Literature DB >> 34078155

The Silk Vista Baby - The UK experience.

P Bhogal1, Hld Makalanda1, K Wong1, P Keston2, J Downer2, J C Du Plessis2, A Nania2, D Simonato3, M Fuschi3, W Chong4, S O'Reilly5, I Rennie5.   

Abstract

BACKGROUND: The Silk Vista Baby (SVB) flow diverter (FDS) is the only FDS deliverable via a 0.017 inch microcatheter and is specifically designed for the distal vasculature. We sought to evaluate the safety and efficacy of the SVB.
MATERIALS AND METHODS: We performed a retrospective review to identify SVB cases at 4 tertiary neurosurgical centres within the U.K. Clinical, procedural, angiographic and follow-up data were collected.
RESULTS: We identified 60 patients (35 female, 58%) of average age 54 ± 10.5 (range 30-72) with 61 aneurysms, 50 (81.9%) located in the anterior circulation. The majority of the aneurysms treated were unruptured (46, 75.4%) and saccular (46, 75.4%). Dome size was 6.2 ± 6.2 mm (range 1-36mm) and parent vessel diameter was 2.3 ± 0.4 mm (range 1.2-3.3 mm).An average number of 1.07 devices were implanted. Coils or other devices were implanted in 14 aneurysms (23.3%). At last angiographic follow-up (n = 55), 7.5 ± 4.2 months post-procedure, 32 aneurysms (57.1%) were graded as RRC I, 7 (12.5%) RRC II, and 17 RRC III (30.4%).Clinical complications, excluding death, were seen in 4 patients (6.8%) including 1 delayed aneurysm rupture and 3 symptomatic ischaemic events. Only one patient had permanent morbidity (mRS 1). 3 patients died during follow-up (5.1%); 2 deaths were related to the aneurysms (3.4%) - one ruptured dissecting MCA aneurysm, and one giant partially thrombosed posterior circulation aneurysm. 93% of patients were mRS ≤ 2 at last follow-up.
CONCLUSION: The SVB has high rates of technical success and an acceptable safety profile. Distal aneurysms may occlude slower due to relative oversizing of the devices.

Entities:  

Keywords:  SVB; aneurysm; flow diversion

Mesh:

Substances:

Year:  2021        PMID: 34078155      PMCID: PMC9131493          DOI: 10.1177/15910199211024061

Source DB:  PubMed          Journal:  Interv Neuroradiol        ISSN: 1591-0199            Impact factor:   1.764


  54 in total

1.  Flow diversion treatment of complex bifurcation aneurysms beyond the circle of Willis: a single-center series with special emphasis on covered cortical branches and perforating arteries.

Authors:  Matthias Gawlitza; Anne-Christine Januel; Philippe Tall; Fabrice Bonneville; Christophe Cognard
Journal:  J Neurointerv Surg       Date:  2015-04-15       Impact factor: 5.836

2.  Flow diverters at and beyond the level of the circle of willis for the treatment of intracranial aneurysms.

Authors:  Silvia Pistocchi; Raphaël Blanc; Bruno Bartolini; Michel Piotin
Journal:  Stroke       Date:  2012-01-26       Impact factor: 7.914

3.  Variable porosity of the pipeline embolization device in straight and curved vessels: a guide for optimal deployment strategy.

Authors:  M Shapiro; E Raz; T Becske; P K Nelson
Journal:  AJNR Am J Neuroradiol       Date:  2013-09-26       Impact factor: 3.825

4.  The Silk Vista Baby: Initial experience and report of two cases.

Authors:  Pervinder Bhogal; Ken Wong; Christopher Uff; John Wadley; Hegoda Ld Makalanda
Journal:  Interv Neuroradiol       Date:  2019-03-31       Impact factor: 1.610

5.  Flow diversion treatment for acutely ruptured aneurysms.

Authors:  Christian A Taschner; Hieronymus D Boogaarts; Michelle F M Ten Brinck; Maike Jäger; Joost de Vries; J André Grotenhuis; René Aquarius; Svein H Mørkve; Riitta Rautio; Jussi Numminen; Rahul Raj; Ajay K Wakhloo; Ajit S Puri
Journal:  J Neurointerv Surg       Date:  2019-08-24       Impact factor: 5.836

6.  Surface-modified flow diverter p48-MW-HPC: Preliminary clinical experience in 28 patients treated in two centers.

Authors:  Laurent Pierot; Sébastien Soize; Matteo Cappucci; Pierre-François Manceau; Roberto Riva; Omer Faruk Eker
Journal:  J Neuroradiol       Date:  2020-12-17       Impact factor: 3.447

7.  Flow Diversion in Middle Cerebral Artery Aneurysms: Is It Really an All-Purpose Treatment?

Authors:  Osman Melih Topcuoglu; Erol Akgul; Ergun Daglioglu; Elif Dilara Topcuoglu; Ahmet Peker; Ilkay Akmangit; Deniz Belen; Anil Arat
Journal:  World Neurosurg       Date:  2015-12-23       Impact factor: 2.104

8.  Pipeline Flex embolisation device with Shield Technology for the treatment of patients with intracranial aneurysms: periprocedural and 6 month outcomes.

Authors:  James Yeomans; Lilian Sandu; Anand Sastry
Journal:  Neuroradiol J       Date:  2020-10-22

9.  Flow diverter stents with hydrophilic polymer coating for the treatment of acutely ruptured aneurysms using single antiplatelet therapy: Preliminary experience.

Authors:  Giuseppe Guzzardi; Andrea Galbiati; Carmelo Stanca; Bruno Del Sette; Andrea Paladini; Christian Cossandi; Alessandro Carriero
Journal:  Interv Neuroradiol       Date:  2020-08-16       Impact factor: 1.610

10.  Use of flow diverters in the treatment of unruptured saccular aneurysms of the anterior cerebral artery.

Authors:  P Bhogal; R Martinez Moreno; O Ganslandt; H Bäzner; H Henkes; M Aguilar Perez
Journal:  J Neurointerv Surg       Date:  2016-10-27       Impact factor: 5.836

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.