| Literature DB >> 34085364 |
Naoki Kinoshita1, Hiroshi Oguchi2, Morihito Shimizu3, Eiji Kidoya1, Hiroki Shioura4, Hirohiko Kimura4.
Abstract
PURPOSE: Medical physicists use a suitable detector connected to an electrometer to measure radiotherapy beams. Each detector and electrometer has a lifetime (due to physical deterioration of detector components and electrical characteristic deterioration in electronic electrometer components), long-term stability [according to IEC 60731:2011, ≤0.5% (reference-class dosimeter)], and calibration frequency [according to Muir et al. (J Appl Clin Med Phys. 2017; 18:182-190), generally 2 years]; thus, physicists should check the electrometer and detector separately. However, to the best of our knowledge, only one study (Blad et al., Phys Med Biol. 1998; 43:2385-2391) has reported checking the electrometer independently from the detector. The present study conducts performance checks on electrometers separately from the detector in clinical settings, using an electrometer equipped with a direct current (DC) generator (EMF 521R) capable of injecting DC (effective range: ±20 pA to ±20 nA) into itself or another electrometer.Entities:
Keywords: clinical electrometer check; electrometer equipped with direct-current generator; nonlinearity; repeatability
Mesh:
Year: 2021 PMID: 34085364 PMCID: PMC8292692 DOI: 10.1002/acm2.13312
Source DB: PubMed Journal: J Appl Clin Med Phys ISSN: 1526-9914 Impact factor: 2.102
Fig. 1Image of EMF 521R used in this work. This electrometer can transfer constant currents to itself [panel (b)] or to another electrometer [panel (c)] through output‐1 and −2 [panel (a)] for electrometer performance checks.
Uncertainty budget (k = 1) for charges generated from the DC generator.
| Source | Output‐1 (%) | Output‐2 (%) |
|---|---|---|
| 1. In‐house standard device calibrations | 0.08 | 0.08 |
| 2. Current calibration with in‐house electrometer | 0.03 | 0.03 |
| 3. Direct current at a radiotherapy facility | 0.12 | 0.05 |
| Combined uncertainty ( | 0.15 | 0.10 |
Characteristics of the six electrometers checked in this work.
| Electrometer | 1 | 2 | 3 | 4 | 5 | 6 |
|---|---|---|---|---|---|---|
| NF in op‐amp | Capacitor | Resistor | Resistor | Capacitor | Resistor | Resistor |
| Range | L; H | Single | L; H | Single | L; H | L; M; H |
| ER |
L: ±1.0 pA to ±10 nA H: ±0.4 nA to ±2 µA | ±4 pA to ±20 nA |
L: ±0.4 pA to ±1 nA* H: ±0.4 nA to ±500 nA* | ±2 pC to ±10 mC* |
L: ±0.4 pA to ±500 pA* H: ±0.4 nA to ±500 nA* |
L: ±0.4 pA to ±250 pA* M: ±20 pA to ±25 nA* H: ±2 nA to ±2.5 μA* |
| EIV |
L: ±1.0 pA to ±10 nA H: ±0.4 nA to ±2 µA | ±4 pA to ±20 nA |
L: ±0.4 pA to ±1 nA* H: ±0.4 nA to ±500 nA* | ±2 pC to ±10 mC* |
L: ±0.4 pA to ±500 pA* H: ±0.4 nA to ±500 nA* |
L: ±0.4 pA to ±250 pA* M: ±20 pA to ±25 A* H: ±2 nA to ±2.5 μA* |
Abbreviations: NF in op‐amp, negative feedback in operational amplifier; Capacitor, auto‐discharge integrating capacitor; Range, measuring range; L, Low; M, middle; H, high; ER, effective range of readings; and EIV, effective range of indicated values. Effective ranges of readings and indicated values for the six electrometers corresponded to those described in IEC 60731:2011, with four exceptions (electrometers 3–6). For these, the performances (e.g., repeatability and nonlinearity) were guaranteed in the user’s manuals to be within values marked with *.
Ranges of DC currents for nonlinearity checks performed in this work.
| Electrometer | Range | Range of input currents (nA) | RP (nA) |
|---|---|---|---|
| 1 | Low | ±0.02 to ±10 | ±5 |
| 2 | Single | ±0.02 to ±20 | ±10 |
| 3 | Low | ±0.02 to ±1 | ±0.5 |
| 4 | Single | ±0.02 to ±20 | ±20 |
| 5 | Low | ±0.02 to ±0.5 | ±0.25 |
| 6 | Low | ±0.02 to ±0.25 | ±0.125 |
| Middle | ±0.02 to ±20 | ±10 |
Abbreviation: RP, reference point.
Fig. 2Nonlinearities for measured current from the DC generator: (a) from output‐1 and (b) from output‐2. The y‐axis shows the nonlinearity (%); the x‐axis indicates the injected currents on a logarithmic scale.
Uncertainty budget (k = 1) for measured charges.
| Source and type of uncertainty | Output 1 (%) | Output 2 (%) |
|---|---|---|
| 1. Charges generated from the DC generator | 0.15 | 0.10 |
| 2. | 0.08 | 0.08 |
| 3. Electrometer reading (EMF 521R SN: X004) | 0.05 | 0.05 |
| Combined uncertainty ( | 0.18 | 0.14 |
Source 1 is quoted from Table 1. Source 2 was obtained from its calibration certificate issued by the ANTM. Source 3 was derived from the electrometer performances of the EMF 521R, which are listed in IEC 60731:2011.
Repeatability for the six electrometers.
| Electrometer | Range | Injected current (pA) | Mean reading (pA) | Standard deviation (%) |
|---|---|---|---|---|
| 1 | Low | −20 | −19.98 | 0.04 |
| +20 | +20.00 | 0.05 | ||
| High | −400 | −399.8 | 0.05 | |
| +400 | +399.6 | 0.05 | ||
| 2 | Single | −20 | −20.00 | 0.01 |
| +20 | +20.01 | 0.01 | ||
| 3 | Low | −20 | −20.00 | 0.01 |
| +20 | +20.04 | 0.02 | ||
| High | −400 | −401.6 | 0.07 | |
| +400 | +399.7 | 0.09 | ||
| 4 | Single | −20 | −19.97 | 0.03 |
| +20 | +19.96 | 0.02 | ||
| 5 | Low | −20 | −20.03 | 0.05 |
| +20 | +20.03 | 0.05 | ||
| High | −400 | −400.4 | 0.02 | |
| +400 | +401 | 0.02 | ||
| 6 | Low | −20 | −19.98 | 0.005 |
| +20 | +20.01 | 0.05 | ||
| Middle | −20 | −20.04 | 0.04 | |
| +20 | +19.99 | 0.03 | ||
| High | −2000 | −2000.5 | 0.05 | |
| +2000 | +2000.4 | 0.06 |
Mean readings and standard deviations were calculated for constant currents injected 10 times successively.
Fig. 3Nonlinearities for the six electrometers when (a) negative and (b) positive currents generated from the DC generator were injected. The y‐axis shows the nonlinearity (%); the x‐axis indicates the injected currents on a logarithmic scale.