| Literature DB >> 34065802 |
Gennaro Auletta1, Annamaria Franzè1, Carla Laria1, Carmine Piccolo1, Carmine Papa1, Pasquale Riccardi2, Davide Pisani3, Angelo Sarnelli1, Valeria Del Vecchio1, Rita Malesci1, Elio Marciano1.
Abstract
BACKGROUND: The aim of this study was to compare, in users of bimodal cochlear implants, the performance obtained using their own hearing aids (adjusted with the standard NAL-NL1 fitting formula) with the performance using the Phonak Naìda Link Ultra Power hearing aid adjusted with both NAL-NL1 and a new bimodal system (Adaptive Phonak Digital Bimodal (APDB)) developed by Advanced Bionics and Phonak Corporations.Entities:
Keywords: APDB; bimodal system; cochlear implant; hearing aid fitting
Year: 2021 PMID: 34065802 PMCID: PMC8161435 DOI: 10.3390/audiolres11020018
Source DB: PubMed Journal: Audiol Res ISSN: 2039-4330
Detailed demographic information.
| ID | Age (years) | Etiology | Age at Diagnosis (years) | Duration of CI Use (years) | PTA (dB) | Previous HA | Previous Fitting Formula |
|---|---|---|---|---|---|---|---|
| S01 | 23 | Rubella | 2 | 6 | 96.25 | GN Resound Enzo | NAL-NL1 |
| S02 | 23 | Rubella | 2 | 9 | 96.25 | GN Resound Enzo | NAL-NL1 |
| S03 | 9 | Usher | 1 | 1 | 81.25 | Phonak Naida Sky UP Q70 | NAL-NL1 |
| S04 | 34 | Connexin 26 | 4 | 2 | 85 | Phonak Naida Sky UP Q70 | NAL-NL1 |
| S05 | 12 | Unknown | 4 | 3 | 97.5 | Phonak Naida Sky UP Q70 | NAL-NL1 |
| S06 | 6 | Connexin 26 | 2 | 2 | 95 | Phonak Naida Sky UP Q70 | NAL-NL1 |
| S07 | 28 | Unknown | 3 | 8 | 107.5 | Phonak Naida Sky UP Q70 | NAL-NL1 |
| S08 | 77 | Noise | unknown | 3 | 105 | GN Resound Enzo | NAL-NL1 |
| S09 | 10 | Connexin 26 | 2 | 3 | 91.25 | Phonak Naida Sky UP Q70 | NAL-NL1 |
| S10 | 10 | Connexin 26 | 1 | 6 | 101.25 | Phonak Naida Sky UP Q70 | NAL-NL1 |
| S11 | 20 | Unknown | 2 | 10 | 103.75 | Phonak Naida Sky UP Q70 | NAL-NL1 |
Figure 1Speech intelligibility scores in % correct for the different noise conditions and different HA models and fitting approaches. Asterisks denote statistical significance.
Median of the individual speech intelligibility improvements. Each entry represents the improvement of the fitting condition indicated in the respective row over the fitting condition indicated in the respective column.
| Quiet | 0 dB SNR | −5 dB SNR | ||||
|---|---|---|---|---|---|---|
| NAL-NL1 | Own HA | NAL-NL1 | Own HA | NAL-NL1 | Own HA | |
| APDB | 5% | 5% | 20% | 30% | 20% | 20% |
| NAL-NL1 | 0% | 10% | 0% | |||