Evangelos Terpos1, Ioannis Ntanasis-Stathopoulos1, Miha Skvarč2.
Abstract
The precise diagnosis of COVID-19 is of outmost importance in order to effectively treat patients and prevent SARS-CoV-2 transmission. Herein, we evaluated the sensitivity and specificity of the COVID-19 Antigen Detection Kit (Colloidal Gold-CG) compared with PCR in nasopharyngeal and nasal samples. A total of 114 positive and 244 negative nasopharyngeal specimens confirmed by PCR were used in this comparative study. When the PCR positive Cycle Threshold (Ct) value was ≤25, CG sensitivity was 100%. When the PCR positive Ct value was ≤33, CG sensitivity was 99%. When the PCR positive Ct value was ≤40, CG sensitivity was 89.47%. Regarding nasal swabs, a total of 109 positive and 250 negative specimens confirmed by PCR were used. When the PCR positive Ct value was ≤25, CG sensitivity was 100%. When the PCR positive Ct value was ≤33, CG sensitivity was 96.12%. When the PCR positive Ct value was ≤37, CG sensitivity was 91.74%. Specificity was above 99% regardless of the Ct value of PCR positivity for both nasopharyngeal and nasal specimens. Overall, the CG showed high sensitivity and specificity when the PCR Ct value was less than 33. Therefore, CG can be used for screening early in the disease course. Confirmatory PCR is essential when a false negative result is suspected.Entities:
Keywords: COVID-19; SARS-CoV-2; antigen test; clinical validation; nasal swab; nasopharyngeal swab; nucleocapsid
Year: 2021 PMID: 34070844 PMCID: PMC8229208 DOI: 10.3390/diagnostics11060995
Source DB: PubMed Journal: Diagnostics (Basel) ISSN: 2075-4418
Test results and analysis of nasopharyngeal swab specimens with Ct values of PCR positivity less than 25.
|
| Nasopharyngeal Swab Specimen | ||
|---|---|---|---|
| Positive (PCR < 25) | PCR Negative | Total | |
| AT Positive | 70 | 1 | 71 |
| AT Negative | 0 | 243 | 243 |
| Total | 70 | 244 | 314 |
| Sensitivity | 100.00% | ||
| 95% CI | 98.79–100.00% | ||
| Specificity | 99.59% | ||
| 95% CI | 98.07–99.91% | ||
| Accuracy | 99.68% | ||
AT: antigen test; CI: confidence interval.
Test results and analysis of nasopharyngeal swab specimens with Ct values of PCR positivity less than 33.
|
| Nasopharyngeal Swab Specimen | ||
|---|---|---|---|
| Positive (PCR < 33) | PCR Negative | Total | |
| AT Positive | 99 | 1 | 100 |
| AT Negative | 1 | 243 | 244 |
| Total | 100 | 244 | 344 |
| Sensitivity | 99.00% | ||
| 95% CI | 97.28–99.64% | ||
| Specificity | 99.59% | ||
| 95% CI | 98.18–99.91% | ||
| Accuracy | 99.42% | ||
AT: antigen test; CI: confidence interval.
Test results and analysis of nasopharyngeal swab specimens with Ct values of PCR positivity less than 40.
|
| Nasopharyngeal Swab Specimen | ||
|---|---|---|---|
| Positive (PCR < 40) | PCR Negative | Total | |
| AT Positive | 102 | 1 | 103 |
| AT Negative | 12 | 243 | 255 |
| Total | 114 | 244 | 358 |
| Sensitivity | 89.47% | ||
| 95% CI | 85.86–92.24% | ||
| Specificity | 99.59% | ||
| 95% CI | 98.22–99.91% | ||
| Accuracy | 96.37% | ||
AT: antigen test; CI: confidence interval.
Test results and analysis of nasal swab specimens with Ct values of PCR positivity less than 25.
|
| Nasal Swab Specimen | ||
|---|---|---|---|
| Positive (PCR < 25) | PCR Negative | Total | |
| AT Positive | 61 | 0 | 61 |
| AT Negative | 0 | 250 | 250 |
| Total | 61 | 250 | 311 |
| Sensitivity | 100.00% | ||
| 95% CI | 94.08–100.00% | ||
| Specificity | 100% | ||
| 95% CI | 98.49–100.00% | ||
| Accuracy | 98.78% | ||
| 95% CI | 98.78–100.00% | ||
AT: antigen test; CI: confidence interval.
Test results and analysis of nasal swab specimens with Ct values of PCR positivity less than 33.
|
| Nasal Swab Specimen | ||
|---|---|---|---|
| Positive (PCR < 33) | PCR Negative | Total | |
| AT Positive | 99 | 0 | 99 |
| AT Negative | 4 | 250 | 254 |
| Total | 103 | 250 | 353 |
| Sensitivity | 96.12% | ||
| 95% CI | 90.43–98.49% | ||
| Specificity | 100.00% | ||
| 95% CI | 98.49–100.00% | ||
| Accuracy | 98.87% | ||
| 95% CI | 97.12–99.56% | ||
AT: antigen test; CI: confidence interval.
Test results and analysis of nasal swab specimens with Ct values of PCR positivity less than 37.
|
| Nasal Swab Specimen | ||
|---|---|---|---|
| Positive (PCR < 37) | PCR Negative | Total | |
| AT Positive | 100 | 0 | 100 |
| AT Negative | 9 | 250 | 259 |
| Total | 109 | 250 | 359 |
| Sensitivity | 91.74% | ||
| 95% CI | 85.05–95.60% | ||
| Specificity | 100.00% | ||
| 95% CI | 98.49–100.00% | ||
| Accuracy | 97.49% | ||
| 95% CI | 95.30–98.68% | ||
AT: antigen test; CI: confidence interval.