and 4C present the S-D Channel DOT sensitivity profile at 20mm, 30mm, and 50mm separation distance, respectively. Age groups were plotted as individual bars. The figures help to visualize age-group differences in the fluence values at each sampling depth interval. Additional figures were also produced for the variance of the S-D Channel DOT fluence (Supporting Information Figure 5) and the half width half maximum (HWHM) locations (Supporting Information Figure 7). We have referenced these additional figures in the Main Text. In response to Reviewer 1 Comment 7, we have also summarized the key information obtained from the visualizations after each section of the Results. We have chosen to keep our original figures in the Main Text, but we will be willing to move additional figures from the Supporting Information to the Main Text based on the reviewer’s suggestion. 4. The title is ambiguous. What is the focus of this title? Is it age-related fluence changes or source-detector separation. We have now changed the title to “Age-related changes in diffuse optical tomography sensitivity profiles in infancy. This is to emphasize the age-related fluence changes. 5. In the abstract, in statement ‘the peak of the sensitivity function was largest at the smallest separation distance and decreased as source-detector distance increased’, which brain region does the sensitivity indicate? Sensitivity in skin, skull or brain?’ Thank you for pointing out the lack of specification. We did not assess DOT sensitivity profiles in each of the tissue types separately as done in published studies (e.g. Strangman, Li, Zhang, 2013; Strangman, Zhang, Li, 2014). Hence, we cannot make a specific statement on whether the sensitivity profiles are for the skin, skull, or brain. We have acknowledged the limitation in the Discussion (“Limitation” section). We now have added a specification in the sentence: “The peak of the sensitivity function in the head was largest at the smallest separation distance and decreased as source-detector distance increased.” 6. In the 3rd paragraph, last sentence ‘PPL represent the penetration depths of the photon measured in the medium’. It sounds to be an inaccurate description for PPL. PPL should be the 'photon trajectory length in a specific tissue'. Can you clarify how you computed PPL in this manuscript. Thank you for raising the issue. We agree that the construct computed in the current paper does not represent “photon trajectory length in a specific tissue”, as defined in the literature we cited (Mansouri et al., 2010; Strangman et al, 2014; Wang et al, 2019). We defined the construct in the “Depth Sensitivity Profile by Source-Detector Separation Distances” section in the Materials and Methods. It is the distance (depth) from the channel location to the voxel with the S-D channel DOT fluence estimation. We computed this construct to display the changes in fluence as photons traveled though the head tissues by separation distances and age groups (as displayed in Figure 4 and 5). We are referring this as “sampling depth”, in order to provide an abbreviated term for the figures and for ease of descriptions. We have stated the abbreviated name in the “S-D Channel DOT Sensitivity Profile by Source-Detector Separation Distances” section. In the same section, we provided more details of the computations: “For each participant and each channel with the target separation distance (e.g. 400 channels with 10mm separation distance), we computed the S-D channel DOT fluence value at each sampling depth. The fluence values were then averaged across channels by sampling depth (1 to 100), target separation distance, and participant. The fluence values were then averaged by separation distance and age group”. Figure 5B, 5C, and 5D displays the averaged fluence values as a function of the sampling depth for the 20mm, 30mm, 50mm target separation distance by age groups. We have removed “PPL” from the third paragraph. We changed the sentences to: “DOT sensitivity is quantified as each location at which a photon traverses through a given tissue region, or “sampling depth”. It represents the depth of the photon measured in the head medium.”. We have changed “PPL” to “sampling depth” in the rest of the manuscript, Figure 4, 5, and 6, as well as Supporting Information. 7. In Fig. 1B and C, a brain model with dense and sparse densities is shown. Which mesh did you use in the manuscript? The dense or the sparse one? Why are these two meshes shown? We apologize for the confusion. As in our response to Comment #5 from the Reviewer 1, we used the dense “segmented FE mesh” in the current study to locate points on the scalp that were closest to the optode positions. We have removed the sparse mesh in Figure 1 and all related information from the Main Text and the Supporting Information. 8. In Fig. 1A, can the author add the legend showing the mapping between color->tissue type? Also, it looks to me the dura is pretty thick and the CSF (dark orange) is not filling between brain and skull? What is the dark blue color in the figure, which appears both outside and inside of the brain? Thank you for the suggestion, and we apologize for the confusion. We have now added a legend and edited the caption for Figure 1 A. The dura is the darker blue ribbon underneath the skull (lighter blue). The CSF (orange) is inside of the dura. The dark blue is the gray matter and the purple is the white matter. 9. In section ‘Scalp Location’, 1st paragraph, can you add citations for 10-10 and 10-5 electrode systems? We have added a citation for the 10-10 and 10-5 systems: “The locations for the 10-10 and 10-5 systems (Jurcak, Tsuzuki, & Dan, 2007) were constructed on each head MRI volume. 10. In section ‘Source-Detector Channels, All Electrode Pairs’, it is confusing to use electrodes here while you are placing light source or optode in the position. Please just call it source or detector and use the same names throughout the manuscript. Thank you for pointing out the confusion. We have changed the “all electrode pairs” to “all optode pairs”. We have changed “electrode(s)” to “optode(s) as appropriate in response to Reviewer 1 and Reviewer 2’s comments. 11. In section ‘Photon Migration Simulations’, while the author admitted the optical properties differ considerably across studies and it is likely different values are needed for adults versus development samples, the author did not give a reason or method how these optical properties are compiled in Table 2. This needs to be more clear and discussed in limitations. Thank you for the suggestion. We have now provided references and reasons for our selection of the optical property for each tissue type in Table 2. We have now discussed the potential issues with using the same optical property settings across age groups in the first paragraph of “Limitations”. We pointed out that “the impact of optical property settings for extracerebral tissues on photon migration might differ between infant and adult head models due to the age-related differences in the thickness of these tissue media (Brigadoi & Cooper, 2015; Dehaes et al., 2011; Fu & Richards, under review; Fukui, Ajichi, & Okada, 2003).” We cited Brigadoi and Cooper (2015)’s validation results indicating that variations in scalp and CSF optical properties within the common ranges might not have substantial impacts on DOT sensitivity estimations using adult and infant head models. However, we argued that “future validation studies are needed to ensure that the age differences in DOT sensitivity profiles are over and above the age-related variations in optical properties of extracerebral and brain tissue layers”. 12. In Table 2, the wavelength of these optical properties is not specified. In addition, citations should be added for the optical properties, showing from which papers these values are derived. Also, the absorption of muscle seems to be low because muscle is highly absorbing. The nasal cavity, which should be air, has an absorption coefficient of 0.0101 mm^-1. What is the reason for this? Please refer to the paper: Selective photobiomodulation for emotion regulation: model-based dosimetry study. Thank you for providing reference. The wavelength of the optical properties in the current paper is 690nm. This information is specified in the legend of Table 2. We have now provided the citations and reasons for the optical property selected for each tissue type in Table 2. We did differentiate the different non-brain tissue types: skin, dura, muscle, and nasal cavity. We agree with the reviewer that different optical properties should have been set for skin, muscle, and nasal cavity (as in Cassano et al., 2019). We have acknowledged this limitation in the Discussion. Our optical property input for the nasal cavity could also be modified. However, we expect the effect of the setting to the fluence estimation is minimal, as the photon is unlikely to reach to the nasal cavity before exiting the head media or being absorbed by the media. 13. In section ‘DOT Sensitivity Analyses’, last sentence, there is a statement ‘This figure shows a similar monotonic decrease in fluence strength as a function of partial path length’. However, which PPL are you referring to? And PPL and penetration depth should not be the same thing. The author should stick with one naming in the manuscript. Thank you for pointing out the issue with the use of terminology. We have changed the naming “PPL” to “sampling depth”. Please refer to our response for Comment 6 for more detailed explanations. We described the computation of sampling depth in the following section “Depth Sensitivity profile by Source-Detector Separation Distances”. In the description of Figure 3, we changed the sentence to: “Figure 3 shows the S-D Channel DOT fluence plotted on an MRI for a single source-detector channel. This figure shows a similar monotonic decrease in fluence strength as photons traveled deeper into the head tissues.” We agree that “penetration depth” and “sampling depth” are not the same construct. We have changed “penetration depth” to “sampling depth” when describing our findings (please also see our response to Comment 2. 14. In Fig. 3, the S-D Channel DOT fluence is a 3-D distribution. But In Fig. 4 and 5, it becomes a scalar that changes depending on PPL. How did you convert it to a value? Also, int Fig. 4 and 5, the S-D Channel DOT fluence does not have a unit. We have clarified in the “DOT Sensitivity Analyses” section of the Materials and Methods: “(S-D Channel DOT fluence)..is a unitless measure that represents the sensitivity of the DOT measure for detecting changes in optical properties at a given point inside the head medium (PMDF; Brigadoi, Aljabar, Kuklisova-Murgasova, Arridge, & Cooper, 2014; Brigadoi & Cooper, 2015; or "3-point Green's function"; Strangman et al., 2013). Figure 3 shows displays the S-D channel DOT fluence distribution at a channel location for all voxels in the head model space. In the next paragraph, we have now specified: “The S-D Channel DOT fluence was extracted voxel by voxel. We recorded the fluence value and the distance (depth) from the channel location to the voxel with the fluence estimation. We refer the distance from the channel to the location of the fluence value inside the head model as “sampling depth” for short. The S-D channel DOT sensitivity profile was defined as the fluence value at the voxel as a function of the sampling depth.” Figure 4 and 5 displayed the S-D channel DOT sensitivity profiles across source-detector separation distances. 15. As mentioned in the major issues, the figures from Fig. 4 to 7 are very difficult to read. It is difficult to extract effective information from these figures. These should be re-organized. We have now added more detailed descriptions of the variables in the figure captions. As our response to Reviewer 1, we have now added summaries of the key findings displayed in Figure 4 to Figure 7 in the Results section. We have also explained our rationales for these visualizations in our response to Comment 3. 16. In section ‘The Shape of S-D Channel DOT Fluence Sensitivity Profiles’, last sentence ‘The larger HWHM location values for the youngest age group implies the DOT sensitivity to deeper cortical areas for the youngest ages.’ However, in Fig. 7B, the HWHM for age 4 to 24 years old is the highest, which conflicts with the claim that the youngest age group has DOT sensitivity in deeper cortical areas. The same statement was made in the 5th sentence in Discussion. Please clarify. We apologize for the confusion. We have now stated in the “Shape of S-D Channel DOT Sensitivity Profile”: “The mean HWHM locations across all separation distances were greater in the youngest infants (2-week through 4.5 month old age groups, Figure 7A and 7B solid line, aqua color), children and adults (Figure 7A and 7B green dashed line)”. We also stated that “The larger HWHM location values imply the DOT sensitivity to deeper cortical areas for the youngest infants, children, and adults.” In the first paragraph of the Discussion, we added: “The shape of S-D channel DOT sensitivity profile was similar between the younger infant groups and child and adult groups.” We placed greater emphasis on discussing the differences in S-D Channel DOT sensitivity profiles across separation distances among infant groups. The child and adult groups were used as reference groups to show broad differences between infants and older ages. 17. In Discussion, while the findings about sensitivity change and separations for infancy are given, the insights on what may cause these changes are not given. While these may not be included in this manuscript, they can be future directions. Thank you for the suggestion. We have now discussed the possibility that “developmental changes in the morphology of the head and the brain may contribute to the age-related differences in S-D channel DOT sensitivity profiles” (the third paragraph of the “S-D Channel DOT Sensitivity Profile by Source-Detector Separation Distances” section in the Discussion. We acknowledged in the end of the paragraph that “a direction for future research is to characterize morphological changes in specific tissue layers that are linked to age-related differences in S-D channel DOT sensitivity profiles across separation distances”. Submitted filename: Response to Reviewers_RR1_022221.docx Click here for additional data file. 10 May 2021 Age-related Changes in Diffuse Optical Tomography Sensitivity Profiles in Infancy PONE-D-20-39767R1 Dear Dr. Fu, We’re pleased to inform you that your manuscript has been judged scientifically suitable for publication and will be formally accepted for publication once it meets all outstanding technical requirements. Within one week, you’ll receive an e-mail detailing the required amendments. When these have been addressed, you’ll receive a formal acceptance letter and your manuscript will be scheduled for publication. An invoice for payment will follow shortly after the formal acceptance. To ensure an efficient process, please log into Editorial Manager at http://www.editorialmanager.com/pone/, click the 'Update My Information' link at the top of the page, and double check that your user information is up-to-date. If you have any billing related questions, please contact our Author Billing department directly at authorbilling@plos.org. If your institution or institutions have a press office, please notify them about your upcoming paper to help maximize its impact. If they’ll be preparing press materials, please inform our press team as soon as possible -- no later than 48 hours after receiving the formal acceptance. Your manuscript will remain under strict press embargo until 2 pm Eastern Time on the date of publication. For more information, please contact onepress@plos.org. Kind regards, Andrea Farina Academic Editor PLOS ONE Additional Editor Comments (optional): I kindly ask you to edit Fig.3 specifying a colorbar and the source-detector separation related to the simulation. Reviewers' comments: Reviewer's Responses to Questions Comments to the Author 1. If the authors have adequately addressed your comments raised in a previous round of review and you feel that this manuscript is now acceptable for publication, you may indicate that here to bypass the “Comments to the Author” section, enter your conflict of interest statement in the “Confidential to Editor” section, and submit your "Accept" recommendation. Reviewer #1: All comments have been addressed Reviewer #2: All comments have been addressed ********** 2. Is the manuscript technically sound, and do the data support the conclusions? The manuscript must describe a technically sound piece of scientific research with data that supports the conclusions. Experiments must have been conducted rigorously, with appropriate controls, replication, and sample sizes. The conclusions must be drawn appropriately based on the data presented. Reviewer #1: Yes Reviewer #2: Yes ********** 3. Has the statistical analysis been performed appropriately and rigorously? Reviewer #1: Yes Reviewer #2: Yes ********** 4. Have the authors made all data underlying the findings in their manuscript fully available? The PLOS Data policy requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception (please refer to the Data Availability Statement in the manuscript PDF file). The data should be provided as part of the manuscript or its supporting information, or deposited to a public repository. For example, in addition to summary statistics, the data points behind means, medians and variance measures should be available. If there are restrictions on publicly sharing data—e.g. participant privacy or use of data from a third party—those must be specified. Reviewer #1: Yes Reviewer #2: Yes ********** 5. Is the manuscript presented in an intelligible fashion and written in standard English? PLOS ONE does not copyedit accepted manuscripts, so the language in submitted articles must be clear, correct, and unambiguous. Any typographical or grammatical errors should be corrected at revision, so please note any specific errors here. Reviewer #1: Yes Reviewer #2: Yes ********** 6. Review Comments to the Author Please use the space provided to explain your answers to the questions above. You may also include additional comments for the author, including concerns about dual publication, research ethics, or publication ethics. (Please upload your review as an attachment if it exceeds 20,000 characters) Reviewer #1: The authors have addressed all comments by reviewers thoroughly, and they have improved the manuscript clearly. The manuscript now provides a very nice and well written description of the research. I recommend the manuscript to be accepted for publication in its current form. Reviewer #2: Age-related sensitivity profile in infancy is a very valuable topic for current fNIRS study. All my earlier comments have been addressed in detail. I recommend to accept this manuscript for the journal. ********** 7. PLOS authors have the option to publish the peer review history of their article (what does this mean?). If published, this will include your full peer review and any attached files. If you choose “no”, your identity will remain anonymous but your review may still be made public. Do you want your identity to be public for this peer review? For information about this choice, including consent withdrawal, please see our Privacy Policy. Reviewer #1: No Reviewer #2: No 31 May 2021 PONE-D-20-39767R1 Age-related Changes in Diffuse Optical Tomography Sensitivity Profiles in Infancy Dear Dr. Fu: I'm pleased to inform you that your manuscript has been deemed suitable for publication in PLOS ONE. Congratulations! Your manuscript is now with our production department. If your institution or institutions have a press office, please let them know about your upcoming paper now to help maximize its impact. If they'll be preparing press materials, please inform our press team within the next 48 hours. Your manuscript will remain under strict press embargo until 2 pm Eastern Time on the date of publication. For more information please contact onepress@plos.org. If we can help with anything else, please email us at plosone@plos.org. Thank you for submitting your work to PLOS ONE and supporting open access. Kind regards, PLOS ONE Editorial Office Staff on behalf of Dr. Andrea Farina Academic Editor PLOS ONE
  53 in total

1.  Volumetric diffuse optical tomography of brain activity.

Authors:  Joseph P Culver; Andrew M Siegel; Jonathan J Stott; David A Boas
Journal:  Opt Lett       Date:  2003-11-01       Impact factor: 3.776

2.  Longitudinal development of cortical and subcortical gray matter from birth to 2 years.

Authors:  John H Gilmore; Feng Shi; Sandra L Woolson; Rebecca C Knickmeyer; Sarah J Short; Weili Lin; Hongtu Zhu; Robert M Hamer; Martin Styner; Dinggang Shen
Journal:  Cereb Cortex       Date:  2011-11-22       Impact factor: 5.357

3.  The Toast++ software suite for forward and inverse modeling in optical tomography.

Authors:  Martin Schweiger; Simon Arridge
Journal:  J Biomed Opt       Date:  2014-04       Impact factor: 3.170

4.  A database of age-appropriate average MRI templates.

Authors:  John E Richards; Carmen Sanchez; Michelle Phillips-Meek; Wanze Xie
Journal:  Neuroimage       Date:  2015-05-02       Impact factor: 6.556

5.  N4ITK: improved N3 bias correction.

Authors:  Nicholas J Tustison; Brian B Avants; Philip A Cook; Yuanjie Zheng; Alexander Egan; Paul A Yushkevich; James C Gee
Journal:  IEEE Trans Med Imaging       Date:  2010-04-08       Impact factor: 10.048

6.  Validating atlas-guided DOT: a comparison of diffuse optical tomography informed by atlas and subject-specific anatomies.

Authors:  Robert J Cooper; Matteo Caffini; Jay Dubb; Qianqian Fang; Anna Custo; Daisuke Tsuzuki; Bruce Fischl; William Wells; Ippeita Dan; David A Boas
Journal:  Neuroimage       Date:  2012-05-23       Impact factor: 6.556

7.  Neural correlates of own- and other-race face recognition in children: a functional near-infrared spectroscopy study.

Authors:  Xiao Pan Ding; Genyue Fu; Kang Lee
Journal:  Neuroimage       Date:  2013-07-25       Impact factor: 6.556

8.  Age-specific MRI brain and head templates for healthy adults from 20 through 89 years of age.

Authors:  Paul T Fillmore; Michelle C Phillips-Meek; John E Richards
Journal:  Front Aging Neurosci       Date:  2015-04-08       Impact factor: 5.750

9.  fNIRS Optodes' Location Decider (fOLD): a toolbox for probe arrangement guided by brain regions-of-interest.

Authors:  Guilherme Augusto Zimeo Morais; Joana Bisol Balardin; João Ricardo Sato
Journal:  Sci Rep       Date:  2018-02-20       Impact factor: 4.379

10.  High-density diffuse optical tomography for imaging human brain function.

Authors:  Muriah D Wheelock; Joseph P Culver; Adam T Eggebrecht
Journal:  Rev Sci Instrum       Date:  2019-05       Impact factor: 1.523

View more
  4 in total

1.  Age-related changes in diffuse optical tomography sensitivity profiles from childhood to adulthood.

Authors:  Xiaoxue Fu; John E Richards
Journal:  J Biomed Opt       Date:  2022-07       Impact factor: 3.758

2.  Examining the Role of Socioeconomic Status and Maternal Sensitivity in Predicting Functional Brain Network Connectivity in 5-Month-Old Infants.

Authors:  Johanna R Chajes; Jessica A Stern; Caroline M Kelsey; Tobias Grossmann
Journal:  Front Neurosci       Date:  2022-06-10       Impact factor: 5.152

3.  Investigating developmental changes in scalp-to-cortex correspondence using diffuse optical tomography sensitivity in infancy.

Authors:  Xiaoxue Fu; John E Richards
Journal:  Neurophotonics       Date:  2021-07-24       Impact factor: 3.593

4.  Objective Numerical Evaluation of Diffuse, Optically Reconstructed Images Using Structural Similarity Index.

Authors:  Vicky Mudeng; Minseok Kim; Se-Woon Choe
Journal:  Biosensors (Basel)       Date:  2021-12-08
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.